• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Sensor cables

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Sensor cablessee related information
Date Initiated by FirmFebruary 24, 2011
Date PostedApril 11, 2011
Recall Status1 Terminated 3 on September 09, 2011
Recall NumberZ-1941-2011
Recall Event ID 58098
510(K)NumberK040050 
Product Classification System, image processing, radiological - Product Code LLZ
ProductTransmitter Sensor Cables, Part Numbers: 1002008, 1004587, 1007914-NAV, GE Healthcare Surgery, Salt Lake City, UT 84116. Cables are used with InstraTrak Navigation Systems, Models IT3000, IT2500, IT2500+, IT3500, IT3500+. InstaTrak System is intended as an aid to the surgeon for precisely locating anatomical structures anywhere on the human body during either open or percutaneous procedures. It is indicated for any medical condition that may benefit from the use of stereotactic surgery and which provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, visible on medical images such as CT, MR, or X-ray.
Code Information All serial numbers.
FEI Number 1720753
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City UT 84116-2862
For Additional Information Contact
801-517-6717
Manufacturer Reason
for Recall
Sensor cables may suffer material degradation when exposed to certain sterilization procedures.
FDA Determined
Cause 2
Component design/selection
ActionGE OEC Medical Systems, Inc sent an Urgent Recall Notice dated March 1, 2011, by certified letter on February 24, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised if their Transmitter Sensor Cables or Receiver Sensor Cables have been exposed to any STERRAD sterilizer other than STERRAD 1005, discontinue use of these sterilizers for this product. The Transmitter Sensor Cables should be inspected for signs of material degradation. Further information is available at 800-874-7378.
Quantity in Commerce708 cables
DistributionWorldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI and WV and the countries of Austria, Canada, China, Denmark, England, France, Germany, Hong Kong, India, Ireland, Italy, Japan, San Salvador, Slovakia, Spain, and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LLZ
-
-