 Class 2 Device Recall Ultraflex Esophageal Stent Systemsee related information
Date Initiated by FirmMay 17, 2010
Date PostedJune 25, 2010
Recall Status1 Terminated 3 on October 20, 2011
Recall NumberZ-1862-2010
Recall Event ID 55784
510(K)NumberK032930 
Product Classification Prosthesis, esophageal - Product Code ESW
ProductUltraflex Esophageal Stent System. Covered - Distal Release - 18/23/12/9. Catalog Number M00513740. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Intended for maintaining esophageal luminal patency in esophageal structures caused by intrinsic and/or extrinsic malignant tumors only and occlusion of concurrent esophageal fistulas.
Code Information Lot Numbers: 11717547 11724997 11730506 11733143 11735857 11744473 11747850 11748115 11748703 11749504 11755828 11757441 11766371 11771487 11773688 11782909 11784694 11788206 11791252 11793665 11800088 11810542 11813053 11814272 11823572 11827356 11830766 11842301 11844882 11846141 11846188 11846324 11846957 11846958 11860236 11863086 11872420 11874726 11875667 11883102 11889562 11895930 11897032 11901171 11931570 11941341 11942972 11945162 11946520 11951127 11969421 11982094 11985488 11991796 11992174 11996653 12002493 12011513 12025015 12025155 12026256 12027554 12031723 12034907 12072087 12073098 12087523 12104000 12114312 12120992 12127636 12144382 12149504 12155217 12156134 12159842 12160987 12162782 12163562 12164645 12176654 12177726 12179420 12180186 12186092 12194918 12202044 12203137 12219719 12221575 12223039 12224998 12225622 12230470 12242300 12246956 12254187 12256519 12260117 12266399 12267859 12269995 12270505 12278319 12310298 12315601 12317781 12318554 12319015 12320698 12322872 12333093 12334478 12339506 12348305 12351417 12354385 12356602 12370142 12371627 12375358 12378613 12384525 12386325 12387668 12396150 12440416 12444651 12448647 12450801 12457782 12492988 12498133 12501390 12513016 12520688 12534067 12551273 12566989 12583267 12600950 12621171 12645078 12645845 12657395 12669445 12673338 12703450 12740197 12741282 12741506 12743202 12757005 12763032 12827702 12829801 12829966 12830765 12830767 12832989 12844419 12849607 12850057 12859668 12864107 12865389 12869468 12870028 12888380 12889011 12898412 12908635 12912946 12928517  
FEI Number 3005099803
Recalling Firm/
Manufacturer
Boston Scientific Corporation
100 Boston Scientific Way
Marlborough MA 01752-1234
For Additional Information Contactsame
508-683-8000
Manufacturer Reason
for Recall
Suture related device failures during deployment and repositioning of the stent
FDA Determined
Cause 2
Device Design
ActionBoston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
DistributionWorldwide Distribution - Only US distributed product is affected by recall
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.