| | Class 2 Device Recall Philips Azurion |  |
| Date Initiated by Firm | June 17, 2026 |
| Date Posted | August 05, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2892-2026 |
| Recall Event ID |
99313 |
| 510(K)Number | K172822 K181830 K200917 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Azurion under the following System Descriptions and corresponding model numbers:
1. Azurion 3 M12; Model Numbers: 722063, 722221.
2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280.
3. Azurion 5 M12; Model Number: 722227.
4. Azurion 5 M20; Model Numbers: 722228, 722281.
5. Azurion 7 B12; Model Numbers: 722067, 722225.
6. Azurion 7 B20; Model Numbers: 722068, 722226.
7. Azurion 7 M12; Model Numbers: 722078, 722223.
8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282.
Interventional Fluoroscopic X-Ray System. |
| Code Information |
1. Model Numbers (UDI Numbers): 722063 ((01)00884838085275(21)56, (01)00884838085275(21)96, (01)00884838085275(21)98, (01)00884838085275(21)80, (01)00884838085275(21)146, (01)00884838085275(21)3, (01)00884838085275(21)22, (01)00884838085275(21)153, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)30, (01)00884838085275(21)19, (01)00884838085275(21)121, (01)00884838085275(21)18, (01)00884838085275(21)16, (01)00884838085275(21)14, (01)00884838085275(21)17, (01)00884838085275(21)53, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)41, (01)00884838085275(21)104, (01)00884838085275(21)151, (01)00884838085275(21)137, (01)00884838085275(21)73, (01)00884838085275(21)150, (01)00884838085275(21)105, (01)00884838085275(21)129, (01)00884838085275(21)59, (01)00884838085275(21)154, (01)00884838085275(21)75, (01)00884838085275(21)94, (01)00884838085275(21)26, (01)00884838085275(21)106, (01)00884838085275(21)135, (01)00884838085275(21)123, (01)00884838085275(21)89, (01)00884838085275(21)79, (01)00884838085275(21)133, (01)00884838085275(21)81, (01)00884838085275(21)140, (01)00884838085275(21)52, (01)00884838085275(21)48, (01)00884838085275(21)148, (01)00884838085275(21)55), 722221 ((01)00884838099203(21)93, (01)00884838099203(21)108, (01)00884838099203(21)101, (01)00884838099203(21)124, (01)00884838099203(21)4, (01)00884838099203(21)106, (01)00884838099203(21)151, (01)00884838099203(21)71, (01)00884838099203(21)105, (01)00884838099203(21)60, (01)00884838099203(21)33, (01)00884838099203(21)55, (01)00884838099203(21)44, (01)00884838099203(21)74, (01)00884838099203(21)142, (01)00884838099203(21)144, (01)00884838099203(21)147, (01)00884838099203(21)96, (01)00884838099203(21)92, (01)00884838099203(21)78, (01)00884838099203(21)23, (01)00884838099203(21)133, (01)00884838099203(21)167, (01)00884838099203(21)26, (01)00884838099203(21)15, (01)00884838099203(21)163, (01)00884838099203(21)67, (01)00884838099203(21)81, (01)00884838099203(21)22).
2. Model Numbers (UDI Numbers): 722064 ((01)00884838085282(21)258, (01)00884838085282(21)229, (01)00884838085282(21)291, (01)00884838085282(21)270, (01)00884838085282(21)64, (01)00884838085282(21)218, (01)00884838085282(21)335, (01)00884838085282(21)513, (01)00884838085282(21)297, (01)00884838085282(21)169, (01)00884838085282(21)479, (01)00884838085282(21)359, (01)00884838085282(21)220, (01)00884838085282(21)375, (01)00884838085282(21)173, (01)00884838085282(21)502, (01)00884838085282(21)157, (01)00884838085282(21)130, (01)00884838085282(21)60, (01)00884838085282(21)263, (01)00884838085282(21)264, (01)00884838085282(21)253, (01)00884838085282(21)254, (01)00884838085282(21)280, (01)00884838085282(21)102, (01)00884838085282(21)461, (01)00884838085282(21)510, (01)00884838085282(21)345, (01)00884838085282(21)185, (01)00884838085282(21)193, (01)00884838085282(21)271, (01)00884838085282(21)214, (01)00884838085282(21)212, (01)00884838085282(21)266, (01)00884838085282(21)213, (01)00884838085282(21)180, (01)00884838085282(21)431, (01)00884838085282(21)354, (01)00884838085282(21)374, (01)00884838085282(21)106, (01)00884838085282(21)58, (01)00884838085282(21)288, (01)00884838085282(21)231, (01)00884838085282(21)287, (01)00884838085282(21)150, (01)00884838085282(21)380, (01)00884838085282(21)233, (01)00884838085282(21)230, (01)00884838085282(21)44, (01)00884838085282(21)197, (01)00884838085282(21)176, (01)00884838085282(21)149, (01)00884838085282(21)289, (01)00884838085282(21)448, (01)00884838085282(21)82, (01)00884838085282(21)323, (01)00884838085282(21)35, (01)00884838085282(21)55, (01)00884838085282(21)45, (01)00884838085282(21)40, (01)00884838085282(21)125, (01)00884838085282(21)484, (01)00884838085282(21)36, (01)00884838085282(21)104, (01)00884838085282(21)155, (01)00884838085282(21)422, (01)00884838085282(21)132, (01)00884838085282(21)67, (01)00884838085282(21)439, (01)00884838085282(21)446, (01)00884838085282(21)138, (01)00884838085282(21)293, (01)00884838085282(21)292, (01)00884838085282(21)379, (01)00884838085282(21)321, (01)00884838085282(21)520, (01)00884838085282(21)519, (01)00884838085282(21)208, (01)00884838085282(21)462, (01)00884838085282(21)342, (01)00884838085282(21)222, (01)00884838085282(21)391, (01)00884838085282(21)98, (01)00884838085282(21)441, (01)00884838085282(21)320, (01)00884838085282(21)494, (01)00884838085282(21)87, (01)00884838085282(21)109, (01)00884838085282(21)311, (01)00884838085282(21)442, (01)00884838085282(21)485, (01)00884838085282(21)121, (01)00884838085282(21)68, (01)00884838085282(21)384, (01)00884838085282(21)228, (01)00884838085282(21)91, (01)00884838085282(21)252, (01)00884838085282(21)73, (01)00884838085282(21)74, (01)00884838085282(21)124, (01)00884838085282(21)191, (01)00884838085282(21)122, (01)00884838085282(21)421, (01)00884838085282(21)400, (01)00884838085282(21)480, (01)00884838085282(21)387, (01)00884838085282(21)11, (01)00884838085282(21)48, (01)00884838085282(21)147, (01)00884838085282(21)509, (01)00884838085282(21)454, (01)00884838085282(21)478, (01)00884838085282(21)262, (01)00884838085282(21)491, (01)00884838085282(21)133, (01)00884838085282(21)337, (01)00884838085282(21)275, (01)00884838085282(21)468, (01)00884838085282(21)412, (01)00884838085282(21)414, (01)00884838085282(21)405, (01)00884838085282(21)396, (01)00884838085282(21)247, (01)00884838085282(21)310, (01)00884838085282(21)406, (01)00884838085282(21)517, (01)00884838085282(21)432, (01)00884838085282(21)504, (01)00884838085282(21)457, (01)00884838085282(21)426, (01)00884838085282(21)304, (01)00884838085282(21)469, (01)00884838085282(21)340, (01)00884838085282(21)393, (01)00884838085282(21)343, (01)00884838085282(21)458, (01)00884838085282(21)444, (01)00884838085282(21)508, (01)00884838085282(21)445, (01)00884838085282(21)395, (01)00884838085282(21)385, (01)00884838085282(21)514, (01)00884838085282(21)236, (01)00884838085282(21)471, (01)00884838085282(21)401, (01)00884838085282(21)234, (01)00884838085282(21)503, (01)00884838085282(21)394, (01)00884838085282(21)407, (01)00884838085282(21)470, (01)00884838085282(21)498, (01)00884838085282(21)413, (01)00884838085282(21)440, (01)00884838085282(21)463, (01)00884838085282(21)330, (01)00884838085282(21)475, (01)00884838085282(21)473, (01)00884838085282(21)308, (01)00884838085282(21)225, (01)00884838085282(21)483, (01)00884838085282(21)474, (01)00884838085282(21)344, (01)00884838085282(21)389, (01)00884838085282(21)492, (01)00884838085282(21)324, (01)00884838085282(21)265, (01)00884838085282(21)235, (01)00884838085282(21)496, (01)00884838085282(21)427, (01)00884838085282(21)341, (01)00884838085282(21)118, (01)00884838085282(21)174, (01)00884838085282(21)452, (01)00884838085282(21)168, (01)00884838085282(21)447, (01)00884838085282(21)198, (01)00884838085282(21)312, (01)00884838085282(21)403, (01)00884838085282(21)386, (01)00884838085282(21)397, (01)00884838085282(21)500, (01)00884838085282(21)226, (01)00884838085282(21)408, (01)00884838085282(21)43, (01)00884838085282(21)488, (01)00884838085282(21)416, (01)00884838085282(21)12, (01)00884838085282(21)285, (01)00884838085282(21)515, (01)00884838085282(21)476, (01)00884838085282(21)449, (01)00884838085282(21)477, (01)00884838085282(21)338, (01)00884838085282(21)215, (01)00884838085282(21)144, (01)00884838085282(21)227, (01)00884838085282(21)295, (01)00884838085282(21)294, (01)00884838085282(21)319, (01)00884838085282(21)126, (01)00884838085282(21)92, (01)00884838085282(21)175, (01)00884838085282(21)372, (01)00884838085282(21)429, (01)00884838085282(21)283, (01)00884838085282(21)357, (01)00884838085282(21)437, (01)00884838085282(21)315, (01)00884838085282(21)134, (01)00884838085282(21)255, (01)00884838085282(21)192, (01)00884838085282(21)116, (01)00884838085282(21)221, (01)00884838085282(21)117, (01)00884838085282(21)96, (01)00884838085282(21)135, (01)00884838085282(21)72), 722222 ((01)00884838099210(21)254, (01)00884838099210(21)226, (01)00884838099210(21)55, (01)00884838099210(21)44, (01)00884838099210(21)262, (01)00884838099210(21)37, (01)00884838099210(21)318, (01)00884838099210(21)269, (01)00884838099210(21)316, (01)00884838099210(21)307, (01)00884838099210(21)336, (01)00884838099210(21)174, (01)00884838099210(21)230, (01)00884838099210(21)182, (01)00884838099210(21)312, (01)00884838099210(21)172, (01)00884838099210(21)151, (01)00884838099210(21)138, (01)00884838099210(21)277, (01)00884838099210(21)224, (01)00884838099210(21)118, (01)00884838099210(21)119, (01)00884838099210(21)102, (01)00884838099210(21)88, (01)00884838099210(21)209, (01)00884838099210(21)145, (01)00884838099210(21)210, (01)00884838099210(21)89, (01)00884838099210(21)108, (01)00884838099210(21)132, (01)00884838099210(21)48, (01)00884838099210(21)4, (01)00884838099210(21)305, (01)00884838099210(21)80, (01)00884838099210(21)83, (01)00884838099210(21)47, (01)00884838099210(21)134, (01)00884838099210(21)222, (01)00884838099210(21)266, (01)00884838099210(21)245, (01)00884838099210(21)177, (01)00884838099210(21)11, (01)00884838099210(21)9, (01)00884838099210(21)66, (01)00884838099210(21)50, (01)00884838099210(21)353, (01)00884838099210(21)278, (01)00884838099210(21)82, (01)00884838099210(21)76, (01)00884838099210(21)319, (01)00884838099210(21)193, (01)00884838099210(21)106, (01)00884838099210(21)178, (01)00884838099210(21)117, (01)00884838099210(21)213, (01)00884838099210(21)215, (01)00884838099210(21)104, (01)00884838099210(21)141, (01)00884838099210(21)142, (01)00884838099210(21)252, (01)00884838099210(21)250, (01)00884838099210(21)251, (01)00884838099210(21)183, (01)00884838099210(21)311, (01)00884838099210(21)228, (01)00884838099210(21)322, (01)00884838099210(21)189, (01)00884838099210(21)297, (01)00884838099210(21)190, (01)00884838099210(21)313, (01)00884838099210(21)169, (01)00884838099210(21)219, (01)00884838099210(21)247, (01)00884838099210(21)282, (01)00884838099210(21)217, (01)00884838099210(21)143, (01)00884838099210(21)349, (01)00884838099210(21)225, (01)00884838099210(21)162, (01)00884838099210(21)333, (01)00884838099210(21)185, (01)00884838099210(21)166, (01)00884838099210(21)161, (01)00884838099210(21)16, (01)00884838099210(21)115, (01)00884838099210(21)54, (01)00884838099210(21)173, (01)00884838099210(21)90, (01)00884838099210(21)292, (01)00884838099210(21)291, (01)00884838099210(21)290, (01)00884838099210(21)295, (01)00884838099210(21)288, (01)00884838099210(21)289, (01)00884838099210(21)294, (01)00884838099210(21)186, (01)00884838099210(21)181, (01)00884838099210(21)296, (01)00884838099210(21)315, (01)00884838099210(21)184, (01)00884838099210(21)180, (01)00884838099210(21)284, (01)00884838099210(21)214, (01)00884838099210(21)116, (01)00884838099210(21)276, (01)00884838099210(21)73, (01)00884838099210(21)157, (01)00884838099210(21)114, (01)00884838099210(21)135, (01)00884838099210(21)159, (01)00884838099210(21)158, (01)00884838099210(21)78, (01)00884838099210(21)125, (01)00884838099210(21)218, (01)00884838099210(21)126), 722280 ((01)00884838103276(21)57, (01)00884838103276(21)43, (01)00884838103276(21)179, (01)00884838103276(21)65, (01)00884838103276(21)76, (01)00884838103276(21)167, (01)00884838103276(21)165, (01)00884838103276(21)80, (01)00884838103276(21)128, (01)00884838103276(21)158, (01)00884838103276(21)149, (01)00884838103276(21)44, (01)00884838103276(21)100, (01)00884838103276(21)164, (01)00884838103276(21)66, (01)00884838103276(21)135, (01)00884838103276(21)175, (01)00884838103276(21)112, (01)00884838103276(21)34, (01)00884838103276(21)79, (01)00884838103276(21)130, (01)00884838103276(21)48, (01)00884838103276(21)52, (01)00884838103276(21)141, (01)00884838103276(21)131, (01)00884838103276(21)37, (01)00884838103276(21)83, (01)00884838103276(21)87, (01)00884838103276(21)90, (01)00884838103276(21)105, (01)00884838103276(21)23, (01)00884838103276(21)30, (01)00884838103276(21)96, (01)00884838103276(21)107, (01)00884838103276(21)171, (01)00884838103276(21)118, (01)00884838103276(21)21, (01)00884838103276(21)147, (01)00884838103276(21)133, (01)00884838103276(21)38, (01)00884838103276(21)150, (01)00884838103276(21)69, (01)00884838103276(21)148, (01)00884838103276(21)98, (01)00884838103276(21)127, (01)00884838103276(21)144, (01)00884838103276(21)143, (01)00884838103276(21)142, (01)00884838103276(21)71, (01)00884838103276(21)146, (01)00884838103276(21)152, (01)00884838103276(21)40, (01)00884838103276(21)17, (01)00884838103276(21)187, (01)00884838103276(21)46, (01)00884838103276(21)91, (01)00884838103276(21)97, (01)00884838103276(21)106, (01)00884838103276(21)120, (01)00884838103276(21)45, (01)00884838103276(21)92, (01)00884838103276(21)54, (01)00884838103276(21)126, (01)00884838103276(21)19, (01)00884838103276(21)93, (01)00884838103276(21)24, (01)00884838103276(21)129, (01)00884838103276(21)74, (01)00884838103276(21)88, (01)00884838103276(21)85, (01)00884838103276(21)170, (01)00884838103276(21)102, (01)00884838103276(21)124, (01)00884838103276(21)123, (01)00884838103276(21)125, (01)00884838103276(21)151, (01)00884838103276(21)39, (01)00884838103276(21)101, (01)00884838103276(21)41, (01)00884838103276(21)42, (01)00884838103276(21)53, (01)00884838103276(21)108, (01)00884838103276(21)119, (01)00884838103276(21)82, (01)00884838103276(21)33, (01)00884838103276(21)20, (01)00884838103276(21)63, (01)00884838103276(21)84, (01)00884838103276(21)145, (01)00884838103276(21)36, (01)00884838103276(21)59, (01)00884838103276(21)50, (01)00884838103276(21)110, (01)00884838103276(21)140, (01)00884838103276(21)117, (01)00884838103276(21)49, (01)00884838103276(21)60, (01)00884838103276(21)103, (01)00884838103276(21)72, (01)00884838103276(21)160, (01)00884838103276(21)81, (01)00884838103276(21)114, (01)00884838103276(21)137, (01)00884838103276(21)58, (01)00884838103276(21)73, (01)00884838103276(21)35, (01)00884838103276(21)22, (01)00884838103276(21)94).
3. Model Numbers (UDI Numbers): 722227 ((01)00884838099227(21)135, (01)00884838099227(21)88, (01)00884838099227(21)127, (01)00884838099227(21)128, (01)00884838099227(21)179, (01)00884838099227(21)246, (01)00884838099227(21)84, (01)00884838099227(21)29, (01)00884838099227(21)108, (01)00884838099227(21)62, (01)00884838099227(21)25, (01)00884838099227(21)104, (01)00884838099227(21)109, (01)00884838099227(21)249, (01)00884838099227(21)215, (01)00884838099227(21)167, (01)00884838099227(21)83, (01)00884838099227(21)195, (01)00884838099227(21)116, (01)00884838099227(21)153, (01)00884838099227(21)262, (01)00884838099227(21)113, (01)00884838099227(21)110, (01)00884838099227(21)219, (01)00884838099227(21)256, (01)00884838099227(21)90, (01)00884838099227(21)69, (01)00884838099227(21)200, (01)00884838099227(21)242, (01)00884838099227(21)136, (01)00884838099227(21)137, (01)00884838099227(21)294, (01)00884838099227(21)95, (01)00884838099227(21)165, (01)00884838099227(21)16, (01)00884838099227(21)12, (01)00884838099227(21)257, (01)00884838099227(21)248, (01)00884838099227(21)106, (01)00884838099227(21)26, (01)00884838099227(21)44, (01)00884838099227(21)130, (01)00884838099227(21)58, (01)00884838099227(21)125, (01)00884838099227(21)184, (01)00884838099227(21)57, (01)00884838099227(21)259, (01)00884838099227(21)131).
4. Model Numbers (UDI Numbers): 722228 ((01)00884838099234(21)49, (01)00884838099234(21)159, (01)00884838099234(21)273, (01)00884838099234(21)132, (01)00884838099234(21)47, (01)00884838099234(21)87, (01)00884838099234(21)38, (01)00884838099234(21)32, (01)00884838099234(21)25, (01)00884838099234(21)23, (01)00884838099234(21)62, (01)00884838099234(21)48, (01)00884838099234(21)153, (01)00884838099234(21)431, (01)00884838099234(21)113, (01)00884838099234(21)127, (01)00884838099234(21)91, (01)00884838099234(21)529, (01)00884838099234(21)145, (01)00884838099234(21)539, (01)00884838099234(21)308, (01)00884838099234(21)335, (01)00884838099234(21)333, (01)00884838099234(21)44, (01)00884838099234(21)418, (01)00884838099234(21)525, (01)00884838099234(21)280, (01)00884838099234(21)387, (01)00884838099234(21)471, (01)00884838099234(21)428, (01)00884838099234(21)465, (01)00884838099234(21)103, (01)00884838099234(21)176, (01)00884838099234(21)102, (01)00884838099234(21)178, (01)00884838099234(21)109, (01)00884838099234(21)16, (01)00884838099234(21)234, (01)00884838099234(21)463, (01)00884838099234(21)71, (01)00884838099234(21)317, (01)00884838099234(21)517, (01)00884838099234(21)203, (01)00884838099234(21)260, (01)00884838099234(21)107, (01)00884838099234(21)101, (01)00884838099234(21)99, (01)00884838099234(21)100, (01)00884838099234(21)108, (01)00884838099234(21)318, (01)00884838099234(21)75, (01)00884838099234(21)78, (01)00884838099234(21)531, (01)00884838099234(21)370, (01)00884838099234(21)464, (01)00884838099234(21)138, (01)00884838099234(21)496, (01)00884838099234(21)312, (01)00884838099234(21)223, (01)00884838099234(21)218, (01)00884838099234(21)513, (01)00884838099234(21)442, (01)00884838099234(21)174, (01)00884838099234(21)548, (01)00884838099234(21)84, (01)00884838099234(21)454, (01)00884838099234(21)346, (01)00884838099234(21)327, (01)00884838099234(21)196, (01)00884838099234(21)415, (01)00884838099234(21)118, (01)00884838099234(21)452, (01)00884838099234(21)179, (01)00884838099234(21)331, (01)00884838099234(21)297, (01)00884838099234(21)503, (01)00884838099234(21)412, (01)00884838099234(21)427, (01)00884838099234(21)488, (01)00884838099234(21)500, (01)00884838099234(21)365, (01)00884838099234(21)443, (01)00884838099234(21)90, (01)00884838099234(21)498, (01)00884838099234(21)284, (01)00884838099234(21)201, (01)00884838099234(21)171, (01)00884838099234(21)229, (01)00884838099234(21)332, (01)00884838099234(21)501, (01)00884838099234(21)306, (01)00884838099234(21)429, (01)00884838099234(21)56, (01)00884838099234(21)257, (01)00884838099234(21)430, (01)00884838099234(21)360, (01)00884838099234(21)352, (01)00884838099234(21)329, (01)00884838099234(21)197, (01)00884838099234(21)366, (01)00884838099234(21)345, (01)00884838099234(21)302, (01)00884838099234(21)166, (01)00884838099234(21)432, (01)00884838099234(21)361, (01)00884838099234(21)305, (01)00884838099234(21)542, (01)00884838099234(21)313, (01)00884838099234(21)220, (01)00884838099234(21)325, (01)00884838099234(21)405, (01)00884838099234(21)433, (01)00884838099234(21)342, (01)00884838099234(21)341, (01)00884838099234(21)184, (01)00884838099234(21)424, (01)00884838099234(21)425, (01)00884838099234(21)426, (01)00884838099234(21)497, (01)00884838099234(21)258, (01)00884838099234(21)183, (01)00884838099234(21)195, (01)00884838099234(21)314, (01)00884838099234(21)282, (01)00884838099234(21)402, (01)00884838099234(21)439, (01)00884838099234(21)421, (01)00884838099234(21)440, (01)00884838099234(21)446, (01)00884838099234(21)472, (01)00884838099234(21)320, (01)00884838099234(21)281, (01)00884838099234(21)480, (01)00884838099234(21)364, (01)00884838099234(21)182, (01)00884838099234(21)448, (01)00884838099234(21)283, (01)00884838099234(21)291, (01)00884838099234(21)473, (01)00884838099234(21)321, (01)00884838099234(21)476, (01)00884838099234(21)409, (01)00884838099234(21)376, (01)00884838099234(21)384, (01)00884838099234(21)534, (01)00884838099234(21)362, (01)00884838099234(21)489, (01)00884838099234(21)303, (01)00884838099234(21)413, (01)00884838099234(21)512, (01)00884838099234(21)404, (01)00884838099234(21)437, (01)00884838099234(21)336, (01)00884838099234(21)200, (01)00884838099234(21)344, (01)00884838099234(21)504, (01)00884838099234(21)94, (01)00884838099234(21)494, (01)00884838099234(21)408, (01)00884838099234(21)483, (01)00884838099234(21)449, (01)00884838099234(21)337, (01)00884838099234(21)438, (01)00884838099234(21)453, (01)00884838099234(21)106, (01)00884838099234(21)311, (01)00884838099234(21)324, (01)00884838099234(21)459, (01)00884838099234(21)477, (01)00884838099234(21)330, (01)00884838099234(21)386, (01)00884838099234(21)288, (01)00884838099234(21)167, (01)00884838099234(21)390, (01)00884838099234(21)261, (01)00884838099234(21)334, (01)00884838099234(21)304, (01)00884838099234(21)350, (01)00884838099234(21)212, (01)00884838099234(21)340, (01)00884838099234(21)515, (01)00884838099234(21)292, (01)00884838099234(21)173, (01)00884838099234(21)543, (01)00884838099234(21)217, (01)00884838099234(21)160, (01)00884838099234(21)82, (01)00884838099234(21)133, (01)00884838099234(21)208, (01)00884838099234(21)165, (01)00884838099234(21)378, (01)00884838099234(21)164, (01)00884838099234(21)228, (01)00884838099234(21)401, (01)00884838099234(21)161, (01)00884838099234(21)147, (01)00884838099234(21)146, (01)00884838099234(21)170, (01)00884838099234(21)295, (01)00884838099234(21)290, (01)00884838099234(21)316, (01)00884838099234(21)207), 722281 ((01)00884838110564(21)55, (01)00884838110564(21)21, (01)00884838110564(21)28, (01)00884838110564(21)27, (01)00884838110564(21)60, (01)00884838110564(21)24, (01)00884838110564(21)48, (01)00884838110564(21)17, (01)00884838110564(21)124, (01)00884838110564(21)64, (01)00884838110564(21)54, (01)00884838110564(21)19, (01)00884838110564(21)26, (01)00884838110564(21)80, (01)00884838110564(21)11, (01)00884838110564(21)9, (01)00884838110564(21)35, (01)00884838110564(21)76, (01)00884838110564(21)294, (01)00884838110564(21)229, (01)00884838110564(21)14, (01)00884838110564(21)33, (01)00884838110564(21)10, (01)00884838110564(21)284, (01)00884838110564(21)57, (01)00884838110564(21)50, (01)00884838110564(21)13, (01)00884838110564(21)34, (01)00884838110564(21)300, (01)00884838110564(21)231, (01)00884838110564(21)75, (01)00884838110564(21)22, (01)00884838110564(21)29, (01)00884838110564(21)31, (01)00884838110564(21)30, (01)00884838110564(21)36).
5. Model Numbers (UDI Numbers): 722067 ((01)00884838085350(21)42, (01)00884838085350(21)221, (01)00884838085350(21)278, (01)00884838085350(21)141, (01)00884838085350(21)133, (01)00884838085350(21)233, (01)00884838085350(21)146, (01)00884838085350(21)256, (01)00884838085350(21)270, (01)00884838085350(21)238, (01)00884838085350(21)187, (01)00884838085350(21)154, (01)00884838085350(21)157, (01)00884838085350(21)251, (01)00884838085350(21)222, (01)00884838085350(21)262, (01)00884838085350(21)29, (01)00884838085350(21)59, (01)00884838085350(21)209, (01)00884838085350(21)260, (01)00884838085350(21)95, (01)00884838085350(21)274, (01)00884838085350(21)218, (01)00884838085350(21)231, (01)00884838085350(21)169, (01)00884838085350(21)275, (01)00884838085350(21)265, (01)00884838085350(21)190, (01)00884838085350(21)203, (01)00884838085350(21)138, (01)00884838085350(21)53, (01)00884838085350(21)140, (01)00884838085350(21)266, (01)00884838085350(21)235, (01)00884838085350(21)272, (01)00884838085350(21)58, (01)00884838085350(21)47, (01)00884838085350(21)273, (01)00884838085350(21)64, (01)00884838085350(21)276, (01)00884838085350(21)232, (01)00884838085350(21)271, (01)00884838085350(21)35, (01)00884838085350(21)207, (01)00884838085350(21)226, (01)00884838085350(21)102, (01)00884838085350(21)105, (01)00884838085350(21)159, (01)00884838085350(21)170, (01)00884838085350(21)193, (01)00884838085350(21)118, (01)00884838085350(21)117, (01)00884838085350(21)85, (01)00884838085350(21)119), 722225 ((01)00884838099265(21)110, (01)00884838099265(21)37, (01)00884838099265(21)157, (01)00884838099265(21)284, (01)00884838099265(21)232, (01)00884838099265(21)230, (01)00884838099265(21)269, (01)00884838099265(21)345, (01)00884838099265(21)296, (01)00884838099265(21)374, (01)00884838099265(21)246, (01)00884838099265(21)128, (01)00884838099265(21)241, (01)00884838099265(21)180, (01)00884838099265(21)367, (01)00884838099265(21)292, (01)00884838099265(21)145, (01)00884838099265(21)306, (01)00884838099265(21)338, (01)00884838099265(21)314, (01)00884838099265(21)83, (01)00884838099265(21)422, (01)00884838099265(21)267, (01)00884838099265(21)223, (01)00884838099265(21)173, (01)00884838099265(21)8, (01)00884838099265(21)301, (01)00884838099265(21)146, (01)00884838099265(21)276, (01)00884838099265(21)109, (01)00884838099265(21)189, (01)00884838099265(21)200, (01)00884838099265(21)372, (01)00884838099265(21)312, (01)00884838099265(21)320, (01)00884838099265(21)393, (01)00884838099265(21)206, (01)00884838099265(21)327, (01)00884838099265(21)96, (01)00884838099265(21)446, (01)00884838099265(21)302, (01)00884838099265(21)242, (01)00884838099265(21)187, (01)00884838099265(21)169, (01)00884838099265(21)135, (01)00884838099265(21)134, (01)00884838099265(21)395, (01)00884838099265(21)116, (01)00884838099265(21)107, (01)00884838099265(21)133, (01)00884838099265(21)313, (01)00884838099265(21)346, (01)00884838099265(21)351, (01)00884838099265(21)349, (01)00884838099265(21)1, (01)00884838099265(21)234, (01)00884838099265(21)218, (01)00884838099265(21)342, (01)00884838099265(21)396, (01)00884838099265(21)348, (01)00884838099265(21)386, (01)00884838099265(21)213, (01)00884838099265(21)125, (01)00884838099265(21)190, (01)00884838099265(21)193, (01)00884838099265(21)341, (01)00884838099265(21)61, (01)00884838099265(21)221, (01)00884838099265(21)154, (01)00884838099265(21)67, (01)00884838099265(21)278, (01)00884838099265(21)162, (01)00884838099265(21)42).
6. Model Numbers (UDI Numbers): 722068 ((01)00884838085367(21)335, (01)00884838085367(21)205, (01)00884838085367(21)494, (01)00884838085367(21)564, (01)00884838085367(21)247, (01)00884838085367(21)85, (01)00884838085367(21)505, (01)00884838085367(21)271, (01)00884838085367(21)443, (01)00884838085367(21)354, (01)00884838085367(21)446, (01)00884838085367(21)73, (01)00884838085367(21)406, (01)00884838085367(21)528, (01)00884838085367(21)350, (01)00884838085367(21)462, (01)00884838085367(21)520, (01)00884838085367(21)469, (01)00884838085367(21)114, (01)00884838085367(21)477, (01)00884838085367(21)107, (01)00884838085367(21)282, (01)00884838085367(21)276, (01)00884838085367(21)608, (01)00884838085367(21)473, (01)00884838085367(21)302, (01)00884838085367(21)126, (01)00884838085367(21)468, (01)00884838085367(21)454, (01)00884838085367(21)410, (01)00884838085367(21)231, (01)00884838085367(21)548, (01)00884838085367(21)495, (01)00884838085367(21)167, (01)00884838085367(21)537, (01)00884838085367(21)289, (01)00884838085367(21)233, (01)00884838085367(21)173, (01)00884838085367(21)536, (01)00884838085367(21)204, (01)00884838085367(21)237, (01)00884838085367(21)133, (01)00884838085367(21)307, (01)00884838085367(21)577, (01)00884838085367(21)210, (01)00884838085367(21)258, (01)00884838085367(21)199, (01)00884838085367(21)480, (01)00884838085367(21)423, (01)00884838085367(21)509, (01)00884838085367(21)246, (01)00884838085367(21)614, (01)00884838085367(21)257, (01)00884838085367(21)369, (01)00884838085367(21)508, (01)00884838085367(21)97, (01)00884838085367(21)303, (01)00884838085367(21)48, (01)00884838085367(21)44, (01)00884838085367(21)550, (01)00884838085367(21)460, (01)00884838085367(21)551, (01)00884838085367(21)478, (01)00884838085367(21)522, (01)00884838085367(21)238, (01)00884838085367(21)360, (01)00884838085367(21)192, (01)00884838085367(21)447, (01)00884838085367(21)570, (01)00884838085367(21)148, (01)00884838085367(21)560, (01)00884838085367(21)318, (01)00884838085367(21)518, (01)00884838085367(21)206, (01)00884838085367(21)521, (01)00884838085367(21)534, (01)00884838085367(21)292, (01)00884838085367(21)591, (01)00884838085367(21)359, (01)00884838085367(21)387, (01)00884838085367(21)128, (01)00884838085367(21)129, (01)00884838085367(21)396, (01)00884838085367(21)334, (01)00884838085367(21)425, (01)00884838085367(21)471, (01)00884838085367(21)540, (01)00884838085367(21)584, (01)00884838085367(21)553, (01)00884838085367(21)602, (01)00884838085367(21)240, (01)00884838085367(21)617, (01)00884838085367(21)604, (01)00884838085367(21)593, (01)00884838085367(21)589, (01)00884838085367(21)526, (01)00884838085367(21)559, (01)00884838085367(21)612, (01)00884838085367(21)556, (01)00884838085367(21)557, (01)00884838085367(21)574, (01)00884838085367(21)603, (01)00884838085367(21)601, (01)00884838085367(21)609, (01)00884838085367(21)616, (01)00884838085367(21)597, (01)00884838085367(21)605, (01)00884838085367(21)575.(01)00884838085367(21)516, (01)00884838085367(21)624, (01)00884838085367(21)599, (01)00884838085367(21)563, (01)00884838085367(21)590, (01)00884838085367(21)600, (01)00884838085367(21)475, (01)00884838085367(21)349, (01)00884838085367(21)519, (01)00884838085367(21)234, (01)00884838085367(21)195, (01)00884838085367(21)532, (01)00884838085367(21)533, (01)00884838085367(21)398, (01)00884838085367(21)618, (01)00884838085367(21)549, (01)00884838085367(21)249, (01)00884838085367(21)531, (01)00884838085367(21)290, (01)00884838085367(21)326, (01)00884838085367(21)286, (01)00884838085367(21)113, (01)00884838085367(21)124, (01)00884838085367(21)362, (01)00884838085367(21)317, (01)00884838085367(21)122, (01)00884838085367(21)100, (01)00884838085367(21)105, (01)00884838085367(21)315, (01)00884838085367(21)436), 722226 ((01)00884838099272(21)209, (01)00884838099272(21)675, (01)00884838099272(21)113, (01)00884838099272(21)593, (01)00884838099272(21)687, (01)00884838099272(21)280, (01)00884838099272(21)884, (01)00884838099272(21)98, (01)00884838099272(21)343, (01)00884838099272(21)180, (01)00884838099272(21)554, (01)00884838099272(21)337, (01)00884838099272(21)400, (01)00884838099272(21)757, (01)00884838099272(21)326, (01)00884838099272(21)262, (01)00884838099272(21)467, (01)00884838099272(21)379, (01)00884838099272(21)398, (01)00884838099272(21)786, (01)00884838099272(21)462, (01)00884838099272(21)458, (01)00884838099272(21)880, (01)00884838099272(21)693, (01)00884838099272(21)510, (01)00884838099272(21)882, (01)00884838099272(21)21, (01)00884838099272(21)367, (01)00884838099272(21)184, (01)00884838099272(21)179, (01)00884838099272(21)250, (01)00884838099272(21)178, (01)00884838099272(21)674, (01)00884838099272(21)575, (01)00884838099272(21)368, (01)00884838099272(21)950, (01)00884838099272(21)712, (01)00884838099272(21)829, (01)00884838099272(21)413, (01)00884838099272(21)249, (01)00884838099272(21)576, (01)00884838099272(21)738, (01)00884838099272(21)708, (01)00884838099272(21)717, (01)00884838099272(21)697, (01)00884838099272(21)665, (01)00884838099272(21)764, (01)00884838099272(21)386, (01)00884838099272(21)721, (01)00884838099272(21)232, (01)00884838099272(21)789, (01)00884838099272(21)682, (01)00884838099272(21)698, (01)00884838099272(21)331, (01)00884838099272(21)378, (01)00884838099272(21)808, (01)00884838099272(21)949, (01)00884838099272(21)718, (01)00884838099272(21)663, (01)00884838099272(21)195, (01)00884838099272(21)412, (01)00884838099272(21)715, (01)00884838099272(21)775, (01)00884838099272(21)282, (01)00884838099272(21)617, (01)00884838099272(21)958, (01)00884838099272(21)221, (01)00884838099272(21)222, (01)00884838099272(21)604, (01)00884838099272(21)430, (01)00884838099272(21)810, (01)00884838099272(21)314, (01)00884838099272(21)561, (01)00884838099272(21)762, (01)00884838099272(21)359, (01)00884838099272(21)978, (01)00884838099272(21)836, (01)00884838099272(21)835, (01)00884838099272(21)795, (01)00884838099272(21)671, (01)00884838099272(21)800, (01)00884838099272(21)799, (01)00884838099272(21)899, (01)00884838099272(21)838, (01)00884838099272(21)335, (01)00884838099272(21)1039, (01)00884838099272(21)203, (01)00884838099272(21)731, (01)00884838099272(21)725, (01)00884838099272(21)701, (01)00884838099272(21)967, (01)00884838099272(21)501, (01)00884838099272(21)119, (01)00884838099272(21)927, (01)00884838099272(21)732, (01)00884838099272(21)974, (01)00884838099272(21)919, (01)00884838099272(21)605, (01)00884838099272(21)686, (01)00884838099272(21)230, (01)00884838099272(21)523, (01)00884838099272(21)327, (01)00884838099272(21)477, (01)00884838099272(21)298, (01)00884838099272(21)521, (01)00884838099272(21)522, (01)00884838099272(21)520, (01)00884838099272(21)259, (01)00884838099272(21)142, (01)00884838099272(21)263, (01)00884838099272(21)422, (01)00884838099272(21)17, (01)00884838099272(21)36, (01)00884838099272(21)164, (01)00884838099272(21)752, (01)00884838099272(21)47, (01)00884838099272(21)544, (01)00884838099272(21)906, (01)00884838099272(21)289, (01)00884838099272(21)238, (01)00884838099272(21)299, (01)00884838099272(21)404, (01)00884838099272(21)284, (01)00884838099272(21)283, (01)00884838099272(21)401, (01)00884838099272(21)833, (01)00884838099272(21)924, (01)00884838099272(21)224, (01)00884838099272(21)233, (01)00884838099272(21)579, (01)00884838099272(21)395, (01)00884838099272(21)444, (01)00884838099272(21)710, (01)00884838099272(21)820, (01)00884838099272(21)235, (01)00884838099272(21)727, (01)00884838099272(21)621, (01)00884838099272(21)777, (01)00884838099272(21)486, (01)00884838099272(21)481, (01)00884838099272(21)431, (01)00884838099272(21)765, (01)00884838099272(21)696, (01)00884838099272(21)451, (01)00884838099272(21)438, (01)00884838099272(21)630, (01)00884838099272(21)854, (01)00884838099272(21)888, (01)00884838099272(21)596, (01)00884838099272(21)645, (01)00884838099272(21)739, (01)00884838099272(21)736, (01)00884838099272(21)963, (01)00884838099272(21)527, (01)00884838099272(21)251, (01)00884838099272(21)553, (01)00884838099272(21)574, (01)00884838099272(21)741, (01)00884838099272(21)748, (01)00884838099272(21)740, (01)00884838099272(21)901, (01)00884838099272(21)900, (01)00884838099272(21)887, (01)00884838099272(21)632, (01)00884838099272(21)1156, (01)00884838099272(21)737, (01)00884838099272(21)662, (01)00884838099272(21)825, (01)00884838099272(21)744, (01)00884838099272(21)877, (01)00884838099272(21)549, (01)00884838099272(21)648, (01)00884838099272(21)831, (01)00884838099272(21)751, (01)00884838099272(21)830, (01)00884838099272(21)860, (01)00884838099272(21)798, (01)00884838099272(21)603, (01)00884838099272(21)577, (01)00884838099272(21)886, (01)00884838099272(21)773, (01)00884838099272(21)716, (01)00884838099272(21)672, (01)00884838099272(21)903, (01)00884838099272(21)669, (01)00884838099272(21)747, (01)00884838099272(21)813, (01)00884838099272(21)580, (01)00884838099272(21)896, (01)00884838099272(21)636, (01)00884838099272(21)742, (01)00884838099272(21)902, (01)00884838099272(21)867, (01)00884838099272(21)872, (01)00884838099272(21)418, (01)00884838099272(21)745, (01)00884838099272(21)815, (01)00884838099272(21)953, (01)00884838099272(21)729, (01)00884838099272(21)650, (01)00884838099272(21)844, (01)00884838099272(21)276, (01)00884838099272(21)855, (01)00884838099272(21)68, (01)00884838099272(21)870, (01)00884838099272(21)811, (01)00884838099272(21)236, (01)00884838099272(21)652).
7. Model Numbers (UDI Numbers): 722078 ((01)00884838085251(21)585, (01)00884838085251(21)613, (01)00884838085251(21)224, (01)00884838085251(21)479, (01)00884838085251(21)278, (01)00884838085251(21)921, (01)00884838085251(21)165, (01)00884838085251(21)657, (01)00884838085251(21)576, (01)00884838085251(21)645, (01)00884838085251(21)379, (01)00884838085251(21)166, (01)00884838085251(21)258, (01)00884838085251(21)257, (01)00884838085251(21)143, (01)00884838085251(21)661, (01)00884838085251(21)520, (01)00884838085251(21)535, (01)00884838085251(21)170, (01)00884838085251(21)298, (01)00884838085251(21)725, (01)00884838085251(21)728, (01)00884838085251(21)906, (01)00884838085251(21)603, (01)00884838085251(21)719, (01)00884838085251(21)163, (01)00884838085251(21)271, (01)00884838085251(21)412, (01)00884838085251(21)748, (01)00884838085251(21)309, (01)00884838085251(21)635, (01)00884838085251(21)232, (01)00884838085251(21)612, (01)00884838085251(21)986, (01)00884838085251(21)135, (01)00884838085251(21)589, (01)00884838085251(21)738, (01)00884838085251(21)343, (01)00884838085251(21)344, (01)00884838085251(21)943, (01)00884838085251(21)898, (01)00884838085251(21)482, (01)00884838085251(21)693, (01)00884838085251(21)694, (01)00884838085251(21)265, (01)00884838085251(21)511, (01)00884838085251(21)746, (01)00884838085251(21)266, (01)00884838085251(21)178, (01)00884838085251(21)179, (01)00884838085251(21)526, (01)00884838085251(21)264, (01)00884838085251(21)287, (01)00884838085251(21)1003, (01)00884838085251(21)1036, (01)00884838085251(21)934, (01)00884838085251(21)929, (01)00884838085251(21)836, (01)00884838085251(21)835, (01)00884838085251(21)975, (01)00884838085251(21)681, (01)00884838085251(21)1071, (01)00884838085251(21)1070, (01)00884838085251(21)1069, (01)00884838085251(21)1072, (01)00884838085251(21)1061, (01)00884838085251(21)849, (01)00884838085251(21)1058, (01)00884838085251(21)1068, (01)00884838085251(21)688, (01)00884838085251(21)20, (01)00884838085251(21)846, (01)00884838085251(21)424, (01)00884838085251(21)800, (01)00884838085251(21)156, (01)00884838085251(21)1039, (01)00884838085251(21)847, (01)00884838085251(21)702, (01)00884838085251(21)683, (01)00884838085251(21)541, (01)00884838085251(21)213, (01)00884838085251(21)1025, (01)00884838085251(21)578, (01)00884838085251(21)912, (01)00884838085251(21)504, (01)00884838085251(21)419, (01)00884838085251(21)855, (01)00884838085251(21)918, (01)00884838085251(21)919, (01)00884838085251(21)976, (01)00884838085251(21)27, (01)00884838085251(21)512, (01)00884838085251(21)718, (01)00884838085251(21)1037, (01)00884838085251(21)839, (01)00884838085251(21)354, (01)00884838085251(21)957, (01)00884838085251(21)471, (01)00884838085251(21)357, (01)00884838085251(21)952, (01)00884838085251(21)652, (01)00884838085251(21)1005, (01)00884838085251(21)349, (01)00884838085251(21)236, (01)00884838085251(21)831, (01)00884838085251(21)501, (01)00884838085251(21)841, (01)00884838085251(21)905, (01)00884838085251(21)682, (01)00884838085251(21)843, (01)00884838085251(21)830, (01)00884838085251(21)323, (01)00884838085251(21)365, (01)00884838085251(21)615, (01)00884838085251(21)667, (01)00884838085251(21)749, (01)00884838085251(21)407, (01)00884838085251(21)505, (01)00884838085251(21)408, (01)00884838085251(21)752, (01)00884838085251(21)138, (01)00884838085251(21)252, (01)00884838085251(21)426, (01)00884838085251(21)169, (01)00884838085251(21)650, (01)00884838085251(21)351, (01)00884838085251(21)442, (01)00884838085251(21)968, (01)00884838085251(21)750, (01)00884838085251(21)926, (01)00884838085251(21)597, (01)00884838085251(21)899, (01)00884838085251(21)978, (01)00884838085251(21)1033, (01)00884838085251(21)1035, (01)00884838085251(21)1009, (01)00884838085251(21)1008, (01)00884838085251(21)1047, (01)00884838085251(21)984, (01)00884838085251(21)989, (01)00884838085251(21)987, (01)00884838085251(21)907, (01)00884838085251(21)1081, (01)00884838085251(21)1004, (01)00884838085251(21)1086, (01)00884838085251(21)1085, (01)00884838085251(21)1084, (01)00884838085251(21)1048, (01)00884838085251(21)1059, (01)00884838085251(21)1034, (01)00884838085251(21)1080, (01)00884838085251(21)875, (01)00884838085251(21)1026, (01)00884838085251(21)1077, (01)00884838085251(21)717, (01)00884838085251(21)716, (01)00884838085251(21)1015, (01)00884838085251(21)1045, (01)00884838085251(21)945, (01)00884838085251(21)947, (01)00884838085251(21)942, (01)00884838085251(21)948, (01)00884838085251(21)940, (01)00884838085251(21)946, (01)00884838085251(21)1052, (01)00884838085251(21)971, (01)00884838085251(21)1062, (01)00884838085251(21)1063, (01)00884838085251(21)994, (01)00884838085251(21)701, (01)00884838085251(21)993, (01)00884838085251(21)1075, (01)00884838085251(21)1046, (01)00884838085251(21)970, (01)00884838085251(21)798, (01)00884838085251(21)1074, (01)00884838085251(21)856, (01)00884838085251(21)710, (01)00884838085251(21)1011, (01)00884838085251(21)1065, (01)00884838085251(21)1067, (01)00884838085251(21)1066, (01)00884838085251(21)1064, (01)00884838085251(21)1000, (01)00884838085251(21)1056, (01)00884838085251(21)992, (01)00884838085251(21)1057, (01)00884838085251(21)965, (01)00884838085251(21)964, (01)00884838085251(21)1043, (01)00884838085251(21)735, (01)00884838085251(21)736, (01)00884838085251(21)820, (01)00884838085251(21)819, (01)00884838085251(21)1049, (01)00884838085251(21)1073, (01)00884838085251(21)1079, (01)00884838085251(21)381, (01)00884838085251(21)593, (01)00884838085251(21)776, (01)00884838085251(21)775, (01)00884838085251(21)421, (01)00884838085251(21)760, (01)00884838085251(21)614, (01)00884838085251(21)864, (01)00884838085251(21)837, (01)00884838085251(21)458, (01)00884838085251(21)991, (01)00884838085251(21)28, (01)00884838085251(21)29, (01)00884838085251(21)363, (01)00884838085251(21)472, (01)00884838085251(21)782, (01)00884838085251(21)961, (01)00884838085251(21)643, (01)00884838085251(21)731, (01)00884838085251(21)447, (01)00884838085251(21)540, (01)00884838085251(21)619, (01)00884838085251(21)527, (01)00884838085251(21)892, (01)00884838085251(21)124, (01)00884838085251(21)164, (01)00884838085251(21)972, (01)00884838085251(21)1032, (01)00884838085251(21)1051, (01)00884838085251(21)575, (01)00884838085251(21)939, (01)00884838085251(21)869, (01)00884838085251(21)874, (01)00884838085251(21)871, (01)00884838085251(21)551, (01)00884838085251(21)1013, (01)00884838085251(21)685, (01)00884838085251(21)198), 722223 ((01)00884838099241(21)533, (01)00884838099241(21)189, (01)00884838099241(21)77, (01)00884838099241(21)439, (01)00884838099241(21)1088, (01)00884838099241(21)210, (01)00884838099241(21)206, (01)00884838099241(21)336, (01)00884838099241(21)87, (01)00884838099241(21)256, (01)00884838099241(21)746, (01)00884838099241(21)829, (01)00884838099241(21)883, (01)00884838099241(21)889, (01)00884838099241(21)665, (01)00884838099241(21)105, (01)00884838099241(21)96, (01)00884838099241(21)211, (01)00884838099241(21)208, (01)00884838099241(21)130, (01)00884838099241(21)888, (01)00884838099241(21)435, (01)00884838099241(21)598, (01)00884838099241(21)119, (01)00884838099241(21)527, (01)00884838099241(21)348, (01)00884838099241(21)132, (01)00884838099241(21)836, (01)00884838099241(21)587, (01)00884838099241(21)812, (01)00884838099241(21)904, (01)00884838099241(21)322, (01)00884838099241(21)875, (01)00884838099241(21)591, (01)00884838099241(21)756, (01)00884838099241(21)1016, (01)00884838099241(21)694, (01)00884838099241(21)537, (01)00884838099241(21)1055, (01)00884838099241(21)134, (01)00884838099241(21)615, (01)00884838099241(21)393, (01)00884838099241(21)434, (01)00884838099241(21)550, (01)00884838099241(21)612, (01)00884838099241(21)848, (01)00884838099241(21)918, (01)00884838099241(21)518, (01)00884838099241(21)683, (01)00884838099241(21)302, (01)00884838099241(21)1033, (01)00884838099241(21)137, (01)00884838099241(21)482, (01)00884838099241(21)919, (01)00884838099241(21)868, (01)00884838099241(21)446, (01)00884838099241(21)823, (01)00884838099241(21)1104, (01)00884838099241(21)719, (01)00884838099241(21)480, (01)00884838099241(21)519, (01)00884838099241(21)522, (01)00884838099241(21)520, (01)00884838099241(21)858, (01)00884838099241(21)523, (01)00884838099241(21)524, (01)00884838099241(21)526, (01)00884838099241(21)521, (01)00884838099241(21)525, (01)00884838099241(21)723, (01)00884838099241(21)821, (01)00884838099241(21)862, (01)00884838099241(21)1041, (01)00884838099241(21)78, (01)00884838099241(21)202, (01)00884838099241(21)535, (01)00884838099241(21)40, (01)00884838099241(21)8, (01)00884838099241(21)492, (01)00884838099241(21)64, (01)00884838099241(21)635, (01)00884838099241(21)924, (01)00884838099241(21)438, (01)00884838099241(21)437, (01)00884838099241(21)436, (01)00884838099241(21)270, (01)00884838099241(21)468, (01)00884838099241(21)376, (01)00884838099241(21)450, (01)00884838099241(21)451, (01)00884838099241(21)505, (01)00884838099241(21)650, (01)00884838099241(21)795, (01)00884838099241(21)793, (01)00884838099241(21)851, (01)00884838099241(21)894, (01)00884838099241(21)179, (01)00884838099241(21)445, (01)00884838099241(21)791, (01)00884838099241(21)807, (01)00884838099241(21)257, (01)00884838099241(21)266, (01)00884838099241(21)460, (01)00884838099241(21)194, (01)00884838099241(21)954, (01)00884838099241(21)684, (01)00884838099241(21)1092, (01)00884838099241(21)48, (01)00884838099241(21)583, (01)00884838099241(21)413, (01)00884838099241(21)600, (01)00884838099241(21)6, (01)00884838099241(21)876, (01)00884838099241(21)419, (01)00884838099241(21)666, (01)00884838099241(21)30, (01)00884838099241(21)787, (01)00884838099241(21)986, (01)00884838099241(21)682, (01)00884838099241(21)721, (01)00884838099241(21)773, (01)00884838099241(21)443, (01)00884838099241(21)82, (01)00884838099241(21)860, (01)00884838099241(21)278, (01)00884838099241(21)534, (01)00884838099241(21)1073, (01)00884838099241(21)613, (01)00884838099241(21)917, (01)00884838099241(21)611, (01)00884838099241(21)1122, (01)00884838099241(21)692, (01)00884838099241(21)996, (01)00884838099241(21)997, (01)00884838099241(21)652, (01)00884838099241(21)900, (01)00884838099241(21)614, (01)00884838099241(21)517, (01)00884838099241(21)991, (01)00884838099241(21)990, (01)00884838099241(21)957, (01)00884838099241(21)780, (01)00884838099241(21)781, (01)00884838099241(21)688, (01)00884838099241(21)686, (01)00884838099241(21)687, (01)00884838099241(21)689, (01)00884838099241(21)634, (01)00884838099241(21)979, (01)00884838099241(21)843, (01)00884838099241(21)651, (01)00884838099241(21)678, (01)00884838099241(21)782, (01)00884838099241(21)987, (01)00884838099241(21)798, (01)00884838099241(21)825, (01)00884838099241(21)797, (01)00884838099241(21)604, (01)00884838099241(21)838, (01)00884838099241(21)762, (01)00884838099241(21)758, (01)00884838099241(21)1057, (01)00884838099241(21)767, (01)00884838099241(21)763, (01)00884838099241(21)765, (01)00884838099241(21)764, (01)00884838099241(21)953, (01)00884838099241(21)644, (01)00884838099241(21)421, (01)00884838099241(21)792, (01)00884838099241(21)760, (01)00884838099241(21)255, (01)00884838099241(21)536, (01)00884838099241(21)80, (01)00884838099241(21)74, (01)00884838099241(21)72, (01)00884838099241(21)473, (01)00884838099241(21)625, (01)00884838099241(21)241, (01)00884838099241(21)1007, (01)00884838099241(21)656, (01)00884838099241(21)407, (01)00884838099241(21)24, (01)00884838099241(21)649, (01)00884838099241(21)902, (01)00884838099241(21)34, (01)00884838099241(21)1013, (01)00884838099241(21)1014, (01)00884838099241(21)1026, (01)00884838099241(21)578, (01)00884838099241(21)808, (01)00884838099241(21)913, (01)00884838099241(21)510, (01)00884838099241(21)958, (01)00884838099241(21)1208, (01)00884838099241(21)879, (01)00884838099241(21)901, (01)00884838099241(21)319, (01)00884838099241(21)320, (01)00884838099241(21)300, (01)00884838099241(21)276, (01)00884838099241(21)493, (01)00884838099241(21)1062, (01)00884838099241(21)1053, (01)00884838099241(21)653).
8. Model Numbers (UDI Numbers): 722079 ((01)00884838085268(21)703904, (01)00884838085268(21)890, (01)00884838085268(21)246, (01)00884838085268(21)905, (01)00884838085268(21)848, (01)00884838085268(21)809, (01)00884838085268(21)1032, (01)00884838085268(21)820, (01)00884838085268(21)1211, (01)00884838085268(21)1213, (01)00884838085268(21)47, (01)00884838085268(21)1365, (01)00884838085268(21)856, (01)00884838085268(21)703540, (01)00884838085268(21)296, (01)00884838085268(21)991, (01)00884838085268(21)574, (01)00884838085268(21)703336, (01)00884838085268(21)842, (01)00884838085268(21)884, (01)00884838085268(21)753, (01)00884838085268(21)703493, (01)00884838085268(21)407, (01)00884838085268(21)877, (01)00884838085268(21)703395, (01)00884838085268(21)703323, (01)00884838085268(21)1108, (01)00884838085268(21)1414, (01)00884838085268(21)703404, (01)00884838085268(21)1355, (01)00884838085268(21)703387, (01)00884838085268(21)863, (01)00884838085268(21)1378, (01)00884838085268(21)520, (01)00884838085268(21)465, (01)00884838085268(21)674, (01)00884838085268(21)361, (01)00884838085268(21)1021, (01)00884838085268(21)608, (01)00884838085268(21)611, (01)00884838085268(21)895, (01)00884838085268(21)883, (01)00884838085268(21)525, (01)00884838085268(21)885, (01)00884838085268(21)843, (01)00884838085268(21)703528, (01)00884838085268(21)754, (01)00884838085268(21)599, (01)00884838085268(21)878, (01)00884838085268(21)703310, (01)00884838085268(21)1285, (01)00884838085268(21)703497, (01)00884838085268(21)730, (01)00884838085268(21)658, (01)00884838085268(21)621, (01)00884838085268(21)795, (01)00884838085268(21)1289, (01)00884838085268(21)605, (01)00884838085268(21)34, (01)00884838085268(21)1222, (01)00884838085268(21)769, (01)00884838085268(21)938, (01)00884838085268(21)1311, (01)00884838085268(21)538, (01)00884838085268(21)989, (01)00884838085268(21)946, (01)00884838085268(21)888, (01)00884838085268(21)402, (01)00884838085268(21)1042, (01)00884838085268(21)703425, (01)00884838085268(21)703532, (01)00884838085268(21)1358, (01)00884838085268(21)1360, (01)00884838085268(21)1359, (01)00884838085268(21)326, (01)00884838085268(21)927, (01)00884838085268(21)1163, (01)00884838085268(21)416, (01)00884838085268(21)424, (01)00884838085268(21)436, (01)00884838085268(21)910, (01)00884838085268(21)1242, (01)00884838085268(21)1064, (01)00884838085268(21)1156, (01)00884838085268(21)1049, (01)00884838085268(21)1361, (01)00884838085268(21)543, (01)00884838085268(21)1200, (01)00884838085268(21)921, (01)00884838085268(21)30, (01)00884838085268(21)366, (01)00884838085268(21)228, (01)00884838085268(21)1187, (01)00884838085268(21)247, (01)00884838085268(21)537, (01)00884838085268(21)777, (01)00884838085268(21)703317, (01)00884838085268(21)746, (01)00884838085268(21)668, (01)00884838085268(21)706, (01)00884838085268(21)837, (01)00884838085268(21)212, (01)00884838085268(21)56, (01)00884838085268(21)703584, (01)00884838085268(21)1381, (01)00884838085268(21)1059, (01)00884838085268(21)1382, (01)00884838085268(21)1246, (01)00884838085268(21)703345, (01)00884838085268(21)60, (01)00884838085268(21)1100, (01)00884838085268(21)327, (01)00884838085268(21)1201, (01)00884838085268(21)702, (01)00884838085268(21)284, (01)00884838085268(21)1374, (01)00884838085268(21)703813, (01)00884838085268(21)704301, (01)00884838085268(21)780, (01)00884838085268(21)704114, (01)00884838085268(21)703719, (01)00884838085268(21)703565, (01)00884838085268(21)1280, (01)00884838085268(21)214, (01)00884838085268(21)704304, (01)00884838085268(21)704059, (01)00884838085268(21)943, (01)00884838085268(21)337, (01)00884838085268(21)703552, (01)00884838085268(21)704184, (01)00884838085268(21)704240, (01)00884838085268(21)704242, (01)00884838085268(21)703662, (01)00884838085268(21)703838, (01)00884838085268(21)703823, (01)00884838085268(21)703670, (01)00884838085268(21)1026, (01)00884838085268(21)380, (01)00884838085268(21)377, (01)00884838085268(21)703355, (01)00884838085268(21)703654, (01)00884838085268(21)703870, (01)00884838085268(21)662, (01)00884838085268(21)703449, (01)00884838085268(21)1405, (01)00884838085268(21)395, (01)00884838085268(21)433, (01)00884838085268(21)703699, (01)00884838085268(21)451, (01)00884838085268(21)258, (01)00884838085268(21)704300, (01)00884838085268(21)704299, (01)00884838085268(21)1306, (01)00884838085268(21)703999, (01)00884838085268(21)704129, (01)00884838085268(21)1172, (01)00884838085268(21)1319, (01)00884838085268(21)703712, (01)00884838085268(21)31, (01)00884838085268(21)704295, (01)00884838085268(21)703702, (01)00884838085268(21)703326, (01)00884838085268(21)703655, (01)00884838085268(21)703939, (01)00884838085268(21)703802, (01)00884838085268(21)703897, (01)00884838085268(21)704266, (01)00884838085268(21)704203, (01)00884838085268(21)703953, (01)00884838085268(21)703986, (01)00884838085268(21)1053, (01)00884838085268(21)703600, (01)00884838085268(21)703710, (01)00884838085268(21)704026, (01)00884838085268(21)703753, (01)00884838085268(21)703967, (01)00884838085268(21)704112, (01)00884838085268(21)984, (01)00884838085268(21)703992, (01)00884838085268(21)703807, (01)00884838085268(21)765, (01)00884838085268(21)703529, (01)00884838085268(21)704154, (01)00884838085268(21)778, (01)00884838085268(21)704037, (01)00884838085268(21)704290, (01)00884838085268(21)703669, (01)00884838085268(21)1429, (01)00884838085268(21)49, (01)00884838085268(21)704209, (01)00884838085268(21)703619, (01)00884838085268(21)704142, (01)00884838085268(21)825, (01)00884838085268(21)600, (01)00884838085268(21)452, (01)00884838085268(21)1020, (01)00884838085268(21)703358, (01)00884838085268(21)703487, (01)00884838085268(21)704092, (01)00884838085268(21)578, (01)00884838085268(21)703572, (01)00884838085268(21)982, (01)00884838085268(21)703583, (01)00884838085268(21)1106, (01)00884838085268(21)703855, (01)00884838085268(21)704127, (01)00884838085268(21)703348, (01)00884838085268(21)703872, (01)00884838085268(21)855, (01)00884838085268(21)1030, (01)00884838085268(21)704199, (01)00884838085268(21)703489, (01)00884838085268(21)703959, (01)00884838085268(21)913, (01)00884838085268(21)304, (01)00884838085268(21)916, (01)00884838085268(21)704284, (01)00884838085268(21)303, (01)00884838085268(21)704101, (01)00884838085268(21)305, (01)00884838085268(21)1194, (01)00884838085268(21)703961, (01)00884838085268(21)993, (01)00884838085268(21)347, (01)00884838085268(21)1074, (01)00884838085268(21)44, (01)00884838085268(21)703477, (01)00884838085268(21)257, (01)00884838085268(21)661, (01)00884838085268(21)704146, (01)00884838085268(21)703982, (01)00884838085268(21)704211, (01)00884838085268(21)703812, (01)00884838085268(21)1122, (01)00884838085268(21)703423, (01)00884838085268(21)1278, (01)00884838085268(21)703544, (01)00884838085268(21)958, (01)00884838085268(21)703740, (01)00884838085268(21)542, (01)00884838085268(21)348, (01)00884838085268(21)703867, (01)00884838085268(21)1154, (01)00884838085268(21)703965, (01)00884838085268(21)704239, (01)00884838085268(21)703993, (01)00884838085268(21)703877, (01)00884838085268(21)703542, (01)00884838085268(21)704173, (01)00884838085268(21)551, (01)00884838085268(21)704085, (01)00884838085268(21)704079, (01)00884838085268(21)485, (01)00884838085268(21)704289, (01)00884838085268(21)596, (01)00884838085268(21)1052, (01)00884838085268(21)703804, (01)00884838085268(21)703906, (01)00884838085268(21)703596, (01)00884838085268(21)703937, (01)00884838085268(21)703972, (01)00884838085268(21)1288, (01)00884838085268(21)703954, (01)00884838085268(21)1183, (01)00884838085268(21)704262, (01)00884838085268(21)703948, (01)00884838085268(21)703556, (01)00884838085268(21)902, (01)00884838085268(21)571, (01)00884838085268(21)1422, (01)00884838085268(21)457, (01)00884838085268(21)703754, (01)00884838085268(21)703569, (01)00884838085268(21)704081, (01)00884838085268(21)703854, (01)00884838085268(21)704103, (01)00884838085268(21)704019, (01)00884838085268(21)1119, (01)00884838085268(21)960, (01)00884838085268(21)704023, (01)00884838085268(21)704119, (01)00884838085268(21)1015, (01)00884838085268(21)703614, (01)00884838085268(21)1076, (01)00884838085268(21)704030, (01)00884838085268(21)704126, (01)00884838085268(21)956, (01)00884838085268(21)703819, (01)00884838085268(21)704036, (01)00884838085268(21)703690, (01)00884838085268(21)703375, (01)00884838085268(21)703648, (01)00884838085268(21)703709, (01)00884838085268(21)704207, (01)00884838085268(21)703728, (01)00884838085268(21)703826, (01)00884838085268(21)381, (01)00884838085268(21)630, (01)00884838085268(21)1318, (01)00884838085268(21)703794, (01)00884838085268(21)901, (01)00884838085268(21)703574, (01)00884838085268(21)762, (01)00884838085268(21)703743, (01)00884838085268(21)703632, (01)00884838085268(21)703913, (01)00884838085268(21)703889, (01)00884838085268(21)703700, (01)00884838085268(21)704246, (01)00884838085268(21)704054, (01)00884838085268(21)703606, (01)00884838085268(21)703473, (01)00884838085268(21)703346, (01)00884838085268(21)703750, (01)00884838085268(21)703990, (01)00884838085268(21)703834, (01)00884838085268(21)703727, (01)00884838085268(21)965, (01)00884838085268(21)498, (01)00884838085268(21)789, (01)00884838085268(21)703482, (01)00884838085268(21)703693, (01)00884838085268(21)703533, (01)00884838085268(21)967, (01)00884838085268(21)475, (01)00884838085268(21)963, (01)00884838085268(21)704261, (01)00884838085268(21)704067, (01)00884838085268(21)349, (01)00884838085268(21)784, (01)00884838085268(21)704274, (01)00884838085268(21)704144, (01)00884838085268(21)704168, (01)00884838085268(21)704167, (01)00884838085268(21)642, (01)00884838085268(21)698, (01)00884838085268(21)1210, (01)00884838085268(21)643, (01)00884838085268(21)704070, (01)00884838085268(21)450, (01)00884838085268(21)703629, (01)00884838085268(21)703973, (01)00884838085268(21)704110, (01)00884838085268(21)703701, (01)00884838085268(21)703628, (01)00884838085268(21)703955, (01)00884838085268(21)703665, (01)00884838085268(21)1367, (01)00884838085268(21)703899, (01)00884838085268(21)704202, (01)00884838085268(21)703828, (01)00884838085268(21)703641, (01)00884838085268(21)703711, (01)00884838085268(21)703651, (01)00884838085268(21)703856, (01)00884838085268(21)703842, (01)00884838085268(21)703724, (01)00884838085268(21)704269, (01)00884838085268(21)704074, (01)00884838085268(21)704108, (01)00884838085268(21)703821, (01)00884838085268(21)704134, (01)00884838085268(21)704058, (01)00884838085268(21)703586, (01)00884838085268(21)704029, (01)00884838085268(21)703979, (01)00884838085268(21)703777, (01)00884838085268(21)703741, (01)00884838085268(21)704174, (01)00884838085268(21)704041, (01)00884838085268(21)704237, (01)00884838085268(21)703650, (01)00884838085268(21)703646, (01)00884838085268(21)704021, (01)00884838085268(21)703817, (01)00884838085268(21)703908, (01)00884838085268(21)703909, (01)00884838085268(21)704031, (01)00884838085268(21)703726, (01)00884838085268(21)704271, (01)00884838085268(21)703756, (01)00884838085268(21)703618, (01)00884838085268(21)704060, (01)00884838085268(21)704120, (01)00884838085268(21)704017, (01)00884838085268(21)703988, (01)00884838085268(21)703573, (01)00884838085268(21)703800, (01)00884838085268(21)703997, (01)00884838085268(21)704073, (01)00884838085268(21)703674, (01)00884838085268(21)704291, (01)00884838085268(21)703673, (01)00884838085268(21)703591, (01)00884838085268(21)704002, (01)00884838085268(21)703733, (01)00884838085268(21)703735, (01)00884838085268(21)703940, (01)00884838085268(21)703757, (01)00884838085268(21)703751, (01)00884838085268(21)703883, (01)00884838085268(21)704139, (01)00884838085268(21)704157, (01)00884838085268(21)704293, (01)00884838085268(21)704279, (01)00884838085268(21)704280, (01)00884838085268(21)704024, (01)00884838085268(21)703912, (01)00884838085268(21)703911, (01)00884838085268(21)703605, (01)00884838085268(21)703797, (01)00884838085268(21)703969, (01)00884838085268(21)704022, (01)00884838085268(21)704039, (01)00884838085268(21)703471, (01)00884838085268(21)704095, (01)00884838085268(21)703768, (01)00884838085268(21)703779, (01)00884838085268(21)703663, (01)00884838085268(21)704138, (01)00884838085268(21)703844, (01)00884838085268(21)703604, (01)00884838085268(21)703830, (01)00884838085268(21)704247, (01)00884838085268(21)703707, (01)00884838085268(21)703734, (01)00884838085268(21)703731, (01)00884838085268(21)704286, (01)00884838085268(21)704008, (01)00884838085268(21)704221, (01)00884838085268(21)703799, (01)00884838085268(21)703576, (01)00884838085268(21)704009, (01)00884838085268(21)704069, (01)00884838085268(21)703840, (01)00884838085268(21)703708, (01)00884838085268(21)704080, (01)00884838085268(21)703998, (01)00884838085268(21)704040, (01)00884838085268(21)704012, (01)00884838085268(21)703603, (01)00884838085268(21)704130, (01)00884838085268(21)703786, (01)00884838085268(21)704131, (01)00884838085268(21)704078, (01)00884838085268(21)704077, (01)00884838085268(21)703630, (01)00884838085268(21)704066, (01)00884838085268(21)703983, (01)00884838085268(21)703507, (01)00884838085268(21)704001, (01)00884838085268(21)703645, (01)00884838085268(21)703919, (01)00884838085268(21)704088, (01)00884838085268(21)703616, (01)00884838085268(21)704201, (01)00884838085268(21)704133, (01)00884838085268(21)703796, (01)00884838085268(21)703858, (01)00884838085268(21)703860, (01)00884838085268(21)703857, (01)00884838085268(21)703859, (01)00884838085268(21)703863, (01)00884838085268(21)703780, (01)00884838085268(21)704166, (01)00884838085268(21)703640, (01)00884838085268(21)703642, (01)00884838085268(21)704093, (01)00884838085268(21)704254, (01)00884838085268(21)704259, (01)00884838085268(21)703680, (01)00884838085268(21)703878, (01)00884838085268(21)703901, (01)00884838085268(21)704020, (01)00884838085268(21)703884, (01)00884838085268(21)703958, (01)00884838085268(21)704091, (01)00884838085268(21)1376, (01)00884838085268(21)704263, (01)00884838085268(21)977, (01)00884838085268(21)1093, (01)00884838085268(21)703975, (01)00884838085268(21)703688, (01)00884838085268(21)704113, (01)00884838085268(21)704090, (01)00884838085268(21)704147, (01)00884838085268(21)703571, (01)00884838085268(21)703657, (01)00884838085268(21)703547, (01)00884838085268(21)703771, (01)00884838085268(21)703703, (01)00884838085268(21)703887, (01)00884838085268(21)703781, (01)00884838085268(21)703923, (01)00884838085268(21)704111, (01)00884838085268(21)703805, (01)00884838085268(21)704217, (01)00884838085268(21)704124, (01)00884838085268(21)703585, (01)00884838085268(21)704094, (01)00884838085268(21)703759, (01)00884838085268(21)703639, (01)00884838085268(21)703627, (01)00884838085268(21)704064, (01)00884838085268(21)704050, (01)00884838085268(21)704255, (01)00884838085268(21)703803, (01)00884838085268(21)704175, (01)00884838085268(21)703891, (01)00884838085268(21)703895, (01)00884838085268(21)704181, (01)00884838085268(21)1263, (01)00884838085268(21)703894, (01)00884838085268(21)704180, (01)00884838085268(21)703918, (01)00884838085268(21)704007, (01)00884838085268(21)703921, (01)00884838085268(21)703789, (01)00884838085268(21)704075, (01)00884838085268(21)703827, (01)00884838085268(21)703587, (01)00884838085268(21)704056, (01)00884838085268(21)704122, (01)00884838085268(21)703876, (01)00884838085268(21)703989, (01)00884838085268(21)704046, (01)00884838085268(21)703950, (01)00884838085268(21)703924, (01)00884838085268(21)704223, (01)00884838085268(21)703851, (01)00884838085268(21)704051, (01)00884838085268(21)703934, (01)00884838085268(21)1287, (01)00884838085268(21)704216, (01)00884838085268(21)704267, (01)00884838085268(21)704177, (01)00884838085268(21)703825, (01)00884838085268(21)704115, (01)00884838085268(21)704044, (01)00884838085268(21)704038, (01)00884838085268(21)703809, (01)00884838085268(21)703836, (01)00884838085268(21)703932, (01)00884838085268(21)703661, (01)00884838085268(21)703676, (01)00884838085268(21)704152, (01)00884838085268(21)704132, (01)00884838085268(21)703548, (01)00884838085268(21)704200, (01)00884838085268(21)703666, (01)00884838085268(21)703880, (01)00884838085268(21)703696, (01)00884838085268(21)703837, (01)00884838085268(21)704270, (01)00884838085268(21)703798, (01)00884838085268(21)703818, (01)00884838085268(21)704096, (01)00884838085268(21)704151, (01)00884838085268(21)703850, (01)00884838085268(21)703725, (01)00884838085268(21)704178, (01)00884838085268(21)704161, (01)00884838085268(21)703871, (01)00884838085268(21)704187, (01)00884838085268(21)703977, (01)00884838085268(21)704082, (01)00884838085268(21)704089, (01)00884838085268(21)703888, (01)00884838085268(21)703928, (01)00884838085268(21)703929, (01)00884838085268(21)703931, (01)00884838085268(21)703930, (01)00884838085268(21)703974, (01)00884838085268(21)703881, (01)00884838085268(21)704252, (01)00884838085268(21)703941, (01)00884838085268(21)703996, (01)00884838085268(21)703917, (01)00884838085268(21)703667, (01)00884838085268(21)703668, (01)00884838085268(21)704153, (01)00884838085268(21)703617, (01)00884838085268(21)704169, (01)00884838085268(21)704219, (01)00884838085268(21)704275, (01)00884838085268(21)703625, (01)00884838085268(21)703949, (01)00884838085268(21)704121, (01)00884838085268(21)704104, (01)00884838085268(21)704128, (01)00884838085268(21)704065, (01)00884838085268(21)704192, (01)00884838085268(21)1228, (01)00884838085268(21)704116, (01)00884838085268(21)704102, (01)00884838085268(21)703898, (01)00884838085268(21)703636, (01)00884838085268(21)703843, (01)00884838085268(21)704264, (01)00884838085268(21)703691, (01)00884838085268(21)703692, (01)00884838085268(21)703664, (01)00884838085268(21)703555, (01)00884838085268(21)703601, (01)00884838085268(21)703633, (01)00884838085268(21)704218, (01)00884838085268(21)704109, (01)00884838085268(21)1224, (01)00884838085268(21)1225, (01)00884838085268(21)704125, (01)00884838085268(21)703684, (01)00884838085268(21)703742, (01)00884838085268(21)703723, (01)00884838085268(21)704148, (01)00884838085268(21)703681, (01)00884838085268(21)703778, (01)00884838085268(21)703635, (01)00884838085268(21)704206, (01)00884838085268(21)704198, (01)00884838085268(21)704197, (01)00884838085268(21)704196, (01)00884838085268(21)703910, (01)00884838085268(21)703553, (01)00884838085268(21)703946, (01)00884838085268(21)704265, (01)00884838085268(21)704189, (01)00884838085268(21)703951, (01)00884838085268(21)704098, (01)00884838085268(21)703539, (01)00884838085268(21)703541, (01)00884838085268(21)703678, (01)00884838085268(21)1034, (01)00884838085268(21)703900, (01)00884838085268(21)704162, (01)00884838085268(21)704159, (01)00884838085268(21)703893, (01)00884838085268(21)703761, (01)00884838085268(21)786, (01)00884838085268(21)390, (01)00884838085268(21)398, (01)00884838085268(21)703653, (01)00884838085268(21)1253, (01)00884838085268(21)703414, (01)00884838085268(21)941, (01)00884838085268(21)392, (01)00884838085268(21)544, (01)00884838085268(21)703505, (01)00884838085268(21)1199, (01)00884838085268(21)703658, (01)00884838085268(21)704296, (01)00884838085268(21)704135, (01)00884838085268(21)704149, (01)00884838085268(21)704034, (01)00884838085268(21)1332, (01)00884838085268(21)703873, (01)00884838085268(21)704010, (01)00884838085268(21)704057, (01)00884838085268(21)1357, (01)00884838085268(21)703459, (01)00884838085268(21)703944, (01)00884838085268(21)952, (01)00884838085268(21)676, (01)00884838085268(21)703892, (01)00884838085268(21)703720, (01)00884838085268(21)410, (01)00884838085268(21)703985, (01)00884838085268(21)906, (01)00884838085268(21)1349, (01)00884838085268(21)703579, (01)00884838085268(21)703810, (01)00884838085268(21)704303, (01)00884838085268(21)704172, (01)00884838085268(21)703694, (01)00884838085268(21)703874, (01)00884838085268(21)1188, (01)00884838085268(21)764, (01)00884838085268(21)704015, (01)00884838085268(21)703675, (01)00884838085268(21)703970, (01)00884838085268(21)704117, (01)00884838085268(21)704084, (01)00884838085268(21)703956, (01)00884838085268(21)704213, (01)00884838085268(21)276, (01)00884838085268(21)704035, (01)00884838085268(21)971, (01)00884838085268(21)704188, (01)00884838085268(21)703814, (01)00884838085268(21)703717, (01)00884838085268(21)703920, (01)00884838085268(21)703925, (01)00884838085268(21)703922, (01)00884838085268(21)704016, (01)00884838085268(21)703729, (01)00884838085268(21)704243, (01)00884838085268(21)308, (01)00884838085268(21)775, (01)00884838085268(21)24, (01)00884838085268(21)703952, (01)00884838085268(21)704018, (01)00884838085268(21)703957, (01)00884838085268(21)704032, (01)00884838085268(21)704033, (01)00884838085268(21)703695, (01)00884838085268(21)703475, (01)00884838085268(21)703947, (01)00884838085268(21)704063, (01)00884838085268(21)703890, (01)00884838085268(21)703589, (01)00884838085268(21)704204, (01)00884838085268(21)703536, (01)00884838085268(21)703849, (01)00884838085268(21)704287, (01)00884838085268(21)703659, (01)00884838085268(21)704025, (01)00884838085268(21)703926, (01)00884838085268(21)703978, (01)00884838085268(21)704260, (01)00884838085268(21)703991, (01)00884838085268(21)703987, (01)00884838085268(21)704250, (01)00884838085268(21)704195, (01)00884838085268(21)703468, (01)00884838085268(21)703467, (01)00884838085268(21)703652, (01)00884838085268(21)703714, (01)00884838085268(21)703481, (01)00884838085268(21)419, (01)00884838085268(21)724, (01)00884838085268(21)1057, (01)00884838085268(21)771, (01)00884838085268(21)1339, (01)00884838085268(21)1176, (01)00884838085268(21)703490, (01)00884838085268(21)703631, (01)00884838085268(21)1025, (01)00884838085268(21)1336, (01)00884838085268(21)703339, (01)00884838085268(21)1161, (01)00884838085268(21)703352, (01)00884838085268(21)1131, (01)00884838085268(21)1091, (01)00884838085268(21)1178, (01)00884838085268(21)703531, (01)00884838085268(21)703762, (01)00884838085268(21)703763, (01)00884838085268(21)703706, (01)00884838085268(21)703705, (01)00884838085268(21)703554, (01)00884838085268(21)1058, (01)00884838085268(21)703787, (01)00884838085268(21)1075, (01)00884838085268(21)703431, (01)00884838085268(21)827, (01)00884838085268(21)1024, (01)00884838085268(21)703410, (01)00884838085268(21)844, (01)00884838085268(21)1094, (01)00884838085268(21)1069, (01)00884838085268(21)703522, (01)00884838085268(21)703354, (01)00884838085268(21)703590, (01)00884838085268(21)1180, (01)00884838085268(21)1067, (01)00884838085268(21)723, (01)00884838085268(21)703795, (01)00884838085268(21)703570, (01)00884838085268(21)703454, (01)00884838085268(21)1173, (01)00884838085268(21)703422, (01)00884838085268(21)703518, (01)00884838085268(21)1404, (01)00884838085268(21)703484, (01)00884838085268(21)703325, (01)00884838085268(21)568, (01)00884838085268(21)1207, (01)00884838085268(21)703865, (01)00884838085268(21)703403, (01)00884838085268(21)1068, (01)00884838085268(21)703437, (01)00884838085268(21)1071, (01)00884838085268(21)612, (01)00884838085268(21)703373, (01)00884838085268(21)277, (01)00884838085268(21)917, (01)00884838085268(21)487, (01)00884838085268(21)1145, (01)00884838085268(21)929), 722224 ((01)00884838099258(21)450, (01)00884838099258(21)409, (01)00884838099258(21)261, (01)00884838099258(21)1083, (01)00884838099258(21)475, (01)00884838099258(21)588, (01)00884838099258(21)2241, (01)00884838099258(21)421, (01)00884838099258(21)288, (01)00884838099258(21)2520, (01)00884838099258(21)91, (01)00884838099258(21)1692, (01)00884838099258(21)2432, (01)00884838099258(21)287, (01)00884838099258(21)2421, (01)00884838099258(21)904, (01)00884838099258(21)3063, (01)00884838099258(21)1150, (01)00884838099258(21)901, (01)00884838099258(21)514, (01)00884838099258(21)553, (01)00884838099258(21)2390, (01)00884838099258(21)618, (01)00884838099258(21)617, (01)00884838099258(21)859, (01)00884838099258(21)611, (01)00884838099258(21)2759, (01)00884838099258(21)2856, (01)00884838099258(21)565, (01)00884838099258(21)224, (01)00884838099258(21)2723, (01)00884838099258(21)2911, (01)00884838099258(21)1039, (01)00884838099258(21)1679, (01)00884838099258(21)3048, (01)00884838099258(21)3046, (01)00884838099258(21)580, (01)00884838099258(21)850, (01)00884838099258(21)1117, (01)00884838099258(21)1082, (01)00884838099258(21)3283, (01)00884838099258(21)1488, (01)00884838099258(21)966, (01)00884838099258(21)1940, (01)00884838099258(21)2656, (01)00884838099258(21)867, (01)00884838099258(21)244, (01)00884838099258(21)1975, (01)00884838099258(21)428, (01)00884838099258(21)2831, (01)00884838099258(21)2331, (01)00884838099258(21)56, (01)00884838099258(21)2058, (01)00884838099258(21)564, (01)00884838099258(21)856, (01)00884838099258(21)1047, (01)00884838099258(21)432, (01)00884838099258(21)439, (01)00884838099258(21)792, (01)00884838099258(21)595, (01)00884838099258(21)1038, (01)00884838099258(21)965, (01)00884838099258(21)963, (01)00884838099258(21)151, (01)00884838099258(21)657, (01)00884838099258(21)3145, (01)00884838099258(21)215, (01)00884838099258(21)770, (01)00884838099258(21)478, (01)00884838099258(21)516, (01)00884838099258(21)1482, (01)00884838099258(21)3525, (01)00884838099258(21)552, (01)00884838099258(21)1175, (01)00884838099258(21)373, (01)00884838099258(21)339, (01)00884838099258(21)1606, (01)00884838099258(21)2549, (01)00884838099258(21)2435, (01)00884838099258(21)2045, (01)00884838099258(21)1960, (01)00884838099258(21)587, (01)00884838099258(21)112, (01)00884838099258(21)2548, (01)00884838099258(21)2377, (01)00884838099258(21)730, (01)00884838099258(21)492, (01)00884838099258(21)2226, (01)00884838099258(21)1935, (01)00884838099258(21)662, (01)00884838099258(21)78, (01)00884838099258(21)2452, (01)00884838099258(21)252, (01)00884838099258(21)1563, (01)00884838099258(21)543, (01)00884838099258(21)351, (01)00884838099258(21)2453, (01)00884838099258(21)300, (01)00884838099258(21)2647, (01)00884838099258(21)2279, (01)00884838099258(21)2220, (01)00884838099258(21)857, (01)00884838099258(21)1041, (01)00884838099258(21)839, (01)00884838099258(21)999, (01)00884838099258(21)982, (01)00884838099258(21)462, (01)00884838099258(21)99, (01)00884838099258(21)2655, (01)00884838099258(21)188, (01)00884838099258(21)2369, (01)00884838099258(21)2301, (01)00884838099258(21)2130, (01)00884838099258(21)165, (01)00884838099258(21)1551, (01)00884838099258(21)627, (01)00884838099258(21)1383, (01)00884838099258(21)2965, (01)00884838099258(21)2967, (01)00884838099258(21)1999, (01)00884838099258(21)28, (01)00884838099258(21)2199, (01)00884838099258(21)746, (01)00884838099258(21)2340, (01)00884838099258(21)80, (01)00884838099258(21)169, (01)00884838099258(21)53, (01)00884838099258(21)332, (01)00884838099258(21)845, (01)00884838099258(21)1499, (01)00884838099258(21)404, (01)00884838099258(21)2957, (01)00884838099258(21)253, (01)00884838099258(21)1243, (01)00884838099258(21)1161, (01)00884838099258(21)2367, (01)00884838099258(21)844, (01)00884838099258(21)979, (01)00884838099258(21)787, (01)00884838099258(21)441, (01)00884838099258(21)84, (01)00884838099258(21)949, (01)00884838099258(21)350, (01)00884838099258(21)793, (01)00884838099258(21)581, (01)00884838099258(21)1827, (01)00884838099258(21)3401, (01)00884838099258(21)2733, (01)00884838099258(21)643, (01)00884838099258(21)1527, (01)00884838099258(21)503, (01)00884838099258(21)1335, (01)00884838099258(21)185, (01)00884838099258(21)379, (01)00884838099258(21)1156, (01)00884838099258(21)875, (01)00884838099258(21)874, (01)00884838099258(21)2667, (01)00884838099258(21)1622, (01)00884838099258(21)352, (01)00884838099258(21)257, (01)00884838099258(21)1568, (01)00884838099258(21)2852, (01)00884838099258(21)2931, (01)00884838099258(21)1764, (01)00884838099258(21)2742, (01)00884838099258(21)1988, (01)00884838099258(21)1974, (01)00884838099258(21)1851, (01)00884838099258(21)2804, (01)00884838099258(21)2933, (01)00884838099258(21)1627, (01)00884838099258(21)1629, (01)00884838099258(21)1632, (01)00884838099258(21)2777, (01)00884838099258(21)1709, (01)00884838099258(21)2550, (01)00884838099258(21)1705, (01)00884838099258(21)1943, (01)00884838099258(21)1942, (01)00884838099258(21)1802, (01)00884838099258(21)1216, (01)00884838099258(21)2772, (01)00884838099258(21)63, (01)00884838099258(21)1670, (01)00884838099258(21)2089, (01)00884838099258(21)1686, (01)00884838099258(21)2186, (01)00884838099258(21)2546, (01)00884838099258(21)3086, (01)00884838099258(21)2513, (01)00884838099258(21)3105, (01)00884838099258(21)1312, (01)00884838099258(21)942, (01)00884838099258(21)1326, (01)00884838099258(21)1609, (01)00884838099258(21)937, (01)00884838099258(21)2245, (01)00884838099258(21)1330, (01)00884838099258(21)3099, (01)00884838099258(21)801, (01)00884838099258(21)1350, (01)00884838099258(21)2373, (01)00884838099258(21)1349, (01)00884838099258(21)1369, (01)00884838099258(21)1970, (01)00884838099258(21)776, (01)00884838099258(21)2571, (01)00884838099258(21)2146, (01)00884838099258(21)799, (01)00884838099258(21)800, (01)00884838099258(21)1747, (01)00884838099258(21)1385, (01)00884838099258(21)2405, (01)00884838099258(21)1204, (01)00884838099258(21)1529, (01)00884838099258(21)2060, (01)00884838099258(21)2479, (01)00884838099258(21)2044, (01)00884838099258(21)2025, (01)00884838099258(21)1151, (01)00884838099258(21)1164, (01)00884838099258(21)2509, (01)00884838099258(21)2588, (01)00884838099258(21)1449, (01)00884838099258(21)765, (01)00884838099258(21)501, (01)00884838099258(21)546, (01)00884838099258(21)1475, (01)00884838099258(21)2586, (01)00884838099258(21)1470, (01)00884838099258(21)2496, (01)00884838099258(21)3079, (01)00884838099258(21)778, (01)00884838099258(21)2464, (01)00884838099258(21)2585, (01)00884838099258(21)1668, (01)00884838099258(21)2178, (01)00884838099258(21)2959, (01)00884838099258(21)828, (01)00884838099258(21)1502, (01)00884838099258(21)2515, (01)00884838099258(21)1044, (01)00884838099258(21)1344, (01)00884838099258(21)1619, (01)00884838099258(21)2238, (01)00884838099258(21)720, (01)00884838099258(21)1581, (01)00884838099258(21)1138, (01)00884838099258(21)2732, (01)00884838099258(21)1949, (01)00884838099258(21)675, (01)00884838099258(21)2111, (01)00884838099258(21)3005, (01)00884838099258(21)548, (01)00884838099258(21)1194, (01)00884838099258(21)1067, (01)00884838099258(21)1129, (01)00884838099258(21)2467, (01)00884838099258(21)1104, (01)00884838099258(21)499, (01)00884838099258(21)530, (01)00884838099258(21)2560, (01)00884838099258(21)1002, (01)00884838099258(21)1704, (01)00884838099258(21)2592, (01)00884838099258(21)3183, (01)00884838099258(21)1368, (01)00884838099258(21)3061, (01)00884838099258(21)3227, (01)00884838099258(21)1522, (01)00884838099258(21)2394, (01)00884838099258(21)2320, (01)00884838099258(21)887, (01)00884838099258(21)2709, (01)00884838099258(21)1474, (01)00884838099258(21)532, (01)00884838099258(21)596, (01)00884838099258(21)1306, (01)00884838099258(21)1647, (01)00884838099258(21)1153, (01)00884838099258(21)1991, (01)00884838099258(21)2865, (01)00884838099258(21)2864, (01)00884838099258(21)1281, (01)00884838099258(21)2923, (01)00884838099258(21)2866, (01)00884838099258(21)1280, (01)00884838099258(21)62, (01)00884838099258(21)37, (01)00884838099258(21)2947, (01)00884838099258(21)2269, (01)00884838099258(21)2705, (01)00884838099258(21)2745, (01)00884838099258(21)2457, (01)00884838099258(21)2527, (01)00884838099258(21)1327, (01)00884838099258(21)708, (01)00884838099258(21)706, (01)00884838099258(21)707, (01)00884838099258(21)1785, (01)00884838099258(21)1862, (01)00884838099258(21)575, (01)00884838099258(21)1549, (01)00884838099258(21)2943, (01)00884838099258(21)1427, (01)00884838099258(21)2261, (01)00884838099258(21)1406, (01)00884838099258(21)2217, (01)00884838099258(21)171, (01)00884838099258(21)2310, (01)00884838099258(21)1125, (01)00884838099258(21)827, (01)00884838099258(21)48, (01)00884838099258(21)1050, (01)00884838099258(21)1348, (01)00884838099258(21)1051, (01)00884838099258(21)1052, (01)00884838099258(21)1053, (01)00884838099258(21)2277, (01)00884838099258(21)1080, (01)00884838099258(21)1054, (01)00884838099258(21)658, (01)00884838099258(21)383, (01)00884838099258(21)2751, (01)00884838099258(21)2605, (01)00884838099258(21)1092, (01)00884838099258(21)2534, (01)00884838099258(21)145, (01)00884838099258(21)35, (01)00884838099258(21)2715, (01)00884838099258(21)2692, (01)00884838099258(21)2953, (01)00884838099258(21)3236, (01)00884838099258(21)1840, (01)00884838099258(21)3111, (01)00884838099258(21)1289, (01)00884838099258(21)264, (01)00884838099258(21)2073, (01)00884838099258(21)1919, (01)00884838099258(21)1667, (01)00884838099258(21)2431, (01)00884838099258(21)2064, (01)00884838099258(21)2416, (01)00884838099258(21)1184, (01)00884838099258(21)2734, (01)00884838099258(21)2862, (01)00884838099258(21)2768, (01)00884838099258(21)2767, (01)00884838099258(21)1411, (01)00884838099258(21)3197, (01)00884838099258(21)768, (01)00884838099258(21)449, (01)00884838099258(21)448, (01)00884838099258(21)842, (01)00884838099258(21)520, (01)00884838099258(21)1463, (01)00884838099258(21)2097, (01)00884838099258(21)358, (01)00884838099258(21)2224, (01)00884838099258(21)3118, (01)00884838099258(21)489, (01)00884838099258(21)2760, (01)00884838099258(21)1742, (01)00884838099258(21)2476, (01)00884838099258(21)752, (01)00884838099258(21)2975, (01)00884838099258(21)1114, (01)00884838099258(21)2429, (01)00884838099258(21)1545, (01)00884838099258(21)1119, (01)00884838099258(21)2927, (01)00884838099258(21)2741, (01)00884838099258(21)2652, (01)00884838099258(21)2713, (01)00884838099258(21)635, (01)00884838099258(21)2538, (01)00884838099258(21)2157, (01)00884838099258(21)551, (01)00884838099258(21)1365, (01)00884838099258(21)1366, (01)00884838099258(21)32, (01)00884838099258(21)1760, (01)00884838099258(21)2107, (01)00884838099258(21)2268, (01)00884838099258(21)1272, (01)00884838099258(21)460, (01)00884838099258(21)2781, (01)00884838099258(21)990, (01)00884838099258(21)2779, (01)00884838099258(21)1446, (01)00884838099258(21)716, (01)00884838099258(21)1173, (01)00884838099258(21)1684, (01)00884838099258(21)1498, (01)00884838099258(21)2643, (01)00884838099258(21)410, (01)00884838099258(21)504, (01)00884838099258(21)459, (01)00884838099258(21)2444, (01)00884838099258(21)1076, (01)00884838099258(21)2286, (01)00884838099258(21)2360, (01)00884838099258(21)2674, (01)00884838099258(21)3655, (01)00884838099258(21)2780, (01)00884838099258(21)412, (01)00884838099258(21)2191, (01)00884838099258(21)1921, (01)00884838099258(21)508, (01)00884838099258(21)804, (01)00884838099258(21)2936, (01)00884838099258(21)1881, (01)00884838099258(21)1831, (01)00884838099258(21)1040, (01)00884838099258(21)2272, (01)00884838099258(21)1676, (01)00884838099258(21)2845, (01)00884838099258(21)1355, (01)00884838099258(21)1205, (01)00884838099258(21)1761, (01)00884838099258(21)2468, (01)00884838099258(21)1633, (01)00884838099258(21)85, (01)00884838099258(21)3234, (01)00884838099258(21)985, (01)00884838099258(21)2446, (01)00884838099258(21)1809, (01)00884838099258(21)2678, (01)00884838099258(21)1420, (01)00884838099258(21)1948, (01)00884838099258(21)2590, (01)00884838099258(21)1434, (01)00884838099258(21)2816, (01)00884838099258(21)2312, (01)00884838099258(21)2296, (01)00884838099258(21)2204, (01)00884838099258(21)1673, (01)00884838099258(21)2357, (01)00884838099258(21)1986, (01)00884838099258(21)1665, (01)00884838099258(21)1824, (01)00884838099258(21)1655, (01)00884838099258(21)2887, (01)00884838099258(21)2061, (01)00884838099258(21)1927, (01)00884838099258(21)2402, (01)00884838099258(21)2904, (01)00884838099258(21)1553, (01)00884838099258(21)2380, (01)00884838099258(21)1413, (01)00884838099258(21)1127, (01)00884838099258(21)2123, (01)00884838099258(21)1732, (01)00884838099258(21)1193, (01)00884838099258(21)1771, (01)00884838099258(21)1448, (01)00884838099258(21)2791, (01)00884838099258(21)1788, (01)00884838099258(21)1144, (01)00884838099258(21)2632, (01)00884838099258(21)1500, (01)00884838099258(21)2334, (01)00884838099258(21)2956, (01)00884838099258(21)1513, (01)00884838099258(21)1198, (01)00884838099258(21)2660, (01)00884838099258(21)1819, (01)00884838099258(21)1440, (01)00884838099258(21)1285, (01)00884838099258(21)1178, (01)00884838099258(21)2650, (01)00884838099258(21)1607, (01)00884838099258(21)1418, (01)00884838099258(21)2471, (01)00884838099258(21)3115, (01)00884838099258(21)1631, (01)00884838099258(21)2115, (01)00884838099258(21)2875, (01)00884838099258(21)2666, (01)00884838099258(21)2212, (01)00884838099258(21)1599, (01)00884838099258(21)2634, (01)00884838099258(21)1389, (01)00884838099258(21)2274, (01)00884838099258(21)2940, (01)00884838099258(21)2941, (01)00884838099258(21)1457, (01)00884838099258(21)1489, (01)00884838099258(21)1624, (01)00884838099258(21)2085, (01)00884838099258(21)1497, (01)00884838099258(21)1946, (01)00884838099258(21)1971, (01)00884838099258(21)1990, (01)00884838099258(21)3002, (01)00884838099258(21)1445, (01)00884838099258(21)2796, (01)00884838099258(21)2024, (01)00884838099258(21)2225, (01)00884838099258(21)2702, (01)00884838099258(21)1674, (01)00884838099258(21)2282, (01)00884838099258(21)3140, (01)00884838099258(21)3102, (01)00884838099258(21)2408, (01)00884838099258(21)2822, (01)00884838099258(21)1941, (01)00884838099258(21)2607, (01)00884838099258(21)1825, (01)00884838099258(21)3128, (01)00884838099258(21)2388, (01)00884838099258(21)2387, (01)00884838099258(21)1520, (01)00884838099258(21)1696, (01)00884838099258(21)2216, (01)00884838099258(21)1708, (01)00884838099258(21)2894, (01)00884838099258(21)2019, (01)00884838099258(21)2020, (01)00884838099258(21)3103, (01)00884838099258(21)3054, (01)00884838099258(21)1729, (01)00884838099258(21)3080, (01)00884838099258(21)2821, (01)00884838099258(21)2407, (01)00884838099258(21)1548, (01)00884838099258(21)1979, (01)00884838099258(21)2492, (01)00884838099258(21)2491, (01)00884838099258(21)2773, (01)00884838099258(21)2774, (01)00884838099258(21)2664, (01)00884838099258(21)1202, (01)00884838099258(21)2372, (01)00884838099258(21)1836, (01)00884838099258(21)2213, (01)00884838099258(21)3041, (01)00884838099258(21)1416, (01)00884838099258(21)1754, (01)00884838099258(21)2333, (01)00884838099258(21)2040, (01)00884838099258(21)1658, (01)00884838099258(21)2169, (01)00884838099258(21)1756, (01)00884838099258(21)1152, (01)00884838099258(21)2726, (01)00884838099258(21)2761, (01)00884838099258(21)2808, (01)00884838099258(21)2763, (01)00884838099258(21)2807, (01)00884838099258(21)2762, (01)00884838099258(21)1292, (01)00884838099258(21)1384, (01)00884838099258(21)2173, (01)00884838099258(21)1822, (01)00884838099258(21)1641, (01)00884838099258(21)3110, (01)00884838099258(21)1291, (01)00884838099258(21)1442, (01)00884838099258(21)3186, (01)00884838099258(21)2545, (01)00884838099258(21)2624, (01)00884838099258(21)2697, (01)00884838099258(21)1532, (01)00884838099258(21)2695, (01)00884838099258(21)2118, (01)00884838099258(21)2338, (01)00884838099258(21)2825, (01)00884838099258(21)2823, (01)00884838099258(21)2337, (01)00884838099258(21)2290, (01)00884838099258(21)1749, (01)00884838099258(21)2109, (01)00884838099258(21)1439, (01)00884838099258(21)2639, (01)00884838099258(21)1130, (01)00884838099258(21)2766, (01)00884838099258(21)1992, (01)00884838099258(21)1195, (01)00884838099258(21)2249, (01)00884838099258(21)2890, (01)00884838099258(21)2067, (01)00884838099258(21)2065, (01)00884838099258(21)2826, (01)00884838099258(21)1752, (01)00884838099258(21)1755, (01)00884838099258(21)2677, (01)00884838099258(21)2382, (01)00884838099258(21)2785, (01)00884838099258(21)1438, (01)00884838099258(21)1996, (01)00884838099258(21)1844, (01)00884838099258(21)1848, (01)00884838099258(21)1989, (01)00884838099258(21)1841, (01)00884838099258(21)2059, (01)00884838099258(21)1121, (01)00884838099258(21)772, (01)00884838099258(21)773, (01)00884838099258(21)2273, (01)00884838099258(21)2295, (01)00884838099258(21)2381, (01)00884838099258(21)2092, (01)00884838099258(21)2023, (01)00884838099258(21)2053, (01)00884838099258(21)1166, (01)00884838099258(21)1070, (01)00884838099258(21)2200, (01)00884838099258(21)1849, (01)00884838099258(21)2428, (01)00884838099258(21)1503, (01)00884838099258(21)1846, (01)00884838099258(21)1847, (01)00884838099258(21)2497, (01)00884838099258(21)2313, (01)00884838099258(21)2470, (01)00884838099258(21)1983, (01)00884838099258(21)2379, (01)00884838099258(21)2934, (01)00884838099258(21)2066, (01)00884838099258(21)1972, (01)00884838099258(21)2755, (01)00884838099258(21)1630, (01)00884838099258(21)2543, (01)00884838099258(21)2292, (01)00884838099258(21)1746, (01)00884838099258(21)2992, (01)00884838099258(21)1675, (01)00884838099258(21)2256, (01)00884838099258(21)1843, (01)00884838099258(21)2158, (01)00884838099258(21)2798, (01)00884838099258(21)1135, (01)00884838099258(21)1415, (01)00884838099258(21)2113, (01)00884838099258(21)1293, (01)00884838099258(21)2117, (01)00884838099258(21)1390, (01)00884838099258(21)2824, (01)00884838099258(21)2701, (01)00884838099258(21)1123, (01)00884838099258(21)2236, (01)00884838099258(21)2815, (01)00884838099258(21)2794, (01)00884838099258(21)1744, (01)00884838099258(21)2461, (01)00884838099258(21)2110, (01)00884838099258(21)2610, (01)00884838099258(21)1818, (01)00884838099258(21)2041, (01)00884838099258(21)2304, (01)00884838099258(21)2393, (01)00884838099258(21)2694, (01)00884838099258(21)1833, (01)00884838099258(21)2090, (01)00884838099258(21)2463, (01)00884838099258(21)2601, (01)00884838099258(21)2602, (01)00884838099258(21)2898, (01)00884838099258(21)2966, (01)00884838099258(21)1437, (01)00884838099258(21)2147, (01)00884838099258(21)1419, (01)00884838099258(21)2136, (01)00884838099258(21)1858, (01)00884838099258(21)1945, (01)00884838099258(21)2551, (01)00884838099258(21)1947, (01)00884838099258(21)1423, (01)00884838099258(21)1770, (01)00884838099258(21)1944, (01)00884838099258(21)1361, (01)00884838099258(21)2897, (01)00884838099258(21)2942, (01)00884838099258(21)1417, (01)00884838099258(21)2179, (01)00884838099258(21)1857, (01)00884838099258(21)1441, (01)00884838099258(21)1748, (01)00884838099258(21)1987, (01)00884838099258(21)2318, (01)00884838099258(21)1842, (01)00884838099258(21)2922, (01)00884838099258(21)1820, (01)00884838099258(21)2392, (01)00884838099258(21)1531, (01)00884838099258(21)3059, (01)00884838099258(21)2516, (01)00884838099258(21)2093, (01)00884838099258(21)1464, (01)00884838099258(21)1867, (01)00884838099258(21)2095, (01)00884838099258(21)2411, (01)00884838099258(21)1191, (01)00884838099258(21)2893, (01)00884838099258(21)2094, (01)00884838099258(21)2412, (01)00884838099258(21)2410, (01)00884838099258(21)2012, (01)00884838099258(21)2203, (01)00884838099258(21)1920, (01)00884838099258(21)2914, (01)00884838099258(21)1977, (01)00884838099258(21)1695, (01)00884838099258(21)1721, (01)00884838099258(21)1610, (01)00884838099258(21)2033, (01)00884838099258(21)2418, (01)00884838099258(21)2108, (01)00884838099258(21)1136, (01)00884838099258(21)2206, (01)00884838099258(21)1616, (01)00884838099258(21)1414, (01)00884838099258(21)1451, (01)00884838099258(21)2895, (01)00884838099258(21)2827, (01)00884838099258(21)1769, (01)00884838099258(21)1778, (01)00884838099258(21)2851, (01)00884838099258(21)2964, (01)00884838099258(21)1137, (01)00884838099258(21)660, (01)00884838099258(21)2542, (01)00884838099258(21)1016, (01)00884838099258(21)68, (01)00884838099258(21)2999, (01)00884838099258(21)2848, (01)00884838099258(21)529, (01)00884838099258(21)528, (01)00884838099258(21)293, (01)00884838099258(21)2361, (01)00884838099258(21)2307, (01)00884838099258(21)1261, (01)00884838099258(21)2638, (01)00884838099258(21)221, (01)00884838099258(21)1772, (01)00884838099258(21)2375, (01)00884838099258(21)1034, (01)00884838099258(21)970, (01)00884838099258(21)1671, (01)00884838099258(21)1834, (01)00884838099258(21)207, (01)00884838099258(21)1868, (01)00884838099258(21)3013, (01)00884838099258(21)1733, (01)00884838099258(21)2814, (01)00884838099258(21)375, (01)00884838099258(21)1617, (01)00884838099258(21)2163, (01)00884838099258(21)1530, (01)00884838099258(21)1626, (01)00884838099258(21)122, (01)00884838099258(21)2443, (01)00884838099258(21)2597, (01)00884838099258(21)2048, (01)00884838099258(21)2584, (01)00884838099258(21)978, (01)00884838099258(21)1401, (01)00884838099258(21)97, (01)00884838099258(21)2344, (01)00884838099258(21)2756, (01)00884838099258(21)1618, (01)00884838099258(21)1284, (01)00884838099258(21)2309, (01)00884838099258(21)2670, (01)00884838099258(21)1358, (01)00884838099258(21)393, (01)00884838099258(21)2252, (01)00884838099258(21)2625, (01)00884838099258(21)2716, (01)00884838099258(21)2137, (01)00884838099258(21)1148, (01)00884838099258(21)2568, (01)00884838099258(21)2482, (01)00884838099258(21)2944, (01)00884838099258(21)1248, (01)00884838099258(21)2833, (01)00884838099258(21)2161, (01)00884838099258(21)2612, (01)00884838099258(21)1751, (01)00884838099258(21)1845, (01)00884838099258(21)1528, (01)00884838099258(21)2455, (01)00884838099258(21)2553, (01)00884838099258(21)1501, (01)00884838099258(21)2554, (01)00884838099258(21)1290, (01)00884838099258(21)1683, (01)00884838099258(21)1741, (01)00884838099258(21)2935, (01)00884838099258(21)3139, (01)00884838099258(21)2757, (01)00884838099258(21)1823, (01)00884838099258(21)1131, (01)00884838099258(21)2008, (01)00884838099258(21)2207, (01)00884838099258(21)2112, (01)00884838099258(21)1859, (01)00884838099258(21)1412, (01)00884838099258(21)2371, (01)00884838099258(21)1373, (01)00884838099258(21)1480, (01)00884838099258(21)688, (01)00884838099258(21)1227, (01)00884838099258(21)1022, (01)00884838099258(21)697, (01)00884838099258(21)699, (01)00884838099258(21)1428, (01)00884838099258(21)1242, (01)00884838099258(21)2517, (01)00884838099258(21)2909, (01)00884838099258(21)1207, (01)00884838099258(21)705, (01)00884838099258(21)1021, (01)00884838099258(21)1176, (01)00884838099258(21)1465, (01)00884838099258(21)759, (01)00884838099258(21)1068, (01)00884838099258(21)1496, (01)00884838099258(21)1552, (01)00884838099258(21)1792, (01)00884838099258(21)168, (01)00884838099258(21)197, (01)00884838099258(21)83, (01)00884838099258(21)212), 722282 ((01)00884838110571(21)57, (01)00884838110571(21)7, (01)00884838110571(21)42, (01)00884838110571(21)26, (01)00884838110571(21)91, (01)00884838110571(21)172, (01)00884838110571(21)61, (01)00884838110571(21)97, (01)00884838110571(21)11, (01)00884838110571(21)88, (01)00884838110571(21)10, (01)00884838110571(21)72, (01)00884838110571(21)8, (01)00884838110571(21)15).
|
| FEI Number |
3042175844
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 6
Best
Netherlands
|
Manufacturer Reason for Recall | Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Philips notified consignees on about 06/17/2026 via letter sent certified mail titled, URGENT Medical Device Correction. Consignees were instructed to circulate the notification letter to all users of the system for awareness and to follow the instructions. In the case motorized movements are unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2) and rotational (3) stand movements remain available. Note: Transversal stand movements (2) are only available for ceiling-mounted systems with the FlexMove option. For floor-mounted systems, only rotational (3) stand movements remain available. The table longitudinal position (4) can be adjusted manually, when the table is not tilted.
Additionally, consignees were instructed to provide a copy of the notification letter to any organizations to which affected units have been transferred to, to keep the notification letter with system documentation until the correction is complete and complete and return the provided response form.
Philips will inspect all affected systems and replace HDDs that are not according to the specifications, as well as correct configurations where different HDD types are used within a single system. Philips expects to start the implementation of this correction in September 2026. |
| Quantity in Commerce | 3,341 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|