 Class 2 Device Recall ECG Lead Set 60 in SOC to Snapsee related information
Date Initiated by FirmOctober 15, 2007
Date PostedApril 24, 2008
Recall Status1 Terminated 3 on September 01, 2009
Recall NumberZ-1090-2008
Recall Event ID 45545
510(K)NumberK922106 
Product Classification ventilation accessories - Product Code GWO
ProductECG Lead Set 60 in SOC to Snap model number 8049-1. Filter, mask, and/or accessories for sleep therapy and ventilation devices, Respironics, Inc., Murrysville, PA 15668.
Code Information Lot Numbers 071010, 071011, and 071012.
FEI Number 2518422
Recalling Firm/
Manufacturer
Respironics, Inc.
1001 Murry Ridge Ln
Murrysville PA 15668-8550
For Additional Information ContactKatherine dePadua
724-387-7770
Manufacturer Reason
for Recall
Foreign material: Glass fragments may present in the plastic bag material used to ship components.
FDA Determined
Cause 2
Environmental control
ActionThe recalling firm issued a Recall Notification dated 10/18/07 to its customers. Replacement products were to be sent.
Quantity in Commerce20950 units
DistributionThe products were shipped nationwide to medical facilities.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = GWO
-
-