Part B: Supplementary Information Sheet (SIS) |
FR Recognition List Number
|
034
|
Date of Entry 01/30/2014
|
FR Recognition Number
|
11-238
|
Standard | |
ASTM F2033-12 Standard Specification for Total Hip Joint Prosthesis and Hip Endoprosthesis Bearing Surfaces Made of Metallic, Ceramic, and Polymeric Mateirals |
|
Scope/Abstract1.1 This specification covers the requirements for the mating bearing surfaces of total hip prostheses and hip endoprostheses. More specifically, this specification covers hip joint replacement of the ball-and-socket configuration.
1.2 This specification covers the sphericity, surface finish requirements, and dimensional tolerances for the spherical articulating metallic or ceramic femoral heads of total hip joint prostheses.
1.3 This specification covers the sphericity, surface finish requirements, and dimensional tolerances for the spherical concave mating surface of metallic and ceramic acetabular components, including the inner polymeric bearing surface of bipolar heads, and the surface finish requirements and dimensional tolerances for the spherical concave mating surface of polymeric acetabular components.
1.4 This specification covers the sphericity and surface finish requirements for the spherical metallic or ceramic femoral heads of hip endoprostheses, and the outer bearing surface of bipolar heads.
1.5 This specification is intended for standard practice regarding the design of total hip joint bearing surfaces. Additionally, the tolerances imposed on the polymeric portion of the bearing surface are intentionally large due to temperature-induced changes and other manufacturing concerns. Some manufacturing methods or designs may intentionally reduce the diameter of the polymeric bearing to more closely mate with the diameter of the head. |
|
Extent of Recognition
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
|
|
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number |
Device Name |
Device Class |
Product Code |
§888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented |
Class 2 |
OQI |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish |
Class 2 |
MAY |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented |
Class 2 |
LZO |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate |
Class 2 |
MEH |
|
Unclassified |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented |
Class 3 |
LPF |
Unclassified |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented |
Class 3 |
MRA |
§888.3300 |
Prosthesis, Hip, Constrained, Metal |
Class 3 |
KXD |
§888.3330 |
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) |
Class 3 |
KWA |
§888.3350 |
Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented |
Class 2 |
OQH |
§888.3350 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented |
Class 2 |
JDI |
§888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented |
Class 2 |
OQG |
§888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous |
Class 2 |
MBL |
§888.3360 |
Prosthesis, Hip, Femoral Component, Cemented, Metal |
Class 2 |
JDG |
§888.3360 |
Prosthesis, Hip, Hemi-, Femoral, Metal |
Class 2 |
KWL |
§888.3360 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented |
Class 2 |
LWJ |
§888.3360 |
Prosthesis, Upper Femoral |
Class 2 |
JDD |
§888.3390 |
Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented |
Class 2 |
KWY |
§888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer |
Class 2 |
KWZ |
§888.3320 |
Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) |
Class 3 |
JDL |
§888.3340 |
Prosthesis, Hip, Semi-Constrained, Composite/Metal |
Class 2 |
KMC |
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance Document for the Preparation of Premarket Notification for Ceramic Ball Hip Systems, January 10, 1995 |
|