Quick Links: Skip to main page content
Skip to Search
Skip to Topics Menu
Skip to Common Links
U.S. Food & Drug Administration
Follow FDA
En Español
Search FDA
Home
Food
Drugs
Medical Devices
Radiation-Emitting Products
Vaccines, Blood & Biologics
Animal & Veterinary
Cosmetics
Tobacco Products
TPLC - Total Product Life Cycle
FDA Home
medical devices
databases
-
510(k)
|
DeNovo
|
Registration & Listing
|
Adverse Events
|
Recalls
|
PMA
|
HDE
|
Classification
|
Standards
CFR Title 21
|
Radiation-Emitting Products
|
X-Ray Assembler
|
Medsun Reports
|
CLIA
|
TPLC
New Search
show TPLC since
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
Back to Search Results
Device
implant, hearing, active, middle ear, partially implanted
Product Code
MPV
Device Class
3
Premarket Approvals (PMA)
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
2
4
3
1
2
1
1
2
0
0
0
1
0
1
0
0
MDR Year
MDR Reports
MDR Events
2014
112
112
2015
80
80
2016
65
65
2017
44
44
2018
53
53
2019
41
41
2020
54
54
2021
69
69
2022
57
57
2023
68
68
2024
16
16
Device Problems
MDRs with this Device Problem
Events in those MDRs
Output Problem
193
193
Device Operates Differently Than Expected
192
192
Adverse Event Without Identified Device or Use Problem
116
116
No Device Output
104
104
Migration
79
79
Insufficient Information
62
62
Device Appears to Trigger Rejection
60
60
Break
7
7
Extrusion
7
7
Improper or Incorrect Procedure or Method
6
6
Device Dislodged or Dislocated
5
5
Appropriate Term/Code Not Available
5
5
Noise, Audible
4
4
Device Stops Intermittently
4
4
Device Inoperable
3
3
Lack of Effect
3
3
Device Handling Problem
3
3
Positioning Problem
3
3
Therapeutic or Diagnostic Output Failure
3
3
Malposition of Device
3
3
Intermittent Continuity
2
2
Degraded
1
1
Mechanical Problem
1
1
Migration or Expulsion of Device
1
1
Circuit Failure
1
1
Air Leak
1
1
No Audible Alarm
1
1
Fitting Problem
1
1
No Audible Prompt/Feedback
1
1
Cut In Material
1
1
Ambient Noise Problem
1
1
Capturing Problem
1
1
Escape
1
1
Improper Device Output
1
1
Output below Specifications
1
1
Intermittent Energy Output
1
1
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
Failure of Implant
450
450
Hearing Loss
69
69
Skin Tears
28
28
No Code Available
26
26
Tissue Damage
25
25
No Information
24
24
No Known Impact Or Consequence To Patient
22
22
Insufficient Information
21
21
Hearing Impairment
19
19
Unspecified Infection
18
18
No Clinical Signs, Symptoms or Conditions
16
16
Tissue Breakdown
15
15
Therapeutic Response, Decreased
15
15
Therapeutic Effects, Unexpected
8
8
Pain
7
7
Bacterial Infection
7
7
Head Injury
6
6
Implant Pain
5
5
Total Hearing Loss
5
5
Fall
4
4
Wound Dehiscence
4
4
Complaint, Ill-Defined
4
4
Pressure Sores
3
3
Deafness
3
3
Erosion
2
2
Headache
2
2
Idioventricular Rhythm
2
2
Discomfort
2
2
Impaired Healing
2
2
Necrosis
2
2
Skin Infection
2
2
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available
2
2
Shock from Patient Lead(s)
1
1
Partial Hearing Loss
1
1
Undesired Nerve Stimulation
1
1
Scar Tissue
1
1
Swelling
1
1
Vertigo
1
1
Perforation of Vessels
1
1
Blurred Vision
1
1
No Consequences Or Impact To Patient
1
1
Meningitis
1
1
Patient Problem/Medical Problem
1
1
Tinnitus
1
1
Hematoma
1
1
Hemorrhage/Bleeding
1
1
Inflammation
1
1
Irritation
1
1
Granuloma
1
1
-
-