• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Ovation Hip Stem, REF 2010104, Offset Stem Inserter Neutral, NonSterile, Ortho Development

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Ovation Hip Stem, REF 2010104, Offset Stem Inserter Neutral, NonSterile, Ortho Developmentsee related information
Date Initiated by FirmJanuary 11, 2011
Date PostedMarch 16, 2011
Recall Status1 Terminated 3 on November 04, 2011
Recall NumberZ-1673-2011
Recall Event ID 57747
510(K)NumberK062775 
Product Classification Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
ProductOvation Hip Stem, REF 201-0104, Offset Stem Inserter Neutral, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. The hip stem inserter is an instrument used to seat the hip stem in the femur. There is a notch in the top of the hip stem, and the hip stem inserter has a corresponding tip at the end of the instrument. The tip of the inserter is designed to fit into the notch of the hip stem, as a guide. Once this is inserted, the surgeon will tap on the opposite end of the inserter with a hammer to seat the press-fitted stem assembly.
Code Information All lots
Recalling Firm/
Manufacturer
Ortho Development Corporation
12187 South Business Park Dr
Draper UT 84020
For Additional Information Contact
801-619-3450
Manufacturer Reason
for Recall
Component of total hip arthroplasty may break during surgical procedure.
FDA Determined
Cause 2
Process design
ActionOrtho Development Corporation notified Field representatives of the issue and conditions that could lead to failure by e-mail on December 14, 2010. Redesigned product is being produced for replacement. For questions regarding this recall call 801-619-3450.
Quantity in Commerce33 units
DistributionNationwide Distribution including AZ, CA, CO, FL, MI, NV, OR, and UT.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LZO
-
-