| | Class 2 Device Recall UHR |  |
| Date Initiated by Firm | October 26, 2007 |
| Date Posted | March 20, 2008 |
| Recall Status1 |
Terminated 3 on January 27, 2010 |
| Recall Number | Z-0636-2008 |
| Recall Event ID |
45854 |
| 510(K)Number | K800207 |
| Product Classification |
Hip Prosthesis Component - Product Code KWL
|
| Product | Stryker Orthopaedics; UHR Universal Bipolar Component; 28 x 50 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430 |
| Code Information |
Ref No: Description: Lot Code: UH1-50-28 UHR BIPOLAR 28X50mm 0L7MHD UH1-50-28 UHR BIPOLAR 28X50mm 3JDMKA UH1-50-28 UHR BIPOLAR 28X50mm N4NMRA UH1-50-28 1E6MEA UH1-50-28 1EEMPD UH1-50-28 4E4MMD UH1-50-28 685MTA UH1-50-28 6W9MPD UH1-50-28 805MPD UH1-50-28 8HMMJA UH1-50-28 8XNMPD UH1-50-28 E9EMPD UH1-50-28 EKPMEA UH1-50-28 HHYMDA UH1-50-28 HLEMEA UH1-50-28 K5AMND |
| FEI Number |
3003070421
|
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah
NJ
07430-2002
|
| For Additional Information Contact | Rita Intorella 201-831-5825 |
Manufacturer Reason for Recall | Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head. |
FDA Determined Cause 2 | Process control |
| Action | Stryker sent Urgent Product Hold letters to US branches/agencies and hospitals on October 26, 2007 advising them to quarantine the product from these lots, and Urgent Product Recall letters on November 2, 2007 advising them to return the devices to Stryker. International distributors were sent an advisory on November 6, 2007. |
| Quantity in Commerce | 360 total |
| Distribution | Worldwide Distribution - USA, Canada,
China, India, Australia, Korea, Singapore, Taiwan, Sweden, Germany, Netherlands, France, Spain, Italy, Poland, South Africa, and England. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|