 Class 1 Device Recall All Med Medical 20ga x 1.5" RA Hub Needlesee related information
Date Initiated by FirmAugust 30, 2010
Date PostedSeptember 29, 2010
Recall Status1 Terminated 3 on January 31, 2013
Recall NumberZ-2542-2010
Recall Event ID 56548
510(K)NumberK891919 
Product Classification Hypodermic Single Lumen Needle - Product Code FMI
ProductAll Med Medical 20ga x 1.5" RA Hub Needle Model #M2015H Reorder #55-2040 Intended to be used to inject fluids into, withdraw fluids from inside the vascular space through a subcutaneously implanted port.
Code Information Lot code: 5H058A 5J007A 6H041A 7J029A 8D029A 8K055A 8K058A 8L034A 
FEI Number 1221765
Recalling Firm/
Manufacturer
Multi-Med, Inc.
26 Victoria Court
Keene NH 03431
For Additional Information Contactsame
603-357-8733
Manufacturer Reason
for Recall
Huber needles may core and result in port leakage or emboli being flushed our of the port.
FDA Determined
Cause 2
Device Design
ActionMulti-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
Quantity in Commerce4,921 units
DistributionWorldwide Distribution -- USA and Mexico.
Total Product Life CycleTPLC Device Report