 Class 2 Device Recall LifeStent FlexStar Stent Systemsee related information
Date Initiated by FirmAugust 29, 2008
Date PostedOctober 10, 2008
Recall Status1 Terminated 3 on April 06, 2012
Recall NumberZ-0042-2009
Recall Event ID 49405
510(K)NumberK053404 
Product Classification Catheter - Product Code FGE
ProductEdwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards Lifesciences LLC Irvine, CA 92614, Model EX060603CD. The LifeStent FlexStar self-expanding biliary stent system is intended for use in the palliation of malignant strictures (neoplasms) in the biliary tree.
Code Information Lot Numbers: 58289987, 58298660, R58315412, 58333466, 58338746, 58347645, 58407267, 58419030, 58423549, 58432229, RFR6K0269,  FR6H0213, and  FR6K0346
FEI Number 2020394
Recalling Firm/
Manufacturer
Bard Peripheral Vascular Inc
1625 W 3rd St
Tempe AZ 85281-2438
For Additional Information Contact
480-894-9515
Manufacturer Reason
for Recall
Some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible.
FDA Determined
Cause 2
Process control
ActionThe customer letters for the tip to sheath gap were sent on 8/29/2008 with return receipt. The letter advises that C.R. Bard has determined that some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible. Per the letter, the firm's analysis has determined that a gap between the tip of the delivery system and the delivery system sheath may be associated with difficult advancement to the treatment site. Users are instructed to inspect all devices to verify that there is not a gap between the tip and sheath (orange colored guidewire lumen is visible), prior to use and to not use any devices that exhibit a tip to sheath gap condition. If devices that exhibit a tip to sheath gap are identified, customers are advised to call Customer Service at 1-800-321-4254 to return the device and a replacement will be provided at no cost. The letter also advises that the Instructions for Use for the LifeStent FlexStar System will be updated to include the inspection step which is provided in the letter.
Quantity in Commerce825 units
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FGE
-
-