|
Class 2 Device Recall DePuy DNP Anatomic Fracture Repair System |
 |
Date Initiated by Firm |
December 23, 2008 |
Date Posted |
February 11, 2009 |
Recall Status1 |
Terminated 3 on October 18, 2010 |
Recall Number |
Z-0983-2009 |
Recall Event ID |
50700 |
510(K)Number |
K052248
|
Product Classification |
Plate, fixation, bone - Product Code HRS
|
Product |
DePuy DNP Anatomic Fracture Repair System, DNP Anatomic Left, Depuy, Warsaw, IN; REF DNPAL.
The device is used in the treatment of fractures and osteotomies of the distal radius.
|
Code Information |
All lots. |
Recalling Firm/ Manufacturer |
Depuy Orthopaedics, Inc. 700 Orthopaedic Dr Warsaw IN 46582-3994
|
For Additional Information Contact |
888-783-7156
|
Manufacturer Reason for Recall |
The DNP nail plate may not attach to the alignment guide (jig) instrument, making it difficult to insert the nail plate into position or to drill needed holes accurately.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm's sales force was instructed via "Urgent Device Correction" letter issued December 29, 2008 to inspect all of these instruments in their possession, and those instruments at their hospital customers, and to leave a copy of the letter with each hospital customer who has the instruments on site.
The letter describes the issue and corrective actions for customers. Please direct questions about this recall to Depuy Orthopaedics, Inc. at 888-783-7156. |
Quantity in Commerce |
Approx. 3,250 of all products |
Distribution |
Nationwide Distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HRS and Original Applicant = HAND INNOVATIONS, LLC.
|
|
|
|