Class 2 Device Recall Pyxis Anesthesia System 3500 |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
December 24, 2008 |
Date Posted |
September 06, 2009 |
Recall Status1 |
Terminated 3 on March 28, 2012 |
Recall Number |
Z-1883-2009 |
Recall Event ID |
52767 |
Product Classification |
Pyxis Anesthesia System 3500 - Product Code BRY
|
Product |
Pyxis Anesthesia System 3500, Cardinal Health, Clinical Technologies and Services, San Diego, CA.
The Pyxis Anesthesia Station 3500 is a medication dispensing product that provides medication and inventory management for the anesthesiologist at the point of care within the operating room. The system tracks and stores controlled substances, medications and supplies for anesthesia providers and automates verification of patient usage records, documents waste, manages inventory levels and tracks utilization. |
Code Information |
Serial Numbers: 710170646, 710170647, 710172138, 710172139, 710172227, 710172228, 710172716, 710172717, 710172883, 710172884, 710173038, 8000030317, 8000100629, 8000119658, 8000120032, 8000120033, 8000120034, 8000120036, 8000120037, 8000120038, 8000120040, 8000120042, 8000120945, 8000120946, 12838303, 12838304, 12838305, 12838306, 12838307, 12838308, 12838309, 12838311, 12838312, 12838313, 12838314, 12838315, 12780360, 12780361, 12780362, 12810275, 12810277, 12810278, 12810280, 12810282, 12810284, 12810286, 12810288, 12847974, 12847975, 12847976, 12847977, 12852780, 12852781, 12852782, 12852783, 12853714, 12853715, 12853716, 12847978, 12813196, 12835697, 12835698, 12858993, 12720368, 12720369, 12720371, 12753364, 12753365, 12753366, 12753368, 12753369, 12753370, 12753371, 12753372, 12753374, 12753727, 12753728, 12753729, 12753730, 12753732, 12753733, 12753734, 12813195, 12813197, 12804383, 12804384, 12804385, 12804386, 12804387, 12804388, 12804389, 12804390, 12804391, 12804392, 12862538, 12862541, 12862543, 12862544, 12862546, 12862547, 12862548, 12862549, 12862550, 12862551, 12862552, 12862553, 12862554, 12862555, 12862556, 12862557, 12862558, 12862559, 12862560, 12862561, 12862562, 12862563, 12862564, 12876732, 12655533, 12655534, 12655536, 12655537, 12655538, 12655539, 12655540, 12874928, 12867424, 12867425, 12867426, 12867427, 12867428, 12867430, 12867431, 12867433, 12867434, 12838367, 12853706, 12853707, 12853708, 12888255, 12720368, 12720369, 12720371, 12592245, 12874930, 12795130, 12795131, 12795132, 12795133, 12795134, 12795135, 12795136, 12795137, 12795138, 12843701, 12843702, 12843703, 12843753, 12843754, 12843755, 12843756, 12843757, 12843758, 12843759, 12843760, 12783138, 12783139, 12783140, 12874932, 12563061, 12563062, 12563063, 12563064, 12856793, 12874371, 12874372, 12874373, 12874374, 12874375, 12874376, 12874377, 12874378, 12874379, 12874380, 12874381, 12874382, 12874383, 12874384, 12819556, 12819557, 12819558, 12819559, 12819560, 12819561, 12786488, 12786489, 12786490, 12843768, 12843770, 12843771, 12843772, 12868966, 12868972, 12870869, 12870870, 12870871, 12870872, 12870873, 12870874, 12870875, 12870876, 12870877, 12870878, 12870879, 12807345, 12871437, 12872230, 12811867, 12811868, 12843749, 12843750, 12810158, 12834022, 12511905, 12511906, 12511907, 12511908, 12511909, 12514284, 12515162, 12515163, 12515164, 12515165, 12515166, 12515167, 12515168, 12515169, 12517084, 12517085, 12517086, 12517087, 12517088, 12517089, 12517090, 12517091, 12517092, 12549935, 12549937, 12549939, 12549942, 12549944, 12549946, 12552472, 12552473, 12552474, 12552475, 12552476, 12552477, 12759450, 12587758, 12587763, 12587765, 12587771, 12595383, 12595386, 12595389, 12595392, 12596725, 12600728, 12600729, 12600730, 12600731, 12600732, 12600734, 12600736, 12600737, 12600741, 12814051, 12814052, 12814053, 12814054, 12814055, 12814056, 12814057, 12814058, 12534219, 12534220, 12534222, 12534225, 12534229, 12816644, 12816645, 12816646, 12827011, 12827012, 12827013, 12827014, 12827015, 12827016, 12827017, 12827018, 12827019, 12827021, 12827023, 12827024, 12827025, 12827026, 12827027, 12827028, 12847184, 12847185, 12847186, 12847187, 12847188, 12847189, 12847190, 12847191, 12847192, 12847193, 12847194, 12795140, 12838351, 12838352, 12546145, 12806536, 12587761, 12587762, 12592236, 12592239, 12592246, 12595384, 12595385, 12595398, 12596715, 12596721, 12808661, 12808663, 12808664, 12808665, 12808666, 12808667, 12752926, 12752927, 12752928, 12752930, 12753353, 12753354, 12753355, 12753356, 12753357, 12753358, 12753359, 12753360, 12753361, 12753362, 12753363, 12852784, 12852785, 12854988, 12854989, 12854990, 12854991, 12854992, 12854993, 12854994, 12854995, 12854996, 12854997, 12854998, 12763732, 12763737, 12787545, 12787546, 12787547, 12787548, 12787550, 12787551, 12792520, 12793260, 12793261, 12847182, 12847183, 12822094, 12822095, 12822096, 12601449, 12795866, 12795867, 12795868, 12795869, 12795870, 12795871, 12795872, 12795873, 12795874, 12795875, 12795876, 12803181, 12835700, 12835701, 12835702, 12835703, 12835704, 12835705, 12835706, 12835707, 12835708, 12835709, 12835710, 12835711, 12835712, 12556158, 12556159, 12556160, 12556161, 12556162, 12556163, 12556164, 12556165, 12556166, 12556167, 12556168, 12559460, 12559461, 12753735, 12753736, 12754540, 12754544, 12754545, 12754547, 12754551, 12754552, 12754554, 12822757, 12822758, 12822759, 12822760, 12822763, 12822769, 12822770, 12822771, 12822772, 12822773, 12822775, 12822776, 12822777, 12822778, 12822779, 12822780, 12823498, 12823499, 12823500, 12823502, 12823503, 12823504, 12823505, 12823506, 12823507, 12823508, 12823509, 12875428, 12834957, 12834958, 12834959, 12834960, 12834961, 12834962, 12822053, 12822054, 12822055, 12822056, 12804393, 12833156, 12833157, 12833158, 12833159, 12834025, 12834026, 12847967, 12847968, 12847969, 12579903, 12579904, 12579905, 12579907, 12579911, 12579912, 12579913, 12579914, 12579915, 12579921, 12579922, 12579923, 12579924, 12579925, 12585045, 12585046, 12585047, 12585048, 12585049, 12585050, 12585051, 12585052, 12585053, 12585054, 12585058, 12585060, 12585061, 12597965, 12619056, 12638091, 12638092, 12638093, 12662280, 12810152, 12810153, 12810154, 12811742, 12811743, 12811744, 12811745, 12876733, 12601175, 12601176, 12526773, 12526774, 12526775, 12786259, 12786260, 12786261, 12786262, 12786264, 12786265, 12786266, 12786267, 12786268, 12786269, 12786270, 12786271, 12592234, 12592241, 12592242, 12860198, 12863837, 12863838, 12863839, 12863840, 12583173, 12583173, 12583174, 12583175, 12872243, 12872244, 12872245, 12872246, 12872248, 12872249, 12872250, 12872252, 12872253, 12872255, 12872256, 12872257, 12872258, 12873760, 12761945, 12761946, 12761947, 12761948, 12761949, 12761950, 12761951, 12763244, 12763244, 12765330, 12765337, 12765338, 12765339, 12791937, 12850276, 12850277, 12850278, 12850279, 12850280, 12850282, 12850283, 12851321, 12851326, 12851328, 12851329, 12851330, 12851331, 12851334, 12851347, 12851348, 12853709, 12853710, 12853711, 12853712, 12853713, 12782360, 12782361, 12791938, 12850275, 12850276, 12850277, 12850278, 12850279, 12850280, 12850282, 12850283, 12568466, 12580045, 12582587, 12582589, 12833067, 12833068, 12559458, 12559459, 12858992, 12858994, 12858996, 12858999, 12859005, 12859007, 12859008, 12859009, 12859010, 12859012, 12859013, 12861135, 12843747, 12853836, 12838368, 12838369, 12816647, 12816648, 12817410, 12817411, 12817412, 12817413, 12817414, 12817415, 12873755, 12858995, 12858997, 12858998, 12859000 12859001, 12859002, 12859003, 12847970, 12847971, 12847972, 12857866, 12857868, 12857869 12857870, 12860199, 12860201, 12860203, 12860204, 12860205, 12860206, 12860207, 12860209, 12860211, 12860213, 12860214, 12860215, 12873042, 12873043, 12873044, 12873045, 12791875, 12791876, 12791877, 12791878, 12791879, 12791880, 12791881, 12791882, 12791883, 12791884, 12791885, 12811001, 12811002, 12811003, 12811004, 12811005, 12811006, 12811007, 12811008, 12811009, 12811011, 12851320, 12851322, 12851323, 12851325, 12851327, 12822768, 12830021, 12830022, 12830023, 12830024, 12830025, 12830026, 12830027, 12830028, 12830029, 12830030, 12833044, 12846390, 12846391, 12846392, 12846393, 12846394, 12846395, 12846396, 12846397, 12846398, 12846400, 12846401, 12847966, 12798023, 12798024, 12798025, 12798026, 12798027, 12806530, 12806531, 12806532, 12806533, 12806534, 12806535, 12797279, 12874316, 12874385, 12874386, 12874387, 12874315, 12806524, 12809423, 12809424, 12809425, 12809426, 12809427, 12809428, 12809429, 12809430, 12809431, 12838342, 12838343, 12838344, 12838345, 12843761, 12874931, 12818210, 12559462, 12600996, 12600997, 12600998, 12600999, 12601000, 12807994, 12807995, 12807996, 12807997, 12807998, 12807999, 12808000, 12808001, 12601447, 12803177, 12803178, 12803179, 12803180, 12803852, 12806525, 12806526, 12806527, 12806528, 12813190, 12813191, 12813192, 12813193, 12815783, 12815784, 12815785, 12815786, 12815787, 12815788, 12815789, 12815790, 12815791, 12815792, 12873037, 12873038, 12873039, 12873040, 12848975, 12848976, 12848977, 12848978, 12848979, 12848980, 12848981, 12848982, 12848983, 12848984, 12848985, 12848986, 12838316, 12838317, 12838318, 12838319, 12820596, 12820597, 12820598, 12820599, 12820600, 12820601, 12820602, 12820603, 12820604, 12820605, 12820606, 12820607, 12820608, 12820609, 12820610, 12820611, 12820612, 12820613, 12820614, 12820615, 12820616, 12820617, 12820618, 12820619, 12820620, 12820621, 12820622, 12820623, 12820624, 12820625, 12820626, 12870867, 12822092, 12822093, 12830896, 12830897, 12830898, 12830899, 12830900, 12830901, 12830902, 12830903, 12830904, 12830905, 12830906, 12830907, 12830908, 12830909, 12830910, 12830911, 12830912, 12833042, 12833043, 12838346, 12803853, 12803854, 12803855, 12803856, 12803857, 12803858, 12803859, 12874925, 12874926, 12874927, 12822767, 12829260, 12829266, 12829267, 12829268, 12829269, 12829270, 12829271, 12829272, 12833045, 12773706, 12773707, 12583172, 12600993, 12600994, 12600995, 12838347, 12838348, 12838349, 12838350, 12556216, 12556217, 12556218, 12556219, 12556220, 12556221, 12556222 12556223, 12556224, 12556225, 12556226, 12556227, 12789734, 12789735, 12856569, 12856570, 12856572, 12856573, 12587764, 12592240, 12592243, 12592244, 12592247, 12592248, 12600738, 12601006, 12601174, 8000120039, 8000120041, 8000121851, 8000121852, 8000121853, 8000121854, 8000121855, 8000121856, 8000121857, 8000121858, 8000121859, 8000121860, 8000121861, 8000121862, 8000121863, 8000121864, 12600727, 12600740, 12601002, 12601004, 12601511, 12794398, 12794399, 12794400, 12794401, 12794402, 12818805, 12818806, 12818807, 12818808, 12818809, 12834965, 12834966, 12541280, 12541281, 12541282, 12541283, 12541284, 12541285, 12851337, 12851338, 12851339, 12851340, 12851341, 12851342, 12851344, 12851345, 12851346, 12488428, 12488431, 12488433, 12488435, 12538087, 12870868, 12795859, 12834012, 12834013, 12834014, 12834015, 12834016, 12834017, 12834018, 12834019, 12834020, 12834021, 12834023, 12834024, 12850288, 12850289, 12850290, 12850291, 12850292, 12850293, 12850294, 12850295, 12850296, 12850297, 12850298, 12850299, 12850300, 12850301, 12850302, 12818798, 12818799, 12818803, 12818804, 12773619, 12773621, 12773622, 12773623, 12773624, 12773625, 12773626, 12868968, 12868969, 12868970, 12868971, 8000118955, 8000118956, 8000118957, 8000118958, 8000118959, 8000118960, 8000118961, 8000118962, 8000118963, 8000118964, 8000118965, 8000118966, 8000118967, 8000118968, 8000118969, 12857851, 12857852, 12857853, 12857851, 12857852, 12857853, 12833160, 12836437, 12836438, 12836439, 12836440, 12836441, 12836442, 12836443, 12856571, 12872247, 12873756, 12873757, 12873758, 12873759, 12873761, 12873762, 12873763, 12873765, 12587284, 12587285, 12589063, 12589067, 12589069, 12589071, 12589072, 12589074, 12589075, 12589076, 12589077, 12589078, 12589079, 12709624, 12709626, 12811013, 12811015, 12797278, 12871438, 12871440, 12871441, 12871442, 12871443, 12871444, 12784008, 12784009, 12784010, 12784011, 12811869, 12752918, 12752919, 12752920, 12752921, 12752922, 12752924, 12752925, 12568468, 12574027, 12582585, 12594132, 12591366, 12591367, 12594129, 12594130, 12594131, 12601172, 12601173, 12601308, 12601309, 12601451, 12798042, 12798043, 12798044, 12798045, 12798046, 12798047, 12798049, 12798050, 12798051, 12798053, 12802151, 12813194, 12774791, 12774792, 12602063, 12583172, 12600993, 12600994, 12600995, 12786263, 12786272, 12843776, 12843777, 12843779, 12843780, 12847973, 12840507, 12840508, 12840509, 12840510, 12840511, 12840512, 12840513, 12840514, 12833046, 12798894 12798895, 12798896, 12798897, 12798898, 12798899, 12810275, 12810277, 12810278, 12810280, 12810282, 12810284, 12810286, 12810288 and 12856794. |
Recalling Firm/ Manufacturer |
Cardinal Health Manufacturing LLC 3750 Torrey View Ct San Diego CA 92130-2622
|
For Additional Information Contact |
858-617-5889
|
Manufacturer Reason for Recall |
Cardinal Health has sent out a Safety Alert Notification to inform its customers that the SR2b Software for the Pyxis Anesthesia System 3500 (PAS 3500) has received several customer reports of system "non-responsiveness" (lock ups) during the use of the product since its release in September 2008.
|
FDA Determined Cause 2 |
Software design |
Action |
The "Safety Alert Notification" letter was mailed to all Pyxis Anesthesia System 3500 customers on December 17, 2008 by registered return receipt mail. The letter was addressed to the Director of Pharmacy at each facility. The mailing packages contain the "Safety Alert Notification" letter and the revised pages of the Pyxis Anesthesia System 3500 User Guide and Safety Manual. The "Safety Alert Notification" letter reinforced to customers how to maintain back-up processes to allow quick access if a system lock-up condition is experienced.
The letter informed customers the situation will be corrected and a solution will be implemented. If you experience any system unresponsiveness during the use of the product, please report the issue to the Cardinal Health Technical Support Center at 1-800-727-4102. If you have any questions regarding this issue, please contact your local field support representative or the Customer Advocacy Department at Customer.Advocacy@cardinalhealth.com. |
Quantity in Commerce |
250 |
Distribution |
Nationwide and Germany. |