| | Class 2 Device Recall Symbia S Series SPECT System |  |
| Date Initiated by Firm | April 22, 2009 |
| Date Posted | August 04, 2009 |
| Recall Status1 |
Terminated 3 on June 21, 2010 |
| Recall Number | Z-1676-2009 |
| Recall Event ID |
52091 |
| 510(K)Number | K041166 |
| Product Classification |
Emission Computed Tomography System - Product Code KPS
|
| Product | Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061
Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy. |
| Code Information |
part 8717741, serial numbers 1326, 1315, 1324, 1330, 1310, 1325, 1336, 1307, 1317, 1331, 1299, 1300, 1301, 1302, 1303, 1304, 1305, 1306, 1308, 1309, 1311, 1312, 1313, 1314, 1316, 1318, 1319, 1320, 1322, 1323, 1327, 1328, 1329, 1332, 1334, 1335, and 1337. |
| FEI Number |
1000514697
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 2501 Barrington Rd Hoffman Estates IL 60195-2061
|
| For Additional Information Contact | Mr. Ron Nolte 847-304-7700 |
Manufacturer Reason for Recall | The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan. |
FDA Determined Cause 2 | Finished device change control |
| Action | Siemens sent Urgent Field Correction Recall letters dated 4/22/09 to the affected users of the Symbia S and Symbia T Systems. The accounts were informed of the potential faulty pinion gear which could crack and cause the gantry to spin freely, damaging the system and potentially injuring a patient during a scan. The customers were advised to stop using the system if they hear any abnormal grinding noise from the gantry. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a replacement of the pinion gear free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter. |
| Quantity in Commerce | 37 units |
| Distribution | Worldwide Distribution -- USA, Australia, Austria, Brazil, Bulgaria, Canada, China, Finland, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Russia, Spain, Taiwan, and Turkey. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|