Part B: Supplementary Information Sheet (SIS) |
FR Recognition List Number
|
036
|
Date of Entry 07/09/2014
|
FR Recognition Number
|
11-279
|
Standard | |
ASTM F2996-13 Standard Practice for Finite Element Analysis (FEA) of Non-Modular Metallic Orthopaedic Hip Femoral Stems |
|
Scope/Abstract1.1 This practice establishes requirements and considerations for the numerical simulation of non-modular (that is, limited to monolithic stems with only a femoral head / trunnion taper interface) metallic orthopaedic hip stems using Finite Element Analysis (FEA) techniques for the estimation of stresses and strains. This standard is only applicable to stresses below the yield strength, as provided in the material certification.
1.2 Purpose-This practice establishes requirements and considerations for the development of finite element models to be used in the evaluation of non-modular metallic orthopaedic hip stem designs for the purpose of prediction of the static implant stresses and strains. This procedure can be used for worst case assessment within a family of implant s to provide efficiencies in the amount of physical testing to be conducted. Recommended procedures for performing model checks and verification are provided to help determine if the analysis follows recommended guidelines. Finally, the recommended content of an engineering report covering the mechanical simulation is presented.
1.3 Limits-This practice is limited in discussion to the static structural analysis of non-modular metallic orthopaedic hip stems (which excludes the prediction of fatigue strength). |
|
Extent of Recognition
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
|
|
Transition Period
FDA recognition of ASTM F2996-13 [Rec# 11-279] will be superseded by recognition of ASTM F2996-20 [Rec# 11-372]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-279] until December 19, 2021. After this transition period, declarations of conformity to [Rec# 11-279] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number |
Device Name |
Device Class |
Product Code |
§888.3330 |
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) |
Class 3 |
KWA |
§888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer |
Class 2 |
KWZ |
§888.3320 |
Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) |
Class 3 |
JDL |
§888.3340 |
Prosthesis, Hip, Semi-Constrained, Composite/Metal |
Class 2 |
KMC |
§888.3350 |
Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented |
Class 2 |
OQH |
§888.3350 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented |
Class 2 |
JDI |
§888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented |
Class 2 |
OQI |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish |
Class 2 |
MAY |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented |
Class 2 |
LZO |
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate |
Class 2 |
MEH |
§888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented |
Class 2 |
OQG |
§888.3358 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented |
Class 2 |
LPH |
§888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous |
Class 2 |
MBL |
§888.3360 |
Prosthesis, Hip, Femoral Component, Cemented, Metal |
Class 2 |
JDG |
§888.3360 |
Prosthesis, Hip, Hemi-, Femoral, Metal |
Class 2 |
KWL |
§888.3360 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented |
Class 2 |
LWJ |
§888.3360 |
Prosthesis, Upper Femoral |
Class 2 |
JDD |
§888.3390 |
Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented |
Class 2 |
KWY |
Unclassified |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented |
Class 3 |
LPF |
Unclassified |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented |
Class 3 |
MRA |
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Non-clinical Information for Femoral Stem Prostheses, September 17, 2007 http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm071275.htm |
|
FDA Technical Contact
|
Standards Development Organization
|
FDA Specialty Task Group (STG)
|
|
*These are provided as examples and others may be applicable. |
|
|