| | Class 2 Device Recall Getinge PowerLED |  |
| Date Initiated by Firm | August 07, 2026 |
| Date Posted | September 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3224-2026 |
| Recall Event ID |
99696 |
| 510(K)Number | K070442 |
| Product Classification |
Light, surgical, ceiling mounted - Product Code FSY
|
| Product | PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm)
Model Numbers: 1. ARD568474114A, 2. ARD567501286; |
| Code Information |
1. Model Number: ARD568474114A;
UDI-DI: N/A;
Serial Numbers: 500190, 500194, 500201, 500203, 500209, 500216, 500218, 500197, 500204, 500210, 500211, 500212, 500213, 500214, 500217, 500206, 500390, 500393, 500397, 500414, 500416, 500417, 500418, 500451, 500266, 500284, 500291, 500293, 500294, 500299, 500300, 500301, 500302, 500306, 500307, 500308, 500117, 500116, 500273, 500274, 500399, 500400, 500402, 500168, 500188, 500192, 500196, 500208, 500219, 500221, 500228, 500233, 500237, 500240, 500247, 500159, 500207, 500215, 500396, 500415, 500419, 500429, 500430, 500433, 500434, 500437, 500439, 500440, 500443, 500478, 500394, 500398, 500403, 500407, 500391, 500404, 500060, 500061, 500062, 500063, 500080, 500081, 500086, 500089, 500094, 500095, 500096, 500099, 500105, 500106, 500111, 500112, 500118, 500119, 500392, 500395, 500405, 500406, 500409, 500427, 500428, 500435, 500436, 500238, 500249, 500251, 500254, 500256, 500258, 500259, 500260, 500261, 500262, 500277, 500278, 500285, 500296, 500304, 500309, 500311, 500312, 500316, 500323, 500328, 500331, 500332, 500338, 500198, 500200, 500067, 500073, 500074, 500076, 500079, 500082, 500069, 500161, 500078, 500051, 500053, 500093, 500100, 500107, 500108, 500113, 500114, 500065, 500070, 500071, 500072, 500075, 500077, 500084, 500085, 500087, 500143, 500145, 500152, 500164, 500167, 500178, 500179, 500253, 500255, 500267, 500268, 500286, 500287, 500292, 500224, 500225, 500229, 500235, 500236, 500128, 500134, 500137, 500148, 500149, 500129, 500173, 500177, 500182, 500184, 500185, 500189, 500195, 500199, 500205, 500169, 500170, 500180, 500142, 500144, 500166, 500172, 500174, 500175, 500181, 500242, 500160, 500162, 500165, 500176, 500157, 500158, 500163, 500171, 500068, 500088, 500131, 500132, 500133, 500151, 500220, 500223, 500226, 500241, 500245, 500248, 500064, 500313, 500315, 500317, 500318, 500320, 500324, 500325, 500326, 500327, 500329, 500330, 500333, 500334, 500335, 500337, 500092, 500097, 500098, 500101, 500102, 500103, 500104, 500109, 500110, 500115, 500120, 500121, 500408, 500359, 500361, 500362, 500363, 500364, 500365, 500368, 500369, 500370, 500372, 500380, 500232, 500239, 500263, 500264, 500269, 500271, 500272, 500373, 500374, 500375, 500386, 500387, 500453, 500454, 500455, 500456, 500461, 500462, 500492, 500493, 500279, 500303, 500305, 500413, 500422, 500425, 500426, 500431, 500438, 500444, 500445, 500446, 500130, 500135, 500136, 500138, 500141, 500146, 500150, 500153, 500154, 500280, 500281, 500282, 500297, 500028, 500029, 500031, 500032, 500033, 500036, 500040, 500041, 500042, 500049, 500056, 500183, 500193, 500257, 500270, 500275, 500276, 500401, 500410, 500411, 500412, 500423, 500424, 500441, 500442, 500447, 500448, 500420, 500421, 500432, 500243, 500037, 500083, 500125, 500126, 500127, 500139, 500140, 500147, 500250, 500252, 500265, 500283, 500288, 500289, 500290, 500295, 500298, 500030, 500034, 500035, 500038, 500039, 500043, 500044, 500045, 500046, 500047, 500048, 500050, 500052, 500054, 500055, 500057, 500227, 500230, 500234, 500244, 500246, 500066, 500222, 500231, 500314, 500319, 500321, 500322, 500336, 500360, 500366, 500367, 500371, 500376, 500377, 500378, 500379, 500381, 500382, 500383, 500384, 500385, 500388, 500186;
2. Model Number: ARD567501286;
UDI-DI: 03700712403478;
Serial Numbers: 1314H000217400, 4014H000279430, 4014H000279432, 4014H000279433, 4014H000279434, 4014H000279437, 4014H000279438, 4314H000288228, 4314H000288229, 4314H000288232, 4314H000288233, 4314H000288234, 4314H000288237, 4314H000288239, 4314H000288257, 4314H000288258, 4314H000288259, 4714H000296081, 4717H000296095, 1315H000382641, 1715H000392775, 2115H000409030, 2115H000409038, 2115H000409040, 2115H000409062, 2113H000123425, 3613H000150570, 2115H000409063, 3015H000436557, 3015H000436569, 3015H000436571, 2115H000409065, 2115H000409067, 2615H000423256, 3015H000436561, 3015H000436566, 3015H000436570, 1314H000215902, 1314H000215904, 0116H000537022, 0716H000208366, 2816H000647716, 0615H000334882, 0615H000334887, 0615H000334891, 0815H000351763, 0815H000351768, 0815H000351780, 1314H000215893, 1314H000215894, 1314H000215896, 1314H000215900, 1314H000215903, 1314H000215907, 1314H000215885, 2114H000236788, 2613H000132823, 2613H000132825, 2613H000132826, 2613H000132827, 2613H000132828, 2614H000244892, 3613H000150571, 3613H000150572, 4014H000279429, 4014H000279431, 4314H000288230, 4314H000288235, 1314H000215886, 1314H000215895, 1314H000215898, 1314H000215899, 1314H000215908, 2114H000236778, 2114H000236779, 2114H000236787, 2114H000236789, 2114H000236790, 2614H000244886, 2115H000409051, 2115H000409058, 3015H000436556, 3015H000436558, 3015H000436568, 4314H000288231, 1314H000215888, 1314H000215889, 1314H000215906, 1314H000215879, 1314H000215880, 1314H000215881, 1314H000215882, 1314H000215890, 1314H000215892, 1314H000215897, 1314H000215901, 2614H000244885, 2113H000123420, 2113H000123421, 2113H000123423, 2113H000123424, 2113H000123426, 2113H000123427, 2113H000123429, 2613H000132824, 2816H000647710, 2614H000244884, 2614H000244889, 2114H000236777, 2114H000236782, 2114H000236785, 2614H000244877, 2614H000244878, 2614H000244883, 2614H000244890, 1314H000215905, 2115H000409039, 0116H000537017, 0116H000537020, 0116H000537021, 0116H000537023, 0116H000537024, 0116H000537025, 0116H000537026, 0615H000334878, 0615H000334881, 0715H000391839, 0716H000261478, 0716H000561475, 0716H000561479, 0716H000561493, 1314H000217399, 2215H000537018, 2615H000423253, 2816H000647709, 2816H000647712, 2816H000647714, 2916H000647713, 3015H000436560, 3015H000436564, 4314H000288260, 4714H000296097, 1314H000215883, 1314H000215884, 1314H000215887, 0815H000351782, 2115H000409029, 2114H000236780, 2114H000236781, 2114H000236783, 2114H000236784, 2114H000236786, 2614H000244879, 2614H000244880, 2614H000244887, 2614H000244888, 2614H000244891, 2115H000409027, 2115H000409055, 2115H000409032, 2115H000409034, 2115H000409053, 2115H000409064, 2115H000409070, 2615H000423252, 2615H000423254, 3015H000436563, 0116H000537019, 0716H000561481, 3613H000150569, 0615H000334888, 0615H000334889, 0716H000561477, 0716H000561480, 0615H000334883, 0615H000334884, 2115H000409068, 2115H000409069, 2615H000423258, 3015H000436559, 3015H000436562, 3015H000436565, 3015H000436567, 4314H000288236, 4314H000288240, 0615H000334885, 0615H000334886, 0615H000334890, 0815H000351765, 0815H000351767, 0815H000351771, 0815H000351776, 0815H000351777, 0815H000351779, 1314H000215891, 2614H000244881, 2614H000244882, 4014H000279436, 4314H000288238, 4314H000288261, 1813H000118088, 1813H000118093, 2613H000132815, 2613H000132818, 3613H000150533, 3613H000150534, 3613H000150537, 3613H000150538, 3613H000150541, 2613H000132811, 3613H000150544, 3613H000150545, 3613H000150546, 3613H000150547, 3613H000150548, 3613H000150565, 3613H000150566, 3613H000150567, 3613H000150568, 0815H000351774, 4014H000279428, 2115H000409054, 0815H000351773, 1315H000382643, 1315H000382644, 2115H000409028, 2115H000409033, 2115H000409035, 2115H000409036, 2115H000409037, 2115H000409052, 2115H000409056, 2115H000409057, 2115H000409059, 2115H000409060, 2115H000409061, 2816H000647715, 0815H000351775, 0815H000351778, 4014H000279435, 3613H000150535, 3613H000150536, 3613H000150539, 3613H000150540, 3613H000150542, 3613H000150543, 2115H000409031, 0815H000351764, 0815H000351766, 0815H000351769, 0815H000351770, 0815H000351772, 0815H000351781, 1813H000118090, 2113H0001231, 2113H000123428, 2613H000132800, 2613H000132801, 2613H000132802, 2613H000132803, 2613H000132804, 2613H000132806, 2613H000132807, 2613H000132808, 2613H000132809, 2613H000132810, 2613H000132812, 2613H000132813, 2613H000132814, 2813H000132805; |
| FEI Number |
3002465150
|
Recalling Firm/ Manufacturer |
Maquet SAS
Parc De Limere
Avenue De La Pomme De Pin; Cs 10008 Ardon
Orleans Cedex 2
France
|
Manufacturer Reason for Recall | The potential for the device to detach
at the junction between the spring arm and suspension arm and fall causing potential injury. |
FDA Determined Cause 2 | Process control |
| Action | On August 7, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers.
Actions to be taken:
1. Identify any affected light(s) in your facility. Do not use affected light(s) until Getinge has completed the inspection and any required repairs. You may continue to use the room where the light is located. The light enclosure must remain closed; use an alternative light source as needed.
2. Forward this notification to anyone in your organization, or any organization Where the affected devices were transferred, who needs to be aware.
3. Complete and return the enclosed response form to recallresponses.qrc@getinge.com. |
| Quantity in Commerce | 667 |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FSY
|
|
|
|