| | Class 2 Device Recall Stryker |  |
| Date Initiated by Firm | July 30, 2026 |
| Date Posted | September 16, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3199-2026 |
| Recall Event ID |
99558 |
| 510(K)Number | K261624 |
| Product Classification |
Prosthesis, hip, hemi-, femoral, metal - Product Code KWL
|
| Product | ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm
C-M079-2-801
ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID
C-M071-6-700
ARC2F ACETABULAR INSERT 50mm OD x 28mm ID 10DEGREE HOOD
C-M071-8-400
BEARING AND AXLE BUSHING ASSY STANDARD ELBOW
C-M110-1-201
BUSHING AND BEARING USE C-MM100253-00
C-MM100253-02
DURACON TIB INSERT LG 12MM MED 14MM LAT
C-M066-3-000
EXETER WINGLESS CENTRALISER FOR C-M082-7-302
C-M082-7-303
EXETER WINGLESS CENTRALIZER XSM
C-M107-5-301
GMRS EXTRA SM ALL-POLY TIBIA SIZE 11mm
C-MM100616-02
GMRS EXTRA SM ALL-POLY TIBIA SIZE 16mm
C-MM100616-03
GMRS EXTRA SM ALL-POLY TIBIA SIZE 8mm
C-MM100616-01
GMRS SML POLY TIBIA 11MM CASE 100520
C-MM100520-03
GMRS SML POLY TIBIA 16MM CASE 100520
C-MM100520-04
GMRS SML POLY TIBIA 8MM CASE 100520
C-MM100520-02
GMRS SML TIBIA 11MM WITH CASE 100411
C-MM100411-03
GMRS SML TIBIA 14MM WITH CASE 100411
C-MM100411-04
GMRS SML TIBIA 8MM WITH CASE 100411
C-MM100411-02
GMRS TIB INSRT SML 10MM WITH CASE 100507
C-MM100507-04
GMRS TIB INSRT SML 16MM WITH CASE 100507
C-MM100507-06
KALI TIBIAL INSERT LARGE RIGHT 8mm
C-M071-6-601
KRH REV TIB INS 16mm THK .655 DIA STEM
C-M113-5-603
MRH KNEE TIBIAL INSERT FOR THIN POST
C-M067-3-103
MRS RADIAL BUSHING FORITH C-M107-7-800
C-M107-7-801
SELF-PRESSURIZING GLENOID SIZE 9
C-M064-4-300
TRIATHLON NO. 9 TIBIAL INSERT 11MM
C-M067-5-003
TRIATHLON NO. 9 TIBIAL INSERT 9MM
C-M067-5-002
TRIDENT ALL POLY CUP 44 X 66MM
C-M066-0-901
TRIDENT CONSTRAINED ACETABULAR INSERT SIZE A X 22mm ID UH1
C-M104-2-002
TRIDENT CONSTRAINED ACETABULAR INSERT SIZE B X 22mm ID UH1
C-M104-2-003
UHR BIPOLAR 22X38MM
UH1-38-22
UHR BIPOLAR 22X40MM
UH1-40-22
UHR BIPOLAR 22X41MM
UH1-41-22
UHR BIPOLAR 22X42MM
UH1-42-22
UHR BIPOLAR 22X43MM
UH1-43-22
UHR BIPOLAR 22X45MM
UH1-45-22
UHR BIPOLAR 26X41mm
UH1-41-26
UHR BIPOLAR 26X42mm
UH1-42-26
UHR BIPOLAR 26X43mm
UH1-43-26
UHR BIPOLAR 26X44mm
UH1-44-26
UHR BIPOLAR 26X45mm
UH1-45-26
UHR BIPOLAR 26X46mm
UH1-46-26
UHR BIPOLAR 26X47mm
UH1-47-26
UHR BIPOLAR 26X48mm
UH1-48-26
UHR BIPOLAR 28X42mm
UH1-42-28
UHR BIPOLAR 28X44mm
UH1-44-28
UHR BIPOLAR 28X45mm
UH1-45-28
UHR BIPOLAR 28X46mm
UH1-46-28
UHR BIPOLAR 28X47mm
UH1-47-28
UHR BIPOLAR 28X48mm
UH1-48-28
Orthopaedic implant components manufactured from UHMWPE (polyethylene) |
| Code Information |
C-M079-2-801
No UDI
Lot G3264034
C-M071-6-700
No UDI
Lot G3043883
C-M071-8-400
No UDI
Lot G3043878
C-M110-1-201
UDI-DI 07613327504811
Lot 573P37
C-MM100253-02
No UDI
Lot N8873H
C-M066-3-000
No UDI
Lot RX06630A
C-M082-7-303
No UDI
Lot L6237
C-M107-5-301
No UDI
Lot 193607
C-MM100616-02
No UDI
Lot 195261
C-MM100616-03
No UDI
Lot 195262
C-MM100616-01
No UDI
Lot 195260
C-MM100520-03
No UDI
Lot 195081
C-MM100520-04
No UDI
Lot 195082
C-MM100520-02
No UDI
Lot 195080
C-MM100411-03
No UDI
Lot 194962
C-MM100411-04
No UDI
Lot 194963
C-MM100411-02
No UDI
Lot 194961
C-MM100507-04
No UDI
Lot LNH795
C-MM100507-06
No UDI
Lot LNH797
C-M071-6-601
No UDI
Lot G3055684
C-M113-5-603
No UDI
Lot 194594
C-M067-3-103
No UDI
Lot MHP0AX
C-M107-7-801
No UDI
Lot 193479
C-M064-4-300
No UDI
Lot RX06443A
C-M067-5-003
No UDI
Lot RX06750D
C-M067-5-002
No UDI
Lot RX06750C
C-M066-0-901
No UDI
Lot RX06609B
C-M104-2-002
No UDI
Lot 192371
C-M104-2-003
No UDI
Lot 192370
UH1-38-22
UDI-DI 07613327016789
Lots 2A01TA
430715
49510K
7E5X23
852J98
D23664
NE659N
UH1-40-22
UDI-DI 07613327016796
Lots 128LXW
16182H
1J5955
1T8203
2137N4
2541YM
2Y52Y2
4E19VD
4K7XYL
577M4L
5D3AVR
746AV4
7V8J0V
828TAR
875M72
8V63LX
9A3W5N
A91P07
DL4D38
LW709R
RA6A1H
VT4K59
X95RTV
UH1-41-22
UDI-DI 07613327016819
Lots 343653
4P1TR7
521MLK
7Y8HEA
975A6Y
AR0KX5
E788E6
KL1YTE
KN0086
P043ME
PL4L5M
UH1-42-22
UDI-DI 07613327016833
Lots 1R2VD1
2E84HD
2X333W
4L5YMY
6J2W2P
7238R2
7883W1
7H2L8M
8K315D
AK2P91
ET3HWR
L16J85
L93PY2
ME1485
MY1N20
P987D0
R92PDW
TM0KR0
X884ET
XD3961
YK39TT
UH1-43-22
UDI-DI 07613327016840
Lots 11138V
23005R
481KR9
516JLV
761T5M
9W3NWV
DA0W51
EK6069
KP0AAD
MD1Y5V
NT68KL
RT8N8H
RY828P
UH1-45-22
UDI-DI 07613327016901
Lots 8Y0KWH
HR869M
R7168K
WR4MT4
UH1-41-26
UDI-DI 07613327016802
Lots 157V1T
1D33YX
1K1E6M
1R5LWT
218RED
254MT3
287D4A
2K7HXW
2X5MTA
3J1A39
3K3P0T
473K4V
5A1T5N
5D6KJ5
5V7H82
6J8XKT
6W1MKA
8R29TH
9T7N8Y
AL5K8M
AW02V8
DT8JEW
H04PHL
JR6R9L
LN70PY
MA6M6E
PE7T6P
PY8PLH
RY49EW
T1497X
V70PNP
VJ6WXM
X82M1D
UH1-42-26
UDI-DI 07613327016826
Lots 186JE6
5Y1RL7
618AMX
7J0XET
MLKT5A
MLL022
MLL05K
R50NW2
UH1-43-26
UDI-DI 07613327016864
Lots 192J0R
1D45L5
1H5V53
2P2A99
2V6D0R
3248V2
334851
386P0V
3N33VT
402RP3
4644KD
4J5Y6H
4L3JPP
5H1A45
5Y2422
626J60
6E33NR
6J30X2
6M2KK8
6V092N
761EJP
7820L1
791NPH
7D0H4K
7K28Y8
7K63EE
7N7WX1
817AJJ
8J6TAL
8L6R73
8T3XDK
983JDR
998H5N
9E631T
A93KWD
AY40L3
DJ105H
E521L6
EX5DA9
HW7XYK
J94M7R
JT565R
JX0L8R
K939KP
KA30HV
KE5770
KP1Y0E
KV3L3H
MD76WR
MJ1DY9
MK2MN3
MLL54K
MLL65N
MLLV64
MR6D42
MW2J8W
N314DK
NP48X0
P147V0
P502R8
P92DV1
RL1TL1
RN8NM1
T66JK4
TA20N9
TT3EJR
TV197Y
TY7YV5
V68NXH
V91EK1
VW74NJ
VX5T02
VY0525
WA0VRE
WT7P1A
X48P1Y
Y63LYJ
Y648WL
YE3W2X
YJ6PHM
YL3HVY
YL6L7K
YX8NTJ
UH1-44-26
UDI-DI 07613327016888
Lots 336652
581XTY
5H1YRH
640Y4R
7Y7H14
9J7T4L
A881HX
JN7TXM
KT4P91
LJ44JX
M52APN
MLL2PT
MLL5X1
MLLD4R
MLLPK1
MLLV2Y
NR773K
NV0NVX
PD7P36
R74D1D
T31P16
VK7JL5
W91ENE
XD45XY
XN8EWL
Y385N4
YN1AJN
UH1-45-26
UDI-DI 07613327016925
Lots 056MNE
059MNE
188AWK
243WTY
352XY0
442K1P
8L58NK
AJ0H2R
AN15H5
HX6PAR
HY2V81
L50T23
MLL0AE
MLL5J4
MLL9TV
MLLKNT
N88PKJ
P31A9Y
RJ8JPR
UH1-46-26
UDI-DI 07613327016949
Lots 1839Y2
1M2AX3
1P2E82
1V5AN9
226AWD
232X83
2465J7
2K0P93
2R3XD6
3614YL
394D8M
3P7091
4V2PVP
5389M0
5N2N75
9X7YK5
A35V7K
DW3J8W
DY85VA
E242NY
H07J22
JV88X1
L04PPV
LY44K5
MLL50M
MLL789
MLLA53
NE45DD
V94NTT
YY5N98
UH1-47-26
UDI-DI 07613327016987
Lots 1Y2K34
21197T
3E0A48
4E7MKW
7K1VLW
A58HN9
J38RE2
J60A4E
LT43X0
M15HKN
MK442Y
MLL2LR
MLL52N
MLLRMW
MLLVRW
MLLWVH
XD724J
UH1-48-26
UDI-DI 07613327017007
Lots 1J8K9L
2Y6JK3
350D6V
3845V3
3D6216
44827N
580Y68
726W7P
7L7TPP
8E7A1J
9V8P94
9Y6HML
H88E74
K54MN4
L783M7
MLL3E9
MLL52W
MP56YN
N00HDA
N26TW7
TR4TT5
W016KY
XJ48DN
YA4YPN
UH1-42-28
UDI-DI 07613327016857
Lots 1R5MDV
275NEA
2V6KMY
4215LV
MLKEHT
MLLJ5X
MLLJK9
MLLV75
UH1-44-28
UDI-DI 07613327016895
Lots 106X12
121A72
123MX4
138KNV
1401YE
1429X2
154N34
1740KE
1830AR
186WEL
1D8LYA
1E4HK2
1E6W32
1E7RRN
1H1P69
1K10H5
1K58A6
1K63LK
1M1W89
1M6JWR
1N6VPP
1R8333
1V4LXR
1X8R7X
1Y2V83
1Y7DPY
201MY5
2116XX
2286V1
25090N
251WMR
2550NN
261Y8Y
287N2R
2D4N58
2X5T7Y
335351
335EXD
391JXR
3K6YR2
444636
4E45R8
4K15YY
4L17DL
4L29A4
4M7J5A
4N2091
4P3L83
4T85VA
4W227H
518WTR
521K2M
5246AR
5325D6
534JYN
555249
558TVA
581YM9
5D1Y5V
5H0EXH
5K0AMY
5P0H6D
5R3H24
5V2RJ0
600TDM
623LA2
6370A6
681Y0M
6885NJ
693MRX
698Y2M
6L4LT7
6N4J52
6T3JAY
6V19AE
6V6DW2
6W0A59
7307MN
758K2A
768V6X
7D5MN2
7L230N
7M6YVP
7X2PH3
7X7XM0
8125RN
867YNX
886V7A
8E6V9P
8M114H
8M4W64
8R438X
8X4PRJ
971KAN
987WJ0
9H673H
9K656K
9V0L0L
A56J67
A85T4Y
A86V5E
AA7VL4
AM3PPW
AW6JML
DD37HD
DH5N6T
DL5E4X
DR69YE
E47VAT
EA6WY1
EK5N6A
EK8KDD
EN17M3
ER2VAV
HW14NW
J477D9
J63M24
J644K6
JE7K73
JT2PYD
K47JDR
KD174L
L548RP
LE83EE
LH10XA
LM74JP
LR3KT5
LV0PP9
M45LKM
MLKMWA
MLL0X7
MLL3T3
MLL3X8
MLL3YN
MLL5V9
MLL6E1
MLL70V
MLLA90
MLLDXM
MLLEEY
MLLLKA
MLLNDM
MLLR0R
MLLY2V
MT2YK4
MT56MX
N082LN
N35W9W
N4476X
N60KXY
N780A7
N80J7M
NR5X7N
NW7H6M
PN3K4T
PV87N1
R20NVJ
RM8HJJ
RY4JDM
T6297W
T73HN0
TP366X
V281XV
V64XWE
V95MVD
VA7769
VL3RJJ
VP8E88
W33XH8
W969RX
WD03AK
WJ4YYD
WK2AHL
WW6MVP
X16M08
XM319D
XN437D
Y703W8
YJ012E
YV08TH
YY4TKX
UH1-45-28
UDI-DI 07613327016918
Lots 1A3013
360RPN
486EJ1
4N85AM
D6348Y
EY829J
MLL1V3
MLL210
MLL24J
MLL44A
MLL5RM
MLL94Y
MLLDXY
MLLJRN
MLLVWK
MLLW3M
MLLY7V
X43WJN
UH1-46-28
UDI-DI 07613327016956
Lots 1804DD
29772K
2E5PX4
3402DJ
3P0NDP
4H0W7N
4J5R3M
4K0NDA
5818XA
5R5H8L
6N77PT
6T493L
7K0L36
7L1KRT
8015WY
802XY1
8P7D00
905HH2
AD0E56
D3227P
E83XET
EV4W50
HT21XV
JP0TKR
L00E8N
M25670
MLL0Y0
MLL1W9
MLL46R
MLL499
MLL602
MLL8Y7
MLLAYE
MLLDY9
MLLJWN
MLLK40
MLLKEH
MLLNE3
MLLRL2
MLLV8T
MLLVTD
MLLXV3
MR613J
NT30DY
P72N17
VE411K
WH517T
XH01PP
Y66HVE
YE5XLP
UH1-47-28
UDI-DI 07613327016970
Lots 2821VP
2N68HT
3682T6
3M0Y7N
4E58YD
661DW6
8K6225
8M60J8
9D22DE
9Y0X3H
A68HK5
AL4LYE
J60DLA
KM65D9
LJ3RRT
MLL8V0
MLL984
MLLH9P
MLLLLD
MLLND6
PV1K03
PW100N
TD5KRD
WX0R02
UH1-48-28
UDI-DI 07613327017014
Lots 1831VA
1M4NYV
1T0L93
1W0E1J
254N4A
294M71
2A2RJ0
2T45AP
3K8VM4
3T6PD5
3Y8DJX
437ANA
448V4T
4623VM
484LE9
4X0M53
5D50WT
5L074T
5T0PX8
6E2MX7
6K5Y7V
6L43JP
7A1HVY
8J8DDN
975VXR
9786P2
9D6M48
9M6KHT
9T2991
9X5MJ3
A228M1
A54PLW
A77V1M
AV5ANE
AX18R8
DK0M6J
DY3LR8
E600KE
EY5P9V
H558EP
H63W11
HP74LE
J72DLP
J76YJ6
JK7JY4
JL1T01
K47WL6
K728DH
KT3K27
L35VMM
M15D88
ML731E
MLL18D
MLL562
MLL6MH
MLL74R
MLLM4T
MLLMWV
MLLNTV
MLLPV9
MLLV38
MLLX14
|
| FEI Number |
3003070421
|
Recalling Firm/ Manufacturer |
Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah
NJ
07430-2006
|
| For Additional Information Contact | Evelyn Ryan 201-831-5000 |
Manufacturer Reason for Recall | Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. |
FDA Determined Cause 2 | Process change control |
| Action | Firm began notifying customers on July 30, 2026, via FedEx 2-day air.
Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm. |
| Quantity in Commerce | 12,156 US |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|