| Class 2 Device Recall iSite PACS with iSite ViewForum Applications | |
Date Initiated by Firm | February 23, 2011 |
Date Posted | April 04, 2011 |
Recall Status1 |
Terminated 3 on August 18, 2011 |
Recall Number | Z-1882-2011 |
Recall Event ID |
58044 |
510(K)Number | K032096 |
Product Classification |
System, image processing, radiological - Product Code LLC
|
Product | iSite PACS with iSite ViewForum Applications, R7.4V6 L1 L2, L3,
Versions: 3.6.96, 3.6.101, 3.6.110, 4.1.96, 4.1.101, and 4.1.110
; Product is manufactured and distributed by Philips Medical Systems, Foster City, CA
ViewForum Applications: iSite ViewForum Applications is an image processing workstation software package designed to run on standard PC hardware. The hardware required is made up of "off-the-shelf" standard computer components. iSite ViewForum Applications software receives image data from medical scanning devices, such as CT and MRI, or from image archives. iSite ViewForum Applications performs viewing, image manipulation, communication, printing, and quantification of images. |
Code Information |
All units of these iSite PACS. |
Recalling Firm/ Manufacturer |
Philips Medical Systems 4100 E 3rd Ave Ste 101 Foster City CA 94404-4819
|
For Additional Information Contact | Victoria Mackinnon 650-293-2624 |
Manufacturer Reason for Recall | A software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements which could cause an incorrect diagnosis and incorrect treatment plan. |
FDA Determined Cause 2 | Software design |
Action | Philips sent an Urgent Field Safety Notice letter dated February 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.
Customers were instructed to assure that all colleagues were aware of the problem.
When using the Volume Vision software, check the correctness of measurements (graphics and numerical values) prior to saving a captured image containing measurements.
Save captured images containing measurements immediately after creating them and before altering the measurements in the original image view.
For any subsequent measurements, create a new capture and save it immediately.
For further information or support concerning this issues, customers should contact their local Philips representative at +1-877-328-2808. |
Distribution | Worldwide Distribution. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LLC
|
|
|
|