| | Class 2 Device Recall ACUSON Sequoia ultra sound systems with Auxiliary 2 MHz Continuous Wve (CW) transducer |  |
| Date Initiated by Firm | March 21, 2009 |
| Date Posted | April 12, 2011 |
| Recall Status1 |
Terminated 3 on March 09, 2012 |
| Recall Number | Z-1961-2011 |
| Recall Event ID |
52928 |
| 510(K)Number | K063085 |
| Product Classification |
System, imaging, pulsed echo, ultrasonic - Product Code ITX
|
| Product | Auxiliary 2 MHz Continuous Wave (CW) transducer, model number 8251046, when used with ACUSON Sequoia ultrasound systems at SW revision 12.X. Manufactured by Siemens Medical Solutions USA, Inc., Ultrasound Business Unit, Mountain View, CA.
The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. It is intended for the following applications: Abdominal, General Imaging and Cardiology for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breast, testes, thyroid, and penis), Neonatal/Adult Transvaginal, Peripheral Vessels and Musculo-skeletal (superficial and conventional). The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes. The Siemens Aux CW transducer is a non-imaging continuous wave transducer indicated for use in transthoracic/cardiac adult and pediatric applications as well as continuous wave spectral Doppler echocardiography. |
| Code Information |
ACUSON Sequoia serial numbers 50026, 50030, 50042, 50043, 50057, 50087, 50108, 50116, 50172, 50202, 50216, 50243, 50252, 50271, 50279, 50295, 50342, 50355, 50368, 50374, 50418, 50425, 50448, 50493, 50497, 50560, 50561, 50604, 50651, 50676, 50680, 50728, 50777, 50788, 50794, 50796, 50806, 50852, 50855, 50893, 50898, 50920, 50938, 50959, 50964, 51020, 51046, 51055, 51087, 51128, 51185, 51213, 51215, 51229, 51243, 51270, 51278, 51290, 51311, 51332, 51350, 51354, 51367, 51389, 51414, 51433, 51446, 51496, 51498, 51514, 51516, 51523, 51524, 51604, 51636, 51639, 51643, 51652, 51740, 51742, 51747, 51757, 51766, 51781, 51792, 51796, 51800, 51811, 51830, 51844, 51846, 51859, 51897, 51899, 51910, 51920, 51926, 51949, 51955, 51971, 51975, 51979, 51980, 51987, 52003, 52006, 52012, 52023, 52040, 52041, 52047, 52059, 52065, 52075, 52087, 52089, 52101, 52111, 52133, 52136, 52143, 52158, 52170, 52191, 52193, 52195, 52200, 52202, 52243, 52252, 52263, 52272, 52274, 52277, 52282, 52285, 52297, 52307, 52314, 52320, 52322, 52327, 52335, 52346, 52365, 52366, 52375, 52394, 52407, 52410, 52452, 52456, 52475, 52491, 52496, 52504, 52505, 52529, 52559, 52566, 52571, 52572, 52578, 52602, 52604, 52611, 52612, 52629, 52642, 52644, 52645, 52652, 52665, 52683, 52685, 52689, 52694, 52700, 52703, 52704, 52705, 52706, 52708, 52722, 52725, 52735, 52740, 52744, 52750, 52753, 52757, 52768, 52774, 52780, 52794, 52795, 52798, 52800, 52801, 52808, 52811, 52829, 52853, 52865, 52868, 52883, 52906, 52908, 52926, 52936, 52954, 52961, 52966, 52968, 52970, 52997, 53008, 53012, 53024, 53031, 53038, 53039, 53055, 53085, 53095, 53096, 53110, 53123, 53129, 53146, 53156, 53163, 53206, 53226, 53231, 53254, 53262, 53268, 53288, 53290, 53300, 53304, 53313, 53329, 53336, 53344, 53355, 53360, 53400, 53408, 53418, 53434, 53457, 53477, 53486, 53491, 53494, 53501, 53510, 53513, 53518, 53524, 53536, 53542, 53551, 53571, 53576, 53577, 53633, 53635, 53644, 53658, 53662, 53678, 53688, 53709, 53717, 53736, 53739, 53742, 53778, 53781, 53817, 53834, 53877, 53893, 53902, 53914, 53923, 53925, 53932, 53948, 53974, 53987, 54008, 54012, 54014, 54047, 54064, 54069, 54074, 54079, 54082, 54090, 54095, 54102, 54108, 54113, 54134, 54147, 54150, 54156, 54166, 54171, 54177, 54195, 54196, 54205, 54247, 54249, 54250, 54270, 54277, 54279, 54291, 54315, 54320, 54328, 54351, 54360, 54361, 54384, 54397, 54420, 54421, 54432, 54447, 54451, 54457, 54481, 54485, 54499, 54500, 54505, 54509, 54513, 54521, 54528, 54531, 54535, 54540, 54550, 54553, 54558, 54559, 54566, 54567, 54569, 54578, 54588, 54604, 54609, 54648, 54665, 54680, 54692, 54697, 54703, 54708, 54711, 54712, 54719, 54728, 54743, 54784, 54786, 54788, 54793, 54797, 54801, 54842, 54846, 54847, 54855, 54859, 54865, 54879, 54883, 54892, 54893, 54896, 54930, 54933, 54938, 54941, 54948, 54960, 54968, 54972, 54991, 54996, 55000, 55001, 55029, 55039, 55043, 55046, 55051, 55068, 55071, 55080, 55087, 55089, 55103, 55105, 55107, 55112, 55114, 55131, 55155, 55164, 55165, 55169, 55172, 55187, 55199, 55223, 55243, 55246, 55258, 55269, 55270, 55277, 55295, 55298, 55306, 55307, 55317, 55325, 55331, 55337, 55343, 55379, 55399, 55403, 55405, 55406, 55408, 55449, 55457, 55471, 55488, 55489, 55491, 55503, 55507, 55512, 55543, 55555, 55578, 55596, 55600, 55604, 55605, 55609, 55625, 55633, 55656, 55667, 55681, 55694, 55698, 55704, 55729, 55735, 55755, 55769, 55774, 55775, 55797, 55798, 55799, 55811, 55813, 55825, 55841, 55844, 55849, 55855, 55859, 55868, 55917, 55921, 55922, 55929, 55933, 55936, 55942, 55946, 55958, 55962, 55966, 55973, 55981, 55982, 56008, 56033, 56038, 56042, 56051, 56052, 56053, 56080, 56098, 56099, 56101, 56108, 56114, 56142, 56146, 56190, 56192, 56229, 56233, 56237, 56240, 56242, 56253, 56269, 56281, 56283, 56287, 56289, 56290, 56291, 56297, 56303, 56305, 56310, 56316, 56318, 56330, 56335, 56347, 56390, 56397, 56400, 56403, 56414, 56416, 56421, 56424, 56439, 56445, 56447, 56452, 56465, 56474, 56484, 56485, 56487, 56500, 56504, 56530, 56532, 56537, 56538, 56564, 56587, 56590, 56602, 56604, 56605, 56618, 56631, 56658, 56664, 56673, 56684, 56703, 56731, 56735, 56742, 56743, 56756, 56763, 56766, 56767, 56772, 56778, 56778, 56787, 56790, 56812, 56856, 56862, 56876, 56934, 56935, 56953, 56957, 56959, 56970, 56985, 56987, 56992, 56996, 57003, 57005, 57074, 57086, 57090, 57090, 57091, 57113, 57119, 57134, 57138, 57144, 57145, 57153, 57155, 57159, 57171, 57190, 57206, 57215, 57219, 57221, 57247, 57252, 57261, 57286, 57288, 57289, 57311, 57314, 57324, 57331, 57332, 57347, 57348, 57353, 57354, 57361, 57371, 57388, 57395, 57402, 57421, 57427, 57442, 57476, 57483, 57484, 57486, 57489, 57514, 57517, 57519, 57529, 57534, 57539, 57542, 57548, 57552, 57560, 57562, 57579, 57596, 57604, 57606, 57607, 57609, 57611, 59004, 59006, 59017, 59019, 59027, 59035, 59039, 59049, 59053, 59059, 59063, 59109, 59112, 59116, 59120, 59128, 59136, 59140, 59189, 59190, 59192, 59198, 59199, 59203, 59205, 59224, 59228, 59239, 59243, 59254, 59259, 59260, 59262, 59263, 59302, 59318, 59322, 59327, 59331, 60012, 60016, 60029, 60031, 60034, 60055, 60058, 60063, 60064, 60065, 60069, 60076, 60077, 60079, 60083, 60084, 60092, 60095, 60102, 60107, 60111, 60118, 60233, 60234, 60248, 60250, 60253, 60306, 60327, 60331, 60332, 60345, 60351, 60357, 60362, 60364, 60429, 60434, 60439, 60441, 60444, 60452, 60528, 60554, 60556, 60559, 60562, 60566, 60574, 60581, 60585, 60592, 60599, 60606, 60608, 60630, 60636, 60643, 60647, 60687, 60705, 60746, 60778, 60780, 60793, 60798, 60808, 60828, 60830, 60833, 60837, 60840, 60841, 60850, 60851, 60854, 60855, 60858, 60859, 60862, 60867, 60868, 60871, 60874, 60890, 60894, 60905, 60909, 60911, 60912, 61027, 61089, 61099, 61100, 61102, 61113, 61130, 61131, 61134, 61136, 61141, 61142, 61158, 61174, 61183, 61185, 61190, 61192, 61193, 61194, 61196, 61207, 61289, 61295, 61298, 61300, 61302, 61308, 61315, 61317, 61318, 61340, 61342, 61351, 61355, 61377, 61395, 61400, 61412, 61416, 61417, 61420, 61422, 61435, 61437, 61475, 61480, 61494, 61506, 61548, 61556, 61560, 61565, 61570, 61581, 61582, 61589, 61657, 61664, 61704, 61721, 61742, 61748, 61749, 61750, 61754, 61776, 61780, 61783, 61786, 61787, 61789, 61790, 61794, 61796, 61798, 61811, 61812, 61816, 61830, 61831, 61842, 61859, 61861, 61869, 61935, 62004, 62034, 62041, 62042, 62045, 62046, 62048, 62056, 62116, 62124, 62192, 62194, 62199, 62202, 62211, 62215, 62219, 62221, 62230, 62241, 62260, 62261, 62262, 62266, 62276, 62364, 62373, 62379, 62380, 62381, 62407, 62421, 62430, 62434, 62521, 62530, 62531, 62536, 62538, 62541, 62548, 62552, 62560, 62563, 62580, 62587, 62686, 62692, 62697, 62698, 62705, 62717, 62721, 62723, 62729, 62731, 62735, 62737, 62777, 62780, 62785, 62790, 62795, 62843, 62848, 62854, 62856, 62859, 62866, 62868, 62872, 62873, 62878, 62935, 63000, 63003, 63012, 63022, 63027, 63095, 63096, 63098, 63106, 63130, 63141, 63157, 63189, 63193, 63200, 63202, 63210, 63216, 63219, 63278, 63279, 63335, 63346, 63350, 63354, 63356, 63357, 63363, 63370, 63371, 63372, 63378, 63386, 63440, 63447, 63460, 63464, 63466, 63467, 63529, 63534, 63540, 63571, 63574, 63576, 63579, 63582, 63586, 63701, 63704, 63705, 63719, 63722, 63731, 63733, 63735, 63749, 63826, 63830, 63839, 63840, 63843, 63847, 63848, 63850, 63857, 63868, 63879, 63883, 63891, 63893, 63894, 63897, 63901, 63905, 63907, 63908, 63915, 63918, 63930, 63939, 63947, 63972, 63976, 63980, 63985, 63986, 64000, 64003, 64010, 64011, 64012, 64017, 64033, 64077, 64084, 64144, 64160, 64163, 64168, 64199, 64206, 64212, 64235, 64243, 64245, 64246, 64248, 64254, 64258, 64260, 64262, 64286, 64292, 64299, 64300, 64304, 64334, 64339, 64351, 64354, 64356, 64365, 64368, 64372, 64374, 64401, 64403, 64408, 64430, 64440, 64478, 64525, 64544, 64591, 64620, 64634, 64636, 64639, 64651, 64676, 64688, 64700, 64701, 64711, 64725, 64737, 64755, 64759, 64792, 64821, 64829, 64830, 64850, 64853, 64857, 64909, 64911, 64920, 64921, 64927, 64933, 64959, 64960, 64993, 64999, 65001, 65038, 65040, 65050, 65052, 65061, 65080, 65102, 65112, 65137, 65144, 65145, 65178, 65182, 65208, 65209, 65286, 65304, 65312, 65344, 65354, 65362, 65486, 65493, 65525, 65527, 65534, 65594, 65598, 65601, 65602, 65604, 65610, 65614, 65617, 65618, 65624, 65626, 65631, 65632, 65633, 65634, 65635, 65636, 65637, 65638, 65640, 65643, 65646, 65650, 65663, 65666, 65667, 65667, 65668, 65672, 65674, 65675, 65682, 65683, 65683, 65686, 65691, 65693, 65694, 65697, 65698, 65702, 65706, 65707, 65711, 65712, 65718, 65720, 65724, 65725, 65728, 65739, 65750, 65755, 65756, 65793, 65798, 65806, 65812, 65815, 65832, 65842, 65843, 65852, 65860, 65869, 65870, 65872, 65909, 65910, 65917, 65919, 65923, 65927, 65953, 65958, 65959, 65963, 65965, 65966, 65969, 65973, 65975, 65981, 65987, 66011, 66019, 66029, 66035, 66041, 66041, 66076, 66105, 66112, 66133, 66134, 66137, 66138, 66140, 66142, 66150, 66190, 66198, 66210, 66211, 66214, 66229, 66236, 66237, 66239, 66242, 66245, 66248, 66249, 66304, 66313, 66320, 66324, 66336, 66339, 66343, 66345, 66352, 66355, 66378, 66381, 66386, 66394, 66401, 66416, 66420, 66426, 66429, 66431, 66449, 66453, 66460, 66467, 66473, 66476, 66477, 66491, 66495, 66496, 66498, 66499, 66508, 66527, 66528, 66531, 66532, 66545, 66548, 66551, 66575, 66577, 66587, 66594, 66595, 66606, 66610, 66615, 66626, 66630, 66639, 66641, 66645, 66646, 66649, 66653, 66669, 66671, 66674, 66678, 66680, 66681, 66692, 66693, 66699, 66703, 66705, 66710, 66715, 66718, 66720, 66728, 66731, 66734, 66736, 66737, 66742, 66744, 66750, 66752, 66753, 66758, 66759, 66768, 66773, 66779, 66780, 66782, 66787, 66795, 66801, 66804, 66807, 66813, 66818, 66819, 66823, 66825, 66829, 66837, 66851, 66867, 66877, 66886, 66899, 66909, 66913, 66917, 66925, 66927, 66932, 66935, 66947, 66955, 66961, 66977, 66978, 67025, 67029, 67029, 67030, 67042, 67050, 67052, 67053, 67059, 67072, 67075, 67076, 67083, 67086, 67090, 67095, 67099, 67107, 67112, 67117, 67117, 67122, 67125, 67127, 67129, 67130, 67148, 67149, 67150, 67151, 67155, 67157, 67163, 67164, 67167, 67172, 67173, 67180, 67181, 67182, 67185, 67192, 67200, 67200, 67201, 67206, 67207, 67210, 67213, 67215, 67216, 67219, 67221, 67226, 67228, 67235, 67239, 67250, 67256, 67264, 67265, 67276, 67277, 67286, 67292, 67297, 67298, 67299, 67305, 67308, 67317, 67318, 67325, 67328, 67333, 67336, 67337, 67339, 67341, 67345, 67348, 67349, 67351, 67352, 67354, 67355, 67357, 67359, 67360, 67363, 67364, 67365, 67366, 67368, 67371, 67373, 67375, 67377, 67381, 67382, 67390, 67394, 67399, 67401, 67402, 67404, 67408, 67419, 67423, 67424, 67425, 67429, 67431, 67436, 67438, 67439, 67443, 67445, 67446, 67448, 67452, 67463, 67471, 67481, 67485, 67491, 67496, 67500, 67511, 67520, 67557, 67562, 67566, 67568, 67572, 67600, 67605, 67609, 67611, 67622, 67626, 67627, 67638, 67639, 67643, 67644, 67645, 67663, 67664, 67672, 67673, 67676, 67682, 67684, 67688, 67691, 67693, 67694, 67697, 67699, 67701, 67702, 67704, 67705, 67708, 67710, 67712, 67715, 67725, 67727, 67728, 67732, 67733, 67734, 67737, 67740, 67758, 67764, 67769, 67778, 67785, 67790, 67793, 67796, 67801, 67806, 67809, 67814, 67817, 67825, 67828, 67830, 67835, 67837, 67838, 67847, 67849, 67852, 67855, 67856, 67867, 67875, 67876, 67877, 67879, 67887, 67890, 67893, 67895, 67897, 67900, 67901, 67903, 67907, 67908, 67909, 67910, 67912, 67913, 67914, 67915, 67916, 67919, 67921, 67922, 67923, 67924, 67925, 67932, 67933, 67936, 67937, 67939, 67942, 67946, 67947, 67952, 67953, 67954, 67957, 67965, 67967, 67968, 67970, 67973, 67976, 67981, 67982, 67983, 67985, 67986, 67987, 67994, 67995, 67997, 67998, 67999, 68001, 68002, 68003, 68004, 68005, 68007, 68009, 68012, 68013, 68015, 68017, 68020, 68021, 68022, 68026, 68027, 68028, 68029, 68030, 68031, 68032, 68033, 68035, 68036, 68037, 68040, 68042, 68044, 68045, 68047, 68050, 68051, 68054, 68055, 68057, 68059, 68060, 68061, 68062, 68063, 68064, 68064, 68067, 68068, 68069, 68070, 68071, 68072, 68073, 68074, 68075, 68077, 68078, 68080, 68083, 68086, 68088, 68090, 68091, 68094, 68096, 68099, 68102, 68103, 68104, 68116, 68120, 68125, 68126, 68128, 68133, 68136, 68139, 68141, 68142, 68144, 68145, 68146, 68150, 68151, 68152, 68153, 68155, 68156, 68159, 68160, 68161, 68167, 68170, 68171, 68174, 68175, 68178, 68180, 68181, 68182, 68183, 68184, 68185, 68186, 68194, 68200, 68201, 68202, 68203, 68206, 68207, 68209, 68210, 68211, 68213, 68214, 68215, 68216, 68217, 68218, 68219, 68220, 68221, 68225, 68233, 68239, 68240, 68240, 68241, 68242, 68244, 68245, 68246, 68247, 68248, 68251, 68252, 68254, 68257, 68258, 68259, 68260, 68261, 68262, 68263, 68264, 68265, 68268, 68271, 68274, 68275, 68277, 68278, 68281, 68283, 68284, 68286, 68289, 68291, 68292, 68295, 68297, 68301, 68302, 68303, 68305, 68309, 68310, 68312, 68317, 68318, 68321, 68323, 68324, 68326, 68328, 68329, 68329, 68330, 68332, 68333, 68334, 68335, 68338, 68339, 68340, 68341, 68342, 68343, 68344, 68345, 68349, 68351, 68353, 68355, 68357, 68358, 68359, 68360, 68361, 68363, 68364, 68365, 68366, 68367, 68368, 68369, 68370, 68373, 68375, 68376, 68377, 68379, 68381, 68386, 68386, 68387, 68389, 68390, 68391, 68395, 68397, 68400, 68402, 68403, 68404, 68405, 68407, 68408, 68409, 68412, 68413, 68415, 68418, 68420, 68421, 68424, 68426, 68427, 68428, 68439, 68440, 68443, 68444, 68448, 68450, 68450, 68453, 68456, 68460, 68468, 68469, 68470, 68471, 68476, 68477, 68482, 68483, 68484, 68486, 68487, 68488, 68490, 68491, 68493, 68495, 68496, 68497, 68499, 68500, 68501, 68502, 68503, 68504, 68506, 68507, 68508, 68510, 68511, 68512, 68513, 68516, 68517, 68518, 68519, 68520, 68521, 68522, 68523, 68524, 68526, 68531, 68533, 68534, 68540, 68542, 68544, 68546, 68547, 68548, 68553, 68554, 68555, 68557, 68559, 68561, 68562, 68563, 68564, 68566, 68569, 68570, 68571, 68572, 68573, 68574, 68577, 68582, 68583, 68585, 68586, 68587, 68592, 68593, 68595, 68596, 68597, 68598, 68599, 68600, 68601, 68603, 68604, 68605, 68606, 68607, 68608, 68609, 68610, 68611, 68612, 68613, 68614, 68615, 68616, 68618, 68619, 68620, 68621, 68622, 68623, 68626, 68628, 68629, 68631, 68632, 68633, 68634, 68636, 68637, 68639, 68640, 68641, 68642, 68643, 68644, 68651, 68652, 68654, 68655, 68656, 68657, 68660, 68661, 68662, 68663, 68665, 68666, 68668, 68669, 68670, 68675, 68676, 68677, 68678, 68679, 68683, 68684, 68685, 68686, 68689, 68690, 68693, 68695, 68696, 68697, 68698, 68699, 68700, 68701, 68702, 68704, 68705, 68706, 68706, 68708, 68711, 68712, 68713, 68714, 68715, 68716, 68719, 68720, 68722, 68723, 68724, 68725, 68727, 68728, 68729, 68731, 68732, 68733, 68734, 68736, 68739, 68741, 68743, 68744, 68748, 68749, 68750, 68751, 68754, 68755, 68756, 68759, 68760, 68761, 68762, 68763, 68764, 68765, 68766, 68767, 68768, 68770, 68771, 68772, 68773, 68775, 68776, 68778, 68779, 68780, 68781, 68782, 68783, 68785, 68787, 68788, 68789, 68790, 68791, 68792, 68793, 68794, 68795, 68796, 68797, 68798, 68799, 68800, 68801, 68802, 68803, 68804, 68805, 68807, 68808, 68811, 68812, 68814, 68815, 68816, 68817, 68818, 68820, 68821, 68822, 68824, 68826, 68827, 68828, 68829, 68831, 68835, 68837, 68838, 68839, 68840, 68842, 68844, 68846, 68847, 68849, 68850, 68852, 68854, 68857, 68858, 68859, 68861, 68862, 68864, 68865, 68873, 68874, 68875, 68876, 68877, 68878, 68879, 68880, 68881, 68882, 68883, 68884, 68886, 68888, 68889, 68890, 68891, 68892, 68893, 68894, 68895, 68896, 68897, 68899, 68900, 68901, 68902, 68904, 68906, 68907, 68908, 68909, 68910, 68912, 68913, 68914, 68917, 68919, 68921, 68923, 68924, 68925, 68926, 68927, 68928, 68929, 68932, 68933, 68934, 68935, 68937, 68938, 68939, 68940, 68941, 68942, 68943, 68948, 68949, 68950, 68951, 68952,
68956, 68957, 68961, 68966, 68967, 68968, 68969, 68970, 68971, 68973, 68977, 68978, 68983, 68984, 68987, 68988, 68989, 68990, 68992, 68996, 68997, 68998, 69000, 69004, 69005, 69006, 69007, 69008, 69009, 69011, 69012, 69013, 69016, 69018, 69019, 69020, 69021, 69023, 69024, 69025, 69026, 69027, 69028, 69029, 69031, 69034, 69036, 69037, 69039, 69040, 69042, 69044, 69045, 69046, 69047, 69048, 69050, 69052, 69053, 69055, 69056, 69057, 69058, 69059, 69060, 69061, 69062, 69063, 69064, 69065, 69066, 69067, 69068, 69069, 69070, 69072, 69073, 69074, 69076, 69077, 69078, 69079, 69080, 69081, 69083, 69085, 69086, 69088, 69089, 69090, 69091, 69092, 69094, 69095, 69096, 69097, 69098, 69101, 69103, 69104, 69105, 69110, 69112, 69113, 69114, 69116, 69117, 69118, 69120, 69121, 69122, 69123, 69124, 69127, 69128, 69129, 69130, 69131, 69132, 69133, 69134, 69135, 69136, 69138, 69143, 69144, 69145, 69146, 69150, 69151, 69152, 69153, 69154, 69155, 69156, 69157, 69158, 69160, 69161, 69162, 69164, 69166, 69167, 69168, 69169, 69170, 69171, 69173, 69174, 69176, 69177, 69179, 69182, 69184, 69185, 69187, 69189, 69190, 69191, 69192, 69193, 69194, 69195, 69196, 69198, 69200, 69201, 69202, 69204, 69205, 69206, 69207, 69208, 69209, 69210, 69212, 69215, 69216, 69217, 69219, 69221, 69222, 69225, 69226, 69228, 69229, 69230, 69231, 69233, 69236, 69242, 69243, 69244, 69245, 69246, 69249, 69253, 69257, 69258, 69264, 69265, 69268, 69269, 69271, 69272, 69274, 69275, 69276, 69277, 69278, 69279, 69280, 69281, 69284, 69285, 69286, 69287, 69288, 69289, 69291, 69292, 69293, 69294, 69295, 69296, 69298, 69299, 69300, 69301, 69302, 69303, 69304, 69305, 69306, 69307, 69308, 69309, 69311, 69312, 69313, 69314, 69315, 69316, 69318, 69319, 69321, 69322, 69323, 69325, 69326, 69327, 69328, 69329, 69330, 69333, 69335, 69337, 69338, 69340, 69341, 69343, 69344, 69346, 69347, 69348, 69349, 69351, 69352, 69353, 69354, 69355, 69356, 69357, 69358, 69359, 69361, 69362, 69364, 69365, 69366, 69367, 69368, 69369, 69371, 69372, 69374, 69375, 69376, 69377, 69378, 69380, 69381, 69382, 69383, 69384, 69385, 69387, 69389, 69390, 69391, 69392, 69393, 69394, 69395, 69397, 69398, 69399, 69400, 69401, 69402, 69403, 69404, 69405, 69406, 69407, 69408, 69409, 69410, 69411, 69416, 69418, 69419, 69423, 69425, 69426, 69428, 69430, 69431, 69432, 69434, 69438, 69439, 69440, 69441, 69445, 69446, 69447, 69449, 69450, 69451, 69452, 69453, 69456, 69459, 69461, 69462, 69464, 69465, 69466, 69467, 69468, 69469, 69470, 69471, 69475, 69482, 69483, 69485, 69486, 69489, 69490, 69491, 69494, 69495, 69502, 69509, 69519, 69520, 69521, 69522, 65677-1, and 65719-1. |
| FEI Number |
2936884
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 1230 Shorebird Way P.O. Box 7393 Mountain View CA 94043
|
| For Additional Information Contact | Kristin Dorrough 650-694-5993 |
Manufacturer Reason for Recall | Overheating. A software bug may cause the transducer to overheat when used with certain transducer models. The transducer can become noticeably hot within a few minutes, which may cause patient discomfort or burns. |
FDA Determined Cause 2 | Software design |
| Action | The firm, SIEMENS, sent a "Customer Safety Advisory Notification" to its customers on 3/21/2009. The notification described the product, problem and actions to be taken. The customers were instructed to disconnect any transducers that may be in an active transducer port and chose another imaging transducer prior to use of the Aux CW transducer. If they do not have another transducer to select other than those listed in the notification, they should close the study in progress and start a new study before using the Aux CW transducer-this procedure should be followed for each new patient and /or exam type. In addition, the customers were instructed to pass this notice on to all those within their organization who need to be aware of this issue until the corrective action is completed. SIEMENS will release a software update that will correct the issue. The customers local service engineer will install the updates on their system as soon as it becomes available.
Should you have any questions regarding this notification, you may contact your local service support person and/or call (650) 969-9112. |
| Quantity in Commerce | 2591 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|