Device
handpiece, contra- and right-angle attachment, dental
Product Code
EGS
Regulation Number
872.4200
Device Class
1
Premarket Reviews
Manufacturer
Decision
AJK ENGINEERING, INC.
SUBSTANTIALLY EQUIVALENT
1
KAVO DENTAL GMBH
SUBSTANTIALLY EQUIVALENT
1
LARES RESEARCH
SUBSTANTIALLY EQUIVALENT
1
MICRO-NX CO., LTD.
SUBSTANTIALLY EQUIVALENT
1
NAKANISHI INC.
SUBSTANTIALLY EQUIVALENT
4
SAESHIN PRECISION CO., LTD.
SUBSTANTIALLY EQUIVALENT
2
MDR Year
MDR Reports
MDR Events
2019
219
219
2020
191
191
2021
294
294
2022
279
279
2023
115
118
2024
27
27
Device Problems
MDRs with this Device Problem
Events in those MDRs
Detachment of Device or Device Component
554
555
Overheating of Device
429
429
Mechanical Problem
117
117
Temperature Problem
84
84
Output Problem
45
45
Dent in Material
45
45
Therapeutic or Diagnostic Output Failure
43
43
Defective Device
35
35
Physical Resistance/Sticking
33
33
Device Handling Problem
18
18
Loose or Intermittent Connection
15
15
Break
14
14
Failure to Align
11
11
Computer Operating System Problem
10
10
Use of Device Problem
9
9
Mechanics Altered
9
9
Incorrect Measurement
8
8
Device Slipped
7
7
Insufficient Information
7
9
Noise, Audible
6
6
Improper or Incorrect Procedure or Method
6
6
Retraction Problem
5
5
Material Separation
5
5
Appropriate Term/Code Not Available
5
5
Data Problem
5
5
Device Fell
3
3
Mechanical Jam
3
3
Fitting Problem
3
3
Calibration Problem
3
3
Defective Component
2
2
Loss of Power
2
2
Energy Output Problem
2
2
Entrapment of Device
2
2
Unintended Movement
2
2
Unexpected Shutdown
2
2
Intermittent Energy Output
1
1
Excessive Heating
1
1
Unintended Electrical Shock
1
1
Device Difficult to Maintain
1
1
Adverse Event Without Identified Device or Use Problem
1
1
Unintended Ejection
1
1
Computer Software Problem
1
1
Failure to Cycle
1
1
Misconnection
1
1
Vibration
1
1
Component Missing
1
1
Device Displays Incorrect Message
1
1
Patient-Device Incompatibility
1
1
Connection Problem
1
1
Device Damaged by Another Device
1
1
Inappropriate or Unexpected Reset
1
1
Material Deformation
1
1
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
No Clinical Signs, Symptoms or Conditions
512
515
Burn(s)
257
257
No Consequences Or Impact To Patient
169
169
Insufficient Information
82
82
Partial thickness (Second Degree) Burn
81
81
Burn, Thermal
79
79
Patient Problem/Medical Problem
66
66
Injury
62
62
Superficial (First Degree) Burn
43
43
No Known Impact Or Consequence To Patient
39
39
Foreign Body In Patient
15
15
No Information
12
12
Pain
9
9
Burning Sensation
4
4
Laceration(s)
3
3
Full thickness (Third Degree) Burn
3
3
Device Embedded In Tissue or Plaque
3
3
Blister
2
2
Skin Burning Sensation
2
2
Tooth Fracture
2
2
Needle Stick/Puncture
2
2
Skin Tears
2
2
Numbness
1
1
Electric Shock
1
1
Eye Injury
1
1
Perforation
1
1
Scar Tissue
1
1
Ulceration
1
1
Unspecified Tissue Injury
1
1
Swelling/ Edema
1
1
No Code Available
1
1
Recalls
Manufacturer
Recall Class
Date Posted
1
Young Dental Manufacturing I, LLC
II
Nov-09-2023
-
-