• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

TPLC - Total Product Life Cycle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search show TPLC since Back to Search Results
Device temporary non-roller type right heart support blood pump
Regulation Description Nonroller-type blood pump.
Definition Providing temporary partial or full right heart circulatory support.
Product CodePYX
Regulation Number 870.4360
Device Class 3

Premarket Approvals (PMA)
2020 2021 2022 2023 2024 2025
8 8 13 4 1 1

MDR Year MDR Reports MDR Events
2020 10 10
2021 7 7
2022 9 9
2023 89 89
2024 1031 1031
2025 35 35

Device Problems MDRs with this Device Problem Events in those MDRs
Adverse Event Without Identified Device or Use Problem 449 449
Restricted Flow rate 160 160
Patient-Device Incompatibility 145 145
Pumping Stopped 84 85
Increase in Pressure 76 76
Positioning Failure 64 64
Failure to Advance 62 62
Deformation Due to Compressive Stress 40 40
Improper Flow or Infusion 39 39
Device Sensing Problem 20 20
Failure to Sense 19 19
Fluid/Blood Leak 16 16
Material Twisted/Bent 16 16
Activation, Positioning or Separation Problem 16 16
Defective Device 16 16
Positioning Problem 13 13
Insufficient Flow or Under Infusion 10 10
Difficult to Insert 10 10
Material Split, Cut or Torn 9 9
Malposition of Device 9 9
Failure to Pump 8 8
Difficult to Advance 8 8
Infusion or Flow Problem 7 7
Difficult or Delayed Positioning 7 7
Detachment of Device or Device Component 6 6
Break 6 6
Pressure Problem 5 5
Failure to Deliver 5 5
Application Program Problem 5 5
Failure to Read Input Signal 5 5
Difficult to Remove 5 5
Material Deformation 5 5
Insufficient Information 4 4
No Pressure 4 4
Migration 4 4
Crack 3 3
Migration or Expulsion of Device 3 3
Defective Component 3 3
Unintended Application Program Shut Down 3 3
Material Separation 2 2
Fracture 2 2
Connection Problem 1 1
Unintended Movement 1 1
Obstruction of Flow 1 1
Device Handling Problem 1 1
Priming Problem 1 1
Optical Problem 1 1
Use of Device Problem 1 1
Failure to Power Up 1 1
Unable to Obtain Readings 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 521 522
Hemorrhage/Bleeding 295 295
Hemolysis 126 126
Thrombosis/Thrombus 115 115
Cardiovascular Insufficiency 61 61
Thrombocytopenia 38 38
Renal Failure 26 26
Ischemia 24 24
Perforation of Vessels 20 20
Hematoma 19 19
Sepsis 19 19
Tachycardia 18 18
Arrhythmia 14 14
Ventricular Fibrillation 14 14
Stroke/CVA 11 11
Cardiac Arrest 10 10
Insufficient Information 9 9
Vascular Dissection 8 8
Renal Impairment 8 8
Hematuria 7 7
Cardiac Perforation 6 6
Atrial Fibrillation 5 5
Great Vessel Perforation 5 5
Bradycardia 4 4
Tricuspid Valve Insufficiency/ Regurgitation 3 3
Asystole 3 3
Ischemia Stroke 3 3
Pericardial Effusion 3 3
Cardiac Tamponade 2 2
Perforation 2 2
Hypovolemia 2 2
Cardiogenic Shock 2 2
Gastrointestinal Hemorrhage 2 2
Air Embolism 2 2
Unspecified Vascular Problem 2 2
Embolism/Embolus 2 2
Multiple Organ Failure 2 2
Device Embedded In Tissue or Plaque 2 2
Hemothorax 2 2
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available 1 1
Hemolytic Anemia 1 1
Obstruction/Occlusion 1 1
Respiratory Failure 1 1
Lactate Dehydrogenase Increased 1 1
Low Cardiac Output 1 1
Unspecified Infection 1 1
Exsanguination 1 1
Disseminated Intravascular Coagulation (DIC) 1 1
Venipuncture 1 1
Pulmonary Edema 1 1

Recalls
Manufacturer Recall Class Date Posted
1 Abiomed, Inc. I Jan-15-2025
2 Abiomed, Inc. II Jan-04-2024
3 Abiomed, Inc. I Aug-03-2023
-
-