• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ABC BendABEAM Handpiece,134003

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ABC BendABEAM Handpiece,134003see related information
Date Initiated by FirmDecember 18, 2008
Date PostedAugust 25, 2010
Recall Status1 Terminated 3 on October 22, 2010
Recall NumberZ-2276-2010
Recall Event ID 52030
510(K)NumberK961505 
Product Classification Electrosurgical, cutting & coagulation & accessories - Product Code GEI
ProductABC Bend-A-Beam Handpiece REF 134003 Handcontrol Malleable Handpiece, 3" (7.6 cm) QTY: 1 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (pouch label) ABC Bend-A-Beam Handpiece REF 134003 Handcontrol Malleable Handpiece, 3" (7.6 cm) QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (case label) Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosurgical units equipped with argon beam coagulation.
Code Information Lot: 0306021 through 0806021
Recalling Firm/
Manufacturer
ConMed Electrosurgery
14603 E Fremont Ave
Centennial CO 80112-4251
For Additional Information Contact
303-269-8824
Manufacturer Reason
for Recall
The internal electrode protrudes in front of the ceramic insulator at the tip of the handpiece.
FDA Determined
Cause 2
Process design
ActionAn Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.
Quantity in CommerceDomestic: 1286 cs (12860 units); Int'l: 1670 cs (16700 units)
DistributionNationwide and International
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GEI
-
-