Class 1 Device Recall Ambulatory Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Ambulatory Infusion Pumpsee related information
Date Initiated by FirmFebruary 28, 2011
Date PostedApril 05, 2011
Recall Status1 Terminated 3 on August 06, 2012
Recall NumberZ-1868-2011
Recall Event ID 57939
510(K)NumberK981816 
Product Classification Pump, infusion - Product Code FRN
ProductCurlin 6000 IOD Ambulatory Infusion Pump Provides a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Code Information 200083, 200091, 200111, 200174, 200181, 200235, 200238, 200256, 200260, 200293, 200300, 200304, 200963, 200972, 200976, 200979, 200992, 200999, 201021, 201050, 201060, 201069, 201093, 201095, 201125, 201130, 201136, 201174, 201176, 201182, 201187, 201188, 201208, 201210, 201211, 201220, 201232, 201233, 201249, 201262, 201270, 201271, 201273, 201285, 201286, 201692, 201692, 201699, 202044, 202386, 202394, 202405, 202405, 202413, 202509, 202565, 202803, 205273, 205355, 205380, 205678, 205702, 205706, 205762, 205800, 205840, 205924, 205935, 205959, 206000, 206001, 206004, 206055, 206076, 206117, 206117, 206120, 206159, 206197, 206221, 206236, 206240, 206241, 206283, 206334, 206345, 206403, 206405, 206407, 206432, 206446, 206447, 206473, 206475, 206477, 206479, 206480, 206497, 206509, 206867, 206900, 206958, 206965, 206965, 206971, 207005, 207029, 207031, 207185, 207195, 207201, 207212, 207230, 207246, 207247, 207265, 207686, 208053, 208054, 208108, 208203, 208236, 208273, 208319, 208320, 208321, 208558, 208567, 208567, 208589, 208662, 208778, 208778, 208975, 209047, 209047, 209069, 209089, 209090, 209091, 209092, 209093, 209094, 209095, 209113, 209114, 209116, 209117, 209118, 209120, 209140, 209145, 209146, 209215, 209215, 209262, 209267, 209269, 209278, 209279, 209280, 209281, 209284, 209286, 209291, 209292, 209292, 209293, 209294, 209295, 209297, 209299, 209299, 209300, 209302, 209303, 209304, 209305, 209306, 209309, 209314, 209318, 209321, 209322, 209323, 209331, 209331, 209333, 209333, 209335, 209338, 209339, 209340, 209340, 209343, 209343, 209346, 209347, 209359 209361 209363 209364 209365 209373 209379 209383 209390 209399 209403 209403 209406 209411 209411 209413 209413 209415 209416 209418 209418 209419 209419 209420 209420 209421 209421 209423 209424 209424 209428 209429 209430 209432 209432 209433 209435 209436 209439 209439 209441 209441 209442 209444 209445 209446 209446 209448 209449 209450 209452 209467 209469 209470 209473 209473 209474 209474 209475 209476 209479 209480 209483 209500 209503 209509 209512 209515 209521 209527 209528 209529 209531 209538 209594 209605 209613 209614 209615 209618 209634 209636 209637 209638 209640 209641 209642 209654 209655 209657 209661 209666 209668 209669 209670 209671 209672 209673 209674 209682 209683 209693 209701 209708 209709 209711 209712 209713 209724 209743 209745 209748 209752 209753 209754 209756 209759 209761 209763 209764 209765 209766 209767 209768 209769 209770 209785 209788 209790 209791 209805 209808 209809 209811 209859 210342 210499 210500 210501 210502 210503 210567 210570 210571 210574 210654 210788 210800 210801 210802 210803 210804 210805 210841 210843 210844 210845 210886 210886 210961 210966 210970 211303 211309 211335 211336 211338 211339 211340 211341 211342 211344 211345 211346 211347 211348 211353 211354 211355 211356 211358 211359 211360 211361 211372 211386 211386 211394 211398 211401 211402 211403 211407 211409 211425 211428 211434 211435 211460 211461 211462 211463 211464 211465 211466 211467 211468 211470 211471 211484 211485 211486 211487 211489 211490 211491 211506 211507 211558 211560 211562 211563 211564 211583 211584 211586 211587 211589 211590 211596 211598 211599 211600 211601 211605 211606 211611 211616 211627 211628 211629 211630 211631 211633 211634 211635 211636 211641 211645 211646 211647 211648 211650 211659 211660 211661 211662 211663 211664 211665 211666 211669 211670 211671 211672 211673 211674 211675 211676 211678 211699 211790 211817 211818 211819 211822 211828 211860 211865 211869 211870 211871 211883 211884 211886 211887 211890 211891 211914 211915 211916 211917 211918 211919 211920 211921 211926 211950 211951 211952 211953 212011 212013 212016 212017 212032 212035 212039 212045 212049 212050 212051 212052 212054 212055 212057 212058 212085 212086 212088 212090 212096 212097 212098 212100 212101 212102 212103 212104 212105 212106 212107 212108 212110 212111 212112 212114 212115 212116 212117 212118 212122 212172 212173 212174 212193 212443 212446 212451 212455 212459 212477 212478 212490 212491 212492 212493 212494 212495 212496 212501 212507 212508 212509 212510 212511 212512 212513 212514 212516 212517 212518 212520 212521 212522 212525 212526 212528 212531 212534 212535 212536 212537 212538 212539 212542 212543 212545 212546 212547 212548 212549 212554 212555 212556 212574 212583 212584 212585 212586 212587 212588 212627 212628 212629 212631 212632 212633 212634 212636 212637 212645 212647 212651 212654 212655 212656 212658 212659 212660 212661 212662 212663 212675 212680 212681 212682 212683 212684 212685 212687 212703 212705 212706 212708 212711 212713 212718 212721 212723 212724 212726 212727 212728 212729 212730 212732 212733 212734 212736 212739 212740 212741 212742 212743 212744 212745 212746 212747 212749 212758 212762 212765 212766 212767 212771 212773 212774 212776 212777 212779 212780 212783 212784 212785 212786 212787 212789 212790 212791 212792 212793 212794 212795 212796 212799 212876 212877 212879 212880 212883 212887 212888 212953 212955 212957 212960 212962 212963 212966 212967 212968 212969 212970 212973 212974 212975 212976 212977 212995 212996 213007 213009 213011 213013 213014 213016 213017 213018 213023 213036 213037 213039 213040 213041 213042 213043 213052 213055 213057 213058 213071 213072 213075 213078 213079 213081 213083 213084 213085 213088 213093 213106 213107 213108 213109 213121 213122 213123 213125 213126 213128 213137 213138 213141 213143 213147 213148 213151 213160 213161 213162 213163 213164 213165 213166 213167 213193 213194 213195 213196 213197 213198 213199 213206 213244 213246 213302 213305 213307 213309 213329 213331 213332 213333 213335 213336 213337 213338 213355 213356 213371 213372 213373 213377 213378 213379 213380 213381 213382 213383 213384 213385 213401 213402 213403 213404 213405 213407 213409 213410 213411 213421 213422 213423 213424 213425 213426 213427 213428 213438 213444 213445 213447 213448 213449 213453 213454 213455 213456 213458 213459 213460 213470 213480 213481 213495 213502 213503 213507 213517 213518 213519 213520 213522 213523 213524 213525 213527 213528 213529 213530 213532 213533 213534 213535 213536 213539 213541 213542 213543 213544 213545 213546 213547 213548 213550 213551 213552 213553 213555 213556 213557 213558 213559 213561 213562 213563 213564 213565 213569 213571 213574 213575 213577 213579 213582 213583 213584 213585 213586 213587 213588 213591 213592 213593 213596 213599 213600 213603 213614 213615 213617 213619 213620 213621 213622 213623 213624 213625 213626 213628 213629 213630 213632 213633 213634 213635 213661 213665 213688 213689 213690 213691 213692 213693 213694 213695 213698 213716 213718 213719 213720 213721 213727 213760 213786 213788 213790 213791 213792 213793 213794 213795 213813 213818 213841 213848 213849 213850 213851 213852 213854 213855 213856 213858 213860 213873 213876 213877 213878 213879 213903 213947 213951 213957 213959 213960 213961 213969 213999 214002 214136 214153 214168 214172 214173 214185 214197 214208 214213 214214 214215 214217 214218 214222 214223 214225 214226 214234 214235 214236 214263 214265 214266 214267 214269 214270 214271 214272 214273 214275 214276 214277 214278 214279 214281 214282 214284 214286 214290 214291 214292 214293 214298 214299 214300 214301 214302 214323 214327 214328 214329 214330 214331 214332 214333 214334 214336 214337 214338 214339 214340 214342 214343 214344 214345 214347 214348 214349 214350 214351 214352 214353 214355 214356 214359 214360 214361 214362 214363 214364 214365 214366 214373 214384 214391 214527 214531 214534 214538 214544 214545 214548 214738 214739 214743 214744 214745 214746 214747 214748 214749 214751 214752 214753 214754 214755 214756 214757 214758 214759 214779 214780 214781 214782 214783 214784 214847 214910 214911 215008 215009 215010 215011 215012 215013 215014 215015 215016 215018 215019 215020 215021 215022 215028 215029 215030 215031 215033 215034 215035 215036 215037 215038 215039 215040 215041 215043 215044 215045 215047 215049 215050 215051 215052 215053 215055 215056 215057 215059 215060 215062 215064 215065 215066 215068 215069 215070 215071 215072 215073 215074 215075 215076 215078 215079 215080 215081 215082 215083 215084 215085 215087 215089 215090 215091 215092 215093 215094 215095 215096 215097 215099 215100 215101 215102 215103 215104 215105 215106 215107 215109 215111 215112 215113 215114 215116 215118 215119 215120 215122 215130 215131 215153 215155 215156 215157 215158 215159 215161 215162 215163 215164 215165 215166 215167 215168 215169 215171 215172 215173 215174 215175 215176 215177 215185 215188 215373 215376 215479 215480 215558 215559 215560 215561 215562 215565 215566 215567 215568 215571 215572 215573 215574 215575 215577 215578 215637 215639 215641 215642 215643 215646 215647 215649 215651 215652 215653 215654 215655 215658 215661 215665 215669 215671 215672 215673 215674 215676 215678 215679 215680 215684 215686 215687 215688 215691 215693 215696 215702 215704 215708 215709 215711 215712 215716 215718 215720 215723 215724 215727 215728 215730 215731 215732 215734 215736 215737 215740 215743 215744 215746 215747 215750 215751 215752 215753 215755 215757 215758 215759 215760 215763 215765 215766 215767 215768 215771 215775 215778 215782 215783 215784 215785 215787 215788 215790 215791 215793 215794 215795 215796 215797 215798 215799 215803 215805 215807 215808 215812 215814 215815 215817 215818 215820 215821 215823 215826 215828 215830 215831 215834 215835 215840 215841 215844 215845 215850 215853 215854 215856 215857 215864 215866 215867 215868 215869 215870 215873 215874 215876 215879 215881 215882 215883 215884 215885 215887 215888 215889 215892 215893 215895 215896 215899 215902 215903 215904 215905 215910 215913 215917 215925 215927 215928 215929 215934 215935 215940 215941 215942 215943 215944 215945 215946 215947 215948 215952 215955 215957 215958 215966 215967 215969 215971 215972 215974 215975 215977 215979 215981 215985 215986 215987 215988 215991 215992 215993 215996 215997 216003 216005 216010 216011 216013 216014 216015 216018 216019 216020 216021 216022 216023 216027 216029 216030 216031 216032 216034 216035 216036 216037 216041 216043 216045 216049 216053 216057 216058 216060 216061 216065 216068 216069 216070 216071 216073 216074 216075 216077 216079 216087 216088 216089 216093 216095 216096 216097 216099 216103 216104 216105 216106 216107 216108 216109 216110 216111 216114 216119 216121 216123 216124 216125 216127 216127 216128 216133 216138 216144 216151 216151 216153 216154 216156 216158 216161 216163 216165 216168 216170 216171 216174 216178 216180 216183 216184 216185 216186 216189 216192 216193 216194 216196 216208 216209 216213 216214 216223 216224 216230 216233 216234 216236 216240 216242 216243 216245 216246 216252 216253 216255 216258 216261 216262 216265 216266 216269 216270 216271 216273 216275 216276 216278 216281 216282 216313 216331 216335 216336 216340 216346 216347 216348 216349 216351 216352 216353 216355 216358 216360 216362 216371 216372 216373 216374 216376 216377 216379 216380 216381 216382 216383 216384 216385 216386 216388 216394 216395 216396 216397 216401 216402 216403 216404 216405 216408 216411 216413 216418 216419 216420 216422 216424 216426 216428 216429 216432 216433 216434 216437 216439 216440 216441 216445 216450 216452 216454 216456 216457 216458 216460 216461 216463 216464 216465 216468 216469 216470 216472 216473 216477 216479 216480 216481 216482 216489 216490 216493 216494 216496 216498 216500 216504 216506 216508 216509 216511 216512 216513 216515 216517 216518 216519 216520 216521 216522 216523 216526 216528 216530 216531 216532 216534 216535 216536 216540 216542 216545 216547 216548 216550 216551 216552 216553 216554 216556 216567 216568 216570 216571 216572 216573 216574 216575 216579 216584 216585 216588 216590 216591 216592 216593 216595 216596 216598 216599 216603 216608 216611 216612 216613 216614 216615 216617 216618 216619 216621 216622 216626 216628 216629 216631 216633 216634 216683 216695 216705 216720 216743 216796 216798 216801 216803 216809 216811 216812 216813 216815 216819 216821 216826 216829 216831 216833 216834 216836 216837 216845 216850 216854 216855 216856 216857 216858 216862 216870 216871 216872 216876 216878 216879 216888 216889 216890 216892 216896 216897 216901 216904 216905 216906 216907 216908 216910 216911 216912 216916 216917 216918 216921 216930 216933 216934 216935 216936 216940 216941 216942 216944 216945 216954 216955 216956 216957 216961 216962 216963 216964 216965 216966 216967 216968 216969 216971 216972 216974 216977 216978 216981 216982 216983 216985 216987 216988 216991 216992 216998 216999 217001 217004 217005 217008 217009 217010 217011 217020 217021 217022 217029 217031 217033 217034 217036 217038 217039 217041 217044 217045 217049 217051 217052 217053 217055 217061 217062 217063 217064 217066 217069 217070 217077 217091 217092 217093 217094 217095 217096 217101 217102 217104 217106 217108 217109 217110 217111 217112 217115 217117 217118 217125 217127 217129 217130 217132 217133 217136 217137 217139 217140 217153 217156 217157 217167 217169 217173 217174 217177 217179 217180 217181 217182 217185 217186 217188 217191 217192 217193 217196 217197 217198 217199 217214 217228 217229 217230 217232 217235 217238 217239 217241 217242 217419 217420 217421 217422 217423 217424 217425 217426 217441 218309 218313 218430 218431 218432 218433 218434 218435 218436 218438 218439 218441 218442 218443 218444 218445 218446 218447 218450 218451 218452 218453 218454 218462 218500 218501 218603 218605 218606 218607 218609 218610 218611 218613 218614 218615 218617 218618 218651 218677 218678 218679 218680 218682 218683 218685 218686 218687 218688 218689 218690 218691 218692 218693 218695 218696 218697 218698 218699 218700 218701 218702 218703 218705 218706 218707 218708 218709 218710 218713 218715 218716 218722 218724 218725 218726 218728 218729 218730 218731 218732 218734 218735 218736 218737 218738 218739 218740 218741 218742 218744 218745 218746 218748 218749 218750 218751 218752 218753 218755 218757 218758 218759 218760 218762 218763 218764 218765 218769 218770 218771 218772 218773 218774 218775 218776 218777 218779 218780 218781 218782 218783 218784 218785 218786 218787 218789 218790 218791 218792 218793 218794 218795 218796 218797 218799 218800 218801 218802 218803 218804 218806 218808 218809 218811 218812 218813 218814 218815 218816 218817 218818 218819 218821 218822 218823 218824 218826 218827 218828 218829 218830 218832 218833 218834 218835 218836 218837 218838 218839 218840 218842 218843 218844 218845 218846 218847 218848 218849 218851 218853 218854 218856 218857 218859 218860 218861 218862 218864 218866 218867 218868 218869 218870 218871 218873 218874 218875 219176 219177 219179 219180 219181 219182 219183 219184 219185 219189 219190 219191 219193 219194 219195 219197 219199 219200 219203 219205 219206 219207 219208 219209 219210 219211 219212 219214 219218 219219 219220 219221 219223 219224 219225 219226 219229 219230 219231 219232 219235 219237 219238 219239 219243 219248 219249 219250 219251 219252 219253 219254 219255 219262 219265 219269 219278 219287 219295 219298 219328 219333 219334 219337 219339 219340 219341 219343 219344 219345 219346 219347 219349 219350 219352 219353 219354 219355 219356 219357 219358 219360 219361 219362 219378 219379 219393 219442 219443 219444 219451 219452 219453 219454 219455 219457 219458 219459 219460 219462 219463 219464 219465 219466 219467 219468 219470 219472 219474 219475 219476 219477 219478 219479 219480 219481 219482 219484 219669 219671 219672 219673 219674 219675 219676 219677 219679 219680 219681 219682 219683 219684 219686 219687 219688 219690 219691 219692 219693 219694 219695 219696 219697 219698 219700 219701 219702 219711 219716 219753 219787 219788 219789 219791 219854 219855 219856 219858 219859 219860 219861 219862 219864 219866 219870 219871 219872 219873 219874 219875 219876 219878 219879 219880 219881 219882 219885 219886 219887 219888 219890 219891 219892 219894 219895 219897 219898 219899 219901 219920 219922 219923 219925 219926 219927 219928 219929 219930 219935 219936 219937 219938 219947 219949 219952 219954 219956 219958 219959 219960 219961 219963 219965 219966 219967 219968 219970 219971 219972 219973 219974 219975 219976 219978 219979 219981 219982 219983 219984 219985 220000 220597 220598 220599 220603 220604 220605 220606 220607 220608 220610 220611 220612 220615 220616 220617 220618 220619 220620 220621 220622 220623 220625 220626 220627 220628 220629 220630 220631 220632 220633 220635 220636 220637 220638 220639 220640 220641 220642 220643 220645 220647 220648 220650 220651 220652 220653 220654 220655 220657 220658 220659 220660 220661 220662 220663 220664 220665 220667 220668 220669 220670 220671 220672 220673 220674 220675 220677 220678 220679 220680 220681 220682 220683 220684 220685 220687 220688 220689 220690 220692 220693 220694 220695 220696 220698 220699 220701 220702 220703 220704 220705 220706 220707 220709 220711 220712 220713 220714 220715 220716 220717 220718 220720 220721 220722 220723 220724 220725 220726 220727 220728 220730 220731 220732 220733 220734 220735 220736 220737 220738 220740 220743 220744 220746 220747 220748 220757 220758 220760 220763 220764 220765 220775 220776 220777 220783 220784 220788 220790 220791 220792 220793 220794 220795 220796 220798 220799 220801 220802 220803 220804 220805 220807 220808 220809 220810 220812 220813 220814 220815 220816 220817 220818 220819 220820 220823 220824 220825 220827 220828 220829 220830 220831 220832 220834 220835 220836 220837 220838 220839 220840 220841 220842 220845 220846 220847 220848 220849 220850 220851 220861 220862 220869 220870 220871 220873 220879 220880 220881 220882 220883 220885 220886 220887 220888 220889 220890 220892 220893 220894 220896 220897 220898 220899 220900 220901 220903 220904 220905 220907 220908 220909 220910 220911 220912 220913 220914 220915 220917 220918 220919 220920 220921 221159 221160 221161 221162 221164 221165 221166 221167 221168 221169 221170 221171 221173 221175 221176 221177 221178 221179 221180 221182 221183 221184 221186 221187 221188 221189 221190 221191 221192 221193 221194 221196 221197 221198 221199 221200 221201 221202 221203 221204 221206 221207 221209 221211 221212 221213 221214 221215 221216 221218 221219 221220 221221 221222 221223 221224 221225 221226 221228 221229 221230 221231 221232 221233 221234 221235 221236 221238 221239 221240 221241 221242 221243 221244 221245 221246 221248 221249 221250 221251 221253 221254 221255 221256 221257 221259 221260 221261 221262 221263 221264 221266 221267 221268 221270 221272 221273 221274 221275 221276 221277 221278 221279 221281 221482 221483 221484 221485 221486 221487 221488 221489 221491 221492 221493 221494 221495 200073 200075 200076 200085 200089 200093 200096 200100 200113 200117 200120 200121 200122 200127 200132 200134 200138 200139 200142 200158 200172 200342 200986 201020 201027 201035 201036 201071 201072 201374 201374 201699 201711 201805 201806 201806 202381 202424 202427 202427 202525 202565 202596 202697 202713 202773 202784 202787 202791 202794 202794 204407 205794 206098 206366 206366 206516 206563 206670 206670 206683 206683 206794 206837 206900 206908 206963 206972 207020 207082 207082 207162 207190 207193 207193 207201 207207 207326 207326 207576 207922 208055 208244 208590 209027 209108 209379 209416 209771 209778 209782 209827 209827 209830 209992 210343 210351 210351 210382 210485 210485 210580 210580 210581 210656 210656 210760 210760 211323 211454 213406 213469 213472 213474 213476 213491 213714 213745 213747 213749 213761 213862 213875 213959 214239 214247 214249 214258 214571 214572 214576 214577 214579 215128 216668 216669 216670 216671 219393 219876 219920 219922 219923 221270 219870 219871 219872 219937 221081 218595 218596 218597 218598 218599 218600 218601 218602 211955 211956 212022 212024 212025 212026 212155 212156 212157 212807 212823 212825 212826 212830 212858 212859 212890 213012 213757 213759 213871 219939 219940 219941 219942 219943 219944 219946 220111 220112 220113 220114 220115 220117 220118 220119 220123 220125 220127 220130 220131 220132 220134 220135 220136 220138 220139 220140 220141 220143 220144 220145 220146 220147 220149 220150 220151 220152 220153 220155 220156 220158 220159 220161 220162 220164 220165 220166 220167 220168 220169 220170 220173 220967 220968 220969 220970 220971 220972 220973 220974 220976 220977 220978 220979 220980 220981 220982 220983 220985 220987 220988 220989 220990 220991 220992 220993 220994 220995 220997 220998 220999 221000 221001 221003 221004 221005 221006 221008 221009 221010 221011 221047 221048 221049 221050 221051 221053 220749 220750 220751 220752 220753 220754 220755 220756 220779 220780 220781 220782 220787 220923 220924 220925 220926 220928 220929 220930 220931 220932 220766 220767 220768 220769 220771 220772 220773 220774 220785 220786 220934 220935 221072 221074 221075 221076 221077 221080 200310 201705 202165 202407 202669 203169 205897 205897 206212 206313 206444 207069 207190 207696 207696 207700 209784 209804 210443 210461 210509 210509 211308 211324 211404 211424 211815 213889 213900 213950 213952 213996 214104 214104 214251 214540 214991 219434 219435 219436 219439 219440 219441 219445 219446 219447 219448 219449 219456 219485 219699 219700 219852 219868 219868 219869 219869 219907 219950 219950 219951 219951 219997 219997 219999 219999 220037 220741 220860 220874 220876 220877 220878 221282 221282 221307 221308 221308 221309 221309 222652 222652 222727 222727  
Recalling Firm/
Manufacturer
Zevex International, Inc.
4314 Zevex Park Ln
Salt Lake City UT 84123
For Additional Information Contact
801-264-1001 Ext. 144
Manufacturer Reason
for Recall
Infusion pumps may exhibit false error codes which render the device inoperable until a recovery process is performed.
FDA Determined
Cause 2
Software design
ActionThe firm, MOOG, sent an "URGENT DEVICE RECALL NOTIFICATION" letter dated February 28, 2011 to its customers. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) Identify the affected Pumps: a) If they have an affected pump, fill out and return the enclosed Business Reply Card within five (5) working days. b) If their pump is not affected by this recall, the pump can be used for its intended purposes. 2) Removal of affected Product: a) remove affected products from service and use non-affected pump b) If you do not have an alternate pump available for use, MOOG recommended that they weigh the risks and benefits for delivering medications using the affected pump versus the risk of not using any pump for each patient WARNING: Affected pumps should not be used to deliver critical medications where a delay in therapy could result in serious adverse health effects, injury or death; and affected pumps should not be used with the neonatal population. 3)Should they Have an Affected pump which exhibits Error Code 45 or is used in therapeutic situations specified in the warning, please contact MOOG Customer Service at (800) 970-2337, Monday thru Friday, 7:00am to 5:00pm MST to arrange for a loaner pump and the return of your affected pump. If you have any additional questions regarding this recall, please contact MOOG's Customers Service Department at (800) 970-2337.
Quantity in Commerce3,407 units
DistributionWorldwide distribution: USA and countries including: Australia, Canada, Costa Rica, Croatia, Dubai, France, Germany, Ireland, Italy, Japan, Jordan, Kuwait, Malaysia, New Zealand, Qatar, Saudi Arabia, Singapore, South Africa, Taiwan, Turkey, UAE, United Kingdom.
Total Product Life CycleTPLC Device Report