|
Class 2 Device Recall Rapid Response HUT Table |
|
Date Initiated by Firm |
February 01, 2010 |
Date Posted |
April 13, 2011 |
Recall Status1 |
Terminated 3 on June 15, 2011 |
Recall Number |
Z-1973-2011 |
Recall Event ID |
58035 |
Product Classification |
Tilting Radiologic Table - Product Code IXR
|
Product |
Rapid Response H.U.T. (Head Up Tilt) Table. Model 1017. Powered electrophysiology tilt table. Table can tilt -15 degrees to 90 degrees (Trendelenburg). Medical Positioning Incorporated.800-593-3246.
Used in clinical and hospital settings by trained radiologists. |
Code Information |
598959, 599042, 599084, 599091, 599144, 599189, 599225, 599278, 599539, 599593, 599616, 599704, 599706, 599708, 599744, 599847, 599862, 599937, 600018, 600032, 600046, 600106, 600107, 600343, 600428, and 600667 located on the side of the table |
Recalling Firm/ Manufacturer |
Medical Positioning Inc 1717 Washington St Kansas City MO 64108-1121
|
For Additional Information Contact |
Joseph J. Hurtig 816-474-1555
|
Manufacturer Reason for Recall |
The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.
|
FDA Determined Cause 2 |
Other |
Action |
An undated Urgent - Voluntary Medical Device Product Correction letter was issued on 2/1/2010 by certified mail with return receipt requested. The letter identified the model and serial numbers of the affected product. The consignees were advised to stop using the table immediately and remove it from service until further notice. Customers were to contact Mr. Joe Hurtig at 1-800-593-3246 or jhurtig@medicalpositioning.com to confirm receipt of the notice and to provide the firm with the correct contact information. Individualized correction packages including replacement parts and instructions were sent to each consignee. The instructions consisted of (1) Rapid Response HUT Motor Replacement Instructions; (2) Control Box Replacement Instructions; and (3) ILD External Power Supply Removal Instructions. |
Quantity in Commerce |
58 total |
Distribution |
Worldwide Distribution -- USA, including Puerto Rico, and Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|