Device system, hypothermia, intravenous, cooling
Product CodeNCX
Regulation Number 870.5900
Device Class 2


Premarket Reviews
ManufacturerDecision
ZOLL CIRCULATION, INC
  SE - WITH LIMITATIONS 1
ZOLL CIRCULATION, INC.
  SE - WITH LIMITATIONS 2

MDR Year MDR Reports MDR Events
2019 438 438
2020 415 415
2021 461 461
2022 511 511
2023 373 373
2024 231 231

Device Problems MDRs with this Device Problem Events in those MDRs
Visual Prompts will not Clear 1113 1113
Fluid/Blood Leak 606 606
Material Integrity Problem 599 599
Use of Device Problem 317 317
Adverse Event Without Identified Device or Use Problem 67 67
Temperature Problem 47 47
Insufficient Cooling 46 46
Unexpected Shutdown 44 44
Failure to Power Up 44 44
Difficult to Remove 27 27
Insufficient Heating 25 25
Difficult to Advance 23 23
Noise, Audible 20 20
No Display/Image 18 18
Power Problem 18 18
Intermittent Loss of Power 13 13
Mechanical Problem 12 12
Unraveled Material 10 10
Leak/Splash 8 8
Display or Visual Feedback Problem 7 7
Complete Loss of Power 7 7
Difficult to Insert 7 7
Detachment of Device or Device Component 7 7
Infusion or Flow Problem 5 5
Material Rupture 5 5
Difficult to Flush 5 5
Display Difficult to Read 5 5
Break 4 4
Backflow 4 4
Material Split, Cut or Torn 4 4
Output Problem 3 3
Gas/Air Leak 3 3
Retraction Problem 3 3
Defective Component 2 2
Positioning Problem 2 2
Material Puncture/Hole 2 2
Crack 2 2
Material Twisted/Bent 2 2
Improper Flow or Infusion 2 2
Excessive Cooling 2 2
Air/Gas in Device 2 2
Appropriate Term/Code Not Available 2 2
No Audible Prompt/Feedback 2 2
Tear, Rip or Hole in Device Packaging 1 1
Material Separation 1 1
Physical Resistance/Sticking 1 1
Electrical Shorting 1 1
Material Protrusion/Extrusion 1 1
Sparking 1 1
Device Alarm System 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 1164 1164
Insufficient Information 460 460
No Consequences Or Impact To Patient 451 451
No Patient Involvement 163 163
No Known Impact Or Consequence To Patient 75 75
Thrombosis/Thrombus 32 32
Cardiac Arrest 15 15
Hemorrhage/Bleeding 11 11
Pulmonary Embolism 9 9
Swelling/ Edema 9 9
Injury 9 9
Thrombus 6 6
Hematoma 5 5
Pneumothorax 5 5
Thrombosis 4 4
Death 4 4
Infiltration into Tissue 4 4
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available 4 4
Vessel Or Plaque, Device Embedded In 3 3
Obstruction/Occlusion 2 2
Cardiac Tamponade 2 2
Unspecified Blood or Lymphatic problem 2 2
Fever 2 2
Abdominal Pain 2 2
Hemothorax 2 2
Failure of Implant 1 1
Inflammation 1 1
Blood Loss 1 1
Chemical Exposure 1 1
Abrasion 1 1
Blister 1 1
Sepsis 1 1
Brain Injury 1 1
Skin Discoloration 1 1
No Code Available 1 1
Ventricular Fibrillation 1 1
Stroke/CVA 1 1
Foreign Body In Patient 1 1
Hypothermia 1 1
Pain 1 1
Ecchymosis 1 1
Bradycardia 1 1
Sudden Cardiac Death 1 1
Air Embolism 1 1
Shock 1 1
Perforation of Vessels 1 1
Convulsion/Seizure 1 1
Myocardial Infarction 1 1
Loss of consciousness 1 1
Heart Block 1 1

Recalls
Manufacturer Recall Class Date Posted
1 ZOLL Circulation, Inc. II Oct-06-2021
-
-