• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medtronic Trillium AFFINITY NT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medtronic Trillium AFFINITY NTsee related information
Date Initiated by FirmMay 07, 2008
Date PostedJuly 30, 2008
Recall Status1 Terminated 3 on December 01, 2008
Recall NumberZ-2005-2008
Recall Event ID 48101
Product Classification cardiovascular bypass oxygenator - Product Code DRY
ProductMedtronic Trillium AFFINITY NT, 511T, Hollow Fiber Oxygenator with Trillium Biopassive Surface, with Plasma Resistant Fiber. MODEL Numbers: 95215, 95216, 95217, 95249, 95250, 95252, 0H46R28, 0M81R41, 0R35R13, 0Y95R22, 1702R51, 1A11R16, 1A19R26, 1B22R5, 1F66R12, 1G47R4, 1K86R3, 1K93R5, 1L71R20, 1M05R12, 1P07R6, 1X24R, 2892R18, 2D64R12, 2E55R9, 2N90R8, 2V85R4, 3B89R12, 3C98R7, 3P11R4, 3R39R2, 3R42R14, 4B46R16, 4E58R16, 4F13R2, 4J01R7, 4Q39R1, 4R68R5, 4R68R6, 4W65R, 4X44R7, 5633R30, 5C89R2, 5D83R, 5R69R1, 5U35R, 9059R20, 9243R19, CB0A12R15, CB2T58R4, CB3L15R5, CB4D99R4, CB4E05R6, CB4J24R2, CB4J67R1, CB4K47R1, CB4K64R5, CB5B53R, CB5C01R2, CB5C82R9, CB5P29R4, CB5Q88R, CB5S97R2, DP38T, HY0W55R17, HY0Y86R17, HY1709R34, HY1C73R16, HY2F13R22, HY2F13R23, HY2J04R2, HY2K12R5, HY2K16R5, HY2K51R14, HY2K60R3, HY2Q31R1, HY2Q32R2, HY2T49R3, HY2U43R5, HY2U98R5, HY2W85R9, HY3B29R13, HY3C53R1, HY3E77R13, HY3H52R13, HY3L42R12, HY3L78R1, HY3M41R4, HY3M43R1, HY3M45R1, HY3Q11R14, HY3T77R2, HY3U59R4, HY4A96R3, HY4B44R, HY4B50R, HY4E05R8, HY4E66R5, HY4V82R2, HY4X31R7, HY4Y15R11, HY4Z59R7, HY4Z62R5, HY5C74R15, HY5D07R, HY5D70R1, HY5L38R1, HY5N41R2, HY5N42R3, HY5P52R, HY5Q00R, HY5S13R, HY5S33R2, JP1Z53R11, M310113Q, M310319A, M310523D, M310525C, M310531G, M310538A, M310543F, M310824V, M310852C, M310854A, M311028K, M311224A, M311818C, M320234E, M320410V, M320608K, M321014K, M321505L, M322318C, M330112G, M330344D, M330607H, M332108C, M332607G, M333405W, M333809B, M334007C, M334007C, M334008D, M334209A, M335615B, M336202V, M337412A, M337414A, M337415A, M337417A, M337418B, M338001P, M340305C, M340615E, M343105G, M380324D, M390206E, M390322C, M390508D, M390824C, M391817L, M391817M, M391908B, M392814A, M392815A, M395216B, M395217B, M395319B, M396010D, M397013G, M397912B, M398309C, M399510A, M399511A, M410811K, M430737A, M443307B, M443822B, M443823B, M447005R, M460524D, M460530B, M460531B, M460532B, M462007B, M470233C, M470237C, M470945S, M470945T, M490750B, M495512K, M580211B, M580416B, M580418E, M720407F, M720409F, M720714C, M720718D, M720720B, M930401B, M930402B, M960116A, M960119A, M973502E, M973901C, M999308D, P391401B, TL0A42R21, TL0W56R7, TL1E95R12, TL1J50R3, TL2C82R4, TL2H89R7, TL2J77R, TL2K45R6, TL2L66R5 TL2L67R, TL2M25R6, TL2P63R1, TL2P93R5, TL2P94R9, TL2R00R3, TL2R66R, TL2S03R4, TL2S21R1, TL2S24R3, TL2S80R24, TL2S80R30, TL2S80R30, TL2T46R7, TL2U42R13, TL2U59R, TL2W21R, TL2W22R5, TL2W90R11, TL2Z00R4, TL2Z34R, TL2Z35R, TL3A08R16, TL3B47R8, TL3C66R4, TL3C69R16, TL3E97R5, TL3E98R7, TL3E99R2, TL3F17R2, TL3F33R1, TL3G03R2, TL3G89R1, TL3H76R12, TL3H85R1, TL3H86R6, TL3K26R8, TL3L73R5, TL3M00R7, TL3M35R14, TL3M92R1, TL3N21R3, TL3N58R1, TL3P08R4, TL3P79R8, TL3R32R4, TL3R79R4, TL3R82R6, TL3R95R3, TL3S25R, TL3S39R3, TL3S63R, TL3T02R2, TL3T11R11, TL3V10R11, TL3V42R6, TL3X14R3, TL3X14R4, TL3X86R5, TL3Y45R1, TL3Y70R1, TL3Y77R, TL3Z02R4, TL3Z07R, TL3Z67R4, TL4A55R19, TL4A76R, TL4A98R1, TL4B15R2, TL4B16R5, TL4C84R, TL4D71R1, TL4E06R1, TL4E30R11, TL4F57R2, TL4F60R, TL4F71R1, TL4F88R6, TL4F97R7, TL4H65R, TL4H98R1, TL4J16R, TL4J32R10, TL4J96R7, TL4K33R3, TL4K78R, TL4K84R8, TL4K88R, TL4K93R, TL4K95R3, TL4L34R4, TL4M16R2, TL4M22R10, TL4M94R5, TL4N20R10, TL4N27R3, TL4N84R, TL4P48R2, TL4P64R4, TL4Q43R3, TL4Q43R4, TL4R10R2, TL4R21R, TL4S97R8, TL4T06R3, TL4U93R7, TL4V76R1, TL4V77R, TL4W07R4, TL4W22R, TL4W94R1, TL4X63R1, TL4X99R8, TL4Y20R2, TL4Z28R2, TL5054R11, TL5263R2, TL5398R19, TL5398R20, TL5A39R3, TL5B82R1, TL5B83R1, TL5C29R2, TL5C35R6, TL5C63R2, TL5C64R3, TL5C97R6, TL5D45R4, TL5D65R, TL5D66R3, TL5E04R, TL5E11R1, TL5F48R, TL5F62R5, TL5F64R4, TL5F96R, TL5G03R2, TL5G04R2, TL5G26R1, TL5G27R, TL5G27R1, TL5G41R2, TL5H55R1, TL5J24R6, TL5J74R2, TL5L27R1, TL5N42R2, TL5N43R3, TL5N53R1, TL5N59R, TL5N60R, TL5N62R4, TL5N79R, TL5N94R, TL5P14R, TL5P59R2, TL5P73R2, TL5P76R2, TL5Q40R1, TL5Q76R3, TL5Q85R2, TL5Q89R1, TL5R08R, TL5R15R2, TL5R18R1, TL5R52R, TL5R57R1, TL5R91R5, TL5S53R, TL5T73R, TL5U33R, TS1259R14, TS1259R15, TS1366R3, TS1366R4, TS1366R5, TS1404R1, TS1414R3, TS1420R
Code Information MODEL # / LOT #: 95215/9834158, 95215/9836054, 95215/9840678, 95215/9841550, 95215/9841559, 95215/9845071, 95215/9847467, 95215/9850444, 95215/9855692, 95215/9856192, 95215/9862423, 95215/9864967, 95215/9866563, 95215/9868270, 95215/9871572, 95215/9874613, 95215/9875139, 95215/9882831, 95215/9884748, 95215/9889245, 95215/9891467, 95215/9892254, 95215/9918684, 95216/9842281, 95216/9846078, 95216/9849122, 95216/9851739, 95216/9858463, 95216/9864836, 95216/9870318, 95216/9870329, 95216/9881577, 95216/9885872, 95216/9890513, 95216/9892011, 95216/9895530, 95216/9896568, 95216/9904481, 95216/9905578, 95216/9905755, 95216/9907437, 95216/9911601, 95216/9912191, 95217/9837473, 95217/9837609, 95217/9837669, 95217/9838938, 95217/9839026, 95217/9842333, 95217/9842382, 95217/9842493, 95217/9844162, 95217/9845850, 95217/9846383, 95217/9847644, 95217/9848942, 95217/9849195, 95217/9849848, 95217/9850346, 95217/9851508, 95217/9852863, 95217/9853060, 95217/9854098, 95217/9856703, 95217/9856833, 95217/9857062, 95217/9858085, 95217/9858293, 95217/9860038, 95217/9861271, 95217/9861502, 95217/9862976, 95217/9863247, 95217/9864837, 95217/9865234, 95217/9865893, 95217/9866024, 95217/9866164, 95217/9868050, 95217/9868760, 95217/9870331, 95217/9870409, 95217/9870650, 95217/9871483, 95217/9872441, 95217/9874427, 95217/9875822, 95217/9877281, 95217/9877367, 95217/9877440, 95217/9877762, 95217/9879316, 95217/9880467, 95217/9880641, 95217/9881610, 95217/9881925, 95217/9881981, 95217/9882840, 95217/9883447, 95217/9883450, 95217/9884377, 95217/9884881, 95217/9886033, 95217/9888892, 95217/9888949, 95217/9888952, 95217/9889923, 95217/9890206, 95217/9891481, 95217/9891930, 95217/9892779, 95217/9894202, 95217/9894438, 95217/9895542, 95217/9895563, 95217/9895750, 95217/9897223, 95217/9897279, 95217/9897429, 95217/9897698, 95217/9898810, 95217/9898944, 95217/9898999, 95217/9900403, 95217/9900493, 95217/9900526, 95217/9900613, 95217/9900980, 95217/9901023, 95217/9901088, 95217/9901471, 95217/9901630, 95217/9901950, 95217/9902927, 95217/9903604, 95217/9904240, 95217/9904417, 95217/9905391, 95217/9905487, 95217/9907004, 95217/9907578, 95217/9907584, 95217/9908483, 95217/9910191, 95217/9910776, 95217/9911401, 95217/9911717, 95217/9912960, 95217/9913072, 95217/9913268, 95217/9920082, 95249/9844242, 95249/9866206, 95249/9875679, 95249/9879355, 95249/9884526, 95249/9899196, 95249/9900455, 95249/9901014, 95249/9901467, 95249/9904646, 95249/9910460, 95249/9918448, 95250/9838923, 95250/9839181, 95250/9845804, 95250/9860022, 95250/9864835, 95250/9868938, 95250/9875809, 95250/9877559, 95250/9880811, 95250/9880821, 95250/9881576, 95250/9889922, 95250/9894165, 95250/9901002, 95250/9903602, 95250/9905673, 95250/9907549, 95250/9907592, 95250/9908172, 95250/9908874, 95250/9913036, 95250/9919854, 95252/9845799, 95252/9907597, 0H46R28/9801904, 0H46R28/9849211, 0H46R28/9874833, 0H46R28/9904051, 0H46R28/9904052, 0M81R41/9856457, 0R35R13/9856482, 0R35R13/9915510, 0Y95R22/9801921, 1702R51/9871082, 1A11R16/9864529, 1A11R16/9898363, 1A11R16/9900836, 1A19R26/9871079, 1B22R5/9852609, 1B22R5/9852610, 1B22R5/9876095, 1F66R12/9852612, 1F66R12/9906222, 1G47R4/9859472, 1K86R3/9812743, 1K86R3/9845548, 1K93R5/9823298, 1K93R5/9898343, 1L71R20/9904065, 1M05R12/9882585, 1P07R6/9849635, 1P07R6/9906224, 1P07R6/9909834, 1X24R/9828429, 1X24R/9882586, 2892R18/9823305, 2D64R12/9849668, 2D64R12/9849669, 2D64R12/9898367, 2E55R9/9842027, 2E55R9/9895126, 2E55R9/9895127, 2N90R8/9898372, 2V85R4/9898358, 3B89R12/9859504, 3C98R7/9849639, 3C98R7/9849640, 3P11R4/9818271, 3P11R4/9894029, 3R39R2/9848260, 3R39R2/9876345, 3R39R2/9911074, 3R42R14/9900843, 3R42R14/9900844, 4B46R16/9849670, 4E58R16/9856481, 4F13R2/9856483, 4J01R7/9891239, 4Q39R1/9864519, 4Q39R1/9906234, 4R68R5/9836720, 4R68R6/9906246, 4W65R/9842003, 4X44R7/9866759, 5633R30/9842026, 5C89R2/9845566, 5D83R/9866760, 5R69R1/9875367, 5R69R1/9895326, 5U35R/9929024, 9059R20/9842010, 9059R20/9900837, 9243R19/9845568, 9243R19/9895319, 9243R19/9909937, CB0A12R15/9818152, CB2T58R4/9869742, CB3L15R5/9853896, CB3L15R5/9904049, CB4D99R4/9915505, CB4E05R6/9828459, CB4E05R6/9874930, CB4J24R2/9906207, CB4J67R1/9895280, CB4K47R1/9909756, CB4K64R5/9856461, CB5B53R/9882599, CB5B53R/9915507, CB5C01R2/9864545, CB5C82R9/9864546, CB5C82R9/9864653, CB5P29R4/9871450, CB5P29R4/9914212, CB5Q88R/9861123, CB5S97R2/9921078, DP38T/9874904, DP38T/9904640, HY0W55R17/9808971, HY0W55R17/9910794, HY0Y86R17/9895142, HY0Y86R17/9895324, HY1709R34/9914044, HY1C73R16/9856469, HY2F13R22/9836729, HY2F13R23/9916965, HY2J04R2/9856414, HY2K12R5/9900680, HY2K16R5/9861559, HY2K16R5/9897791, HY2K51R14/9853921, HY2K51R14/9871066, HY2K51R14/9897797, HY2K60R3/9828457, HY2Q31R1/9856468, HY2Q32R2/9909790, HY2T49R3/9898330, HY2U43R5/9895125, HY2U98R5/9909926, HY2W85R9/9849675, HY2W85R9/9849676, HY2W85R9/9849677, HY3B29R13/9895322, HY3C53R1/9871041, HY3E77R13/9849120, HY3H52R13/9895169, HY3L42R12/9866761, HY3L42R12/9886081, HY3L78R1/9923377, HY3M41R4/9849658, HY3M43R1/9828413, HY3M45R1/9864484, HY3Q11R14/9895138, HY3Q11R14/9895139, HY3T77R2/9896464, HY3T77R2/9898435, HY3U59R4/9865471, HY3U59R4/9923417, HY4A96R3/9828227, HY4B44R/9874912, HY4B50R/9864521, HY4E05R8/9909957, HY4E66R5/9855421, HY4V82R2/9882601, HY4X31R7/9914047, HY4Y15R11/9897011, HY4Y15R11/9907958, HY4Z59R7/9902612, HY4Z62R5/9882303, HY5C74R15/9871083, HY5D07R/9864530, HY5D70R1/9853933, HY5D70R1/9909930, HY5L38R1/9904073, HY5L38R1/9918472, HY5N41R2/9869677, HY5N42R3/9900846, HY5P52R/9877621, HY5P52R/9902248, HY5Q00R/9857173, HY5S13R/9911027, HY5S33R2/9922456, JP1Z53R11/9860355, M310113Q/200986523, M310113Q/201017351, M310319A/201004514, M310523D/200960594, M310523D/200983800, M310523D/201014186, M310525C/200776830, M310525C/200914593, M310525C/200999719, M310531G/200951787, M310538A/200767906, M310538A/200998230, M310543F/200933365, M310543F/200968691, M310543F/200996499, M310824V/200986605, M310824V/201017488, M310852C/200946515, M310854A/200946514, M311028K/200937915, M311028K/200971865, M311028K/201025580, M311224A/200970165, M311224A/200996653, M311818C/200979740, M320234E/200951680, M320410V/201011466, M320608K/200991864, M320608K/201021498, M321014K/200995749, M321505L/200973472, M322318C/200945750, M322318C/200973463, M330112G/200983044, M330344D/200973338, M330607H/200978376, M332108C/201004516, M332108C/201029676, M332607G/200996493, M333405W/200986410, M333809B/200972405, M334007C/200926570, M334007C/200974974, M334007C/201033042, M334008D/200962670, M334008D/200978372, M334209A/200974961, M334209A/200984671, M334209A/201016613, M335615B/200995752, M336202V/200978364, M336202V/201011474, M337412A/200962585, M337412A/201017359, M337414A/200962660, M337414A/201020119, M337414A/201029347, M337415A/200950040, M337415A/200977462, M337417A/200984626, M337418B/200984630, M337418B/201016607, M338001P/200981099, M338001P/201024768, M340305C/200936579, M340615E/200972408, M343105G/200979741, M380324D/200978556, M380324D/201012978, M390206E/200948497, M390206E/200973465, M390322C/200957117, M390508D/200971872, M390824C/200933406, M390824C/201017510, M391817L/200948854, M391817M/200974967, M391908B/200945741, M391908B/200990138, M392814A/200977455, M392814A/201001630, M392814A/201034510, M392815A/200950055, M395216B/200962721, M395216B/200985176, M395217B/200933306, M395217B/200971868, M395319B/200973265, M396010D/200970167, M397013G/200950190, M397013G/201001823, M397912B/200935634, M397912B/201049318, M398309C/200984061, M398309C/201014558, M399510A/200975056, M399511A/200964869, M399511A/201017354, M410811K/200935071, M410811L/200977489, M430737A/200973327, M430737A/201054543, M443307B/201039509, M443822B/200976415, M443822B/201054699, M443823B/200976416, M443823B/201054695, M447005R/200978551, M460524D/200949744, M460530B/200949742, M460530B/200966981, M460531B/200999992, M460532B/200926957, M460532B/200966932, M460532B/200999981, M462007B/200991873, M462007B/201021499, M470233C/200974975, M470233C/201029764, M470237C/200975131, M470945S/200933318, M470945S/200971849, M470945T/201001926, M490750B/200938518, M495512K/200987075, M495512K/201004645, M580211B/200957101, M580211B/201036281, M580416B/200945726, M580416B/201024718, M580418E/200983048, M720407F/200968561, M720409F/200938479, M720409F/200974976, M720714C/200938466, M720714C/200973329, M720714C/201001639, M720718D/200968698, M720718D/200973667, M720720B/200964933, M930401B/200922137, M930402B/200922135, M960116A/200979559, M960119A/200979558, M973502E/200984001, M973502E/201013844, M973901C/200813720, M999308D/200948517, M999308D/200974969, P391401B/200953133, P391401B/201054693, TL0A42R21/9882593, TL0W56R7/9868069, TL1E95R12/9909898, TL1J50R3/9658714, TL2C82R4/9923403, TL2H89R7/9898365, TL2J77R/9841963, TL2K45R6/9845562, TL2K45R6/9900831, TL2L66R5/9895285, TL2L66R5/9909791, TL2L67R/9898332, TL2M25R6/9854038, TL2P63R1/9856415, TL2P93R5/9904074, TL2P94R9/9906240, TL2R00R3/9856416, TL2R00R3/9909792, TL2R66R/9874913, TL2S03R4/9906233, TL2S03R4/9909890, TL2S21R1/9836684, TL2S24R3/9874935, TL2S80R24/9853927, TL2S80R30/9868071, TL2S80R30/9885490, TL2S80R30/9897798, TL2T46R7/9915448, TL2U42R13/9849687, TL2U42R13/9898371, TL2U59R/9895286, TL2W21R/9849620, TL2W22R5/9923378, TL2W90R11/9879897, TL2Z00R4/9895111, TL2Z34R/9909795, TL2Z35R/9915484, TL3A08R16/9891237, TL3B47R8/9841965, TL3C66R4/9864486, TL3C66R4/9864487, TL3C69R16/9828450, TL3E97R5/9845541, TL3E98R7/9921079, TL3E99R2/9885607, TL3F17R2/9898451, TL3F33R1/9919253, TL3G03R2/9871067, TL3G89R1/9593332, TL3G89R1/9898334, TL3H76R12/9846723, TL3H85R1/9891217, TL3H86R6/9911088, TL3K26R8/9809022, TL3L73R5/9895287, TL3L73R5/9923379, TL3M00R7/9879882, TL3M35R14/9863156, TL3M35R14/9868110, TL3M92R1/9911060, TL3N21R3/9864488, TL3N58R1/9926687, TL3P08R4/9849753, TL3P08R4/9923415, TL3P79R8/9875092, TL3P79R8/9910788, TL3R32R4/9808976, TL3R32R4/9871042, TL3R79R4/9887326, TL3R82R6/9856417, TL3R82R6/9856418, TL3R95R3/9915500, TL3S25R/9887067, TL3S39R3/9925910, TL3S63R/9898336, TL3T02R2/9812851, TL3T11R11/9900024, TL3T11R11/9900024, TL3V10R11/9874853, TL3V42R6/9904031, TL3V42R6/9914075, TL3V42R6/9919254, TL3X14R3/9856420, TL3X14R4/9909961, TL3X86R5/9906215, TL3Y45R1/9881136, TL3Y45R1/9891218, TL3Y70R1/9924750, TL3Y77R/9864490, TL3Y77R/9898337, TL3Y77R/9900811, TL3Z02R4/9887333, TL3Z07R/9852606, TL3Z07R/9895270, TL3Z67R4/9845542, TL4A55R19/9874941, TL4A76R/9866737, TL4A98R1/9856421, TL4A98R1/9856422, TL4A98R1/9909804, TL4B15R2/9818238, TL4B16R5/9859493, TL4C84R/9859466, TL4D71R1/9895289, TL4E06R1/9904071, TL4E30R11/9879885, TL4F57R2/9919260, TL4F60R/9834322, TL4F71R1/9859497, TL4F71R1/9891235, TL4F88R6/9871078, TL4F97R7/9864517, TL4H65R/9885464, TL4H65R/9909805, TL4H98R1/9909892, TL4J16R/9909806, TL4J32R10/9845571, TL4J32R10/9906244, TL4J32R10/9909956, TL4J96R7/9895309, TL4J96R7/9906235, TL4J96R7/9909902, TL4K33R3/9864974, TL4K33R3/9883915, TL4K33R3/9895124, TL4K33R3/9929052, TL4K78R/9914187, TL4K84R8/9911087, TL4K88R/9856456, TL4K93R/9895307, TL4K93R/9909889, TL4K95R3/9874933, TL4L34R4/9930801, TL4M16R2/9841994, TL4M16R2/9841995, TL4M16R2/9911076, TL4M22R10/9859501, TL4M94R5/9854039, TL4N20R10/9896593, TL4N27R3/9871071, TL4N84R/9849655, TL4N84R/9909908, TL4P48R2/9895320, TL4P64R4/9849671, TL4P64R4/9907954, TL4Q43R3/9852368, TL4Q43R4/9898373, TL4R10R2/9845570, TL4R10R2/9879894, TL4R21R/9914198, TL4R21R/9931609, TL4S97R8/9859509, TL4T06R3/9871074, TL4U93R7/9911026, TL4V76R1/9923408, TL4V77R/9903794, TL4W07R4/9913292, TL4W22R/9871068, TL4W22R/9895313, TL4W94R1/9900834, TL4X63R1/9882598, TL4X99R8/9852634, TL4Y20R2/9895131, TL4Z28R2/9885485, TL4Z28R2/9887337, TL4Z28R2/9909940, TL5054R11/9856424, TL5263R2/9909906, TL5398R19/9843924, TL5398R20/9895328, TL5A39R3/9864541, TL5B82R1/9849665, TL5B82R1/9859499, TL5B82R1/9879887, TL5B83R1/9859500, TL5B83R1/9879888, TL5C29R2/9866092, TL5C29R2/9916514, TL5C35R6/9900842, TL5C63R2/9841958, TL5C63R2/9859389, TL5C63R2/9862430, TL5C63R2/9915511, TL5C64R3/9904062, TL5C97R6/9895137, TL5D45R4/9871075, TL5D65R/9890576, TL5D66R3/9834377, TL5D66R3/9859510, TL5D66R3/9859511, TL5D66R3/9927936, TL5E04R/9911079, TL5E04R/9929680, TL5E11R1/9921081, TL5F48R/9836721, TL5F48R/9900839, TL5F62R5/9896938, TL5F64R4/9896939, TL5F96R/9896940, TL5G03R2/9896941, TL5G04R2/9896942, TL5G26R1/9931614, TL5G27R/9893749, TL5G27R1/9910921, TL5G41R2/9879902, TL5H55R1/9858542, TL5H55R1/9885489, TL5J24R6/9868594, TL5J24R6/9870786, TL5J74R2/9891246, TL5J74R2/9904064, TL5L27R1/9828462, TL5N42R2/9869678, TL5N43R3/9865432, TL5N43R3/9914137, TL5N53R1/9921857, TL5N59R/9869738, TL5N60R/9869785, TL5N60R/9926703, TL5N62R4/9864809, TL5N62R4/9887341, TL5N62R4/9894785, TL5N79R/9911085, TL5N94R/9874148, TL5P14R/9883936, TL5P59R2/9896929, TL5P73R2/9925064, TL5P76R2/9932315, TL5Q40R1/9879468, TL5Q76R3/9930859, TL5Q76R3/9932577, TL5Q85R2/9913293, TL5Q89R1/9892383, TL5Q89R1/9926708, TL5R08R/9875101, TL5R15R2/9878911, TL5R18R1/9923206, TL5R52R/9894022, TL5R57R1/9918893, TL5R91R5/9918898, TL5R91R5/9931616, TL5S53R/9892387, TL5S53R/9910061, TL5T73R/9918921, TL5U33R/9932314, TS1259R14/0000678918, TS1259R14/0000680574, TS1259R14/0000698748, TS1259R15/0000717577, TS1366R3/0000666582, TS1366R3/0000666584, TS1366R4/0000674278, TS1366R4/0000674279, TS1366R4/0000676748, TS1366R4/0000690365, TS1366R4/0000690367, TS1366R4/0000711275, TS1366R4/0000711278, TS1366R4/0000711281, TS1366R4/0000714758, TS1366R4/0000714759, TS1366R5/0000679864, TS1366R5/0000679865, TS1366R5/0000679877, TS1366R5/0000679878, TS1366R5/0000680617, TS1366R5/0000683344, TS1366R5/0000683365, TS1366R5/0000685091, TS1366R5/0000685092, TS1366R5/0000687732, TS1366R5/0000687733, TS1366R5/0000693076, TS1366R5/0000693081, TS1366R5/0000694811, TS1366R5/0000697858, TS1366R5/0000697859, TS1366R5/0000699550, TS1366R5/0000699552, TS1366R5/0000703198, TS1366R5/0000704023, TS1366R5/0000705433, TS1366R5/0000705434, TS1366R5/0000709159, TS1366R5/0000709160, TS1366R5/0000717459, TS1366R5/0000717579, TS1366R5/0000721710, TS1366R5/0000721714, TS1366R5/0000724038, TS1366R5/0000724040, TS1366R5/0000727453, TS1404R1/0000702270, TS1404R1/0000706954, TS1404R1/0000713786, TS1404R1/0000714751, TS1404R1/0000717143, TS1404R1/0000717431, TS1404R1/0000720102, TS1404R1/0000721485, TS1405R1/0000725135, TS1405R1/0000727254, TS1406R1/0000708248, TS1414R3/0000694899, TS1420R/0000671368, TS1420R/0000676756, TS1420R/0000679997, TS1420R/0000680698, TS1420R/0000686918, TS1420R/0000689296, TS1420R/0000689303, TS1420R/0000690357, TS1420R/0000700945, TS1420R/0000704416, TS1420R/0000704575, TS1420R/0000707196, TS1420R/0000708343, TS1420R/0000712408, TS1420R/0000716405, TS1420R/0000720477, TS1420R/0000724108, TS1420R/0000727448, TS1423R3/0000672120, TS1423R3/0000674975, TS1423R3/0000676009, TS1423R3/0000683458, TS1423R3/0000692018, TS1423R3/0000697723, TS1423R3/0000697734, TS1423R3/0000700801, TS1423R3/0000704367, TS1423R3/0000707214, TS1423R3/0000720398, TS1423R3/0000727391, TS1434R/0000687709, TS1434R/0000692805, TS1434R/0000699993, TS1434R/0000712504, TS1434R/0000725183, TS1434R1/0000717260, TS1434R1/0000717582, TS1434R1/0000718578, TS1434R1/0000721812, TS1434R2/0000678999, TS1434R2/0000681548, TS1434R2/0000687716, TS1434R2/0000712336, TS1434R2/0000714756, TS1434R2/0000723102, TS1434R2/0000723732, TS1434R2/0000725187, TS1434R2/0000727458, TS1436R2/0000677992, TS1436R2/0000698843, TS1436R2/0000720585, TS1441R1/0000679020, TS1441R1/0000693083, TS1441R1/0000701007, TS1441R1/0000714931, TS1441R1/0000725137, TS1442R1/0000703329, TS1442R1/0000704007, TS1464R1/0000674905, TS1464R1/0000700813, TS1470R/0000664062, TS1470R/0000676148, TS1470R/0000676810, TS1470R/0000678000, TS1470R/0000689134, TS1470R/0000692914, TS1470R/0000692981, TS1470R/0000694076, TS1470R/0000710419, TS1476R/0000691107, TS1476R/0000705312, TS1476R/0000711280, TS1476R/0000721321, TS1477R3/0000664920, TS1480R3/0000682643, TS1480R5/0000682651, TS1480R5/0000685096, TS1480R5/0000689355, TS1480R5/0000698716, TS1480R5/0000704319, TS1480R5/0000710146, TS1480R5/0000716164, TS1487R/0000663088, TS1487R/0000666583, TS1487R/0000666630, TS1487R/0000678776, TS1487R/0000678783, TS1487R/0000678801, TS1487R1/0000687729, TS1487R1/0000687731, TS1487R1/0000690238, TS1487R1/0000691281, TS1487R1/0000692990, TS1487R1/0000697608, TS1487R1/0000717379, TS1487R2/0000699592, TS1487R2/0000706950, TS1487R2/0000708267, TS1487R2/0000712299, TS1487R2/0000713681, TS1487R2/0000721526, TS1489R/0000718453, TS1491R/0000685752, TS1491R/0000685754, TS8008R26/0000675071, TS8008R26/0000708169, TS8008R26/0000714871, TS8024R18/0000705493, TS8040R28/0000666588, TS8040R28/0000674368, TS8040R28/0000676750, TS8040R29/0000672038, TS8040R29/0000679875, TS8040R29/0000683433, TS8040R29/0000685093, TS8040R29/0000690113, TS8040R29/0000690366, TS8040R29/0000693079, TS8040R29/0000694813, TS8040R29/0000697860, TS8040R29/0000699551, TS8040R29/0000703230, TS8040R29/0000705431, TS8040R29/0000709006, TS8040R29/0000711274, TS8040R29/0000714757, TS8040R29/0000717461, TS8040R29/0000721712, TS8040R29/0000724041, TS8041R20/0000678901, TS8041R20/0000680561, TS8041R21/0000689220, TS8093R12/0000713994, TS8148R15/0000687704, TS8148R16/0000689307, TS8148R16/0000695319, TS8148R16/0000711323, TS8148R16/0000712300, TS8148R16/0000713895, TS8148R16/0000718475, TS8148R16/0000722562, TS8151R9/0000673946, TS8151R9/0000677841, TS8151R9/0000685846, TS8151R9/0000690190, TS8151R9/0000692240, TS8151R9/0000696181, TS8151R9/0000705370, TS8151R9/0000711440, TS8151R9/0000718570, TS8151R9/0000722592, TS8163R10/0000665624, TS8163R10/0000666650, TS8163R11/0000672251, TS8163R11/0000672834, TS8163R11/0000676653, TS8163R11/0000679964, TS8163R11/0000682493, TS8163R11/0000685839, TS8163R12/0000689004, TS8163R12/0000689978, TS8163R12/0000690490, TS8163R12/0000692238, TS8163R12/0000696205, TS8163R12/0000697740, TS8163R12/0000702224, TS8163R12/0000707102, TS8163R12/0000708272, TS8163R12/0000709201, TS8163R12/0000711251, TS8163R12/0000717014, TS8163R12/0000717015, TS8163R12/0000722920, TS8163R12/0000724873, TS8178R8/0000676646, TS8178R8/0000684077, TS8178R8/0000686010, TS8178R8/0000689240, TS8182R15/0000685793, TS8182R15/0000686030, TS8182R15/0000704422, TS8183R10/0000676650, TS8183R10/0000681387, TS8183R10/0000686020, TS8183R10/0000693138, TS8183R10/0000697660, TS8183R10/0000703123, TS8183R10/0000707104, TS8183R10/0000709012, TS8183R10/0000712314, TS8183R10/0000718376, TS8183R10/0000722940, TS8183R10/0000725096, TS8183R10/0000726183, TS8183R9/0000672138, TS8183R9/0000673930, TS8184R15/0000672061, TS8184R15/0000673924, TS8184R16/0000684081, TS8184R16/0000684083, TS8184R16/0000685763, TS8184R16/0000687711, TS8184R16/0000690156, TS8184R16/0000692966, TS8184R16/0000696108, TS8184R16/0000700721, TS8184R16/0000707030, TS8184R16/0000708997, TS8184R16/0000712312, TS8184R16/0000718372, TS8184R16/0000724871, TS8212R5/0000667107, TS8212R5/0000672190, TS8212R5/0000673921, TS8212R5/0000675074, TS8212R5/0000677755, TS8212R5/0000679826, TS8212R5/0000680582, TS8212R5/0000683308, TS8212R5/0000684936, TS8212R6/0000687727, TS8212R6/0000689379, TS8212R6/0000691242, TS8212R6/0000692012, TS8212R6/0000693784, TS8212R6/0000699591, TS8212R6/0000700636, TS8212R6/0000704357, TS8212R6/0000705311, TS8212R6/0000707975, TS8212R6/0000711462, TS8212R6/0000714659, TS8212R6/0000717141, TS8212R6/0000720220, TS8212R6/0000722455, TS8212R6/0000726154, TS8212R6/0000727385, TS8258R2/0000686006, TS8258R3/0000675072, TS8258R3/0000696184, TS8258R3/0000702234, TS8267R3/0000690451, TS8267R4/0000711409, TS8287R3/0000665661, TS8287R3/0000673927, TS8287R4/0000686951, TS8287R4/0000692017, TS8287R4/0000703002, TS8287R4/0000713560, TS8287R4/0000720183, TS8296R9/0000689212, TS8296R9/0000694040, TS8296R9/0000700867, TS8296R9/0000712412, TS8296R9/0000713824, TS8309R4/0000667327, TS8309R5/0000676110, TS8309R5/0000684159, TS8309R5/0000691354, TS8309R5/0000699633, TS8309R5/0000709010, TS8309R5/0000715985, TS8309R5/0000721464, TS8311R1/0000685019, TS8311R1/0000721524, TS8311R1/0000722574, TS8323R6/0000686933, TS8323R6/0000708172, TS8328R8/0000675082, TS8328R8/0000676691, TS8328R8/0000677937, TS8328R8/0000681483, TS8328R8/0000681485, TS8328R8/0000684167, TS8328R8/0000685857, TS8328R8/0000686930, TS8328R8/0000704402, TS8328R8/0000708271, TS8328R9/0000716362, TS8328R9/0000718476, TS8329R6/0000675972, TS8329R6/0000678643, TS8329R6/0000682496, TS8329R6/0000690153, TS8329R6/0000694609, TS8329R6/0000698829, TS8329R6/0000713937, TS8329R6/0000720242, TS8329R6/0000727366, TS8339R3/0000685016, TS8339R3/0000709246, TS8339R3/0000714930, TS8339R3/0000720450, TS8340R6/0000678804, TS8340R6/0000691166, TS8340R6/0000718471, TS8340R6/0000726245, TS8345R1/0000665622, TS8345R2/0000682432, TS8345R2/0000697658, TS8345R2/0000713763, TS8348R5/0000672917, TS8348R6/0000679862, TS8348R6/0000692188, TS8348R6/0000693951, TS8348R6/0000700637, TS8348R6/0000702900, TS8348R6/0000708230, TS8348R6/0000711369, TS8348R6/0000714782, TS8348R6/0000720328, TS8348R6/0000726226, TS8348R6/0000727365, TS8349R7/0000685097, TS8349R7/0000713789, TS8350R4/0000666627, TS8350R4/0000674282, TS8350R6/0000678857, TS8350R6/0000685025, TS8350R6/0000707078, TS8350R6/0000711382, TS8350R6/0000718473, TS8350R6/0000725003, TS8357R2/0000674242, TS8357R4/0000677770, TS8357R4/0000679825, TS8357R4/0000693921, TS8357R4/0000714008, TS8357R4/0000723733, TS8357R4/0000724050, TS8363R3/0000685840, TS8363R3/0000716000, TS8363R5/0000689009, TS8363R5/0000694043, TS8363R5/0000722560, TS8363R5/0000723918, TS8365R5/0000681505, TS8365R5/0000689964, TS8365R5/0000712247, TS8365R5/0000713977, TS8376R3/0000685786, TS8376R3/0000685788, TS8376R3/0000690097, TS8376R4/0000695350, TS8376R4/0000695356, TS8376R4/0000698742, TS8376R4/0000708404, TS8376R4/0000708405, TS8376R4/0000716494, TS8376R4/0000716498, TS8376R4/0000721487, TS8376R4/0000724870, TS8380R4/0000672239, TS8380R4/0000675975, TS8380R4/0000680565, TS8380R4/0000684934, TS8380R4/0000685790, TS8380R5/0000687712, TS8380R5/0000688054, TS8380R5/0000688084, TS8380R5/0000693074, TS8380R5/0000694815, TS8380R5/0000696107, TS8380R5/0000700717, TS8380R5/0000703228, TS8380R6/0000704550, TS8380R6/0000708944, TS8380R6/0000712517, TS8380R6/0000718406, TS8380R6/0000721782, TS8381R6/0000683469, TS8381R6/0000683470, TS8381R6/0000687082, TS8381R6/0000691322, TS8381R6/0000692011, TS8381R6/0000701974, TS8381R6/0000720415, TS8381R6/0000723734, TS8381R6/0000725138, TS8382R5/0000674951, TS8382R5/0000685844, TS8382R6/0000691102, TS8382R6/0000699992, TS8382R6/0000704356, TS8382R6/0000710029, TS8382R6/0000716017, TS8387R5/0000672198, TS8387R5/0000678739, TS8387R5/0000682603, TS8387R5/0000687077, TS8387R5/0000687868, TS8387R5/0000693054, TS8387R5/0000694123, TS8387R5/0000708408, TS8387R5/0000714897, TS8387R5/0000720194, TS8387R5/0000727368, TS8388R3/0000662094, TS8388R3/0000662908, TS8388R3/0000663937, TS8388R3/0000664737, TS8388R3/0000665616, TS8388R3/0000666464, TS8388R3/0000667103, TS8388R3/0000670937, TS8388R3/0000672282, TS8388R3/0000672916, TS8388R3/0000675107, TS8388R3/0000675976, TS8388R3/0000677752, TS8388R3/0000677759, TS8388R3/0000683314, TS8388R4/0000679863, TS8388R4/0000680660, TS8388R4/0000680793, TS8388R4/0000681468, TS8388R4/0000682586, TS8388R4/0000685070, TS8388R4/0000685755, TS8388R4/0000685761, TS8388R4/0000687710, TS8388R4/0000688985, TS8388R4/0000689006, TS8388R4/0000690154, TS8388R4/0000691109, TS8388R4/0000692110, TS8388R4/0000692964, TS8388R4/0000694600, TS8388R4/0000694812, TS8388R4/0000696192, TS8388R4/0000698353, TS8388R4/0000699548, TS8388R4/0000700720, TS8388R4/0000701976, TS8388R4/0000703005, TS8388R4/0000704024, TS8388R4/0000705314, TS8388R4/0000707032, TS8388R4/0000708114, TS8388R4/0000709031, TS8388R4/0000710369, TS8388R4/0000711245, TS8388R4/0000712246, TS8388R4/0000713766, TS8388R4/0000714784, TS8388R4/0000716292, TS8388R4/0000718393, TS8388R4/0000720218, TS8388R4/0000721463, TS8388R4/0000724872, TS8388R4/0000727255, TS8390R3/0000667287, TS8390R3/0000671481, TS8390R4/0000693263, TS8390R5/0000702343, TS8390R5/0000725136, TS8397R5/0000725229, TS9412R4/0000680661, TS9412R4/0000686889, TS9412R4/0000700908, TS9412R4/0000700973, TS9412R4/0000704567, TS9412R4/0000715042, TS9412R4/0000721788, TS9416R1/0000666539, TS9416R2/0000676111, TS9416R2/0000689000, TS9416R2/0000696021, TS9416R2/0000709244, TS9416R2/0000711304, TS9416R2/0000721652, TS9416R2/0000723021, TS9420R4/0000724092, TS9422R35/0000673932, TS9422R35/0000677981, TS9422R35/0000683408, TS9422R35/0000688056, TS9422R35/0000690443, TS9422R35/0000698349, TS9422R35/0000699661, TS9422R35/0000702214, TS9422R35/0000709158, TS9422R35/0000713968, TS9422R35/0000726432, TS9422R35/0000727315, TS9423R2/0000666496, TS9423R2/0000686906, TS9423R3/0000674357, TS9423R3/0000682606, TS9423R3/0000693024, TS9423R3/0000696183, TS9423R3/0000703124, TS9423R3/0000724042, TS9425R38/0000675106, TS9425R38/0000687713, TS9425R38/0000687714, TS9425R38/0000696016, TS9425R38/0000705319, TS9425R38/0000707057, TS9425R38/0000716291, TS9425R38/0000726244, TS9749R27/0000680595, TS9749R27/0000683301, TS9749R27/0000690189, TS9749R27/0000695318, TS9749R27/0000699634, TS9749R27/0000712301, TS9749R27/0000713764, TS9749R27/0000716440, TS9749R27/0000720350, TS9749R27/0000724869, TS9749R27/0000726185, TS9781R11/0000665814, TS9781R12/0000676648, TS9781R12/0000681386, TS9781R12/0000691170, TS9781R12/0000700771, TS9781R12/0000711279, TS9781R12/0000725107, TS9800R30/0000674358, TS9800R30/0000674384, TS9800R30/0000675218, TS9800R30/0000681373, TS9800R30/0000685110, TS9800R30/0000690494, TS9800R30/0000701004, TS9800R30/0000711333, TS9800R30/0000721759, TS9806R11/0000665848, TS9806R11/0000667330, TS9806R11/0000673003, TS9807R23/0000683377, TS9807R23/0000685753, TS9807R23/0000692010, TS9807R23/0000700929, TS9807R23/0000712471, TS9807R23/0000725186, TS9818R23/0000667387, TS9818R24/0000680577, TS9818R24/0000684938, TS9818R24/0000692020, TS9818R24/0000712468, TS9818R24/0000725015, TS9824R16/0000712413, TS9824R16/0000723728, TS9824R16/0000725097, TS9917R27/0000662927, TS9950R21/0000675998, TS9950R21/0000685758, TS9950R21/0000698361, TS9950R21/0000717504, TS9959R12/0000662196, TS9959R12/0000673917, TS9959R12/0000683368, TS9959R12/0000689970, TS9959R13/0000693279, TS9959R13/0000697811, TS9959R13/0000707165, TS9959R13/0000710353, TS9959R13/0000723942, XP41BT/9904032, XP41T/9866739, XP41T/9874914.
FEI Number 1000116158
Recalling Firm/
Manufacturer
Medtronic International Trade Inc
7611 Northland Dr N
Brooklyn Park MN 55428-1088
For Additional Information Contact
763-391-9000
Manufacturer Reason
for Recall
Medtronic is advising that certain Trillium-coated Medtronic cardiopulmonary bypass (CBP) products (e.g., oxygenators, reservoirs, pumps and tubing packs) were manufactured with heparin batches contaminated with oversulfated chondroitin sulfate (OSCS). Based upon currently available information, Medtronic has determined that Medtronic Trillium-coated products can continue to be used in cardiopulm
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedtronic Cardiovascular Customers were sent the May 05, 2008 Urgent Medical Device Corrective Action letter to advise them that certain Trillium-coated cardiopulmonary bypass products were manufactured with heparin which may be contaminated with oversulfated chondroitin sulfate. The letter discussed the risk evaluation and gave recommendations to healthcare professionals for patient care. Affected devices can continue to be used in cardiopulmonary bypass procedures.
Quantity in Commerce59500
DistributionWorldwide Distribution, including USA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-