• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 046 Date of Recognition 12/23/2016 
FR Recognition Number 11-313
Standard
ISO 7207-2 Second edition 2011-07-01
Implants for surgery - Components for partial and total knee joint prostheses - Part 2: Articulating surfaces made of metal, ceramic and plastics materials [Including AMENDMENT 1 (2016)]
Scope/Abstract
This part of ISO 7207 specifies surface finish requirements for the articulating surfaces of total and partial knee joint prostheses classified in ISO 7207-1. This part of ISO 7207 is intended to provide guidance for periodic validation of production processes.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)
Regulation Number Device Name Device Class Product Code
§888.3565 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 2 MBH
§888.3560 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer Class 2 MBV
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive Class 2 OIY
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Class 2 JWH
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Metal Class 3 KRP
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Polymer Class 2 KRQ
§888.3540 Prosthesis, Knee, Patello/Femoral, Semi-Constrained, Cemented, Metal/Polymer Class 2 KRR
§888.3535 Prosthesis, Knee, Femorotibial, Unicompartmental/Unicondylar, Uncemented, Porous-Coated, Metal/Polymer Class 2 NJD
§888.3530 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer Class 2 HRY
§888.3520 Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer Class 2 HSX
§888.3510 Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer Class 2 KRO
§888.3500 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Composite Class 2 KYK
§888.3490 Prosthesis, Knee, Femorotibial, Non-Constrained, Metal/Composite Cemented Class 2 KTX
§888.3490 Prosthesis, Knee, Non-Constrained (Metal-Carbon Reinforced Polyethylene) Cemented Class 2 KMB
§888.3480 Metal Cemented Constrained Femorotibial Knee Prosthesis Class 3 KRN
Unclassified Prosthesis, Knee, Femorotibial, Unicompartmental, Semi-Constrained, Metal/Polymer, Mobile Bearing Class 3 NRA
Unclassified Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Polymer/Metal/Polymer Class 3 LXY
Unclassified Prosthesis, Knee, Patello/Femorotibial, Unconstrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 3 MBD
Unclassified Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing Class 3 NJL
§888.3480 Prosthesis, Knee, Hinged (Metal-Metal) Class 3 HRZ
Relevant FDA Guidance and/or Supportive Publications
Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, January 16, 2003
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Orthopedic
-
-