• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall 25 GA Total Plus Pak

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall 25 GA Total Plus Paksee related information
Date Initiated by FirmApril 29, 2009
Date PostedMay 26, 2009
Recall Status1 Terminated 3 on October 14, 2009
Recall NumberZ-1643-2009
Recall Event ID 51964
510(K)NumberK911808 
Product Classification Ophthalmic Trocar Cannula - Product Code NGY
ProductAlcon 25 GA Total Plus Pak, Catalog/Item #8065750220; for use with ACCURUS Vitrectomy System; distributed by Alcon Laboratories, Inc., Fort Worth, TX 76134.
Code Information Lot #9013991X, 9014291X and 9014509X.
FEI Number 1610287
Recalling Firm/
Manufacturer
Alcon Laboratories, Inc
6201 South Fwy
Fort Worth TX 76134
For Additional Information Contact
817-551-6813
Manufacturer Reason
for Recall
Specific lots of Alcon 25 GA Trocar Blades do not conform to finished product specifications in that oxidation can occur on some of the blades.
FDA Determined
Cause 2
Process control
ActionFirm began notifying consignees of recall by Urgent: Medical Device Recall letter on 04/29/09 (letters sent via FedEx). Consignees instructed to check inventory for affected lots; to complete and return the Medical Device Recall Response Form; to contact Alcon to arrange for return of the devices; that credit will be issued for the recalled product when it is returned; and in the case that there is no inventory remaining, to indicate that on the Response Form and return it as well. Consignees should contact Alcon Technical Consumer Affairs at 1-800-561-6466, Option 8 for technical assistance.
Quantity in Commerce3,523 surgical paks.
DistributionWorldwide Distribution -- USA and Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NGY
-
-