• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 8-474
Standard
ASTM F2886-17
Standard Specification for Metal Injection Molded Cobalt-28Chromium-6Molybdenum Components for Surgical Implant Applications
Scope/Abstract
1.1 This specification covers chemical, mechanical, and metallurgical requirements for metal injection molded (MIM)cobalt-28chromium-6molybdenum components to be used in the manufacture of surgical implants.
1.2 The MIM components covered by this specification may have been densified beyond their as-sintered density by postsinter processing.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 David M. Saylor
  FDA/OC/CDRH/OSEL/DBCMS/
  301-796-2626
  david.saylor@fda.hhs.gov
 Matthew Di Prima
  FDA/OC/CDRH/OSEL/DAM/
  301-796-2507
  matthew.diprima@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Materials
*These are provided as examples and others may be applicable.
-
-