• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Cholesterol Reagent

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Cholesterol Reagentsee related information
Date Initiated by FirmNovember 07, 2002
Date PostedDecember 31, 2002
Recall Status1 Terminated 3 on May 15, 2012
Recall NumberZ-0404-03
Recall Event ID 25081
510(K)NumberK880237 
Product Classification Enzymatic Esterase--Oxidase, Cholesterol - Product Code CHH
ProductElan Diagnostics CHO ATAC Cholesterol Reagent Kit Product Number: 516-250
Code Information Lot Number: 2030000 Exp.Date: 11/30/03
FEI Number 1221157
Recalling Firm/
Manufacturer
Elan Diagnostics, Inc.
2 Thurber Boulevard
Smithfield RI 02917
For Additional Information ContactPeter J. Sacchetti
417-401-233 Ext. 3526
Manufacturer Reason
for Recall
Potential for reporting lower than actual HDL (cholesterol) results due to low QC Recovery
FDA Determined
Cause 2
Other
ActionThe firm notified Distributors by telephone on 11/4/02 and end-users by Certified Mail on 11/4/02. Distributors and End-users are requested to dispose of product and return label for replacement product.
Quantity in Commerce170 Kits
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.