| Date Initiated by Firm | March 16, 2005 |
| Date Posted | June 22, 2005 |
| Recall Status1 |
Terminated 3 on June 30, 2006 |
| Recall Number | Z-0916-05 |
| Recall Event ID |
31822 |
| 510(K)Number | K021786 |
| Product Classification |
Passer, Wire, Orthopedic - Product Code HXI
|
| Product | EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055. |
| Code Information |
Lot Number: 069811; 120055; 120056; 120057; 123827; 125637; 125638. |
| FEI Number |
2242816
|
Recalling Firm/ Manufacturer |
EBI, L.P. 100 Interpace Pkwy Parsippany NJ 07054-1149
|
| For Additional Information Contact | James Pastena 973-299-9300 Ext. 3302 |
Manufacturer Reason for Recall | The wire appears to be soft and the wire bent during insertion. This causes kinks in which the reamer could not pass over the bent/kinked wire. |
FDA Determined Cause 2 | Other |
| Action | The firm distributed a recall letter to their EBI Fracture Sales Representatives and their EBI Trauma Distributors on 3/16/2005. |
| Quantity in Commerce | 280 units |
| Distribution | These devices are distributed nationwide to sales representatives who provide the product directly to the hospital/physician at the time of the surgery. The product was also distributed to Italy and Spain. There are no government accounts. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = HXI
|