| Date Initiated by Firm | August 04, 2006 |
| Date Posted | November 03, 2006 |
| Recall Status1 |
Terminated 3 on December 06, 2006 |
| Recall Number | Z-0119-2007 |
| Recall Event ID |
35934 |
| 510(K)Number | K972525 |
| Product Classification |
mechanical cardiopulmonary resuscitator - Product Code DRM
|
| Product | Michigan Instruments Thumper Mechanical CPR Device; Model 1007. |
| Code Information |
All units with a manual configuration (as opposed to a pneumatic configuration). |
Recalling Firm/ Manufacturer |
Michigan Instruments, Inc. 4717 Talon Court S.E. Grand Rapids MI 49512-5408
|
| For Additional Information Contact | 800-530-9939 Ext. 312 |
Manufacturer Reason for Recall | Failure to initiate compressions when first turned on, if improperly shut down-Operator Manual updated to include proper shut down procedures |
FDA Determined Cause 2 | Other |
| Action | Consignees were sent a letter dated 8/4/06 with additional instructions for use. A second letter was sent on 9/1/06 to correct the new instructions provided in the first letter for page 19 of the instructions. |
| Quantity in Commerce | 1100 |
| Distribution | Worldwide- USA and countries of Brazil, Canada, China, Germany, Greece, Japan, Korea, Macau, Netherlands, Pakistan, Spain, Taiwan and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DRM
|