| | Class 2 Device Recall Captus 3000 Thyroid Uptake System with Portable Stand |  |
| Date Initiated by Firm | September 20, 2010 |
| Date Posted | October 28, 2010 |
| Recall Status1 |
Terminated 3 on July 05, 2012 |
| Recall Number | Z-0167-2011 |
| Recall Event ID |
56975 |
| Product Classification |
Calibrator, dose, radionuclide - Product Code KPT
|
| Product | Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010. Manufactured by Capintec, Inc, Pittsburgh, PA. |
| Code Information |
Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010 |
| FEI Number |
2518443
|
Recalling Firm/ Manufacturer |
Capintec Inc
620 Alpha Dr
Pittsburgh
PA
15238
|
| For Additional Information Contact | Art Hall 201-825-9500 |
Manufacturer Reason for Recall | The spring arm on portable stand may have been over tightened during assembly, which may result in failure. If the spring arm fails, it will not balance the collimator weight, and the arm will drop to the lowest position. |
FDA Determined Cause 2 | Employee error |
| Action | Capintec issued an Alert letter to all customers identifying the affected device, informing them of the problem and what actions are to be taken by customers. Capintec will send a replacement arm for the system, which is to be installed by the customer using included instructions. The firm included a correction acknowledgement form to be completed and returned by the customer.
Customers can contact Capintec concerning this recall at 201 825-9500. |
| Quantity in Commerce | 9 units |
| Distribution | Worldwide Distribution: USA in the states of NC and TN, and in the countries of Pakistan, Paraguay, Portugal, Saudi Arabia, and Spain. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|