| | Class 2 Device Recall EndoWave |  |
| Date Initiated by Firm | March 10, 2008 |
| Date Posted | September 02, 2008 |
| Recall Status1 |
Terminated 3 on May 13, 2011 |
| Recall Number | Z-1752-2008 |
| Recall Event ID |
47245 |
| 510(K)Number | K072507 |
| Product Classification |
Catheter, Continuous flush - Product Code KRA
|
| Product | PT-3B Control Unit - component of the EndoWave Infusion Catheter System. EndoWave Infusion System consists of three main components, a single use EndoWave Infusion Catheter System which is comprised of a Drug Delivery Catheter (CD) and removable Ultrasound Core (USC) and a reusable EKOS PT-3B Control System. Box is labeled EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core |
| Code Information |
Model PT3B; Catalog Number 600-10301; Serial Number: PT3B-001, PT3B-002, PT3B-004, PT3B-006, PT3B-007, PT3B-008, PT3B-009, PT3B-010, PT3B-011, PT3B-012, PT3B-013, PT3B-014, PT3B-017, PT3B-018, PT3B-020, PT3B-021, PT3B-022, PT3B-024, PT3B-025, PT3B-027, PT3B-029, PT3B-030, PT3B-031, PT3B-032, PT3B-033, PT3B-034, PT3B-035, PT3B-037, PT3B-038, PT3B-040, PT3B-041, PT3B-042, PT3B-043, PT3B-044, PT3B-045, PT3B-046, PT3B-047, PT3B-049, PT3B-050, PT3B-051, PT3B-052, PT3B-057, PT3B-058, PT3B-059, PT3B-060, PT3B-062, PT3B-063, PT3B-064, PT3B-065, PT3B-066, PT3B-067, PT3B-068, PT3B-070, PT3B-071, PT3B-072, PT3B-073, PT3B-074, PT3B-075, PT3B-076, PT3B-078, PT3B-079, PT3B-080, PT3B-082, PT3B-084, PT3B-085, PT3B-087, PT3B-088, PT3B-089, PT3B-090, PT3B-091, PT3B-093, PT3B-094, PT3B-095, PT3B-096, PT3B-097, PT3B-098, PT3B-099, PT3B-100, PT3B-102, PT3B-103, PT3B-104, PT3B-108, PT3B-109, PT3B-110, PT3B-111, PT3B-112, PT3B-113, PT3B-114, PT3B-115, PT3B-118, PT3B-119, PT3B-121, PT3B-123, PT3B-124, PT3B-127, PT3B-128, PT3B-130, PT3B-131, PT3B-132, PT3B-133, PT3B-135, PT3B-137, PT3B-138, PT3B-139, PT3B-140, PT3B-141, PT3B-143, PT3B-145, PT3B-146, PT3B-147, PT3B-148, PT3B-150, PT3B-151, PT3B-152, PT3B-153, PT3B-154, PT3B-155, PT3B-156, PT3B-157, PT3B-158, PT3B-159, PT3B-163, PT3B-164, PT3B-165, PT3B-168, PT3B-169, PT3B-172, PT3B-173, PT3B-174, PT3B-175, PT3B-176, PT3B-177, PT3B-178, PT3B-180, PT3B-181, PT3B-183, PT3B-186, PT3B-188, PT3B-192, PT3B-193, PT3B-195, PT3B-196, PT3B-197, PT3B-199, PT3B-201, PT3B-202, PT3B-204, PT3B-205, PT3B-206, PT3B-207, PT3B-208, PT3B-210, PT3B-211, PT3B-212, PT3B-214, PT3B-215, PT3B-217, PT3B-218, PT3B-219, PT3B-221, PT3B-253, PT3B-254, PT3B-255, PT3B-256, PT3B-257, PT3B-258, PT3B-259, PT3B-260, PT3B-261, PT3B-262, PT3B-263, PT3B-264, PT3B-265, PT3B-266, PT3B-267, PT3B-268, PT3B-269, PT3B-270, PT3B-271, PT3B-273, PT3B-274, PT3B-275, PT3B-276, PT3B-280, PT3B-285, PT3B-286, PT3B-287, PT3B-288, PT3B-289, PT3B-291, PT3B-292, PT3B-294, PT3B-295, PT3B-297, PT3B-298, PT3B-308, PT3B-310, PT3B-311, PT3B-312, PT3B-313, PT3B-314, PT3B-315, PT3B-316, PT3B-319, PT3B-320, PT3B-321, PT3B-323, PT3B-326, PT3B-328, PT3B-329, PT3B-332, PT3B-333, PT3B-334, PT3B-335, PT3B-337, PT3B-338, PT3B-339, PT3B-344. |
Recalling Firm/ Manufacturer |
EKOS Corporation 11911 N Creek Pkwy S Bothell WA 98011-8809
|
| For Additional Information Contact | Jocelyn Kersten 425-415-3132 |
Manufacturer Reason for Recall | Potential for unit to burn patient. |
FDA Determined Cause 2 | Software design |
| Action | In March 2008, the firm planned to visit each consignee to install the new software version 01.05.02 and hand deliver letters, dated 02/20/2008, notifying consignees of the reason for the action. The old software version is 01.04.02. The software was upgraded because the Ultrasound Core heats very quickly and if removed from the DDC while operating it may result in burns to the patient or user. The software will turn off the Ultrasound Core automatically if removed from the DDC during operation, and will prompt to restart therapy if ultrasound is turned off during treatment.
On 04/17/2008, the firm issued a new letter with the date of April 2008 informing customers of the issue. In addition the letter indicates that the corrective action, release of the new software is delayed to incorporate further changes, and that customers will be notified when the software is available for their unit. Customers are advised that the device is safe to use when operating according to the Instructions for Use. EKOS contact number 1-888-356-7435 |
| Quantity in Commerce | 218 units |
| Distribution | Products were distributed nationwide to hospitals. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KRA
|
|
|
|