Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 10-113
Standard
ANSI  Z80.30-2018 (R2023)
American National Standard for Ophthalmics - Toric Intraocular Lenses
Scope/Abstract
This standard applies to any monofocal intraocular lens (IOL) whose primary indication is the reduction of astigmatism either with the correction of aphakia or the modification of the refractive power of a phakic eye. It does not include IOLs used to correct presbyopia. This standard addresses the vocabulary, optical properties and test methods, mechanical properties and test methods, labeling, biocompatibility, sterility, shelf-life and transport stability, and clinical investigations necessary for this type of device. As applies to any standard, alternative validated test methods may be used.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§886.3600 Lens, Intraocular, Toric Optics Class 3 MJP
§886.3600 Phakic Toric Intraocular Lens Class 3 QCB
Relevant FDA Guidance and/or Supportive Publications*
Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices - Guidance for Industry and Food and Drug Administration Staff, Issued August 2015.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Bennett Walker
  FDA/OC/CDRH/OPEQ/OHTI/DHTIA/
  301-796-5093
  Bennett.Walker@fda.hhs.gov
Standards Development Organization
ANSI American National Standards Institute https://www.ansi.org/
FDA Specialty Task Group (STG)
Ophthalmic