| Date Initiated by Firm | June 26, 2009 |
| Date Posted | September 14, 2009 |
| Recall Status1 |
Terminated 3 on July 27, 2010 |
| Recall Number | Z-2135-2009 |
| Recall Event ID |
52713 |
| 510(K)Number | K873251 |
| Product Classification |
Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate - Product Code MEH
|
| Product | Stryker Orthopaedics Restoration Gap Plate Screw;
Dia. 6.5 mm, Lnth 20 mm;
Made in USA, Sterile;
Howmedica Osteonics Corp.
325 Corporate Drive
Mahwah, NJ 07430 |
| Code Information |
Catalog number: 2080-0020; GAP PLATE SCREWS T0WMKE 01-Jul-2013 GAP PLATE SCREWS T0XMKE 02-Jul-2013 GAP PLATE SCREWS T10MKE 02-Jul-2013 GAP PLATE SCREWS T0YMKE 02-Jul-2013 GAP PLATE SCREWS 22RMLE 07-Jul-2013 GAP PLATE SCREWS 229MLE 07-Jul-2013 GAP PLATE SCREWS 22PMLE 08-Jul-2013 GAP PLATE SCREWS D01MLE 21-Jul-2013 GAP PLATE SCREWS AYYMLE 23-Jul-2013 GAP PLATE SCREWS PLAMME 27-Aug-2013 GAP PLATE SCREWS PN2MME 27-Aug-2013 GAP PLATE SCREWS PL8MME 28-Aug-2013 GAP PLATE SCREWS PL9MME 28-Aug-2013 GAP PLATE SCREWS PN1MME 28-Aug-2013 GAP PLATE SCREWS 7DXMNE 16-Sep-2013 GAP PLATE SCREWS MET9H4 18-Dec-2013 GAP PLATE SCREWS MET9X8 18-Dec-2013 GAP PLATE SCREWS MET9X7 22-Dec-2013 GAP PLATE SCREWS METL9Y 28-Dec-2013 GAP PLATE SCREWS METLA0 30-Dec-2013 GAP PLATE SCREWS MET9H5 30-Dec-2013 GAP PLATE SCREWS MET9X9 30-Dec-2013 GAP PLATE SCREWS MET9XA 30-Dec-2013 GAP PLATE SCREWS MHA4YT 14-Jan-2014 GAP PLATE SCREWS MHA4YR 18-Jan-2014 GAP PLATE SCREWS MHA55K 18-Jan-2014 GAP PLATE SCREWS MHA55L 18-Jan-2014 GAP PLATE SCREWS MHA55H 18-Jan-2014 |
| FEI Number |
3003070421
|
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp.
325 Corporate Drive
Mahwah
NJ
07430
|
| For Additional Information Contact | Colleen O'Meara 201-972-2100 |
Manufacturer Reason for Recall | Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069.
For further information, contact Stryker Orthopaedics at 1-201-831-5970. |
| Quantity in Commerce | 51,113 total - all sizes. |
| Distribution | Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = MEH
|