|
Class 2 Device Recall OEC Miniview 6800 Mobile fluoroscopic xray system |
|
Date Initiated by Firm |
May 24, 2007 |
Date Posted |
July 03, 2007 |
Recall Status1 |
Terminated 3 on June 02, 2008 |
Recall Number |
Z-0928-2007 |
Recall Event ID |
38272 |
510(K)Number |
K992506
|
Product Classification |
Fluoroscopic x-ray system - Product Code JAA
|
Product |
OEC Miniview 6800 mobile fluoroscopic x-ray system, GE Healthcare, Surgery, Salt Lake City, Utah. |
Code Information |
Serial numbers in the range of 86-0032 to 86-2161 |
Recalling Firm/ Manufacturer |
OEC Medical Systems, Inc 384 Wright Brothers Dr Salt Lake City UT 84116-2862
|
For Additional Information Contact |
801-328-9300
|
Manufacturer Reason for Recall |
X-ray systems did not have the required labeling to indicate Air Kerma Rate and cumulative Air Kerma during fluoroscopic procedures
|
FDA Determined Cause 2 |
Other |
Action |
Consignees were notified by letter on 05/24/2007 and advised that a service representative would be contacting them for installation of the required label and to provide them with an update to their user's manuals. |
Quantity in Commerce |
811 systems |
Distribution |
Worldwide; USA including VA facilities in AK, CA, ID, MD, MI, NY and OR. Military distribution to CA, CO, CT, FL, GA, KS, LA, MS, OH, OK, TX, VA and Iraq. Foreign distribution to Austria, Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan and UK. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAA and Original Applicant = GE DEC MEDICAL SYSTEMS
|
|
|
|