• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall EVOS 3.5MM LATERAL PROXIMAL TIBIA PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EVOS 3.5MM LATERAL PROXIMAL TIBIA PLATEsee related information
Date Initiated by FirmSeptember 06, 2026
Date PostedOctober 30, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0318-2027
Recall Event ID 99904
510(K)NumberK210837 K213126 K170457 
Product Classification Plate, fixation, bone - Product Code HRS
ProductEVOS 3.5MM LATERAL PROXIMAL TIBIA PLATE 6 HOLE LEFT 91MM REF72463106
Code Information Lot: 25MM04429/UDI: 00885556640401
FEI Number 1020279
Recalling Firm/
Manufacturer
Smith & Nephew, Inc.
1450 E Brooks Rd
Memphis TN 38116-1804
For Additional Information ContactDavid Snyder
901-396-2121
Manufacturer Reason
for Recall
Incorrect labeling; 3.5MM Partial Articular Lateral Proximal Tibia Plate 6 hole Left 91MM implant labeled and lasered etched as as 3.5MM Lateral Proximal Tibia Plate 6 hole left 91mm implant
FDA Determined
Cause 2
Labeling Change Control
ActionOn September 3, 2026, Smith & Nephew, Inc. issued a Urgent Medical Device Recall Notice to affected consignees via 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. 1a. 0If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number and name in Section A. 2. If you have no product to return: a. Please put an X in the appropriate box below. b. Complete Section B of the form, sign and send to FieldActions@smith-nephew.com. 3. If you have product to return: a. Please complete Section A - list the product part number, batches, and quantities that you are returning. b. Complete Section B of the form, sign and send to FieldActions@smith-nephew.com. 4. Once the form is received by Smith+Nephew, you will be sent a Return Authorization (RA) number. 5. Return affected product to: Smith+Nephew | Attn: Quality Hold and Field Action Returns Department | 1450 E Brooks Road, Building G | Memphis, TN 38116
Quantity in Commerce7 units
DistributionUS Nationwide distribution in the states of AZ, NV, TX, NM, WA, NY.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = HRS
-
-