| | Class 2 Device Recall TruCore II Disposable Automatic Biopsy Instrument and TrueCore II URO Disposable Automatic Biopsy |  |
| Date Initiated by Firm | July 02, 2010 |
| Date Posted | October 08, 2010 |
| Recall Status1 |
Terminated 3 on June 28, 2011 |
| Recall Number | Z-0025-2011 |
| Recall Event ID |
56767 |
| 510(K)Number | K982960 |
| Product Classification |
Instrument, biopsy - Product Code MJG
|
| Product | Tru-Core II Disposable Automatic Biopsy Instrument and Tru-Core II URO Disposable Automatic Biopsy Instrument. The Tru-Core Disposable Automatic Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, prostate, breast, ect. The recalling lot's for the product are as follows:
Model Number: 763114100X
Lot #'s:
92661WY5, 93211XXV, 00141YVE, 00541ZKX, 010920HL, 01522192, 92791X5I, 93271Y1W, 00201YYE, 00621ZQ3, 011320LH, 015221BD, 92791X7X, 93411Y9O, 00211Z0C, 00671ZRN, 011720Q, 015521CL, 92861XCT, 93421YBJ, 00251Z1T, 00761ZZP, 012420U0, 016221H2, 93001XLG, 93431YBI, 00351Z86, 0081201Y, 0138211F, 016721KJ, 93081XPC, 93571YJO, 00461ZE2, 0088206B, 0141215R, 93141XST, 00061YOV, 00541ZK0, 009620BO, 0146219M.
Model Number: 763114160X
Lot #'s
92661WY7, 00111YPZ, 00671ZRO, 013220Z7, 016221H7, 93131XLS, 00331Z61, 0110201Y, 014125Q, 00882065, 00391Z8Y, 013120YC, 015521CJ.
Model Number: 763114200X
Lot #'s
92661WY0, 00751ZYJ, 012420U1
Model Number: 763116100X
Lot #'s
92791X80, 93491YG9, 00331Z6M, 0083203O, 012420U3, 01672KL, 92791X81, 93501YGA, 00351Z8D, 009620BL, 013120Y9, 92931XDD, 00121YRT, 00431ZE1, 009920D3, 015221AX, 93001XLN, 00131YUG, 00541ZKV, 011020IU, 015221BE, 93081XPH, 00541ZJY, 00601ZOT, 011320LG, 015521CK, 93291Y43, 00251Z1R, 0081201W, 012020QC, 016221H3.
Model Number: 763116160X
Lot #'s
92661WY3, 00882068, 00351Z8A, 00761ZZQ, 012020QB, 016721KK, 92791X89, 00912084, 00471ZG0, 010320FM, 012420TZ, 92861XD4, 01472193, 00541ZK1, 010920HK, 0141215U, 93001XLR, 00151YVC, 00601ZOR, 0110201V, 015521CM, 93291Y40, 00331Z6J, 00701ZUX, 011720OP, 016221HS
Model Number: 763116200X
Lot #'s
121KN
Model Number: 763116200X
Lot #'s
92671X09, 00471ZG1, 00882067, 010320FN, 00211Z09, 00601ZOP, 009920D0, 011020IT
Model Number: 763118100X
Lot #'s
92661WY8, 93491YG7, 00331Z6L, 00621ZQ2, 009920D2, 0138213F, 92791X7Z, 93491YG8, 00351Z88, 00701ZUW, 010320FL, 01472194, 92861XDS, 00061YOT, 00471ZG2, 00761ZZO, 0110201W, 016221H4, 93091XPL, 00131YUF, 00541ZJZ, 0083203P, 012020QE, 93211XXW, 00251Z1S, 00541ZKU, 009620BM, 012420TX.
Model Number: 763118160X
Lot #s
92661WY6, 93271Y1V, 00201YYD, 00471ZG3, 012420U2, 016221GZ, 92671WZJ, 93411Y9P, 00251Z1Q, 00671ZRK, 013120YB, 92671WZK, 93421YBK, 00331Z6N, 00701ZUY, 013220Z9, 92861XD0, 00061YOW, 00351Z8E, 00882069, 015221BB, 93001XLM, 00111YQO, 00431ZE3, 009920D4, 015521CO.
Model Number: 763118200X
Lot #s
92861XCD, 93571YJN, 00331Z6K, 00671ZRM, 011320LF, 01382134, 93001XL6, 0061YOU, 00351Z89,00701ZUZ, 011720OO, 0141215S, 93001XLJ, 00912082, 00391Z8Z, 009620BQ, 012020QA, 01472195, 93081XPJ, 00111YQM, 00471ZG5, 009920D5, 013120Y8, 016221H1, 93211XXX, 00131YUD, 00481ZG4, 010320FJ, 013220Z6, 93400Y9K, 00201YYC, 00621ZQ1, 010920HJ, 0137211I.
Model Number: 763120100X
Lot #s
93001XLV, 00351Z8B, 0088206A, 01472196, 00331Z60, 00601ZOS, 01202QF.
Model Number: 763120160X
Lot #s
92661WY1, 00882066, 00111YQP, 00351Z8C, 013120YD, 01472197, 92861XDG, 00912083, 00211Z08, 00621ZQ5, 013220Z8.
Model Number: 763120200X
Lot #s
00351Z85, 012020QG, 012420TW.
Model Number: 763418200X
Lot #s
92671WZV, 93291Y41, 00351Z87, 00601ZNH, 010320FK, 0141215T, 92671WZW, 00061YOX, 00431ZE0, 0083203N, 0110201S, 015221BC, 92861XDH, 00131YUE, 00491ZIO, 009620BN, 0011720ON, 015521CN, 93081XPR, 00141YVB, 00541ZKT, 009920D1, 0138213E, 016221H6.
Model Number: 763418250X
Lot #s
92751X60, 00671ZRL,
Model Number: 763418250X
Lot #s
92661WY4, 93271Y1N, 00141YVD, 00601ZOQ, 0110201X, 013220ZA, 92751X6P, 93291Y42, 00211Z0B, 00761ZZR, 012020QD, 0137211G, 92781X5G, 93631YJP, 00221Z13, 0081201X, 012420TY, 01382133, 92861XDN, 00912085, 00431ZDZ, 009620BP, 013120YA, 016221H0. |
| Code Information |
The recalling lot's for the product are as follows: Model Number: 763114100X Lot #'s: 92661WY5, 93211XXV, 00141YVE, 00541ZKX, 010920HL, 01522192, 92791X5I, 93271Y1W, 00201YYE, 00621ZQ3, 011320LH, 015221BD, 92791X7X, 93411Y9O, 00211Z0C, 00671ZRN, 011720Q, 015521CL, 92861XCT, 93421YBJ, 00251Z1T, 00761ZZP, 012420U0, 016221H2, 93001XLG, 93431YBI, 00351Z86, 0081201Y, 0138211F, 016721KJ, 93081XPC, 93571YJO, 00461ZE2, 0088206B, 0141215R, 93141XST, 00061YOV, 00541ZK0, 009620BO, 0146219M. Model Number: 763114160X Lot #'s 92661WY7, 00111YPZ, 00671ZRO, 013220Z7, 016221H7, 93131XLS, 00331Z61, 0110201Y, 014125Q, 00882065, 00391Z8Y, 013120YC, 015521CJ. Model Number: 763114200X Lot #'s 92661WY0, 00751ZYJ, 012420U1 Model Number: 763116100X Lot #'s 92791X80, 93491YG9, 00331Z6M, 0083203O, 012420U3, 01672KL, 92791X81, 93501YGA, 00351Z8D, 009620BL, 013120Y9, 92931XDD, 00121YRT, 00431ZE1, 009920D3, 015221AX, 93001XLN, 00131YUG, 00541ZKV, 011020IU, 015221BE, 93081XPH, 00541ZJY, 00601ZOT, 011320LG, 015521CK, 93291Y43, 00251Z1R, 0081201W, 012020QC, 016221H3. Model Number: 763116160X Lot #'s 92661WY3, 00882068, 00351Z8A, 00761ZZQ, 012020QB, 016721KK, 92791X89, 00912084, 00471ZG0, 010320FM, 012420TZ, 92861XD4, 01472193, 00541ZK1, 010920HK, 0141215U, 93001XLR, 00151YVC, 00601ZOR, 0110201V, 015521CM, 93291Y40, 00331Z6J, 00701ZUX, 011720OP, 016221HS Model Number: 763116200X Lot #'s 121KN Model Number: 763116200X Lot #'s 92671X09, 00471ZG1, 00882067, 010320FN, 00211Z09, 00601ZOP, 009920D0, 011020IT Model Number: 763118100X Lot #'s 92661WY8, 93491YG7, 00331Z6L, 00621ZQ2, 009920D2, 0138213F, 92791X7Z, 93491YG8, 00351Z88, 00701ZUW, 010320FL, 01472194, 92861XDS, 00061YOT, 00471ZG2, 00761ZZO, 0110201W, 016221H4, 93091XPL, 00131YUF, 00541ZJZ, 0083203P, 012020QE, 93211XXW, 00251Z1S, 00541ZKU, 009620BM, 012420TX. Model Number: 763118160X Lot #s 92661WY6, 93271Y1V, 00201YYD, 00471ZG3, 012420U2, 016221GZ, 92671WZJ, 93411Y9P, 00251Z1Q, 00671ZRK, 013120YB, 92671WZK, 93421YBK, 00331Z6N, 00701ZUY, 013220Z9, 92861XD0, 00061YOW, 00351Z8E, 00882069, 015221BB, 93001XLM, 00111YQO, 00431ZE3, 009920D4, 015521CO. Model Number: 763118200X Lot #s 92861XCD, 93571YJN, 00331Z6K, 00671ZRM, 011320LF, 01382134, 93001XL6, 0061YOU, 00351Z89,00701ZUZ, 011720OO, 0141215S, 93001XLJ, 00912082, 00391Z8Z, 009620BQ, 012020QA, 01472195, 93081XPJ, 00111YQM, 00471ZG5, 009920D5, 013120Y8, 016221H1, 93211XXX, 00131YUD, 00481ZG4, 010320FJ, 013220Z6, 93400Y9K, 00201YYC, 00621ZQ1, 010920HJ, 0137211I. Model Number: 763120100X Lot #s 93001XLV, 00351Z8B, 0088206A, 01472196, 00331Z60, 00601ZOS, 01202QF. Model Number: 763120160X Lot #s 92661WY1, 00882066, 00111YQP, 00351Z8C, 013120YD, 01472197, 92861XDG, 00912083, 00211Z08, 00621ZQ5, 013220Z8. Model Number: 763120200X Lot #s 00351Z85, 012020QG, 012420TW. Model Number: 763418200X Lot #s 92671WZV, 93291Y41, 00351Z87, 00601ZNH, 010320FK, 0141215T, 92671WZW, 00061YOX, 00431ZE0, 0083203N, 0110201S, 015221BC, 92861XDH, 00131YUE, 00491ZIO, 009620BN, 0011720ON, 015521CN, 93081XPR, 00141YVB, 00541ZKT, 009920D1, 0138213E, 016221H6. Model Number: 763418250X Lot #s 92751X60, 00671ZRL, Model Number: 763418250X Lot #s 92661WY4, 93271Y1N, 00141YVD, 00601ZOQ, 0110201X, 013220ZA, 92751X6P, 93291Y42, 00211Z0B, 00761ZZR, 012020QD, 0137211G, 92781X5G, 93631YJP, 00221Z13, 0081201X, 012420TY, 01382133, 92861XDN, 00912085, 00431ZDZ, 009620BP, 013120YA, 016221H0. |
| FEI Number |
1036710
|
Recalling Firm/ Manufacturer |
Medical Device Technologies, Inc.
3600 Sw 47th Ave
Gainesville
FL
32608-7555
|
| For Additional Information Contact | Jayme Wilson 352-338-0440 Ext. 350 |
Manufacturer Reason for Recall | Medical Device Technologies, Inc., dba Angiotech, Gaineville, Florida is recalling Tru-Core II Disposable Biopsy Devices due to compromised packaging integrity. The Tru-Core Disposable Automatic Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, prostate, breast, ect. The recall is for the following Model Numbers: 763114100X, 763114160X, 763114 |
FDA Determined Cause 2 | Other |
| Action | Angiotech sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated July 2, 2010 and response card to all customers including distributors. The letter described the product, problem and action to be taken by the customers. The customers were instructed to review their current inventory and segregate the affected lot numbers; arrange for return and replacement of the affected product, and complete and return the Customer Acknowledgment Form via fax to the Quality Assurance department at 1-352-338-0662 or 1-800-333-0440.
If you have any questions concerning this notification, contact Shannon Brooks at (352) 338-0440 ext 355 or Katrenia Williams at ext 353. |
| Quantity in Commerce | 11,123 boxes (111,230 units) |
| Distribution | Worldwide distribution: USA and countries including: Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Israel, Japan, New Zealand, Peru, South Africa, , Uruguay, and Venezuela. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = MJG
|
|
|
|