Date Initiated by Firm |
September 27, 2002 |
Date Posted |
February 01, 2003 |
Recall Status1 |
Terminated 3 on January 12, 2005 |
Recall Number |
Z-0526-03 |
Recall Event ID |
25248 |
510(K)Number |
K953910
|
Product Classification |
System, X-Ray, Mobile - Product Code IZL
|
Product |
Philips BV300 Mobile X-Ray System |
Code Information |
Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076 |
Recalling Firm/ Manufacturer |
Philips Medical Systems 22100 Bothell Everett Highway Bothell WA 98041
|
For Additional Information Contact |
Sarah Baxter 425-487-7665
|
Manufacturer Reason for Recall |
The C-bow support can bend and break allowing C-arm to fall from support
|
FDA Determined Cause 2 |
Other |
Action |
The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct. |
Quantity in Commerce |
229 devices |
Distribution |
The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZL and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA, INC.
|