Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 11-322
Standard
ASTM F1541-17
Standard Specification and Test Methods for External Skeletal Fixation Devices
Scope/Abstract
1.1 This specification provides a characterization of the design and mechanical function of external skeletal fixation devices (ESFDs), test methods for characterization of ESFD mechanical properties, and identifies needs for further development of test methods and performance criteria. The ultimate goal is to develop a specification, which defines performance criteria and methods for measurement of performance-related mechanical characteristics of ESFDs and their fixation to bone. It is not the intention of this specification to define levels of performance or case-specific clinical performance of the devices, as insufficient knowledge is available to predict the consequences of the use of any of these devices in individual patients for specific activities of daily living. Furthermore, it is not the intention of this specification to describe or specify specific designs for ESFDs.
1.2 This specification describes ESFDs for surgical fixation of the skeletal system. It provides basic ESFD geometrical definitions, dimensions, classification, and terminology; material specifications; performance definitions; test methods; and characteristics determined to be important to the in-vivo performance of the device.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§888.3040 Pin, Fixation, Smooth Class 2 HTY
§888.3040 Pin, Fixation, Threaded Class 2 JDW
§888.3030 Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component Class 2 KTT
Relevant FDA Guidance and/or Supportive Publications*
Reviewers Guidance Checklist for Orthopedic External Fixation Devices Version #5, Issued February 1997
FDA Technical Contacts
 Jonathan Peck
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIB/
  301-796-5650
  jonathan.peck@fda.hhs.gov
 Christopher Ferreira
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIC/
  301-796-9350
  christopher.ferreira@fda.hhs.gov