• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recallsee related information
Date Initiated by FirmJuly 15, 2010
Date PostedOctober 01, 2010
Recall Status1 Terminated 3 on October 19, 2011
Recall Numberz-2640-2010
Recall Event ID 56652
Product Classification Instrument, Ultrasonic Surgical - Product Code LFL
ProductSonopet Ultrasonic Surgical System, Model UST-2001, Manufacturer Miwatec Co. LTD, Tokyo, Japan.
Code Information 06I0599, 08I0973, 07J0790, 04K0294, 07C0671, 09B1031, 04G0254, 04H0259, 06I0601, 08D0887, 06B0472, 05G0373, 07F0718, 07G0738, 05J0421, 08D0888, 04L0315, 05G0375, 04L0311, 06C0484, 04A0209, 08I0974, 04A0208, 03L0197, 07I0787, 09B1028, 07L0822, 06B0470, 04G0252, 05I0401, 09B1026, 06B0467, 06G556, 03I0166, 05E0378, 05E0351, 05J0422, 04D0237, 05L0448, 08C0872, 07E0703, 07L0825, 08L0998, 07K0811, 08G0937, 05J0423, 06B0468, 08D0885, 08D0891, 08G0935, 08G0936, 07E0700, 07E0701, 06H0576, 04D0230, 04D0231, 06I0603, 07B0655, 07B0653, 08G0932, 04F0242, 04I0262, 08D0889, 03I0168, 08B0855, 07K0807, 08A0847, 08B0858, 06F0543, 07F0721, 07F0719, 07F0720, 07F0732, 04K0293, 04I0279, 06G0569, 06K0618, 06I0586, 04K0292, 05A0341, 07K0808, 07H0766, 07H0767, 07E0714, 04H0261, 05L0444, 05L0446, 07K0810, 07K0812, 07K0813, 07L0830, 08D0893, 07E0705, 06H0574, 06H0577, 06H0578, 06I0588, 08I0985, 08I0972, 09B1035, 08F0914, 07F0731, 04G0253, 07F0722, 07K0805, 04K0295, 07C0672, 07K0806, 08B0857, 07B0652, 07G0739, 05E0379, 07I0785, 07I0786, 06G0567, 04I0282, 08E0907, 07I0788, 07I0789, 04I0271, 06C0488, 06C0487, 04B0194, 08D0886, 06B0473, 06C0486, 09B1027, 04H0260, 06A0456, 06E0532, 07C0670, 06I0602, 06C0489, 06E0530, 03L0196, 04J0283, 06G0568, 03I0170, 06B0475, 03I0171, 08G0934, 07F0717, 08F0925, 08A0843, 03H0165, 06G557, 06G555, 08L0999, 06A0459, 06A0460, 06A0461, 09A1008, 09A1009, 09A1010, 09A1016, 07L0827, 07L0828, 07L0829, 08C0871, 08C0873, 08C0874, 08C0875, 09B1034, 08D0890, 09C1051, 06D0511, 09E1069, 09E1070, 09E1071, 09E1085, 09E1086, 08E0895, 08E0896, 08F0926, 08F0927, 08F0928, 08F0929, 08F0930, 08G0938, 06H0573, 07K0623, 07G0735, 07G0736, 06H0575, 06I0589, 06K0621, 03G0162, 04D0236, 05L0447, 06D0510, 07H0764, 07H0765, 06G558, 06D0512, 07L0824, 09A1017, 08D0877, 06H0572, 06H0571, 06C0490, 06H0579, 08B0859, 08B0854, 08I0970, 06B0469, 06F0545, 09B1032, 09B1033, 08D0892, 09C1049, 09C1050, 07E0704, 08E0905, 08E0906, 06C0491, 07E0699, 06I0590, 05L0445, 07H0763, 06C0492, 07K0622, 07E0702, 06G0554, 07G0740, 08F0915, 07J0791, 07C0669, 06E0534, 07G0741, 06I0587, 03L0195, 06I0600, 05F0381, 07B0656, 08B0860, 04L0313, 07L0823, 08A0842, 08G0933, 05G0374, 07G0742, 06K0620, 07G0747, 07D0688, 04I0272, 05G0390, 05I0400, 05H0397, 06C0485, 08B0856, 07F0716, 07D0684, 06F0544, 08F0916, 07K0809, 03G0163, 06B0474, 05G0389, 07A0643, 08F0912, 05F0380, 02#0065, 04G0251, 03I0172, 04I0273, 04F0244, 07L0832, 08I0984, 09B1025, 06B0471, 08A0848, 06E0531, 06E0533, 08E0908, 08G0939, 04F0245, 06J0604, 07G0743, 07G0745, 05G0396, 05G0395, 05H0398, 06H0570, 07D0685, 02%0075, 02%0076, 04L0312, 07D0687, 05G0392, 03D0134, 02I0109, 02I0110, 03I0169, 07G0748, 03E0161, 07B0657, 06A0457, 04I0280, 04I0281, 04K0291, 07G0746, 08G0931, 07B0654, 07F0730, 06K0619, 05A0342, 05G0391, 03I0167, 04C0223, 06F0542, 03H0164, 08I0971, 08F0913, 06J0617, 03G0148, 03G0149, 03G0150, 04C0222, 04K0310, 04A0205, 04K0309, 03I0173, 02L0119, 03A0124, 02#0057, 02G0092, 01(0010, 01J0024, 02!0041, 01+0028, 01,0034, 01,0032, 01,0033, 09B1030, 09C1047, 09J1103, 09I1104, 09I1105, 09I1106, 09L1107, 09L1108, 09L1109, 09L1111, 09L1112, 09L1113, 09L1129, 09L1130, 09L1131, 09L1132, 09L1133, 09L1134, 09L1135, 09L1136, 09L1137, 09L1138, 09L1139, 09L1140, 09L1141, 09L1142, 09L1143, 09L1144, 09L1145, 09L1146, 09L1147, 09L1152, 09L1153, 09L1154, 09L1155, 09I1094, 09H1076, 08H0955, 09E1079, 02L0120, 09I1096, 02L0114, 09I1100, 09I1099, 09I1102, 02A0038, 03I0153, 03L0190, 09A1024, 09D1053, 03A0133, 06F0546, 06F0547, 04D0232, 05D0336, 05B0344, 09I1101, 02E0073, 09L1110, 09J1126, 04C0224, 04D0235, 04F0241, 04F0243, 04L0314, 07L0833, 09L1151, 09L1148, 09L1149, 09L1150, 09J1116, 09J1117, 09J1118, 09J1119, 09J1120, 09J1121, 09J1122, 09J1123, 09J1124, 09J1125, 09J1127, 09C1045, 07G0744, 07G0749, 06A0458, 05I0399, 07D0683, 07D0686, 07G0756, 09E1019, 09C1041, 09C1042, 02!0039, 02F0080, 02%0074, 01(0002, 01+0026, 0037, 02I0105, 03F0143, 04A0210, 0323, 0368, 06A0465, 08I0969, 07L0831, 09A1018, 09A1023, 06B0477, 06B0478, 09I1114, 09I1115, 07G0733, 07G0734, 08G0940, 08I0975, 07I0782, 07I0781, 09A1007, 09B1029, 08A0845, 06B0476, 09B1036, 09B1037, 09B1038, 09C1048, 08C0876, 08C0870, 07H0770, 06E0529, 08E0902, 08H0956, 07H0771, 07H0772, 04B0216, 09A1021, 09I1091, 07C0668, 06G0566, 07B0651, 07H0762, 05I0407, 02$0071, 08G0941, 06J0605, 01+0029, 01+0030, 07I0783, 08J0986, 08J0994, 09C1039, 09C1040, 08E0904, 08E0903, 07I0784, 04F0247, 04E0234, 04E0233, 04F0246, 04F0248, 04L0317, 04L0318, 04J0284, 04J0285, 02F0078, 02F0079, 03C0125, 03#0126, 03D0156, 03D0157, 03D0158, 08K0996, 08C0869, 08H0951, 05G0394, 05F0362, 02B0055, 05A0343, 05K0434, 06F0551, 02E0072, 03E0159, 04C0217, 05K0433, 07J0799, 07K0826, 09A1020, 08H0952, 07E0624, 06I0591, 05E0352, 05G0377, 02B0053, 08H0959, 08A0839, 08I0966, 08I0977, 09D1057, 08F0910, 07C0667, 07B0658, 05I0408, 07G0750, 09H1065, 06E0539, 06E0527, 08F0911, 05I0413, 07E0706, 02L0113, 07F0725, 04D0225, 05L0440, 09D1059, 03A0121, 07A0644, 05H0385, 06D0513, 05A0335, 07H0774, 04K0297, 07D0691, 07E0696, 09H1090, 07I0780, 07G0753, 02G0089, 08I0980, 08H0950, 07F0724, 02B0052, 04G0263, 04K0302, 06J0606, 01(0007, 08L1006, 05F0359, 07C0678, 07C0666, 09E1088, 03K0186, 02!0046, 07L0836, 05A0328, 02!0043, 02$0067, 02B0054, 03K0187, 02)0108, 08C0865, 01J0018, 08I0981, 05F0366, 04G0264, 05K0431, 05K0432, 03+0184, 07G0757, 07K0815, 09E1080, 04A0201, 05L0439, 05E0349, 07E0713, 02F0084, 06B0480, 08H0949, 06L0628, 08H0954, 06G0561, 05F0363, 04C0219, 06L0629, 06L0634, 02F0077, 07J0793, 08A0840, 08J0989, 05J0424, 04L0325, 07C0663, 04K0303, 07D0682, 07H0775, 06E0526, 05H0383, 08H0953, 03G0152, 04K0305, 10B1166, 07G0759, 07E0711, 04G0250, 06L0633, 06H0580, 09D1061, 01(0006, 05H0406, 08L1003, X7L0821, 07B0661, 07G0752, 07I0779, 05G0376, 09H1073, 09G1063, 01)0012, 02G0090, 05F0354, 02K0112, 03J0174, 07L0820, 08E0898, 06G0560, 05E0350, 06I0598, 06E0525, 07D0681, 01J0022, 04B0211, 06A0449, 08F0921, 05F0371, 08L1002, 03L0189, 04J0288, 05I0414, 05F0358, 07G0754, 02H0104, 06C0497, 08C0882, 04K0298, 07C0675, 04I0278, 08I0982, 07J0798, 08E0901, 05H0403, 05H0388, 07F0729, 04K0306, 08J0991, 07I0777, 07E0712, 08C0866, 08I0968, 03J0178, 03K0179, 06B0479, 06A0450, 05A0329, 06L0641, 08H0947, 05D0345, 02F0085, 08C0867, 05L0441, 07J0795, 06G0553, 07F0727, 08H0958, 06C0508, 04I0277, 07E0709, 09E1082, 08E0900, 06H0583, 06L0635, 02G0093, 07J0792, 06E0524, 04A0198, 05J0416, 06E0541, 06D0516, 05A0327, 05F0365, 09A1013, 06H0585, 04I0275, 04K0296, 02H0103, 03J0175, 08C0878, 09H1075, 04G0257, 07J0797, 04H0269, 07G0758, 02G0096, 05I0410, 02F0081, 02G0094, 06B0482, 05K0438, 06C0494, 07I0776, 09A1014, 07L0834, 01J0025, 05D0338, 09D1055, 08H0946, 06F0549, 06F0550, 09D1058, 05F0357, 05J0417, 05H0387, 07J0803, 07D0680, 01,0031, 06J0610, 06D0514, 06D0515, 01J0021, 05E0355, 08H0948, 06L0636, 08I0965, 02!0040, 10B1168, 02H0102, 05D0340, 07D0679, 06L0630, 01(0008, 06D0509, 02B0049, 02#0058, 02$0068, 02H0101, 07C0673, 09H1074, 02B0051, 09G1046, 02!0042, 04A0204, 04K0300, 04K0301, 05F0367, 08B0861, 04H0270, 04K0304, 04J0289, 08J0992, 09A1012, 09H1072, 06L0637, 02I0106, 06J0615, 05H0386, 06L0642, 08J0990, 07D0693, 02$0069, 06G0552, 03K0181, 01+0027, 05F0356, 05D0339, 08J0988, 07B0650, 07J0802, 03C0138, 05H0404, 04A0203, 05K0435, 03C0136, 04G0258, 07G0755, 01)0011, 06A0464, 06L0627, 07A0647, 03F0132, 08L1001, 02H0100, 05I0409, 02K0111, 02B0056, 07G0760, 08C0879, 04L0321, 03A0122, 08A0851, 02!0044, 09E1084, 07K0804, 04B0213, 06H0582, 07A0646, 04B0199, 02F0088, 05F0364, 07E0698, 09H1066, 09H1089, 06L0631, 06J0609, 06J0607, 03G0145, 03L0193, 06A0451, 04D0227, 01(0003, 06G0563, 03L0192, 07F0723, 09D1060, 01J0023, 02I0107, 03C0137, 05K0430, 03K0182, 03K0183, 01,0035, 02F0083, 01J0020, 08G0944, 06B0481, 06I0596, 08I0967, 08H0960, 08D0883, 02B0050, 07E0708, 06A0454, 01,0036, 01(0005, 06A0453, 08I0964, 09H1077, 06C0495, 06H0584, 07E0707, 04F0240, 07J0796, 06J0612, 08C0880, 08G0943, 06A0466, 06G0559, 04G0249, 04H0267, 03F0142, 06B0483, 03J0176, 06A0455, 09H1067, 07I0778, 06L0625, 03J0177, 05H0382, 08F0917, 05A0334, 09E1087, 05J0418, 07J0801, 06G0565, 02F0086, 02B0048, 05D0346, 01(0009, 07K0816, 06I0593, 07L0835, 08H0957, 07B0660, 05E0353, 05E0348, 07L0819, 06J0611, 05J0427, 04L0324, 01(0004, 08G0945, 03L0191, 02$0066, 06G0562, 06G0564, 09I1097, 08I0979, 02G0095, 05I0412, 06E0537, 09D1054, 05H0402, 03B0128, 06D0517, 01)0014, 01J0016, 05J0420, 05H0384, 04J0287, 06I0592, 08F0918, 04L0322, 04K0299, 07E0710, 06C0503, 06C0507, 06D0522, 06C0501, 06E0535, 06D0520, 06L0640, 06C0506, 06C0499, 06E0536, 06D0521, 06C0504, 06D0518, 06C0505, 05L0443, 06E0538, 06C0500, 07C0676, 06D0519, 06C0498, 06L0639, 06C0502, 08F0920, 05J0426, 05F0369, 04A0207, 07D0690, 04C0218, 02F0082, 05J0428, 05D0347, 04C0221, 01J0017, 08A0841, 03F0131, 08B0853, 02B0047, 06L0638, 08B0863, 05L0442, 04H0265, 02F0087, 06F0548, 07C0674, 09A1011, 04D0229, 04B0215, 06E0540, 02G0098, 03B0127, 04C0220, 03B0129, 02G0097, 03G0151, 10B1162, 02C0062, 02C0063, 02#0060, 02#0061, 02#0064, 05K0437, 09G1044, 09E1083, 05F0372, 05I0411, 05A0326, 08L1000, 07D0692, 06I0597, 02$0070, 07D0689, 05A0333, 04K0307, 08A0849, 07G0761, 08F0924, 08L1005, 01)0015, 08E0897, 07C0664, 08I0983, 08B0852, 08J0993, 07F0726, 09H1068, 09A1015, 05H0405, 04C0200, 07G0751, 07K0817, 07K0818, 05J0425, 05J0419, 06J0613, 05A0331, 05F0360, 06C0496, 05I0415, 08E0899, 09E1081, 03G0146, 09D1056, 09A1022, 08F0923, 03K0180, 07J0794, 06I0595, 03G0144, 06E0528, 08B0864, 03G0147, 06I0594, 03K0185, 07D0694, 04J0290, 06C0493, 07J0800, 02#0059, 07K0814, 07L0837, 08C0881, 07E0715, 07D0695, 06J0614, 08L1004, 04D0228, 05A0330, 04K0308, 06D0523, 06L0632, 09G1043, 08G0942, 02H0099, 06A0462, 04G0255, 04G0256, 05K0436, 09D1062, 08J0987, 07B0649, 06J0608, 02!0045, 07H0773, 04B0214, 04B0212, 04I0274, 08I0963, 08I0978, 03A0123, 07B0662, 04H0268, 06A0452, 07C0665, 03C0139, 05A0332, 07C0677, 04F0238, 08A0838, 06A0463, 04L0320, 07A0645, 07B0648, 04A0202, 01J0019, 05K0429, 03F0141, 07B0659, 01)0013, 09I1098, 04J0286, 01(0001, 04F0239, 03L0188, 06H0581, 02G0091, 07E0697, 10B1167, 09G1064, 04I0276, 04A0206, 04D0226, 09I1095, 05D0337, 08F0919, 03B0130, 07G0737, 04H0266, 07F0728, 08A0850, 05F0370, 08B0862, 05F0361, 08D0884, 08F0922, 06L0626, 10B1172, 10B1170, 10B1159, 10B1165, 10B1164, 10B1169, 10B1160, 10B1173, 03B0140, 10B1157, 10B1163, 10B1161, 10B1174, 10B1158, 10B1171, 08A0844, 02K0115, 02K0116, 02K0117, 02K0118, 04L0319, 04L0316, 03H0135, 05G0393, 07H0768, 07H0769, 08J0995, 08E0909, 09G1078, 09I1092, 09I1093 and 03E0160.
FEI Number 1811755
Recalling Firm/
Manufacturer
Stryker Instruments Div. of Stryker Corporation
4100 E. Milham Ave.
Kalamazoo MI 49001
For Additional Information ContactJennifer Mars
269-323-7700
Manufacturer Reason
for Recall
Stryker Instruments acquired the Sonopet Ultrasonic Surgical System from Mutoh co. Ltd on 3/29/2010. Stryker determined that the Sonopet consode did not meet the IEC-60601-1 standard. Stryker decided to recall the Sonopet consode.
FDA Determined
Cause 2
Device Design
ActionThe firm to send URGENT MEDICAL DEVICE RECALL NOTIFICATIONS to customers dated 7/15/2010. The customers were instructed to continue to use the device. Stryker technicians were dispatched to replace a board with a new board.
Quantity in Commerce1161
DistributionWordwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-