Date Initiated by Firm |
August 28, 2009 |
Date Posted |
November 09, 2009 |
Recall Status1 |
Terminated 3 on May 12, 2010 |
Recall Number |
Z-0141-2010 |
Recall Event ID |
53137 |
Product Classification |
Vertebral Biopsy set component - Product Code FCG
|
Product |
Cardinal Health, Craig Calibrated Cannula, overall length 5-3/8" (13.7 cm), catalog #OS5000-005, sold separately, or as a component of the Craig Vertebral Biopsy Set, catalog #OS5000 |
Code Information |
catalog #OS5000-005, lot number K08 |
Recalling Firm/ Manufacturer |
Cardinal Health 1430 Waukegan Rd Mc Gaw Park IL 60085-6726
|
For Additional Information Contact |
Michele B. Donatich R.N. 847-473-1500
|
Manufacturer Reason for Recall |
The Craig Calibrated Cannula does not interface with the toothed cutter properly, which may prevent the toothed cutter from properly advancing through the cannula to the biopsy site.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Cardinal Health sent Urgent Product Recall letters dated 8/28/09 to all customers who received the product during the recalled period on the same day via UPS Next Day Air. The accounts were informed of the problem with the cannula, and were instructed to functionally test the cannula by passing the toothed cutter within it to ensure the cutter advances freely into the cannula. If any restriction is noticed, the product should be segregated and returned to Cardinal Health. The accounts were requested to complete and return by fax the enclosed acknowledgement form, indicating the number of affected cannulas being returned. Any questions were directed to V. Mueller Customer Service at 800-323-9088. |
Quantity in Commerce |
19 units |
Distribution |
Nationwide distribution. |