• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Medtronic SIGMA 100/200/300 Series IPGs

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Medtronic SIGMA 100/200/300 Series IPGssee related information
Date Initiated by FirmMay 18, 2009
Date PostedJune 11, 2009
Recall Status1 Terminated 3 on March 28, 2012
Recall NumberZ-1517-2009
Recall Event ID 52076
PMA NumberP980035S002 P890003S065 
Product Classification Implantable Pacemakere Pulse-Generator - Product Code DXY
ProductMedtronic SIGMA 100/200/300 Series IPGs, Dual Chamber atrial sensing, ventricular sensing and pacing Pacemaker, model SVDD303.
Code Information Model SVDD303, serial numbers: PJD183402H, PJD183414H, PJD183415H, PJD183422H, PJD183425H, PJD183426H, PJD183430H, PJD183437H, PJD183440H, PJD183443H, PJD183447H, PJD186130H, PJD186131H, PJD186132H, PJD186133H, PJD186134H, PJD186137H, PJD186138H, PJD186140H, PJD186141H, PJD186142H, PJD186143H, PJD186145H, PJD186147H, PJD186149H, PJD186151H, PJD186152H, PJD186155H, PJD186157H, PJD186158H, PJD186159H, PJD186160H, PJD186165H, PJD186167H, PJD186169H, PJD186172H, PJD186173H, PJD186174H, PJD186175H, PJD186176H, PJD186178H, PJD186179H, PJD186182H, PJD186185H, PJD186186H, PJD186188H, PJD186189H, PJD186192H, PJD186193H, PJD186195H, PJD186196H, PJD188418H, PJD188420H, PJD188421H, PJD188423H, PJD188424H, PJD188425H, PJD188427H, PJD188428H, PJD188429H, PJD188430H, PJD188431H, PJD188432H, PJD188433H, PJD188434H, PJD188435H, PJD188438H, PJD188441H, PJD188442H, PJD188443H, PJD188444H, PJD188445H, PJD188446H, PJD188447H, PJD188448H, PJD188449H, PJD188450H, PJD188451H, PJD188452H, PJD188453H, PJD188454H, PJD188456H, PJD188457H, PJD188458H, PJD188459H, PJD188460H, PJD188463H, PJD188464H, PJD188465H, PJD188466H, PJD188467H, PJD188469H, PJD188470H, PJD188471H, PJD188472H, PJD188473H, PJD188475H, PJD188476H, PJD188477H, PJD188478H, PJD188480H, PJD188482H, PJD188483H, PJD188484H, PJD188485H, PJD188486H, PJD188487H, PJD188488H, PJD188489H, PJD188490H, PJD188493H, PJD188495H, PJD188496H, PJD188497H, PJD188499H, PJD188500H, PJD188501H, PJD188502H, PJD188503H, PJD188504H, PJD188506H, PJD188507H, PJD188508H, PJD188509H, PJD188510H, PJD188511H, PJD188512H, PJD188513H, PJD188515H, PJD188516H, PJD188517H, PJD188519H, PJD201652H, PJD201654H, PJD201657H, PJD201658H, PJD201659H, PJD201660H, PJD201662H, PJD201665H, PJD201666H, PJD201667H, PJD201669H, PJD201670H, PJD201671H, PJD201675H, PJD201676H, PJD201677H, PJD201679H, PJD201681H, PJD201682H, PJD201687H, PJD201688H, PJD201689H, PJD201690H, PJD201694H, PJD203257H, PJD203258H, PJD203259H, PJD203261H, PJD203262H, PJD203264H, PJD203265H, PJD203266H, PJD203267H, PJD203269H, PJD203270H, PJD203271H, PJD203272H, PJD203273H, PJD203274H, PJD203275H, PJD203276H, PJD203277H, PJD203278H, PJD203279H, PJD205935H, PJD205937H, PJD205939H, PJD205940H, PJD205942H, PJD205943H, PJD205944H, PJD205945H, PJD205946H, PJD205948H, PJD205949H, PJD205953H, PJD205955H, PJD205957H, PJD205959H, PJD205962H, PJD205966H, PJD205968H, PJD205970H, PJD205973H, PJD205974H, PJD205976H, PJD205977H, PJD205983H, PJD205984H, PJD205985H, PJD205986H, PJD205990H, PJD205992H, PJD207333H, PJD207334H, PJD207341H, PJD207356H, PJD207363H, PJD207371H, PJD207375H, PJD207382H, PJD207390H, PJD207392H, PJD207396H, PJD208691H, PJD208692H, PJD208703H, PJD208706H, PJD208711H, PJD208713H, PJD208714H, PJD208715H, PJD208724H, PJD208732H, PJD208734H, PJD208736H, PJD208739H, PJD208740H, PJD208743H, PJD209427H, PJD430299H, PJD430300H, PJD430301H, PJD430302H, PJD430303H, PJD430304H, PJD430305H, PJD430306H, PJD430307H, PJD430308H, PJD430309H, PJD430310H, PJD430311H, PJD430312H, PJD430313H, PJD430314H, PJD430316H, PJD430317H, PJD430318H, PJD430319H, PJD430320H, PJD430321H, PJD430322H, PJD430323H, PJD430324H, PJD430325H, PJD430326H, PJD430327H, PJD430328H, PJD430329H, PJD430330H, PJD430331H, PJD441378H, PJD441379H, PJD441380H, PJD441382H, PJD441383H, PJD441384H, PJD441385H, PJD441386H, PJD441387H, PJD441388H, PJD441389H, PJD441390H, PJD441392H, PJD441393H, PJD441394H, PJD441396H, PJD442436H, PJD442448H, PJD445371H, PJD445372H, PJD445373H, PJD445376H, PJD445378H, PJD445806H, PJD445807H, PJD445808H, PJD445809H, PJD445810H, PJD445811H, PJD445812H, PJD445814H, PJD445815H, PJD445816H, PJD445817H, PJD445818H, PJD445819H, PJD445820H, PJD445821H, PJD445822H, PJD445824H, PJD445825H, PJD445826H, PJD445827H, PJD448138H, PJD448139H, PJD610348S, PJD610357S, PJD610358S, PJD610789S, PJD610952S, PJD610953S, PJD610962S, PJD610968S, PJD610971S, PJD611308S, PJD611322S, PJD611326S, PJD611332S, PJD611340S, PJD611349S, PJD611351S, PJD611361S, PJD611364S, PJD611367S, PJD611369S, PJD611376S, PJD611965S, PJD612711S, PJD612774S, PJD614078S, PJD614081S, PJD614086S, PJD614087S, PJD675113S, PJD675114S, PJD675115S, PJD675116S, PJD675117S, PJD675118S, PJD675119S, PJD675120S, PJD675121S, PJD675122S, PJD675123S, PJD675124S, PJD675125S, PJD675128S, PJD675130S, PJD675131S, PJD675132S, PJD675133S, PJD675134S, PJD675135S, PJD675136S, PJD675137S, PJD675138S, PJD675140S, PJD675141S, PJD675142S, PJD675143S, PJD675144S, PJD675145S, PJD675146S, PJD675147S, PJD675148S, PJD675149S, PJD675150S, PJD675151S, PJD675153S, PJD675154S, PJD675157S, PJD675158S, PJD675159S, PJD675160S, PJD675161S, PJD675162S, PJD675163S, PJD675164S, PJD675165S, PJD675166S, PJD675167S, PJD675168S, PJD675169S, PJD675170S, PJD675171S, PJD675172S, PJD675173S, PJD675174S, PJD675175S, PJD675176S, PJD675177S, PJD675178S, PJD675179S, PJD675180S, PJD675181S, PJD675182S, PJD675183S, PJD675184S, PJD675185S, PJD675187S, PJD675188S, PJD675189S, PJD675190S, PJD675191S, PJD675192S, PJD675694S, PJD675695S, PJD675696S, PJD675697S, PJD675698S, PJD675699S, PJD675701S, PJD675702S, PJD675703S, PJD675704S, PJD675705S, PJD675706S, PJD675708S, PJD675709S, PJD675710S, PJD675711S, PJD675712S, PJD675713S, PJD675714S, PJD675715S, PJD675717S, PJD675718S, PJD675719S, PJD675720S, PJD675721S, PJD675722S, PJD675723S, PJD675724S, PJD675725S, PJD675727S, PJD675728S, PJD675729S, PJD675730S, PJD675731S, PJD675732S, PJD675734S, PJD675735S, PJD675736S, PJD675739S, PJD675740S, PJD675741S, PJD675743S, PJD675744S, PJD675745S, PJD676147S, PJD676148S, PJD676149S, PJD676150S, PJD676151S, PJD676152S, PJD676153S, PJD676154S, PJD676155S, PJD676156S, PJD676157S, PJD676158S, PJD676159S, PJD676160S, PJD676161S, PJD676162S, PJD676163S, PJD676164S, PJD676165S, PJD676166S, PJD676167S, PJD676168S, PJD676169S, PJD676170S, PJD676171S, PJD676172S, PJD676173S, PJD676174S, PJD676175S, PJD676176S, PJD676177S, PJD676178S, PJD676179S, PJD676180S, PJD676181S, PJD676183S, PJD676184S, PJD676185S, PJD676186S, PJD676187S, PJD676188S, PJD676189S, PJD676190S, PJD676192S, PJD676193S, PJD676194S, PJD676195S, PJD676196S, PJD676712S, PJD676714S, PJD676715S, PJD676716S, PJD676717S, PJD676718S, PJD676719S, PJD676720S, PJD676721S, PJD676722S, PJD676723S, PJD676724S, PJD676725S, PJD676726S, PJD676727S, PJD676728S, PJD676729S, PJD676730S, PJD676731S, PJD676732S, PJD676733S, PJD676734S, PJD676735S, PJD676737S, PJD676738S, PJD676739S, PJD676740S, PJD676741S, PJD676742S, PJD676743S, PJD676744S, PJD676745S, PJD676746S, PJD676747S, PJD676748S, PJD676750S, PJD676751S, PJD676752S, PJD676753S, PJD676754S, PJD676755S, PJD676756S, PJD676758S, PJD676759S, PJD676761S, PJD676762S, PJD676763S, PJD676764S, PJD676765S, PJD676766S, PJD676767S, PJD676768S, PJD676769S, PJD676770S, PJD676771S, PJD676772S, PJD676773S, PJD676774S, PJD676775S, PJD676776S, PJD676777S, PJD676778S, PJD676779S, PJD676780S, PJD676781S, PJD676783S, PJD676784S, PJD676785S, PJD676786S, PJD676787S, PJD676788S, PJD676789S, PJD676790S, PJD676791S, PJD676792S, PJD676793S, PJD676794S, PJD676795S, PJD676798S, PJD676799S, PJD676800S, PJD676801S, PJD676802S, PJD676803S, PJD676805S, PJD676806S, PJD676807S, PJD676808S, PJD676809S, PJD677182S, PJD677184S, PJD677185S, PJD677186S, PJD677187S, PJD677189S, PJD677190S, PJD677192S, PJD677193S, PJD677194S, PJD677195S, PJD677196S, PJD677197S, PJD677198S, PJD677199S, PJD677200S, PJD677201S, PJD677202S, PJD677203S, PJD677205S, PJD677206S, PJD677207S, PJD677208S, PJD677209S, PJD677210S, PJD677212S, PJD677213S, PJD677214S, PJD677215S, PJD677217S, PJD677218S, PJD677219S, PJD677220S, PJD677221S, PJD677222S, PJD677223S, PJD677224S, PJD677225S, PJD677226S, PJD677227S, PJD677228S, PJD677229S, PJD677230S, PJD677231S, PJD677232S, PJD677233S, PJD677234S, PJD677235S, PJD677236S, PJD677237S, PJD677238S, PJD677239S, PJD677240S, PJD677241S, PJD677242S, PJD677243S, PJD677244S, PJD677246S, PJD677247S, PJD677248S, PJD677249S, PJD677250S, PJD677251S, PJD677252S, PJD677253S, PJD677254S, PJD677255S, PJD677256S, PJD677257S, PJD677258S, PJD677259S, PJD677261S, PJD677471S, PJD677472S.PJD677473S, PJD677474S, PJD677475S, PJD677476S, PJD677477S, PJD677478S, PJD677479S, PJD677480S, PJD677481S, PJD677482S, PJD677483S, PJD677484S, PJD677485S, PJD677486S, PJD677487S, PJD677488S, PJD677489S, PJD677490S, PJD677492S, PJD677493S, PJD677494S, PJD677495S, PJD677496S, PJD677497S, PJD677498S, PJD677499S, PJD677945S, PJD678157S, PJD678158S, PJD678159S, PJD678160S, PJD678161S, PJD678162S, PJD678163S, PJD678164S, PJD678165S, PJD678166S, PJD678167S, PJD678168S, PJD678169S, PJD678170S, PJD678171S, PJD678172S, PJD678173S, PJD678174S, PJD678175S, PJD678176S, PJD678177S, PJD678178S, PJD678179S, PJD678180S, PJD678181S, PJD678182S, PJD678183S, PJD678184S, PJD678185S, PJD678186S, PJD678187S, PJD678188S, PJD678189S, PJD678190S, PJD678191S, PJD678192S, PJD678193S, PJD678194S, PJD678195S, PJD678196S, PJD678197S, PJD678198S, PJD678199S, PJD678200S, PJD678201S, PJD678202S, PJD678203S, PJD678204S, PJD678205S, PJD678206S, PJD678207S, PJD678208S, PJD678209S, PJD678210S, PJD678211S, PJD678212S, PJD678213S, PJD678214S, PJD678215S, PJD678216S, PJD678217S, PJD678218S, PJD678219S, PJD678220S, PJD678221S, PJD678222S, PJD678223S, PJD678224S, PJD678225S, PJD678226S, PJD678227S, PJD678228S, PJD679616S, PJD679617S, PJD679619S, PJD679622S, PJD679623S, PJD679624S, PJD679626S, PJD679628S, PJD679629S, PJD679633S, PJD679634S, PJD679636S, PJD679637S, PJD679638S, PJD679639S, PJD679640S, PJD679642S, PJD679644S, PJD679645S, PJD679647S, PJD679649S, PJD679650S, PJD679651S, PJD679652S, PJD679653S, PJD679654S, PJD679656S, PJD679657S, PJD679659S, PJD679660S, PJD679662S, PJD680212S, PJD680226S, PJD680929S, PJD680930S, PJD680933S, PJD680937S, PJD680938S, PJD680939S, PJD680942S, PJD680946S, PJD680949S, PJD680950S, PJD680952S, PJD680958S, PJD680959S, PJD680960S, PJD680962S, PJD680964S, PJD680966S, PJD680972S, PJD680975S, PJD680978S, PJD680981S, PJD680982S, PJD680984S, PJD680987S, PJD680990S, PJD680991S, PJD680993S, PJD680994S, PJD680995S, PJD680997S, PJD680998S, PJD680999S, PJD681001S, PJD681002S, PJD681003S, PJD681004S, PJD681012S, PJD681013S, PJD681016S, PJD681018S, PJD681020S, PJD681024S, PJD681025S, PJD685370S, PJD685371S, PJD685375S, PJD685377S, PJD685379S, PJD685380S, PJD685381S, PJD685382S, PJD685383S, PJD685385S, PJD685388S, PJD685399S, PJD685400S, PJD685402S, PJD685406S, PJD685408S, PJD685409S, PJD685410S, PJD686906S, PJD692195S, PJD692196S, PJD692197S, PJD692198S, PJD692199S, PJD692200S, PJD692201S, PJD692202S, PJD692203S, PJD692204S, PJD692205S, PJD692206S, PJD692207S, PJD692208S, PJD692210S, PJD692211S, PJD692212S, PJD692214S, PJD692215S, PJD692216S, PJD692217S, PJD692218S, PJD692219S, PJD692220S, PJD692221S, PJD692223S, PJD692224S, PJD692225S, PJD692226S, PJD692227S, PJD692228S, PJD692229S, PJD692230S, PJD692231S, PJD692232S, PJD692233S, PJD692234S, PJD692235S, PJD692236S, PJD692237S, PJD692238S, PJD692239S, PJD692240S, PJD692241S, PJD692242S, PJD692244S, PJD692245S, PJD692246S, PJD692247S, PJD692248S, PJD692249S, PJD692250S, PJD692252S, PJD692253S, PJD692254S, PJD692255S, PJD692256S, PJD692257S, PJD692258S, PJD692259S, PJD692260S, PJD692261S, PJD692262S, PJD692264S, PJD692265S, PJD692266S, PJD692267S, PJD692269S, PJD692270S, PJD692272S, PJD692273S, PJD692274S, PJD692275S, PJD692276S, PJD692277S, PJD692279S, PJD692282S, PJD692284S, PJD692286S, PJD692287S, PJD692288S, PJD692289S, PJD692290S, PJD692291S, PJD692293S, PJD692294S, PJD692295S, PJD692296S, PJD692297S, PJD692298S, PJD692299S, PJD692301S, PJD692302S, PJD692303S, PJD692306S, PJD692307S, PJD693032S, PJD693033S, PJD693034S, PJD693035S, PJD693036S, PJD693037S, PJD693038S, PJD693039S, PJD693040S, PJD693041S, PJD693042S, PJD693043S, PJD693044S, PJD693045S, PJD693046S, PJD693047S, PJD693049S, PJD693050S, PJD693051S, PJD693052S, PJD693053S, PJD693054S, PJD693055S, PJD693056S, PJD693057S, PJD693058S, PJD693059S, PJD693060S, PJD693061S, PJD693062S, PJD693063S, PJD693064S, PJD693065S, PJD693066S, PJD693067S, PJD693068S, PJD693069S, PJD693070S, PJD693071S, PJD693072S, PJD693073S, PJD693074S, PJD693075S, PJD693076S, PJD693077S, PJD693078S, PJD693079S, PJD693080S, PJD693081S, PJD693082S, PJD693083S, PJD693084S, PJD693085S, PJD693086S, PJD693087S, PJD693088S, PJD693089S, PJD693090S, PJD693091S, PJD693092S, PJD693094S, PJD693095S, PJD693096S, PJD693097S, PJD693098S, PJD693099S, PJD693100S, PJD693101S, PJD693102S, PJD693103S, PJD693104S, PJD693105S, PJD693106S, PJD693107S, PJD693108S, PJD693109S, PJD693110S, PJD693111S, PJD693112S, PJD693113S, PJD693114S, PJD693115S, PJD693116S, PJD693117S, PJD693118S, PJD693119S, PJD693120S, PJD693121S, PJD693692S, PJD693693S, PJD693694S, PJD693695S, PJD693696S, PJD693697S, PJD693698S, PJD693699S, PJD693700S, PJD693701S, PJD693702S, PJD693703S, PJD693704S, PJD693705S, PJD693706S, PJD693707S, PJD693708S, PJD693709S, PJD693710S, PJD693711S, PJD693712S, PJD693713S, PJD693714S, PJD693715S, PJD693716S, PJD693717S, PJD693718S, PJD693720S, PJD693721S, PJD693722S, PJD693723S, PJD693724S, PJD693725S, PJD693726S, PJD693727S, PJD693728S, PJD693729S, PJD693730S, PJD693731S, PJD693732S, PJD693733S, PJD693734S, PJD693735S, PJD693736S, PJD693737S, PJD693738S, PJD693739S, PJD693740S, PJD693741S, PJD693742S, PJD693743S, PJD693744S, PJD693745S, PJD693746S, PJD693747S, PJD693749S, PJD693750S, PJD693751S, PJD693752S, PJD693754S, PJD693755S, PJD693756S, PJD693757S, PJD693758S, PJD693759S, PJD693760S, PJD693761S, PJD693762S, PJD693763S, PJD693764S, PJD693765S, PJD693766S, PJD693767S, PJD693768S, PJD693770S, PJD693771S, PJD693772S, PJD693773S, PJD693774S, PJD693775S, PJD693776S, PJD693777S, PJD693778S, PJD693779S, PJD693780S, PJD693781S, PJD693782S, PJD693783S, PJD693784S, PJD693785S, PJD693786S, PJD693787S, PJD693788S, PJD693789S, PJD693790S, PJD693791S, PJD694177S, PJD694178S, PJD694179S, PJD694180S, PJD694181S, PJD694182S, PJD694183S, PJD694184S, PJD694185S, PJD694186S, PJD694187S, PJD694188S, PJD694189S, PJD694190S, PJD694191S, PJD694192S, PJD694193S, PJD694194S, PJD694195S, PJD694196S, PJD694197S, PJD694198S, PJD694199S, PJD694200S, PJD694201S, PJD694202S, PJD694203S, PJD694204S, PJD694205S, PJD694206S, PJD694207S, PJD694208S, PJD694209S, PJD694210S, PJD694211S, PJD694212S, PJD694213S, PJD694214S, PJD694215S, PJD694216S, PJD694217S, PJD694218S, PJD694219S, PJD694220S, PJD694221S, PJD694222S, PJD694223S, PJD694224S, PJD694225S, PJD694226S, PJD694711S, PJD694713S, PJD694715S, PJD694716S, PJD694718S, PJD694719S, PJD694732S, PJD694734S, PJD694735S, PJD694736S, PJD694737S, PJD694738S, PJD694739S, PJD694740S, PJD694741S, PJD694742S, PJD694743S, PJD694744S, PJD694745S, PJD694746S, PJD694747S, PJD694748S, PJD694749S, PJD694750S, PJD694751S, PJD694752S, PJD694753S, PJD694754S, PJD694755S, PJD694756S, PJD694757S, PJD694758S, PJD694759S, PJD694760S, PJD694761S, PJD694762S, PJD694763S, PJD694764S, PJD694765S, PJD694766S, PJD694767S, PJD694768S, PJD694769S, PJD694770S, PJD694771S, PJD694772S, PJD694773S, PJD694774S, PJD694775S, PJD694776S, PJD694778S, PJD694780S, PJD694781S, PJD694782S, PJD694783S, PJD694784S, PJD694785S, PJD694786S, PJD694787S, PJD694788S, PJD694789S, PJD694790S, PJD694791S, PJD694793S, PJD694794S, PJD694795S, PJD694796S, PJD694797S, PJD694798S, PJD694799S, PJD694800S, PJD695188S, PJD695189S, PJD695190S, PJD695191S, PJD695192S, PJD695193S, PJD695194S, PJD695195S, PJD695197S, PJD695199S, PJD695200S, PJD695201S, PJD695461S, PJD695462S, PJD695463S, PJD695467S, PJD695468S, PJD695469S, PJD695470S, PJD695471S, PJD695472S, PJD695473S, PJD695475S, PJD695476S, PJD695479S, PJD695480S, PJD695481S, PJD695482S, PJD695484S, PJD695485S, PJD695486S, PJD695488S, PJD695489S, PJD695491S, PJD695492S, PJD695493S, PJD695494S, PJD695495S, PJD695496S, PJD695497S, PJD695498S, PJD695499S, PJD695500S, PJD695501S, PJD695503S, PJD695504S, PJD695506S, PJD695507S, PJD695508S, PJD695509S, PJD695510S, PJD695511S, PJD695512S, PJD695514S, PJD695515S, PJD695516S, PJD695519S, PJD695520S, PJD695521S, PJD695522S, PJD695524S, PJD695525S, PJD695526S, PJD695527S, PJD695529S, PJD695530S, PJD695531S, PJD695535S, PJD695536S, PJD695537S, PJD695539S, PJD695540S, PJD695541S, PJD695542S, PJD695544S, PJD695545S, PJD695546S, PJD695547S, PJD695550S, PJD695552S, PJD695553S, PJD695554S, PJD695557S, PJD695558S, PJD695559S, PJD695562S, PJD695563S, PJD695565S, PJD695566S, PJD695570S, PJD695571S, PJD695572S, PJD696405S, PJD696617S, PJD696696S, PJD696701S, PJD696709S, PJD696739S, PJD696761S, PJD696788S, PJD696792S, PJD696807S, PJD696808S, PJD696809S, PJD696818S, PJD704625S, PJD704627S, PJD704629S, PJD730015S, PJD731580S, PJD731581S, PJD731582S, PJD731583S, PJD731584S, PJD731585S, PJD731586S, PJD731587S, PJD731588S, PJD731590S, PJD731591S, PJD731592S, PJD731593S, PJD731594S, PJD731595S, PJD731596S, PJD731597S, PJD731598S, PJD731599S, PJD731600S, PJD731601S, PJD731602S, PJD731603S, PJD731604S, PJD731605S, PJD731606S, PJD731608S, PJD731609S, PJD731610S, PJD731611S, PJD731612S, PJD731613S, PJD731614S, PJD731615S, PJD731616S, PJD731617S, PJD731619S, PJD731620S, PJD731621S, PJD731622S, PJD731623S, PJD731624S, PJD731625S, PJD731626S, PJD731627S, PJD731628S, PJD731629S, PJD731631S, PJD731632S, PJD731633S, PJD731637S, PJD731639S, PJD731640S, PJD731641S, PJD731642S, PJD731643S, PJD731644S, PJD731645S, PJD731646S, PJD731647S, PJD731649S, PJD731650S, PJD731651S, PJD731652S, PJD731654S, PJD731655S, PJD731656S, PJD731658S, PJD731659S, PJD731660S, PJD731662S, PJD731663S, PJD731664S, PJD731665S, PJD731666S, PJD731668S, PJD732144S, PJD732145S, PJD732146S, PJD732147S, PJD732148S, PJD732149S, PJD732150S, PJD732152S, PJD732153S, PJD732154S, PJD732155S, PJD732156S, PJD732157S, PJD732158S, PJD732161S, PJD732162S, PJD732163S, PJD732164S, PJD732165S, PJD732166S, PJD732168S, PJD732169S, PJD732170S, PJD732171S, PJD732174S, PJD732175S, PJD732176S, PJD732177S, PJD732178S, PJD732179S, PJD732180S, PJD732182S, PJD732185S, PJD732187S, PJD732190S, PJD732191S, PJD732193S, PJD732194S, PJD732195S, PJD732196S, PJD732198S, PJD732199S, PJD732201S, PJD732202S, PJD732204S, PJD732205S, PJD732206S, PJD732209S, PJD732210S, PJD732212S, PJD732213S, PJD732214S, PJD732215S, PJD732216S, PJD732218S, PJD732219S, PJD732220S, PJD732221S, PJD732273S, PJD732277S, PJD732279S, PJD732283S, PJD732284S, PJD732285S, PJD732287S, PJD732288S, PJD732293S, PJD732294S, PJD732295S, PJD732296S, PJD732297S, PJD732298S, PJD732299S, PJD732300S, PJD732301S, PJD733190S, PJD733201S, PJD733214S, PJD733218S, PJD733221S, PJD733238S, PJD733285S, PJD733291S, PJD733299S, PJD733305S, PJD733312S, PJD733321S, PJD738118S, PJD738119S, PJD738120S, PJD738121S, PJD738122S, PJD738123S, PJD738125S, PJD738126S, PJD738127S, PJD738128S, PJD738129S, PJD738130S, PJD738131S, PJD738132S, PJD738134S, PJD738135S, PJD738136S, PJD738137S, PJD738138S, PJD738139S, PJD738140S, PJD738141S, PJD738142S, PJD738143S, PJD738144S, PJD738145S, PJD738146S, PJD738147S, PJD738148S, PJD738149S, PJD738150S, PJD738151S, PJD738153S, PJD738154S, PJD738155S, PJD738157S, PJD738158S, PJD738160S, PJD738161S, PJD738162S, PJD738163S, PJD738164S, PJD738165S, PJD738167S, PJD738168S, PJD738169S, PJD738170S, PJD738171S, PJD738172S, PJD738173S, PJD738174S, PJD738175S, PJD738176S, PJD738177S, PJD738178S, PJD738179S, PJD738180S, PJD738181S, PJD738182S, PJD738183S, PJD738184S, PJD738185S, PJD738186S, PJD738187S, PJD738188S, PJD738189S, PJD738190S, PJD738191S, PJD738192S, PJD738193S, PJD738194S, PJD738195S, PJD738196S, PJD738198S, PJD738199S, PJD738200S, PJD738201S, PJD738202S, PJD738203S, PJD738204S, PJD738205S, PJD738206S, PJD738207S, PJD738209S, PJD738210S, PJD738211S, PJD738212S, PJD738213S, PJD738214S, PJD738215S, PJD738216S, PJD738217S, PJD738218S, PJD738219S, PJD738220S, PJD738221S, PJD738222S, PJD738223S, PJD738224S, PJD738225S, PJD738226S, PJD738227S, PJD738228S, PJD738231S, PJD738233S, PJD738234S, PJD738235S, PJD738236S, PJD738237S, PJD738238S, PJD738239S, PJD738240S, PJD738241S, PJD738242S, PJD738243S, PJD738244S, PJD738245S, PJD738246S, PJD738248S, PJD738249S, PJD738250S, PJD738252S, PJD738253S, PJD738254S, PJD738255S, PJD738256S, PJD738257S, PJD738258S, PJD738259S, PJD738260S, PJD738261S, PJD738262S, PJD738263S.PJD738264S, PJD738265S, PJD738266S, PJD738267S, PJD738268S, PJD738269S, PJD738270S, PJD738273S, PJD738274S, PJD738275S, PJD738276S, PJD738277S, PJD738278S, PJD738279S, PJD738280S, PJD738281S, PJD738282S, PJD738284S, PJD738286S, PJD738288S, PJD738289S, PJD738290S, PJD738291S, PJD738292S, PJD738293S, PJD738294S, PJD738295S, PJD738296S, PJD738299S, PJD738300S, PJD738304S, PJD738305S, PJD738306S, PJD738308S, PJD738309S, PJD738312S, PJD738315S, PJD738316S, PJD738318S, PJD738320S, PJD738325S, PJD738326S, PJD738328S, PJD738330S, PJD738331S, PJD738332S, PJD738333S, PJD738334S, PJD738336S, PJD738337S, PJD738338S, PJD738339S, PJD738342S, PJD738343S, PJD738344S, PJD738346S, PJD738347S, PJD738348S, PJD738349S, PJD738351S, PJD738352S, PJD738353S, PJD738356S, PJD738358S, PJD738362S, PJD738363S, PJD738366S, PJD738368S, PJD738369S, PJD738371S, PJD738374S, PJD738375S, PJD738376S, PJD738378S, PJD738380S, PJD738381S, PJD738382S, PJD738386S, PJD738388S, PJD738391S, PJD738392S, PJD738393S, PJD738395S, PJD738396S, PJD738398S, PJD738399S, PJD738402S, PJD738403S, PJD738404S, PJD738613S, PJD738614S, PJD738615S, PJD738616S, PJD738617S, PJD738618S, PJD738620S, PJD738621S, PJD738622S, PJD738623S, PJD738624S, PJD738625S, PJD738626S, PJD738627S, PJD738628S, PJD738629S, PJD738630S, PJD738631S, PJD738632S, PJD738633S, PJD738634S, PJD738635S, PJD738636S, PJD738637S, PJD738638S, PJD738639S, PJD738640S, PJD738641S, PJD738642S, PJD738643S, PJD738644S, PJD738645S, PJD738646S, PJD738647S, PJD738648S, PJD738649S, PJD738650S, PJD738651S, PJD738652S, PJD738860S, PJD738861S, PJD738862S, PJD738863S, PJD738864S, PJD738865S, PJD738866S, PJD738867S, PJD738868S, PJD738869S, PJD738870S, PJD738871S, PJD738872S, PJD738873S, PJD738874S, PJD738875S, PJD738876S, PJD738877S, PJD738878S, PJD738879S, PJD738880S, PJD738881S, PJD738882S, PJD738885S, PJD738886S, PJD738887S, PJD738888S, PJD738889S, PJD738890S, PJD738892S, PJD738893S, PJD738894S, PJD738895S, PJD738896S, PJD738898S, PJD738900S, PJD739168S, PJD739169S, PJD739170S, PJD739171S, PJD739172S, PJD739173S, PJD739174S, PJD739175S, PJD739176S, PJD739177S, PJD739178S, PJD739179S, PJD739180S, PJD739181S, PJD739182S, PJD739183S, PJD739184S, PJD739185S, PJD739186S, PJD739187S, PJD739188S, PJD739189S, PJD739190S, PJD739191S, PJD739192S, PJD739193S, PJD739196S, PJD739197S, PJD739198S, PJD739199S, PJD739201S, PJD739202S, PJD739203S, PJD739204S, PJD739205S, PJD739206S, PJD739208S, PJD739209S, PJD739211S, PJD739212S, PJD739213S, PJD739214S, PJD739215S, PJD739216S, PJD739217S, PJD739218S, PJD739220S, PJD739221S, PJD739222S, PJD739223S, PJD739224S, PJD739225S, PJD739226S, PJD739227S, PJD739228S, PJD739230S, PJD739232S, PJD739233S, PJD739234S, PJD739235S, PJD739236S, PJD739239S, PJD739240S, PJD739242S, PJD739243S, PJD739244S, PJD739246S, PJD739250S, PJD739251S, PJD739252S, PJD739253S, PJD739254S, PJD739256S, PJD739257S, PJD739258S, PJD739259S, PJD739260S, PJD739261S, PJD739262S, PJD739263S, PJD739264S, PJD739265S, PJD739267S, PJD739268S, PJD739269S, PJD739270S, PJD739272S, PJD739273S, PJD739274S, PJD739275S, PJD739276S, PJD739277S, PJD739279S, PJD739280S, PJD739281S, PJD739283S, PJD739284S, PJD739286S, PJD739288S, PJD739289S, PJD739290S, PJD739293S, PJD739294S, PJD739296S, PJD739297S, PJD739298S, PJD739301S, PJD739302S, PJD739305S, PJD739308S, PJD739310S, PJD739311S, PJD739312S, PJD739313S, PJD739314S, PJD739316S, PJD739320S, PJD739321S, PJD739323S, PJD739324S, PJD739325S, PJD739326S, PJD739647S, PJD739648S, PJD739650S, PJD739651S, PJD739652S, PJD739654S, PJD739655S, PJD739656S, PJD739657S, PJD739658S, PJD739661S, PJD739662S, PJD739665S, PJD739666S, PJD739667S, PJD739668S, PJD739669S, PJD739670S, PJD739671S, PJD739672S, PJD739673S, PJD739674S, PJD739675S, PJD739676S, PJD739677S, PJD739678S, PJD739679S, PJD739680S, PJD739681S, PJD739682S, PJD739683S, PJD739684S, PJD739685S, PJD739686S, PJD739687S, PJD739688S, PJD739689S, PJD739691S, PJD739692S, PJD739693S, PJD739694S, PJD739695S, PJD739696S, PJD739698S, PJD739699S, PJD739700S, PJD739701S, PJD739702S, PJD739703S, PJD739704S, PJD739705S, PJD739706S, PJD739707S, PJD739708S, PJD739709S, PJD739710S, PJD739712S, PJD739713S, PJD739714S, PJD739715S, PJD739716S, PJD739717S, PJD739718S, PJD739720S, PJD739721S, PJD739722S, PJD739723S, PJD739724S, PJD739725S, PJD739726S, PJD739727S, PJD739728S, PJD739729S, PJD739730S, PJD739732S, PJD739733S, PJD739734S, PJD739735S, PJD739736S, PJD739738S, PJD739739S, PJD739740S, PJD739741S, PJD739743S, PJD739744S, PJD740545S, PJD740546S, PJD740547S, PJD740548S, PJD740549S, PJD740550S, PJD740552S, PJD740553S, PJD740554S, PJD740555S, PJD740556S, PJD740557S, PJD740558S, PJD740559S, PJD740560S, PJD740561S, PJD740562S, PJD740563S, PJD740564S, PJD740565S, PJD740566S, PJD740567S, PJD740568S, PJD740569S, PJD740570S, PJD740571S, PJD740572S, PJD740573S, PJD740574S, PJD740575S, PJD740576S, PJD740577S, PJD740579S, PJD740580S, PJD740581S, PJD740582S, PJD740583S, PJD740584S, PJD740585S, PJD740586S, PJD740587S, PJD740588S, PJD740589S, PJD740590S, PJD740592S, PJD740593S, PJD740594S, PJD740595S, PJD740596S, PJD740597S, PJD740598S, PJD740599S, PJD740600S, PJD740601S, PJD740602S, PJD740603S, PJD740604S, PJD740605S, PJD740606S, PJD740607S, PJD740608S, PJD740609S, PJD740611S, PJD740612S, PJD740613S, PJD740614S, PJD740615S, PJD740616S, PJD740617S, PJD740618S, PJD740619S, PJD740620S, PJD740621S, PJD740622S, PJD740623S, PJD740624S, PJD740625S, PJD740626S, PJD740627S, PJD740628S, PJD740629S, PJD740630S, PJD740631S, PJD740632S, PJD740633S, PJD740634S, PJD740635S, PJD740636S, PJD740637S, PJD740638S, PJD740639S, PJD740640S, PJD740641S, PJD740642S, PJD740643S, PJD740974S, PJD740975S, PJD740976S, PJD740977S, PJD740978S, PJD740979S, PJD740981S, PJD740984S, PJD740985S, PJD740986S, PJD740987S, PJD740988S, PJD740989S, PJD740990S, PJD740991S, PJD740992S, PJD740993S, PJD740994S, PJD740995S, PJD740996S, PJD740997S, PJD740998S, PJD740999S, PJD741000S, PJD741001S, PJD741002S, PJD741003S, PJD741004S, PJD741005S, PJD741006S, PJD741007S, PJD741008S, PJD741010S, PJD741011S, PJD741012S, PJD741013S, PJD741015S, PJD741016S, PJD741017S, PJD741018S, PJD741019S, PJD741020S, PJD741021S, PJD741022S, PJD741023S, PJD741025S, PJD741026S, PJD741027S, PJD741029S, PJD741030S, PJD741031S, PJD741032S, PJD741033S, PJD741034S, PJD741035S, PJD741036S, PJD741037S, PJD741038S, PJD741039S, PJD741040S, PJD741041S, PJD741042S, PJD741043S, PJD741044S, PJD741045S, PJD741046S, PJD741048S, PJD741049S, PJD741050S, PJD741051S, PJD741052S, PJD741053S, PJD741054S, PJD741055S, PJD741056S, PJD741057S, PJD741058S, PJD741059S, PJD741060S, PJD741062S, PJD741063S, PJD741065S, PJD741066S, PJD741067S, PJD741068S, PJD741071S, PJD741073S, PJD741074S, PJD741075S, PJD741076S, PJD741077S, PJD741078S, PJD741079S, PJD741080S, PJD741081S, PJD741082S, PJD741083S, PJD741246S, PJD741247S, PJD741251S, PJD741261S, PJD741273S, PJD741281S, PJD741283S, PJD741284S, PJD741285S, PJD741287S, PJD741293S, PJD741299S, PJD741301S, PJD741303S, PJD741305S, PJD741307S, PJD741309S, PJD741660S, PJD741672S, and PJD744349S.   
Recalling Firm/
Manufacturer
Medtronic Inc. Cardiac Rhythm Disease Managment
8200 Coral Sea St. N.E.
Saint Paul MN 55112
For Additional Information Contact
763-514-4000
Manufacturer Reason
for Recall
One or more bond wire pairs will lift or separate from the bonding terminals on the device hybrid. This may present clinically as loss of rate response, premature battery depletion, loss of telemetry, or no output.
FDA Determined
Cause 2
Device Design
ActionMedtronic Sales representative hand delivered an Important Patient Safety Information letter, dated May 2009, to physicians beginning 05/18/09. The letter describes the issues, provides a predicted failure rate for the 3 populations of devices, and provides Patient Management Recommendations. The letter recommended that physicians should advise their patients to seek medical attention immediately if they experience symptoms (e.g., fainting or lightheadedness). It was also recommended that physicians should consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets. Medtronic will offer supplemental device warranty if the device is not already at elective replacement time. The last recommendation was that physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent. The letter also provides Physician and Patient Support. Letters will be mailed to patients with registered devices beginning 05/27/09. Questions should be directed to a local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
Quantity in Commerce2247
DistributionWorldwide Distribution -- USA, including Washington D.C., Guam, Puerto Rico, and Virgin Islands; Albania, Algeria, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latin America, Latvia, Lebanon, Luxembourg, Martinique, Mauritius, Mexico, Morocco, Nepal, Netherlands, Netherlands Antilles, Norway, Oman, Pakistan, Palestinian Territory Occupied, Peru, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Senegal, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Vatican City State (Holy See), and Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = DXY
-
-