• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MAKO PN 204588 Camera

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MAKO PN 204588 Camerasee related information
Date Initiated by FirmOctober 25, 2010
Date PostedMay 20, 2011
Recall Status1 Terminated 3 on November 09, 2011
Recall NumberZ-2232-2011
Recall Event ID 58464
Product Classification Camera, surgical and accessories - Product Code KQM
ProductMAKO PN 204588 Camera 204588 Label With Camera Stand Number ***PN 200294 SN CAM XXX 2011-01.*** Weight 86.2kg.***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90.*** Manufactured in USA*** MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204589 Robot Arm Label***PN 203999 SN ROB XXX 2011-01.***V 100/120/230 A 9.6/8.0/4.2 Hz 50/60***Weight 39.4kg***Class I Equipment Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1:2003 CAN/CSA-C22.2 No. 601-1-M90.***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204588 Label With Guidance Module Number***PN 201251 SN GUD XXX 2011-01***Weight 51.7kg***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, En 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** THE RIO uses an infra-red camera for the purpose of spatial navigation. This camera is present on the Camera Stand and communicates with the Guidance Module to provide the locations of the tracked instruments (i.e. robotic-arm, patient trackers, probes, etc). The information is then used to provide the virtual representation of the patient anatomy and the tools on the monitor, and provides necessary information to the robotic-arm to ensure proper bone preparation.
Code Information Lot #K081867
FEI Number 3005985723
Recalling Firm/
Manufacturer
Mako Surgical Corporation
2555 Davie Rd Ste 110
Davie FL 33317-7424
For Additional Information ContactJim Pomeroy
954-927-2044 Ext. 604
Manufacturer Reason
for Recall
MAKO Surgical Corporation recalled ONE infra-red camera MAKO PN 204854 which was a component of their RIO Robotic Arm Interactive Orthopedic System with SSN P7-03127, on November 22, 2010. This component can function improperly and cause the accuracy of the system to be outside of the specification listed by the manufacturer. The recall was initiated after the camera manufacture, NDI, Waterloo, On
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm, MAKO, sent a "RECALL NOTICE" letter dated November 22, 2010, to its customer. The letter described the product, problem and action to be taken. The customer was instructed to complete and return the attached Acknowledgement Form to their MAKOplasty Specialist (MPS) via mail to: MAKO Surgical Corp., Attn: Jim Pomeroy, 2555 Davie Road, Fort Lauderdale, FL 33317. The letter stated that "It is important that the camera be replaced immediately, and MAKO is currently addressing the situation with the camera manufacturer." In addition, MAKO will replace the camera with a camera that is not affected by the recall and verify the customers system to meet the functional specifications. If you have any questions or concerns, please contact your MPS or Sr. Director of Quality Assurance at 954-927-2044 ext 604, and /or Sr. Director, Customer Support at 954-927-2044 ext. 442.
Quantity in Commerce1
DistributionNationwide distribution: NY only.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-