| | Class 2 Device Recall Medtronic Carmedacoated |  |
| Date Initiated by Firm | May 09, 2008 |
| Date Posted | July 24, 2008 |
| Recall Status1 |
Terminated 3 on May 08, 2012 |
| Recall Number | Z-1919-2008 |
| Recall Event ID |
48100 |
| 510(K)Number | K031700 |
| Product Classification |
device for Cardiopulmonary Bypass - Product Code DTN
|
| Product | Medtronic AFFINITY NT Hollow Fiber Oxygenator with Carmeda BioActive Surface, CB511, with Plasma Resistant Fiber. Single Use gas exchange device for Cardiopulmonary Bypass. Catalog Number: 71166 |
| Code Information |
Catalog / Lot: 71166 / 9821992, 71166 / 9826735. |
Recalling Firm/ Manufacturer |
Medtronic Perfusion Systems 7611 Northland Dr N Brooklyn Park MN 55428-1088
|
| For Additional Information Contact | 763-391-9000 |
Manufacturer Reason for Recall | Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS). |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medtronic issued a Press Release on 05/07/2008. An Urgent Medical Device Recall letter was sent to customers on May 07, 2008. The letter identifies affected devices, explains the situation and requests that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall Certificate acknowledging the receipt of this information. |
| Quantity in Commerce | 84 US, 14 OUS |
| Distribution | Worldwide Distribution: USA, Australia, Belgium, Canada, Czech Republic, France, Italy, Japan, Norway, and Taiwan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|