Device
temporary non-roller type right heart support blood pump
Regulation Description
Nonroller-type blood pump.
Definition
Providing temporary partial or full right heart circulatory support.
Product Code
PYX
Regulation Number
870.4360
Device Class
3
Premarket Approvals (PMA)
2021
2022
2023
2024
2025
2026
8
13
4
1
7
0
MDR Year
MDR Reports
MDR Events
2021
7
7
2022
9
9
2023
90
90
2024
1031
1035
2025
550
552
Device Problems
MDRs with this Device Problem
Events in those MDRs
Adverse Event Without Identified Device or Use Problem
573
577
Device Sensing Problem
266
268
Restricted Flow rate
258
258
Patient-Device Incompatibility
164
164
Increase in Pressure
110
110
Pumping Stopped
104
104
Failure to Advance
83
83
Positioning Failure
75
75
Deformation Due to Compressive Stress
47
47
Improper Flow or Infusion
39
39
Failure to Sense
31
31
Malposition of Device
24
24
Patient Device Interaction Problem
21
21
Fluid/Blood Leak
18
18
Defective Device
17
17
Activation, Positioning or Separation Problem
16
16
Material Twisted/Bent
16
16
Positioning Problem
14
14
Application Program Problem
13
13
Failure to Pump
13
13
Material Split, Cut or Torn
12
12
Difficult to Advance
10
10
Difficult to Insert
10
10
Insufficient Flow or Under Infusion
10
10
Infusion or Flow Problem
9
9
Decrease in Pressure
8
8
Failure to Read Input Signal
7
7
Difficult to Remove
7
7
Difficult or Delayed Positioning
7
7
Pressure Problem
6
6
Detachment of Device or Device Component
6
6
Break
5
5
Failure to Deliver
5
5
Material Deformation
5
5
No Pressure
5
5
Fracture
4
4
Crack
4
4
Migration
4
4
Material Separation
4
4
Unintended Application Program Shut Down
3
3
Defective Component
3
3
Migration or Expulsion of Device
3
3
Use of Device Problem
2
2
Computer Operating System Problem
2
2
Difficult to Open or Close
2
2
Obstruction of Flow
1
1
Failure to Power Up
1
1
Backflow
1
1
Inaccurate Flow Rate
1
1
Unable to Obtain Readings
1
1
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
No Clinical Signs, Symptoms or Conditions
751
753
Hemorrhage/Bleeding
374
378
Hemolysis
194
195
Thrombosis/Thrombus
174
174
Cardiovascular Insufficiency
66
66
Thrombocytopenia
48
48
Renal Failure
41
41
Arrhythmia
34
34
Ischemia
34
35
Unspecified Vascular Problem
33
33
Tachycardia
31
31
Sepsis
27
27
Hematoma
25
25
Perforation of Vessels
22
22
Cardiac Arrest
22
22
Hematuria
21
21
Stroke/CVA
20
20
Ventricular Fibrillation
19
19
Insufficient Information
17
17
Atrial Fibrillation
12
12
Renal Impairment
11
12
Cardiac Perforation
9
9
Low Blood Pressure/ Hypotension
8
8
Vascular Dissection
8
8
Embolism/Embolus
7
7
Multiple Organ Failure
7
7
Fever
7
7
Unspecified Infection
6
6
Pericardial Effusion
6
6
Pulmonary Edema
6
6
Bradycardia
6
6
Great Vessel Perforation
5
5
Asystole
5
5
Hypovolemia
4
4
Ischemia Stroke
4
4
Tricuspid Valve Insufficiency/ Regurgitation
4
4
Cardiogenic Shock
3
3
Perforation
3
3
Inflammation
3
3
Respiratory Failure
3
3
Cardiac Tamponade
2
2
Heart Failure/Congestive Heart Failure
2
2
Anemia
2
2
Exsanguination
2
2
Gastrointestinal Hemorrhage
2
2
Pulmonary Valve Insufficiency/ Regurgitation
2
2
Hemothorax
2
2
Air Embolism
2
2
Foreign Body In Patient
1
1
Hemolytic Anemia
1
1
Recalls
Manufacturer
Recall Class
Date Posted
1
Abiomed, Inc.
I
Jan-15-2025
2
Abiomed, Inc.
II
Jan-04-2024
3
Abiomed, Inc.
I
Aug-03-2023
-
-