| | Class 2 Device Recall Perfusion Tubing Set |  |
| Date Initiated by Firm | October 07, 2010 |
| Date Posted | November 15, 2010 |
| Recall Status1 |
Terminated 3 on December 17, 2010 |
| Recall Number | Z-0384-2011 |
| Recall Event ID |
57052 |
| 510(K)Number | K881330 |
| Product Classification |
Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
|
| Product | Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084502000, Manufactured by Sorin Group USA, Inc. 14401 W. 65th Way Arvada, CO 80004
Indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours. |
| Code Information |
Lot No. 1020100026, 1021700023, 1023000022 |
| FEI Number |
1718850
|
Recalling Firm/ Manufacturer |
Sorin Group USA, Inc. 14401 W 65th Way Arvada CO 80004
|
| For Additional Information Contact | 303-467-6306 |
Manufacturer Reason for Recall | Defective product packaging may compromise product sterility. |
FDA Determined Cause 2 | Packaging process control |
| Action | Sorin Group sent an "URGENT: FIELD SAFETY NOTICE" letter dated October 7, 2010, to all customers. The letter described the product, problem and action to be taken by customer. The customers were instructed to check all affected inventory to exam for holes. Any damaged products were to be returned to Sorin. If no damage was found, product could be used. The customers were also instructed to contact Sorin Cardiopulmonary Customer Service at 1-800-650-2623 to return product and receive replacement; they were to share this information with all personnel who may use this product; and to complete and return the attached Customer Response Form via the self-addressed prepaid envelope provided or fax to (303) 467-6502 as soon as possible.
For further information, contact 1-800-650-2623. |
| Quantity in Commerce | 98 |
| Distribution | Nationwide distribution: CA, FL, IL, IN, KS, MD, MO, NE, NJ, NM, TN, TX, and VA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DWF
|
|
|
|