 Class 2 Device Recall Medtronic InSync II Marquissee related information
Date Initiated by FirmFebruary 09, 2005
Date PostedMarch 11, 2005
Recall Status1 Terminated 3 on December 28, 2006
Recall NumberZ-0604-05
Recall Event ID 31142
PMA NumberP010031S005 
Product Classification Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) - Product Code NIK
ProductMedtronic InSync II Marquis, model 7289 Dual Chamber Implantable Cardioverter Defibrillator (ICD) with Cardiac Resynchronization Therapy devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capability of transmitting via telemetry the programmed values, measured and collected data, and event markers.
Code Information Model 7289 InSync II Marquis (serial numbers listed include all affected devices but may include some unaffected devices. Patients are recommended to check with their doctor for exact serial numbers affected.) S-Series: PRJ600001 - PRJ609639S and PRJ614813 - PRJ614815S
Recalling Firm/
Manufacturer
Medtronic Inc. Cardiac Rhythm Managment
7000 Central Ave Ne
Fridley MN 55432
For Additional Information ContactTim Samsel
763-514-3149
Manufacturer Reason
for Recall
Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.
FDA Determined
Cause 2
Other
ActionMedtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
Quantity in Commerce7235 ONLY distributed within US
Distributionworldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-