Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 038 Date of Entry 01/27/2015 
FR Recognition Number 3-88
Standard
ASTM F2514-08 (Reapproved 2014)
Standard Guide for Finite Element Analysis (FEA) of Metallic Vascular Stents Subjected to Uniform Radial Loading
Scope/Abstract
This guide establishes general requirements and considerations for the development of finite element models used in the evaluation of the performance of a metallic vascular stent design under uniform radial loading. Suggested criteria are provided for evaluating the typical cases of metallic stents under uniform radially oriented and pulsatile loading. Recommended procedures for checking and validating the finite element model(s) are provided as a means to assess the model and analysis results. Finally, the recommended content of an engineering report covering the mechanical simulations is presented.
Extent of Recognition
Complete standard
Transition Period
FDA recognition of ASTM F2514-08 (Reapproved 2014) [Rec# 3-88] will be superseded by recognition of ASTM F2514-21 [Rec# 3-171]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 3-88] until December 18, 2022. After this transition period, declarations of conformity to [Rec# 3-88] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
Unclassified Coronary Drug-Eluting Stent Class 3 NIQ
Unclassified Stent, Carotid Class 3 NIM
Unclassified Stent, Coronary Class 3 MAF
Unclassified Stent, Iliac Class 3 NIO
Unclassified Stent, Renal Class 3 NIN
Unclassified Stent, Superficial Femoral Artery Class 3 NIP
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems, Issued April 2010.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Finn Donaldson
  FDA/OC/CDRH/OPEQ/OHTII/DHTIIC/
  301-796-9579
  Finn.Donaldson@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Cardiovascular
*These are provided as examples and others may be applicable.
-
-