Part B: Supplementary Information Sheet (SIS) |
FR Recognition List Number
|
046
|
Date of Entry 12/23/2016
|
FR Recognition Number
|
14-496
|
Standard | |
ASTM F1608-16 Standard Test Method for Microbial Ranking of Porous Packaging Materials (Exposure Chamber Method) |
|
Scope/Abstract1.1 This test method is used to determine the passage of airborne bacteria through porous materials intended for use in packaging sterile medical devices. This test method is designed to test materials under conditions that result in the detectable passage of bacterial spores through the test material. 1.1.1 A round-robin study was conducted with eleven laboratories participating. Each laboratory tested duplicate samples of six commercially available porous materials to determine the Log Reduction Value (LRV) (see calculation in Section 12). Materials tested under the standard conditions described in this test method returned average values that range from LRV 1.7 to 4.3. 1.1.2 Results of this round-robin study indicate that caution should be used when comparing test data and ranking materials, especially when a small number of sample replicates are used. In addition, further collaborative work (such as described in Practice E691) should be conducted before this test method would be considered adequate for purposes of setting performance standards. |
|
Extent of Recognition
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
FDA recognition of ASTM F1608-16 [Rec# 14-496] will be superseded by recognition of ASTM F1608-21 [Rec# 14-574]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 14-496] until July 9, 2023. After this transition period, declarations of conformity to [Rec# 14-496] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
This standard is relevant for all medical devices where sterilization and packaging are components of the development and manufacture of the product. |
|
Relevant FDA Guidance and/or Supportive Publications*
1. ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
2. ISO 11607-2 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.
3. ISO/TS 16775 First edition 2014-05-15 Packaging for terminally sterilized medical devices - Guidance on the application of ISO 11607-1 and ISO 11607-2.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
FDA Technical Contact
|
Standards Development Organization
|
FDA Specialty Task Group (STG)
|
|
*These are provided as examples and others may be applicable. |