| | Class 1 Device Recall Cardinal Health Presource |  |
| Date Initiated by Firm | September 23, 2026 |
| Date Posted | November 07, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0329-2027 |
| Recall Event ID |
100064 |
| Product Classification |
Cardiovascular procedure kit - Product Code OEZ
|
| Product | Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100,labeled as:
1) OPEN HEART A,B,C, KIT CATALOG # PC69HEECD;
2) OPEN HEART A,B,C, KIT CATALOG # PC69HEEC2;
3) OPEN HEART A,B,C,KIT CATALOG # PC69HEEC3;
4) OPEN HEART A,B,C, KIT CATALOG # PC69HEEC4;
5) OPEN HEART A,B,C CATALOG # PC69HEECD;
6) OPEN HEART A,B,C CATALOG # PC69HEECE;
7) KIT, OPEN HEART A,B,C CATALOG # PC69HEECG;
8) OPEN HEART A,B,C CATALOG # PC69HEEC5;
9) OPEN HEART A,B,C CATALOG # PC69HEEC6;
10) OPEN HEART A,B,C CATALOG # PC69HEEC7;
11) OPEN HEART A,B,C CATALOG # PC69HEEC8;
12) KIT, OPEN HEART A,B,C CATALOG # PC69HEEC9;
13) KIT, OPEN HEART A,B,C CATALOG # PC69HEE10;
14) KIT, OPEN HEART A,B,C CATALOG # PC69HEE11;
15) KIT, OPEN HEART A,B,C CATALOG # PC69HEE12;
16) KIT, OPEN HEART A,B,C CATALOG # PC69HEE13;
17) KIT, OPEN HEART A,B,C CATALOG # PV69HEECA;
18) KIT, OPEN HEART A, B, C CATALOG # PV69HEECB;
19) KIT, OPEN HEART A, B, C CATALOG # PV69HEECC;
20) KIT, OPEN HEART A, B, C CATALOG # PV69HEECF;
21) KIT, OPEN HEART A, B, C CATALOG # PV69HEECG;
22) KIT, OPEN HEART A,B,C CATALOG # PV69HEEC1;
23) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC2;
24) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC3;
25) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC4;
26) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC5;
27) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC6;
28) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC7;
29) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC8;
30) KIT, OPEN HEART A, B, C CATALOG # PV69HEEC9;
31) KIT, OPEN HEART A, B, C CATALOG # PV69HEE10;
32) KIT, OPEN HEART A, B, C CATALOG # PV69HEE11;
33) KIT, OPEN HEART A, B, C CATALOG # PV69HEE12;
34) KIT, OPEN HEART A, B, C CATALOG # PV69HEE13;
35) KIT, OPEN HEART A, B, C CATALOG # PV69HEE14;
36) KIT, OPEN HEART A, B, C CATALOG # PV69HEE15;
37) KIT, OPEN HEART A, B, C CATALOG # PV69HEE16;
38) KIT, OPEN HEART A, B, C CATALOG # PV69HEE17;
39) KIT, OPEN HEART A, B, C CATALOG # PV69HEE18;
40) KIT, OPEN HEART A, B, C CATALOG # PV69HEE19;
41) KIT, OPEN HEART A, B, C CATALOG # PV69HEE20;
42) KIT, OPEN HEART A, B, C CATALOG # PV69HEE21;
43) KIT, OPEN HEART A, B, C CATALOG # PV69HEE23;
44) KIT, OPEN HEART A, B, C CATALOG # PV69HEE24;
45) KIT, OPEN HEART A, B, C CATALOG # PV69HEE25;
46) KIT, OPEN HEART A, B, C CATALOG # PV69HEE26;
47) KIT, OPEN HEART A, B, C CATALOG # PV69HEE27;
48) KIT, OPEN HEART A, B, C CATALOG # PV69HEE28;
49) KIT, OPEN HEART A, B, C CATALOG # PV69HEE29;
50) OPEN HEART MC KIT CATALOG # PC46OHASRA;
51) OPEN HEART MC KIT CATALOG # PC46OHASRB;
52) OPEN HEART MC CATALOG # PC46OHASRC;
53) OPEN HEART MC KIT CATALOG # PC46OHASRC;
54) OPEN HEART MC KIT CATALOG # PC46OHASRD;
55) OPEN HEART MC KIT CATALOG # PC46OHASRE;
56) OPEN HEART MC KIT CATALOG # PC46OHASRF;
57) OPEN HEART MC CATALOG # PC46OHASR1;
58) OPEN HEART MC KIT CATALOG # PC46OHASR2;
59) OPEN HEART-DT, SANTA ROSA KIT CATALOG # PC46OHSRA;
60) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRC;
61) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRD;
62) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRG;
63) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRH;
64) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRI;
65) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRJ;
66) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSRK;
67) OPEN HEART DT,SANTA ROSA KIT CATALOG # PC46OHSR1;
68) OPEN HEART DT, SANTA ROSA KIT CATALOG # PC46OHSR2;
69) OPEN HEART DT,SANTA ROSA KIT CATALOG # PC46OHSR3;
70 |
| Code Information |
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 264455;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 267607;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 267607;
CATALOG # PC69HEEC2, GTIN 887488003736 (ea) 50887488003731 (cs), WO # 276284;
CATALOG # PC69HEEC2, GTIN 887488003736 (ea) 50887488003731 (cs), WO # 276284;
CATALOG # PC69HEEC3, GTIN 887488038431 (ea) 50887488038436 (cs), WO # 284706;
CATALOG # PC69HEEC3, GTIN 887488038431 (ea) 50887488038436 (cs), WO # 284706;
CATALOG # PC69HEEC4, GTIN 10887488071039 (ea) 50887488071037 (cs), WO # 292763;
CATALOG # PC69HEEC4, GTIN 10887488071039 (ea) 50887488071037 (cs), WO # 292763;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 257377;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 265386;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 265386;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 295276;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 295276;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 295276;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 315421;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 315421;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 332659;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 343157;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 343157;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 352928;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 363702;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 363702;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 390767;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 390767;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 413827;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 413827;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 423178;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 438065;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 438065;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 438065;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 493906;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 493906;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 509951;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 509951;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 549544;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 568113;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 594603;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 594603;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 615044;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 615044;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 621754;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 621754;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 637878;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 647064;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 657485;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 659400;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 659400;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 681786;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 699479;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 699479;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 699479;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 715558;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 720808;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 720808;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 720808;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 720808;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 736186;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 736186;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 755170;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 755170;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 767313;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 767313;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 789583;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 789583;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 789583;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 811456;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 841899;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 841899;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 862818;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 862818;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 882686;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 882686;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 904306;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 904306;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 915163;
CATALOG # PC69HEECE, GTIN 10887488677965 (ea) 50887488677963 (cs), WO # 915163;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 116604;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 219042;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 219042;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 244124;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 244124;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 267543;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 267543;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 284139;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 284139;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 299867;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 299867;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 317325;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 317325;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 993974;
CATALOG # PC69HEECG, GTIN 10888439081039 (ea) 50888439081037 (cs), WO # 993974;
CATALOG # PC69HEEC5, GTIN 10887488107851 (ea) 50887488107859 (cs), WO # 234916;
CATALOG # PC69HEEC6, GTIN 10887488115702 (ea) 50887488115700 (cs), WO # 239606;
CATALOG # PC69HEEC7, GTIN 10887488241074 (ea) 50887488241072 (cs), WO # 335103;
CATALOG # PC69HEEC7, GTIN 10887488241074 (ea) 50887488241072 (cs), WO # 335103;
CATALOG # PC69HEEC8, GTIN 10887488549972 (ea) 50887488549970 (cs), WO # 535689;
CATALOG # PC69HEEC8, GTIN 10887488549972 (ea) 50887488549970 (cs), WO # 535689;
CATALOG # PC69HEEC9, GTIN 10888439049572 (ea) 50888439049570 (cs), WO # 942155;
CATALOG # PC69HEEC9, GTIN 10888439049572 (ea) 50888439049570 (cs), WO # 942155;
CATALOG # PC69HEEC9, GTIN 10888439049572 (ea) 50888439049570 (cs), WO # 942155;
CATALOG # PC69HEE10, GTIN 10888439160895 (ea) 50888439160893 (cs), WO # 970963;
CATALOG # PC69HEE11, GTIN 10888439312188 (ea) 50888439312186 (cs), WO # 145518;
CATALOG # PC69HEE11, GTIN 10888439312188 (ea) 50888439312186 (cs), WO # 145518;
CATALOG # PC69HEE11, GTIN 10888439312188 (ea) 50888439312186 (cs), WO # 145518;
CATALOG # PC69HEE12, GTIN 10888439377828 (ea) 50888439377826 (cs), WO # 174911;
CATALOG # PC69HEE12, GTIN 10888439377828 (ea) 50888439377826 (cs), WO # 174911;
CATALOG # PC69HEE13, GTIN 10888439425024 (ea) 50888439425022 (cs), WO # 193147;
CATALOG # PC69HEE13, GTIN 10888439425024 (ea) 50888439425022 (cs), WO # 193147;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 359802;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 359802;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 396436;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 396436;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 414761;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 414761;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 458352;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 458352;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 458352;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 471617;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 486656;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 494014;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 503300;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 514433;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 514433;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 548024;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 555690;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 569263;
CATALOG # PV69HEECA, GTIN 10888439611922 (ea) 50888439611920 (cs), WO # 569263;
CATALOG # PV69HEECB, GTIN 10195594954055 (ea) 50195594954053 (cs), WO # 704095;
CATALOG # PV69HEECB, GTIN 10195594954055 (ea) 50195594954053 (cs), WO # 793174;
CATALOG # PV69HEECB, GTIN 10195594954055 (ea) 50195594954053 (cs), WO # 804446;
CATALOG # PV69HEECB, GTIN 10195594954055 (ea) 50195594954053 (cs), WO # 826830;
CATALOG # PV69HEECB, GTIN 10195594954055 (ea) 50195594954053 (cs), WO # 838634;
CATALOG # PV69HEECC, GTIN 10197106134412 (ea) 50197106134410 (cs), WO # 11276;
CATALOG # PV69HEECC, GTIN 10197106134412 (ea) 50197106134410 (cs), WO # 11276;
CATALOG # PV69HEECC, GTIN 10197106134412 (ea) 50197106134410 (cs), WO # 49621;
CATALOG # PV69HEECC, GTIN 10197106134412 (ea) 50197106134410 (cs), WO # 49621;
CATALOG # PV69HEECC, GTIN 10197106134412 (ea) 50197106134410 (cs), WO # 989242;
CATALOG # PV69HEECF, GTIN 10197106700143 (ea) 50197106700141 (cs), WO # 385626;
CATALOG # PV69HEECF, GTIN 10197106700143 (ea) 50197106700141 (cs), WO # 385626;
CATALOG # PV69HEECF, GTIN 10197106700143 (ea) 50197106700141 (cs), WO # 385626;
CATALOG # PV69HEECF, GTIN 10197106700143 (ea) 50197106700141 (cs), WO # 429495;
CATALOG # PV69HEECF, GTIN 10197106700143 (ea) 50197106700141 (cs), WO # 444224;
CATALOG # PV69HEECG, GTIN 10198956089570 (ea) 50198956089578 (cs), WO # 594407;
CATALOG # PV69HEECG, GTIN 10198956089570 (ea) 50198956089578 (cs), WO # 600326;
CATALOG # PV69HEECG, GTIN 10198956089570 (ea) 50198956089578 (cs), WO # 743987;
CATALOG # PV69HEECG, GTIN 10198956089570 (ea) 50198956089578 (cs), WO # 792247;
CATALOG # PV69HEECG, GTIN 10198956089570 (ea) 50198956089578 (cs), WO # 792247;
CATALOG # PV69HEEC1, GTIN 10888439828672 (ea) 50888439828670 (cs), WO # 435995;
CATALOG # PV69HEEC2, GTIN 10195594001759 (ea) 50195594001757 (cs), WO # 592860;
CATALOG # PV69HEEC2, GTIN 10195594001759 (ea) 50195594001757 (cs), WO # 592860;
CATALOG # PV69HEEC2, GTIN 10195594001759 (ea) 50195594001757 (cs), WO # 592860;
CATALOG # PV69HEEC3, GTIN 10195594115722 (ea) 50195594115720 (cs), WO # 627575;
CATALOG # PV69HEEC3, GTIN 10195594115722 (ea) 50195594115720 (cs), WO # 627575;
CATALOG # PV69HEEC4, GTIN 10195594201203 (ea) 50195594201201 (cs), WO # 647412;
CATALOG # PV69HEEC4, GTIN 10195594201203 (ea) 50195594201201 (cs), WO # 647412;
CATALOG # PV69HEEC4, GTIN 10195594201203 (ea) 50195594201201 (cs), WO # 647412;
CATALOG # PV69HEEC5, GTIN 10195594236328 (ea) 50195594236326 (cs), WO # 652178;
CATALOG # PV69HEEC5, GTIN 10195594236328 (ea) 50195594236326 (cs), WO # 652178;
CATALOG # PV69HEEC6, GTIN 10195594321451 (ea) 50195594321459 (cs), WO # 682948;
CATALOG # PV69HEEC7, GTIN 10195594552732 (ea) 50195594552730 (cs), WO # 771834;
CATALOG # PV69HEEC7, GTIN 10195594552732 (ea) 50195594552730 (cs), WO # 771834;
CATALOG # PV69HEEC8, GTIN 10195594727093 (ea) 50195594727091 (cs), WO # 855372;
CATALOG # PV69HEEC8, GTIN 10195594727093 (ea) 50195594727091 (cs), WO # 855372;
CATALOG # PV69HEEC8, GTIN 10195594727093 (ea) 50195594727091 (cs), WO # 855372;
CATALOG # PV69HEEC8, GTIN 10195594727093 (ea) 50195594727091 (cs), WO # 855372;
CATALOG # PV69HEEC9, GTIN 10195594771324 (ea) 50195594771322 (cs), WO # 876732;
CATALOG # PV69HEE10, GTIN 10195594812256 (ea) 50195594812254 (cs), WO # 899661;
CATALOG # PV69HEE10, GTIN 10195594812256 (ea) 50195594812254 (cs), WO # 899661;
CATALOG # PV69HEE11, GTIN 10197106006917 (ea) 50197106006915 (cs), WO # 927590;
CATALOG # PV69HEE11, GTIN 10197106006917 (ea) 50197106006915 (cs), WO # 927590;
CATALOG # PV69HEE12, GTIN 10197106040270 (ea) 50197106040278 (cs), WO # 943204;
CATALOG # PV69HEE12, GTIN 10197106040270 (ea) 50197106040278 (cs), WO # 943204;
CATALOG # PV69HEE12, GTIN 10197106040270 (ea) 50197106040278 (cs), WO # 943204;
CATALOG # PV69HEE13, GTIN 10197106096116 (ea) 50197106096114 (cs), WO # 969974;
CATALOG # PV69HEE13, GTIN 10197106096116 (ea) 50197106096114 (cs), WO # 969974;
CATALOG # PV69HEE14, GTIN 10197106385029 (ea) 50197106385027 (cs), WO # 93405;
CATALOG # PV69HEE15, GTIN 10197106427750 (ea) 50197106427758 (cs), WO # 117274;
CATALOG # PV69HEE15, GTIN 10197106427750 (ea) 50197106427758 (cs), WO # 117274;
CATALOG # PV69HEE15, GTIN 10197106427750 (ea) 50197106427758 (cs), WO # 117274;
CATALOG # PV69HEE15, GTIN 10197106427750 (ea) 50197106427758 (cs), WO # 117274;
CATALOG # PV69HEE16, GTIN 10197106527139 (ea) 50197106527137 (cs), WO # 187810;
CATALOG # PV69HEE17, GTIN 10197106557686 (ea) 50197106557684 (cs), WO # 208126;
CATALOG # PV69HEE17, GTIN 10197106557686 (ea) 50197106557684 (cs), WO # 208126;
CATALOG # PV69HEE17, GTIN 10197106557686 (ea) 50197106557684 (cs), WO # 208126;
CATALOG # PV69HEE18, GTIN 10197106591338 (ea) 50197106591336 (cs), WO # 228281;
CATALOG # PV69HEE19, GTIN 10197106600894 (ea) 50197106600892 (cs), WO # 237286;
CATALOG # PV69HEE19, GTIN 10197106600894 (ea) 50197106600892 (cs), WO # 237286;
CATALOG # PV69HEE20, GTIN 10197106647172 (ea) 50197106647170 (cs), WO # 264190;
CATALOG # PV69HEE20, GTIN 10197106647172 (ea) 50197106647170 (cs), WO # 264190;
CATALOG # PV69HEE21, GTIN 10197106672778 (ea) 50197106672776 (cs), WO # 279752;
CATALOG # PV69HEE21, GTIN 10197106672778 (ea) 50197106672776 (cs), WO # 279752;
CATALOG # PV69HEE21, GTIN 10197106672778 (ea) 50197106672776 (cs), WO # 279752;
CATALOG # PV69HEE23, GTIN 10198956211407 (ea) 50198956211405 (cs), WO # 624895;
CATALOG # PV69HEE24, GTIN 10198956267404 (ea) 50198956267402 (cs), WO # 663441;
CATALOG # PV69HEE25, GTIN 10198956268074 (ea) 50198956268072 (cs), WO # 663431;
CATALOG # PV69HEE26, GTIN 10198956295896 (ea) 50198956295894 (cs), WO # 685081;
CATALOG # PV69HEE27, GTIN 10198956419520 (ea) 50198956419528 (cs), WO # 744437;
CATALOG # PV69HEE28, GTIN 10198956462922 (ea) 50198956462920 (cs), WO # 762071;
CATALOG # PV69HEE29, GTIN 10198956469662 (ea) 50198956469660 (cs), WO # 775756;
CATALOG # PC46OHASRA, GTIN 382602649444 (ea) 50382602649449 (cs), WO # 734420;
CATALOG # PC46OHASRA, GTIN 382602649444 (ea) 50382602649449 (cs), WO # 734420;
CATALOG # PC46OHASRB, GTIN 382602833904 (ea) 50382602833909 (cs), WO # 757044;
CATALOG # PC46OHASRB, GTIN 382602833904 (ea) 50382602833909 (cs), WO # 757044;
CATALOG # PC46OHASRC, GTIN 382602895544 (ea) 50382602895549 (cs), WO # 760996;
CATALOG # PC46OHASRC, GTIN 382602895544 (ea) 50382602895549 (cs), WO # 776296;
CATALOG # PC46OHASRC, GTIN 382602895544 (ea) 50382602895549 (cs), WO # 785348;
CATALOG # PC46OHASRC, GTIN 382602895544 (ea) 50382602895549 (cs), WO # 785348;
CATALOG # PC46OHASRC, GTIN 382602895544 (ea) 50382602895549 (cs), WO # 789539;
CATALOG # PC46OHASRD, GTIN 382603089607 (ea) 50382603089602 (cs), WO # 793389;
CATALOG # PC46OHASRD, GTIN 382603089607 (ea) 50382603089602 (cs), WO # 793389;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 805027;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 805027;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 805027;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 805027;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 808904;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 808904;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 819655;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 819655;
CATALOG # PC46OHASRE, GTIN 382603153179 (ea) 50382603153174 (cs), WO # 819655;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 827444;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 827444;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 832190;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 832190;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 837239;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 837239;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 846602;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 846602;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 851416;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 851416;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 851416;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 862246;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 868690;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 868690;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 871438;
CATALOG # PC46OHASRF, GTIN 382603375977 (ea) 50382603375972 (cs), WO # 871438;
CATALOG # PC46OHASR1, GTIN 382602961119 (ea) 50382602961114 (cs), WO # 767063;
CATALOG # PC46OHASR1, GTIN 382602961119 (ea) 50382602961114 (cs), WO # 767063;
CATALOG # PC46OHASR2, GTIN 382603609546 (ea) 50382603609541 (cs), WO # 853545;
CATALOG # PC46OHASR2, GTIN 382603609546 (ea) 50382603609541 (cs), WO # 853545;
CATALOG # PC46OHSRA, GTIN 382602388688 (ea) 50382602388683 (cs), WO # 690242;
CATALOG # PC46OHSRA, GTIN 382602388688 (ea) 50382602388683 (cs), WO # 709218;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 731486;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 731486;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 731486;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 731486;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 753249;
CATALOG # PC46OHSRC, GTIN 382602653298 (ea) 50382602653293 (cs), WO # 753249;
CATALOG # PC46OHSRD, GTIN 382602834130 (ea) 50382602834135 (cs), WO # 757470;
CATALOG # PC46OHSRD, GTIN 382602834130 (ea) 50382602834135 (cs), WO # 757470;
CATALOG # PC46OHSRG, GTIN 382602911961 (ea) 50382602911966 (cs), WO # 787827;
CATALOG # PC46OHSRG, GTIN 382602911961 (ea) 50382602911966 (cs), WO # 787827;
CATALOG # PC46OHSRG, GTIN 382602911961 (ea) 50382602911966 (cs), WO # 789548;
CATALOG # PC46OHSRH, GTIN 382603089577 (ea) 50382603089572 (cs), WO # 798585;
CATALOG # PC46OHSRH, GTIN 382603089577 (ea) 50382603089572 (cs), WO # 798585;
CATALOG # PC46OHSRH, GTIN 382603089577 (ea) 50382603089572 (cs), WO # 798585;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 801433;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 813995;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 819870;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 824650;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 824650;
CATALOG # PC46OHSRI, GTIN 382603153155 (ea) 50382603153150 (cs), WO # 824650;
CATALOG # PC46OHSRJ, GTIN 382603375151 (ea) 50382603375156 (cs), WO # 830059;
CATALOG # PC46OHSRJ, GTIN 382603375151 (ea) 50382603375156 (cs), WO # 833155;
CATALOG # PC46OHSRJ, GTIN 382603375151 (ea) 50382603375156 (cs), WO # 833155;
CATALOG # PC46OHSRJ, GTIN 382603375151 (ea) 50382603375156 (cs), WO # 839321;
CATALOG # PC46OHSRJ, GTIN 382603375151 (ea) 50382603375156 (cs), WO # 839321;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 845334;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 845334;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 848470;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 855003;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 855003;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 865974;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 865974;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 868318;
CATALOG # PC46OHSRK, GTIN 382603498447 (ea) 50382603498442 (cs), WO # 875612;
CATALOG # PC46OHSR1, GTIN 382602656312 (ea) 50382602656317 (cs), WO # 723832;
CATALOG # PC46OHSR2, GTIN 382602930870 (ea) 50382602930875 (cs), WO # 764167;
CATALOG # PC46OHSR3, GTIN 382602978810 (ea) 50382602978815 (cs), WO # 770150;
CATALOG # PC46OHSR3, GTIN 382602978810 (ea) 50382602978815 (cs), WO # 770150;
CATALOG # PC46OHSR3, GTIN 382602978810 (ea) 50382602978815 (cs), WO # 770150;
CATALOG # PC46OHSR3, GTIN 382602978810 (ea) 50382602978815 (cs), WO # 770150;
CATALOG # PC69HEECA, GTIN 885425543604 (ea) 50885425543609 (cs), WO # 112827;
CATALOG # PC69HEECA, GTIN 885425543604 (ea) 50885425543609 (cs), WO # 112827;
CATALOG # PC69HEECA, GTIN 885425543604 (ea) 50885425543609 (cs), WO # 116774;
CATALOG # PC69HEECA, GTIN 885425543604 (ea) 50885425543609 (cs), WO # 116774;
CATALOG # PC69HEECA, GTIN 885425543604 (ea) 50885425543609 (cs), WO # 116774;
CATALOG # PC69HEECB, GTIN 885425592350 (ea) 50885425592355 (cs), WO # 134257;
CATALOG # PC69HEECC, GTIN 885425672151 (ea) 50885425672156 (cs), WO # 149655;
CATALOG # PC69HEECC, GTIN 885425672151 (ea) 50885425672156 (cs), WO # 153778;
CATALOG # PC69HEECC, GTIN 885425672151 (ea) 50885425672156 (cs), WO # 159538;
CATALOG # PC69HEECC, GTIN 885425672151 (ea) 50885425672156 (cs), WO # 159538;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 167390;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 167390;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 173105;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 173105;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 177300;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 177300;
CATALOG # PC69HEECD, GTIN 885425752686 (ea) 50885425752681 (cs), WO # 177300;
CATALOG # PC69HEEC1, GTIN 885425742618 (ea) 50885425742613 (cs), WO # 147352;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 908216;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 916243;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 927775;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 927775;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 927775;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 927775;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 941334;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 951164;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 956409;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 956409;
CATALOG # PC69OHVQJ, GTIN 8854250644574 (ea) 58854250644579 (cs), WO # 956409;
CATALOG # PC69OHVQ9, GTIN 8854250943448 (ea) 58854250943443 (cs), WO # 966101;
CATALOG # PC69OHVQ9, GTIN 8854250943448 (ea) 58854250943443 (cs), WO # 966101;
CATALOG # PC6AOHNBA, GTIN 8854250346416 (ea) 58854250346411 (cs), WO # 146857;
CATALOG # PC6AOHNBB, GTIN 382603883120 (ea) 50382603883125 (cs), WO # 153205;
CATALOG # PC6AOHNBB, GTIN 382603883120 (ea) 50382603883125 (cs), WO # 156521;
CATALOG # PC6AOHNBB, GTIN 382603883120 (ea) 50382603883125 (cs), WO # 156521;
CATALOG # PC6AOHNBB, GTIN 382603883120 (ea) 50382603883125 (cs), WO # 156521;
CATALOG # PC69OHVQD, GTIN 382603104799 (ea) 50382603104794 (cs), WO # 990479;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 100196;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 100196;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 103884;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 103884;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 107903;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 107903;
CATALOG # PC69OHVQE, GTIN 382603151564 (ea) 50382603151569 (cs), WO # 995116;
CATALOG # PC69OHVQF, GTIN 382603385563 (ea) 50382603385568 (cs), WO # 111544;
CATALOG # PC69OHVQF, GTIN 382603385563 (ea) 50382603385568 (cs), WO # 111544;
CATALOG # PC69OHVQF, GTIN 382603385563 (ea) 50382603385568 (cs), WO # 117501;
CATALOG # PC69OHVQF, GTIN 382603385563 (ea) 50382603385568 (cs), WO # 117501;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 133284;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 133284;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 133284;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 133284;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 143826;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 143826;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 146083;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 146083;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 146083;
CATALOG # PC69OHVQH, GTIN 382603575322 (ea) 50382603575327 (cs), WO # 151998;
CATALOG # PC69OHVQI, GTIN 8854250005580 (ea) 58854250005585 (cs), WO # 154864;
CATALOG # PC69OHVQI, GTIN 8854250005580 (ea) 58854250005585 (cs), WO # 154864;
CATALOG # PC69OHVQI, GTIN 8854250005580 (ea) 58854250005585 (cs), WO # 159781;
CATALOG # PC69OHVQI, GTIN 8854250005580 (ea) 58854250005585 (cs), WO # 159781;
CATALOG # PC69OHVQI, GTIN 8854250005580 (ea) 58854250005585 (cs), WO # 162415;
CATALOG # PC69OHVQ3, GTIN 382603538877 (ea) 50382603538872 (cs), WO # 121277;
CATALOG # PC69OHVQ3, GTIN 382603538877 (ea) 50382603538872 (cs), WO # 121277;
CATALOG # PC69OHVQ3, GTIN 382603538877 (ea) 50382603538872 (cs), WO # 121277;
CATALOG # PC69OHVQ4, GTIN 382603561936 (ea) 50382603561931 (cs), WO # 123085;
CATALOG # PC69OHVQ4, GTIN 382603561936 (ea) 50382603561931 (cs), WO # 123085;
CATALOG # PC69OHVQ5, GTIN 382603605777 (ea) 50382603605772 (cs), WO # 126995;
CATALOG # PC69OHVQ5, GTIN 382603605777 (ea) 50382603605772 (cs), WO # 126995;
CATALOG # PC69OHVQ6, GTIN 382603623801 (ea) 50382603623806 (cs), WO # 128406;
CATALOG # PC69OHVQ6, GTIN 382603623801 (ea) 50382603623806 (cs), WO # 128406;
CATALOG # PC69OHVQ6, GTIN 382603623801 (ea) 50382603623806 (cs), WO # 128406;
CATALOG # PC69OHVQ7, GTIN 382603860435 (ea) 50382603860430 (cs), WO # 150103;
CATALOG # PC69OHVQ7, GTIN 382603860435 (ea) 50382603860430 (cs), WO # 150103;
CATALOG # PC69OHVQ7, GTIN 382603860435 (ea) 50382603860430 (cs), WO # 150103;
CATALOG # PC69OHVQ8, GTIN 382603952512 (ea) 50382603952517 (cs), WO # 158525;
CATALOG # PC69OHVQ8, GTIN 382603952512 (ea) 50382603952517 (cs), WO # 158525;
CATALOG # PV69HEE33, GTIN 10198956584518 (ea) 50198956584516 (cs), WO # 878931;
CATALOG # PV69HEE32, GTIN 10198956563506 (ea) 50198956563504 (cs), WO # 854949.
|
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Cardinal Health Quality Systems 800-292-9332 |
Manufacturer Reason for Recall | Notification of performance disclosures of kit component. |
FDA Determined Cause 2 | Other |
| Action | Cardinal Health issued a PRODUCT ADVISORY to its consignees on 10/23/2026 via US mail. The notice explained this was not a recall , but rather an advisory notice to alert the clinicians of this correction. The Cardinal Health Correction notice should be shared with all users. Enclosed in the notice was a copy of the letter from BD. The BD notice explained the issue, clinical impact, and provided Clinical Recommendations for Sets with Updated Performance. The recommendations include using an alternative pump if the BD Alaris Pump with identified infusion sets are not appropriate, and if no alternatives are available to utilized with caution enhanced clinical monitoring, and adherence to previously communicated warnings. Also, implement the Risk Mitigation steps outlined in the v12.3-v12.6 BD Alaris Infusion System User Manual, summarized:
Rate and Bolus Accuracy
1) Avoid hanging the solution container below the BD Alaris Pump Module, or higher than 21 inches above the top of the BD Alaris Pump Module.
Time to Alarm (TTA)
1) When loading the set ensure tubing is directly over the pressure sensors.
2) Ensure all clamps are open; there are no kinks or collapses in the tubing outside of the pump; drops are flowing in the drip chamber. Ensure the vent is open if applicable.
3) Do not use Selectable pressure mode with a setting above 50 mmHg for rates less than 10 mL/h.
Post Occlusion Bolus Volume (POBV)
1) When addressing an occlusion (clearing) ensure that the fluid flow to patient is OFF.
2) Disconnect the tubing to relieve excess pressure through a stopcock if present. Weigh the risks of disconnection with the risks of an unintended bolus.
Discontinuation of Sets and Recommendations:
Immediately discard any sets listed within Attachment C: Discontinued Alaris Pump Sets per your facility's procedures.
a. To receive credit for discarded product, please return the completed Customer Response Form attached to this notice.
If a set is included on Attachment C as disc |
| Quantity in Commerce | 9571 units |
| Distribution | US: CA, TX, NC |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|